ivoclar Bluephase G4 100 User manual

Manufacturer:
Ivoclar Vivadent AG
Bendererstrasse 2
9494 Schaan/Liechtenstein
www.ivoclar.com
Rx ONLY
Date information prepared:
2023-03-23/Rev. 1
756420/WE2
Bluephase®G4 100
[en] Instructions for Use 2
LED Polymerization unit
[sv] Bruksanvisning 20
LED Polymerisationsapparat
[da] Brugsanvisning 38
LED Polymeriseringslampe
[fi] Käyttöohjeet 56
LED-valokovettaja
[no] Bruksanvisning 74
LED-polymeriseringsapparat
[nl] Gebruiksaanwijzing 92
LED-polymerisatieapparaat
[el] Oδηγίε Xρήσεω 110
Συσκευή πολυμερισμού LED
[tr] Kullanma Talimatı 128
LED Polimerizasyon Cihazı
[ru] 146
Светодиодный полимеризационный прибор
[pl] Instrukcja stosowania 164
Lampa polimeryzacyjna-LED
Appendix 182

2
Dear Customer
Optimum polymerization is an important requirement for all light-cured materials in
order to consistently produce high-quality restorations. The curing light selected also
plays a decisive role in this respect. Therefore, we would like to thank you for having
purchased Bluephase® G4 100.
Bluephase G4 100 is a high-quality medical device which has been designed according
to the latest standard of science and technology in compliance with the relevant
industry standards.
These Instructions for Use will help you safely start up the device, make full use of
its capabilities and ensure a long service life.
Should you have any further questions, please do not hesitate to contact us.
(see addresses on the reverse page)
Your Ivoclar Team
Introduction

3
Table of Contents
1 Product Overview 4
1.1 List of parts
1.2 Indicators on the charging base
1.3 Indicators on the handpiece
1.4 Operating the curing light
2 Intended Use 7
3 Application 9
3.1 Start-Up
3.2 Operation
4 Maintenance and Cleaning 13
5 What if …? 15
6 Safety Information 16
7 Product Specifications 18
8 Additional Information 19
English

4
1.2 Indicators on the charging base
Charging base:
– Indicator is black: Battery is charged
– Indicator flashes blue: Battery is charging
1.1 List of parts
1 Product Overview
1.1 List of parts
Anti-glare cone
10 mm light guide
Start/Stop button
Handpiece
Handpiece support
Time selection buttonProgram selection button
Charging base
Anti-glare shield Power pack
Power cord
Protective sleeves
Battery

5
1.3 Indicators on the handpiece
Product Overview
Curing program / light intensity
Curing times High Power
Start/Stop button
Time selection button
Error indicator
Battery indicator
High Power / 1,200 mW/cm2
PreCure (preset to 2s) / 950 mW/cm2
20 seconds
15 seconds
10 seconds
Polyvision activated
Program selection button

6
1.4 Operating the curing light
With the handpiece switched on, the current charging status
is shown on the handpiece as follows:
No indicators lit up on the handpiece: Battery sufficiently
charged
Curing capacity of minimum 20 minutes in the High Power
program.
Battery symbol on the handpiece is flashing orange:
Battery weak
Time/intensity can still be set and a polymerization time of
approximately 3 minutes in the High Power program is left.
Place the light into the charging base as soon as possible!
Battery symbol on the handpiece is flashing orange and
a red “x” is shown: Battery completely discharged
The light can no longer be activated and the curing time can
no longer be set.
Product Overview
Start/Stop button
to trigger/abort the
curing process
Time selection button
to select the desired
curing time
Program selection button
to select the desired
curing program
The polyvision function is activated by long pressing (>2 s) the program or time selection button
(see 3.2 Operation). To confirm that polyvision is activated, the handpiece will beep and vibrate
for a short time and the polyvision indicator will light up. Renewed long pressing of the program
or time selection button results in the polyvision function being deactivated; the handpiece will
not vibrate.
Activating/deactivating polyvision

