ASSEMBLY INSTRUCTIONS
4. The e-Gauge Pointer may need to be broached out, to suit gauge
pinion.
5. Fit Target on Target Holder with capacitor in line with the hole provided.
6. Screw the e-Gauge® Target onto the e-Gauge® pointer.
7. The distance between the top of the target and the e-Gauge lens, should not be
more than 4mm. (see picture below)
2 washers are supplied to adjust the height of the Target on the Pointer,
if required
8. The Pointer and Target must be in-line along the axis, with the PCB writing
below the axis line - as per the picture below:
When all correctly in position and at -135° on the gauge, glue the target
in place securely.