7
2 Intended Use
Intended purpose
Polymerization of light-curing dental materials
Patient Target group
– Patients with permanent teeth
– Patients with deciduous teeth
Intended users / Special training
– Dentists (clinical procedure)
– Dental assistants (clinical procedure)
– No special training required
Use
For dental use only.
Description
Bluephase G4 100 is an LED curing light that produces blue light. It is used for the polymerization of light-curing
dental materials directly in the oral cavity of patients.
Indications
None
Areas of application:
Polymerization of light-curing dental materials in the wavelength range of 385-515 nm, including filling materials,
dental adhesives, cavity liners, bases, fissure sealants, temporary restorations, luting materials for brackets and
indirect restorations (e.g. ceramic inlays).
Contraindications
None
Limitations of use
Materials whose polymerization is activated outside the wavelength range of 385– 515 nm (no such materials
known to date). If you are not sure about certain products, please ask the manufacturer of the corresponding
material.
Do not charge or use the appliance near flammable or combustible substances.
Never use without light guide.
The use of a light guide other than the one provided in the delivery form is not admissable.
Using the device stacked on or close to other equipment should be avoided because correct function can be
disrupted. If such use is unavoidable, the device needs to be monitored and checked for correct function.
Portable and mobile high-frequency communication devices may interfere with medical equipment.
The use of mobile phones during operation is not allowed.
Caution – The use of control or adjustment devices or performing procedures other than those specified
herein may result in hazardous radiation exposure.
Never use without eye protection for patients and users.
Warning
This unit should not be used near flammable anaesthetics or mixtures of flammable anaesthetics with air,
oxygen or nitric oxide.
In the case of serious adverse events in connection with the product, please contact Ivoclar Vivadent AG,
Bendererstrasse 2, 9494 Schaan, www.ivoclar.com, and your local health authorities.
Direct exposure to curing light may cause damage to the eyes.

8
Intended Use
Clinical benefit
In combination with restoration materials:
– Reconstruction of chewing function
– Restoration of esthetics
Residual risks
Users should be aware that any dental intervention in the oral cavity involves certain risks. Some of these risks are
listed below:
– As is the case with all high-performance lights, the high light intensity results in a certain heat development.
Prolonged exposure of areas near the pulp and soft tissues may result in irreversible damage.
Signs and symbols in these Instructions for Use
The signs and symbols in these Instructions for Use
facilitate the finding of important points and have
the following meanings:
Symbol Notes
Observe Instructions for Use
Caution
Limitations of use and Warning
Warning symbols and mandatory signs on the device
The signs on the device have the following meaning:
Symbol Notes
Double insulation
(device complies with safety class II)
Protection against electrical shock
(BF type apparatus)
AC voltage
DC voltage
The product must be disposed of
according to the corresponding
national legal requirements.
Recyclable
Caution
Observe Instructions for Use
(Failure to observe the Instructions for Use
may result in a risk to the patient or user.)
Observe Instructions for Use

9
Charging base
Before you switch on the device, make sure that the voltage mentioned on the rating plate complies with your local
power supply. Connect the power cord with the power supply. Make sure that the power cord is easily accessible at
any time and can be easily disconnected from the power supply.
Handpiece
Unpack the handpiece from the packaging and detach the light guide by
pulling it out. Then clean the handpiece and light guide (see Maintenance
and Cleaning). After cleaning, reinsert the light guide.
For reasons of hygiene, we recommend using a disposable protective sleeve
for each patient (see Maintenance and Cleaning). Make sure to fit the
protective sleeve snugly to the light guide. Sleeves enclosed in the delivery
form may be used or sleeves can be purchased per country specific
regulations. Then, attach the anti-glare cone or anti-glare shield to the
light guide.
3.1 Start-Up
Check the delivery for completeness and any possible transportation damage (see List of Parts).
If parts are damaged or missing, contact your Ivoclar representative.
3 Application

10
Gently place the handpiece in the corresponding rest
in the charging base without using any force. If a
hygiene sleeve is used, please remove it before you
charge the battery. If possible, always use the light
with a fully charged battery. This will prolong the
service life.
We therefore recommend placing the handpiece into
the charging base after each patient. If the battery is
fully discharged, the charging time is 2 hours.
Application
Battery
The battery must be fully charged before it is used for the first time! When
fully charged, the battery features a curing capacity of at least 20 minutes.
Slide the battery straight into the handpiece until you hear and feel it click
into place.
3.2 Operation
Disinfect contaminated surfaces of the curing light as well as light guides and anti-glare cones before each use.
Additionally, the light guide can be sterilized using the autoclaves intended for this purpose (see chapter
Maintenance and Cleaning). Furthermore, make sure that the emitted light intensity permits adequate poly-
merization. For that purpose, inspect the light guide for contamination and damage and check the light intensity at
regular intervals (see paragraph Measuring the light intensity).
Selecting the curing program and curing time
Bluephase G4 100 is equipped with 3 selectable curing times and 2 curing programs for different indications.
Use the Time/Program selection button to set the desired curing time and/or the light intensity.
H(High Power Program), 1,200 mW/cm2 *:
The following curing times can be selected in the High Power program: 10, 15 or 20 seconds.
PRE (PreCure Program)*:
The PreCure program is used to tack cure light-curing, adhesive luting composites in order to facilitate the removal
of excess material. If other luting composites are used, the distance from the light guide to the composite must be
increased or several curing cycles must be conducted. The curing time of the PreCure program is preset to 2 seconds
and cannot be altered.
* See Chapter 7 Product Specifications
The PreCure program must not be used for conventional light-curing!
The battery is an expendable part that typically needs to be replaced after
approximately 2.5 years. See battery label for the age of the battery.
Battery: charging status
The current charging status is displayed on the handpiece as described
on page 6.

11
Application
Observe the Instructions for Use of the material applied when selecting the curing time and intensity. The curing
recommendations for composite materials apply to all shades and, if not mentioned otherwise in the Instructions for
Use, to a maximum layer thickness of 2 mm. Generally, these recommendations apply to situations where the emission
window of the light guide is placed directly over the material to be polymerized. Increasing the distance between the
light source and the material will require the curing time to be extended accordingly. For instance, if the distance
to the material is approx. 11 mm, the effective light output is reduced by approx. 50%. In this case, the recommended
curing time has to be doubled.
1) The information provided herein applies to the 10 mm light guide supplied in the delivery form.
2) The information regarding heat development and burn hazards must be taken into consideration (see Safety notes).
Start/Stop button
Time selection button
High Power (1,200 mW/cm2)
PreCure (950 mW/cm2)
Curing times
Polyvision activated
Program selection button

12
Application
Polyvision - Automatic assistant with intelligent anti-glare protection
Bluephase G4 100 features a fully automated "polyvision" assistant for safe polymerization. This
feature enables the curing light to automatically sense if the handpiece is inadvertently moved
from its position during the curing procedure. To prevent any associated reduction in the amount
of energy being transferred, the device will start to vibrate to alert the user to the improper use
and will automatically extend the curing time by 10%. If the handpiece is significantly shifted away from its initial
position (e.g. out of the oral cavity), the light will automatically switch off so that the curing process can be restarted
and carried out correctly.
In addition, polyvision also acts as intelligent anti-glare protection. Polyvision ensures that the curing light cannot be
used outside the mouth. The light can only be activated once the light guide is positioned directly over the material
to be polymerized. This protects the operator and patient from being blinded by the light.
If a protective sleeve is used, this function is not available. Contamination of the light guide may affect the
functionalities of the polyvision assistant. The automatic assistant is there to support the user. It does not obviate
the need for monitoring by the user.
If you do not want to use the assistant, you can deactivate it at any time by long pressing (> 2 seconds) the Time or
Program selection button. The symbol on the handpiece (see 1.3 Indicators on the handpiece) then disappears.
Cure Memory function
The last settings used, together with the combination of curing program and curing time, are automatically saved.
Start/Stop
The light is switched on by means of the Start/Stop button. It is recommended that the emission window of the light
guide is placed directly on the material to be polymerized. Once the selected curing time has elapsed, the curing
program is automatically terminated. If desired, the light can be switched off before the set curing time has elapsed
by pressing the Start/Stop button again.
Acoustic signals
Acoustic signals can be heard for the following functions:
– Start (Stop)
– Every 10 seconds
– Curing time and program change
– Insert battery
– Error message (when anti-glare protection is activated or curing cycle aborted)
Light intensity
The light intensity is maintained at a consistent level during operation. If the supplied 10 mm light guide is used, the
light intensity has been calibrated to 1,200 mW/cm2± 10% in the High Power program. If another light guide than the
one supplied is used, it directly influences the emitted light intensity.
In a light guide with parallel walls (10 mm), the diameter of the light entry and the light emission window is the same.
When using focussing light guides (e.g. Pin-Point light guide 6>2 mm), the diameter of the light entry is larger than
that of the light emission window. The incident blue light is therefore concentrated over a smaller area. In this way,
the emitted light intensity is increased. Pin-Point light guides are suitable for spot-on polymerization, e.g. to fix
veneers before the removal of excess. For complete polymerization, the light guide must be changed.

13
4 Maintenance and Cleaning
For reasons of hygiene, we recommend using a disposable protective sleeve
for each patient. Make sure to fit the protective sleeve closely to the light
guide. Disinfect contaminated surfaces of the device and anti-glare cones
(FD 366/Dürr Dental, Incidin Liquid/Ecolab, CaviCide/Metrex Research).
Sterilize the light guide before each use if disposable protective sleeves are
not used. Make sure that no liquids or other foreign substances enter the
handpiece, charging base and particularly the power pack during cleaning
(risk of electrical shock). Disconnect the charging base from the power
source when cleaning it.
Cleaning the housing
Wipe the handpiece and handpiece holder with a customary aldehyde-free
disinfecting solution. Do not clean with highly aggressive disinfecting
solutions (e.g. solutions based on orange oil or with an ethanol content of
more than 40%), solvents (e.g. acetone), or pointed instruments, which may
damage or scratch the plastic. Clean dirty plastic parts with a soapy solution.
Pre-treating the light guide
Before cleaning and/or disinfecting the light guide, pretreat it. This applies to both automated and manual cleaning
and disinfection:
– Remove substantial contamination immediately after use or 2 hours after that at the latest. For this purpose,
thoroughly rinse the light guide under running water (for at least 10 seconds). Alternatively, use a suitable
aldehyde-free disinfecting solution to remove any adherent blood.
– To remove contamination manually, use a soft brush or soft cloth. Partially polymerized composite can be
removed with alcohol and a plastic spatula, if necessary. Do not use sharp or pointed objects, as they may
scratch the surface.
Cleaning and disinfection:
Machine cleaning and disinfection in a washer-disinfector unit should be preferred.
Mechanical cleaning and disinfection the light guide (Disinfector/CDU (cleaning and disinfection unit))
Machine cleaning and disinfection is possible using e.g. Neodisher® MediClean forte, Dr. Weigert, 0.5 %, according
to the cleaning program, e.g. cleaning 55 °C (131 °F) (+5 °C (41 °F)/–0 °C), 5–10 min, disinfection 90 °C (194 °F)
(+5 °C (41 °F)/–0°C), 5–10 min.
Manually cleaning and disinfecting the light guide
For manual cleaning, place the light guide in an instrument cleaning solution (e.g. ID 212 forte/Dürr Dental) for
the recommended reaction time (15 minutes). Make sure that the light guide is sufficiently immersed in cleaning
solution (cleaning bath with sieve insert and lid). Observe the instructions for use of the disinfectant manufacturer
when using cleaning and disinfectant solution.
When cleaning and disinfecting, please make sure that the agents used are free of:
– organic, mineral and oxidizing acids (the minimum admissible pH value is 5.5)
– alkaline solutions (the maximum admissible pH value is 11)
– oxidizing agent (e.g. hydrogen peroxide)
After the cleaning process, remove the light guide from the solution and thoroughly rinse it under running water
(20 +/– 2 °C / 36 +/– 4 °F ) for at least 10 seconds.

14
Sterilization of the light guide
Thorough cleaning and disinfecting is imperative to ensure that the subsequent sterilization is effective. Use only
autoclave sterilization for this purpose: 3x pre-vacuum, sterilization time (exposure time at sterilization temperature)
is 4 minutes at 134 °C (273 °F); pressure should be 2 bar (29 psi). Use nationally approved sterilization pouches.
Dry the sterilized light guide (10 min) using either the special drying program of your steam autoclave or hot air.
The light guide has been tested for up to 200 sterilization cycles.
Checking the light guide
After that, check the light guide for damage. Hold it against the light. If individual segments appear black, glass
fibres are broken. If this is the case, replace the light guide with a new one. If you can still see signs of contamination
on the light guide, the cleaning and disinfecting procedure must be repeated.
Maintenance and Cleaning

15
5 What if …?
Indicator Causes Error rectification
Red “x” lights up The device is overheated. Allow the device to cool down and try again after a
certain time. If the error persists, please contact your
dealer or your local Service Centre.
Electronic component of the
handpiece is defective.
Remove and reinsert the battery. If the error persits,
please contact your dealer or your local Service Centre.
Red “x” and
baery symbol
light up
Battery empty Place the device in the charging base and charge it.
Battery contacts dirty Remove battery and clean the battery contacts.
The charging base
is not illuminated
during charging
– Power pack not connected or
defective
– Battery fully charged
Check if the power pack is correctly positioned in the
charging base or if the power pack is connected to the
power supply by means of the power cord.
Repair work
The warranty period for Bluephase G4 100 is 3 years from the date of purchase (battery: 1 year). Malfunctions
resulting from faulty material or manufacturing errors are repaired free of charge during the warranty period. The
warranty however does not confer rights to compensation for any material or non-material damage. The apparatus
must only be used for the intended purposes. Any other uses are contraindicated. The manufacturer does not accept
any liability resulting from misuse and warranty claims cannot be accepted in such cases.
This is particularly valid for:
– Damage resulting from improper handling, especially incorrectly stored batteries (see Product Specifications:
Transportation and storage conditions).
– Damage to components resulting from wear under standard operating conditions (e.g. battery).
– Damage resulting from external influences, e.g. blows, drops to the floor.
– Damage resulting from incorrect set-up or installation.
– Damage resulting from connecting the unit to a power supply, the voltage and frequency of which
do not comply with the ones stated on the rating plate.
– Damage resulting from improper repairs or modifications that have not been carried out by certified
Service Centres.
In case of a claim under warranty, the complete apparatus (handpiece, charging base, battery, power cord and power
pack) must be returned, carriage paid, to the dealer or directly to Ivoclar together with the purchase document. Use
the original packaging with the corresponding cardboard inserts for transportation. Repair work may only be carried
out by a certified Ivoclar Vivadent Service Centre. In case of a defect that cannot be rectified, please contact your
dealer or your local Service Centre (see addresses on the reverse side). A clear description of the defect or the
conditions under which the defect occurred will facilitate locating the problem. Please enclose this description when
returning the apparatus.

16
6 Safety Information
– In the case of serious adverse events in connection with the product, please contact Ivoclar Vivadent AG,
Bendererstrasse 2, 9494 Schaan, www.ivoclar.com, and your local health authorities.
– The current Instructions for Use are available in the download section of the Ivoclar Vivadent AG website
(www.ivoclar.com).
– Bluephase G4 100 is an electrical appliance and a medical device which is subject to IEC 60601-1 (EN 60601-1)
and EMC Standard IEC 60601-1-2 (EN 60601-1-2) as well as the Medical Device Regulation (EU) 2017/745 (MDR).
The curing light complies with the relevant EU regulations.
– The curing light has been shipped from the manufacturer in a safe and technically sound condition.
In order to maintain this condition and to ensure risk-free operation, the notes and regulations in these
Instructions for Use must be observed. To prevent damage to equipment and risks for patients, users and third
parties, the following safety instructions must be observed.
Usage and liability
– Bluephase G4 100 must only be employed for its intended use. Any other uses are contraindicated. Do not touch
defective, open devices. Liability cannot be accepted for damage resulting from misuse or failure to observe the
Instructions for Use.
– The user is responsible for testing Bluephase G4 100 for its use and suitability for the intended purposes. This is
particularly important if other equipment is used in the immediate vicinity of the curing light at the same time.
– Use only original spare parts and accessories from Ivoclar. The manufacturer does not accept any liability for
damage resulting from the use of other spare parts or accessories.
– The light guide is an applied part and may warm up to a maximum of 45 °C (113 °F) at the interface to the
handpiece during operation.
– Keep out of the reach of children!
– For use in dentistry only!
Operating voltage
Before switching on, make sure that
a) the voltage indicated on the rating plate complies with the local power supply and
b) the unit has acquired the ambient temperature.
Do not touch the exposed contacts of the connection plug (power pack). If the battery or power pack are used
separately (i.e. during start-up), contact with patients or third parties must be prevented.
Assumption of impaired safety
If it has to be assumed that safe operation is no longer possible, the power must be disconnected and the battery
removed to avoid accidental operation. This may be the case, for example, if the device is visibly damaged or no
longer works correctly. A complete disconnection from the power supply is only ensured when the power cord is
disconnected from the power source. Ensure that the device can be quickly and easily disconnected at any time.
Eye protection
Direct or indirect exposure of the eyes must be prevented. Prolonged exposure to the light is unpleasant for the eyes
and may result in injury. To optimize user safety, the device has been equipped with intelligent anti-glare protection.
For this, the "polyvision" function must be activated (see 3.2. Operation). With the polyvision function activated, the
Bluephase G4 100 automatically senses if the handpiece is outside of the mouth and automatically switches off the
light if it has been activated inadvertently. If a protective sleeve is used, this function is not available.
In addition, we also recommend using the anti-glare protective equipment supplied. Individuals who are sensitive to
light, who take photosensitizing drugs, have undergone eye surgery, or work with the apparatus (or in its vicinity) for
long periods of time, should not be exposed and should wear protective orange goggles that absorb light below a
wavelength of 515 nm. The same applies for patients.

17
Safety Information
Battery
Caution: Use only original spare parts for Bluephase G4 100, particularly Ivoclar Vivadent AG batteries and charging
bases. Do not short circuit battery. Do not touch the battery contacts. Do not store at temperatures above 40 °C / 104 °F
(or 60 °C / 140 °F for a short period). Always store batteries charged. The storage period must not exceed 6 months.
May explode if disposed of in fire.
Please note that lithium-ion batteries may react with explosion, fire and smoke development if handled improperly or
mechanically damaged. Damaged lithium-ion batteries must no longer be used.
The electrolytes and electrolyte fumes released during explosion, fire and smoke development are toxic and
corrosive. Do not touch leaking batteries with bare hands. In case of contact with the eyes or skin, immediately wash
with copious amounts of water.
Avoid inhalation of fumes. In case of indisposition, see a physician immediately. Remove electrolyte residue from
surfaces by washing/wiping with a moist cloth. Wash contaminated pieces of clothing immediately.
Heat development
Generally, the stipulated curing times, particularly in areas near the pulp (adhesives: 10 seconds), must be
observed. Uninterrupted curing times of more than 20 seconds on the same tooth surface, as well as direct
contact with the gingiva, oral mucous membrane or skin have to be prevented. Polymerize indirect restorations at
intermittent intervals of 20 seconds or use external cooling with an air stream. The instructions regarding curing
programs and curing times must be observed (see Selecting the curing program and the curing time). In addition,
the light emission window must remain positioned exactly on the material to be cured during the entire duration of
the curing cycle (e.g. by holding it in place using a finger).
After several curing cycles on the same tooth, there is a risk that the pulp suffers damage caused
by the increased temperature!
Disposal
The product must be disposed of according to the corresponding national legal requirements. The curing
light must not be disposed of as normal houshold waste. Dispose unserviceable batteries and curing lights
according to the corresponding legal requirements in your country. Batteries must not be incinerated.

18
7 Product Specifications
Technical data
Light source Ivoclar Vivadent Polywave® LED
Wavelength range 385–515 nm
Peak 1: 400–410 nm
Peak 2: 450–465 nm
Light intensity High Power program: 1,200 mW/cm2± 10 %
PreCure program: 950 mW/cm2± 10 %
Operation 3 min on / 7 min off (intermittent)
Light guide 10 mm, autoclavable (active surface 0.61 cm2)
Signal transmitter Beeps every 10 seconds and every time the Start/Stop button or the
Time/Program selection button is activated or the anti-glare protection
is enabled or the curing process is aborted
Dimensions of the handpiece
(without light guide)
L = 170 mm, B = 30 mm, H = 30 mm
Weight of the handpiece 135 g (including battery and light guide)
Operating voltage of the handpiece 3.7 VDC with battery
5 VDC with power pack
Operating voltage of the charging base 5 VDC
Power supply Input: 100–240 VAC, 50–60 Hz max. 1 A
Output: 5 VDC / 3 A
Manufacturer: EDAC POWER ELEC.
Type: EM1024B2
Operating conditions Temperature +10 °C bis +35 °C
Relative humidity 30 % to 75 %
Ambient pressure 700 hPa to 1060 hPa
Dimensions of the charging base D = 110 mm, H = 55 mm
Weight of the charging base 155 g
Charging time Approx. 2 hours (with the battery empty)
Power supply of the handpiece Li-ion battery (approx. 20 min. with a new, fully charged battery in the
High Power program)
Transportation and storage conditions Temperature –20 °C to +60 °C
Relative humidity 5 % to 90 %, non-condensing
Ambient pressure 500 hPa to 1060 hPa
The curing light has to be stored in closed, roofed rooms and must not
be exposed to severe jarring.
Battery:
– Do not store at temperatures above 40 °C / 104 °F (or 60 °C / 140 °F for
a short period). Recommended storage temperature 15–30°C / 59–86 °F
– Keep the battery charged and store no longer than 6 months.
Delivery form 1 Charging base with power cord and power pack
1 Handpiece
1 Handpiece support
1 Light guide 10 mm
1 Anti-glare shield
3 Anti-glare cones
1 Pack of sleeves
1 Instructions for Use

19
8 Additional information
Keep material out of the reach of children!
Not all products are available in all countries.
The product has been developed solely for use in dentistry. Processing should be carried out strictly according to the Instructions
for Use. Liability cannot be accepted for damages resulting from failure to observe the Instructions or the stipulated area of use.
The user is responsible for testing the products for their suitability and use for any purpose not explicitly stated in the Instructions.

20
Bäste kund
Optimal ljushärdning är en viktig förutsättning för att genomgående erhålla
högkvalitativa restaurationer med alla ljushärdande material. Val av ljushärdningslampa
spelar också en viktig roll. Därför vill vi tacka dig för ditt köp av Bluephase® G4 100.
Bluephase G4 100 är ett högkvalitativt medicintekniskt hjälpmedel, som har utformats
efter senaste vetenskapliga och tekniska rön, tillsammans med relevanta industriella
normer.
Denna bruksanvisning kommer att hjälpa dig att på ett säkert sätt sätta lampan i bruk,
lära dig att utnyttja dess möjligheter till fullo under hela lampans långa livslängd.
Om du har frågor, tveka inte att kontakta oss.
(se adressen på baksidan)
Ditt Ivoclar Team
Introduktion
Table of contents
Languages: