Iwatsu HV-P60A User manual

High Voltage Probe
HV-P60A
Instruction Manual

Ⓒ2015 IWATSU TEST INSTRUMENTS CORPORATION.All rights reserved.

i
High Voltage Probe HV-P60A
KML096511 A1-712080(N)
Introduction
◇Thank you for purchasing the Iwatsu's instrument. We hope that it
can be used for a prolonged period.
◇We would ask you to thoroughly read this instruction manual before
commencing the operation and keep it at a readily accessible
location for future reference.
◇This instruction manual mainly describes notes on use,
compositions, basic operating procedure and specifications.
To ensure the Safety Operation
To ensure safe operation of this instrument and to prevent injury to
the user or damage to property, read and carefully observe
warnings and cautions in the following sections
Definition of warnings and cautions used in this
manual
Warnings
Incorrect operation or malfunction to observe the
warning may result in death or serious injury.
Cautions
Incorrect operation or malfunction to observe the
caution may result in injury or damage to the instrument
Notes
◇Parts of the contents of this manual may be modified without notice
for improvements in performance and functions.
◇Reproduction or reprinting of the contents of this manual without
prior permission from IWATSU is prohibited.
◇If any question about this instrument arises, contact Iwatsu office or
our sales distributors.
Revision History
◇May 2015: 1st edition

ii
High Voltage Probe HV-P60A
Read the following safety information
Warnings
●Measure in accordance with operating procedure when using
this instrument.
Erroneous operating procedure (refer to “3. Adjustment of Probe” and “4.
Measuring method”.) may result in an electric shock, fire or malfunction.
●When connecting the object to be measured to the nose of the
probe, firmly fix a measuring lead with an accessory Clamp.
Imcomplete clamping may result in an electric shock, fire or malfunction.
●Do not bring a grouding lead into contact with an insulator or
the nose of the probe.
If the grouding lead is brought into contact with the insulator or nose of the
probe, it may result in an electric shock, fire or malfunction.
●Coating of the grouding lead is voltage resistant (approx. 10
kV, but a clip is not and should be used carefully.
●Do not apply a voltage higher than the specified one.
If a voltage higher than the specified is applied, it may result in an electric
shock, fire or malfunction. The maximum input voltage is 60 kV at DC to 20
kHz. If the frequency exceeds 20 kHz, the maximum input voltage will
decrease (see “Fig.7 Maximum Input Voltage vs. Frequency”).
●During or before measurement, make sure that no one or
nothing come into contact around you.
If you touch someone or something during measurement, it may result in an
electric shock.
[Read the next page.]

iii
High Voltage Probe HV-P60A
Read the following safety information
Warnings
●Do not use in an environment with explosive gases.
It may cause an explosion.
●If you notice smoke, foul odor or abnormal noise, stop
measurement immediately.
Continued use under these circumstances may result in electric shock, fire
or malfunction. Immediatelly, turn off the power switch of the object to be
measured and remove this probe from the object to be measured after
confirming the thing that a high voltage has not gone out, and contact
Iwatsu office or our sales distributors for repair. The customer should never
repair the instrument because it is dangerous.
●This instrument is not a waterproof. Make sure no water gets
on or inside the instrument.
If the instrument is used in the wet condition, it may cause an electric
shock, fire or malfunction. If water is allowed in it, immediately turn off the
power switch of the object to be measured and remove this probe from the
object to be measured after confirming the thing that a high voltage has not
gone out, and contact Iwatsu office or our sales distributors for repair.
●Do not use this instrument when the drop of water adheres to
it or the hand is wet.
Otherwise electric shock, fire or malfunction could result.
●Do not place this instrument on an unstable place, such as a
shaky stand or an inclined place.
If this instrument is placed on an unstable place, it may be fallen down or
laid down, causing a personal injury or malfunction. It is recommended to
use a probe stand SK-301 (option). If this instrument is fallen down or
damaged, immediately turn off the power switch of the object to be
measured and remove this probe from the object to be measured after
confirming the thing that a high voltage has not gone out, and contact
Iwatsu office or our sales distributors for repair.
[Read the next page.]

iv
High Voltage Probe HV-P60A
Read the following safety information
Warnings
●Do not put a container that water, medicine, etc. has entered
on this instrument.
If not, an electric shock, fire or malfunction mayresult. If this instrument is wet
with water or medcine, contact Iwatsu office or our sales distributors forrepair.
●Do not directly hold this instrument when using
If the instrument is directly held by the hand, never hold it because it may
result in an electric shock.
●Do not drop the instrument or give strong vibrations or shock
to it.
A strong shock may damage the instrument and result in malfunction or an
electronic shock. When this instrument is dropped or damaged, contact
Iwatsu office or our sales distributors for repair.
●Do not modify or disassemble this instrument.
The repair and remodeling by the customer cause an electric shock,
fire or malfunction. The user cannot repair this instrument. Contact
Iwatsu office or our sales distributors for repair. Moreover, please note
not responding to the repair when remodeling the instrument.
●Do not use this instrument when it has failed.
Using Probe body or cable may be a cause of an electric shock or fire. In
the event of a malfunction, contact Iwatsu office or our sales distributors for
repair.
[Read the next page.]

v
High Voltage Probe HV-P60A
Read the following safety information
Warnings
●Connect the playground of the probe body with the grounding
potential (playground) of the object to be measured.
When the playground of the probe body is connected with potential other
than the playground of the object to be measured, it might cause an
electric shock or malfunction (for the object to be measured, this
instrument, and connected oscilloscope).
●When thunder begins, pull out the power plug of the target or
the oscilloscope to which this instrument is connected from
the outlet. Then, detach this instrument from the oscilloscope
or the object to be measured, and do not use it.
If not, lightning may cause an electric shock, fire, or malfunction.
●When the insulator is contaminated, wipe off contamination
prior to using it.
If the insulator is contaminated, a high voltage may leak and cause an
electric shock, fire or malfunction. Apply a little amount of neutral detergent
to soft cloth and wipe the insulator gently with it. Do not use highly volatile
liquid such as benzene or thinner.
[Read the next page.]

vi
High Voltage Probe HV-P60A
Read the following safety information
Cautions
●Always use or storage this instrument only within the rated operating
range.
The use of this instruction out of the rated operating range could result in
malfunction. The temperature ranges that allow the use of this instrument
are as follows.
・The range of operating or storaged temperature
:-10 ℃to +50 ℃(no dew condensation)
●Do not use or store this instrument in a location with direct sunlight,
high temperature, high humidity, or dew condensation.
An internal temperature might go up, and it cause an electric shock, fire or
malfunction.
●Do not put this instrument in a place exposed to lamp soot or steam
generated by counters, humidifiers, etc.
It may cause fire or malfunction.
●For your safety, remove the oscilloscope and the target to be measured
from this instrument before maintenance. In addition, wipe off with a dry
cloth when water droplets are attached.
If not, an electric shock or malfunction may occur.
●Before transporting this instrument, pack the housed instrument in the
packing material used at the time of purchase, or the equivalent or
better.
This instrument may fail if a large vibration or impact is applied during
transportation, possibly causing an electric shock or fire. If you do not have
a proper packing or shock-absorbing material for transportation, contact
Iwatsu office or our sales distributors. If transportation is to be handled by a
forwarding agent, display "Contains a precision machine" on each side of
the packing box.
●We recommend that you contact Iwatsu office or our sales distributors
for periodic inspection and calibration about once a year.
The measuring instrument might grow by the secular distortion the error
margin of measurements, and not fill a prescribed performance.
(The check and the calibration are received for a fee.)

vii
High Voltage Probe HV-P60A
Checking packed materials
When receiving this instrument, check the packed materials while
referring to the following “Components.” If there is a missing item or
an item damaged during transportation, immediately contact Iwatsu
office or our sales distributors.
Components
・Probe body ・・・・・・・・・1
・Accessories
Output cable ・・・・・・・・・1
Clamp ・・・・・・・・・1
Tip ・・・・・・・・・1
Ground lead ・・・・・・・・・1
Adjusting screwdriver ・・・・・・・・・1
Carrying bag ・・・・・・・・・1
Instruction Manual ・・・・・・・・・1

viii
High Voltage Probe HV-P60A
Management of instrument
When disposing of this instrument, it must be recycled or disposed of
properly in accordance with local laws or regulations. When
disposing of it, request a recycling company to dispose of it in
accordance with local laws or regulations.
Repair and sending instrument to be
repaired
If a malfunction occurs, send this instrument to Iwatsu office or our
sales distributors. Malfunctions that occur during the warranty
period due to our responsibility will be repaired by us free of
charge.
When sending an instrument to be repaired, clearly write the
instrument name, serial number (in the label on the rear of this
instrument), description of the malfunction, and the name, division,
and telephone number of the responsible person.
Use the box with a packing box used when goods is delivered or an
equal cushioning material to avoid the accident in transit when you
send it. If you do not have a proper packing or shock-absorbing
material for transportation, contact Iwatsu office or our sales
distributors.
Cleaning of this instrument
To clean this instrument, wipe it gently with a soft cloth moistened
with a small amount of water or mild detergent. Never use solvents
such as benzene, alcohol, acetone, ether, ketones, thinners, or
gasoline, as they can deform and discolor the case.
Storage
Do not store this instrument in the places below.
• Exposed to direct sunlight
• Dusty place
• Place with corrosive gas
Storage conditions of this instrument are as follows.
• Storage temperature: −10°C to +50°C

1
High Voltage Probe HV-P60A
Contents
Introduction......................................................................................... i
To ensure the Safety Operation......................................................... i
Checking packed materials..............................................................vii
Components......................................................................................vii
Management of instrument.............................................................viii
Repair and sending instrument to be repaired..............................viii
Cleaning of this instrument ............................................................viii
Storage .............................................................................................viii
1. General Description.......................................................................2
2. Name of each part and composition ............................................2
2.1 Probe body and Accessories .................................................. 2
2.2 Carrying bag (accessory) ........................................................ 3
3. Adjustment of Probe......................................................................4
3.1 Preparations ............................................................................. 4
3.2 Adjustment ............................................................................... 5
4. Measuring method.........................................................................7
4.1 Installation and connection..................................................... 7
4.2 Measuring Procedure............................................................. 12
5. Specifications...............................................................................13
5.1 Instrument specifications...................................................... 13
5.2 Appearance............................................................................. 15
Contacts

2
High Voltage Probe HV-P60A
1. General Description
This instrument is a passive prove to be connected to an oscilloscope
to measure a high-voltage waveform. Its frequency bandwidth is DC
to 50 MHz, maximum input voltage is 60 Kv, and attenation ratio is
2000:1.
2. Name of each part and composition
2.1 Probe body and Accessories
This instrument is composed of a probe body of Figure 1(The output
cable is contained.) and various accessories.
Figure 1. Probe body and Accessories
付属品
Screwdriver
Probe body
Tip
Clamp
Ground lead
Matching box part

3
High Voltage Probe HV-P60A
2.2 Carrying bag (accessory)
A probe body and various accessories are bundled to Carrying bag
like Figure 2, and delivered. Moreover, the probe body and the
accessory can be carried by putting it in Carrying bag after packing
things are opened.
Figure 2. Carrying bag (accessory)

4
High Voltage Probe HV-P60A
3. Adjustment of Probe
A waveform adjustment is necessary for this probe. A waveform
adjustment of this probe adjusts the capacity of the variable capacitor
that exists in this probe as the gain steadies for the frequency when
the oscilloscope is combined with this probe. If this adjustment is not
appropriately done, a correct measurement cannot be done.
Adjust the waveform of this probe when this probe is connected and
used for the oscilloscope.
The waveform of this probe is adjusted with the following
preparations according to the procedure.
3.1 Preparations
Prepare the signal generator that can generate the following signal
specifications and the measuring instrument (oscilloscope) used for
the measurement thereafter.
As for the example of the used equipment, it is not the following.
< Signal generator >
・Waveform : Square wave
・Frequency : 1 kHz or less
・Output voltage : 50 V to 100 V
・Rise time : 1μs or less
・Example of the used equipment
Example: KHT-1000 (PMK)
< Oscilloscope >
・Example of the used equipment
Example 1: DS-5400A series
(IWATSU TEST INSTRUMENTS CORPORATION)
Example 2: DS-5100B series
(IWATSU TEST INSTRUMENTS CORPORATION)

5
High Voltage Probe HV-P60A
3.2 Adjustment
This instrument is connected and adjusted according to the following
procedures.
(1) As shown in Figure 3(a), the ground terminal of the probe body is
connected with the terminal GND of the signal generator.
(2) INPUT of the oscilloscope is connected with the connector of the
probe body with Output cable.
(3) The output of the signal generator is connected with the tip of
the probe body
Signal Generator
Figure 3 (a) Connection when probe is adjusted
Oscilloscope
Probe body
Ground terminal
Matching Box
Connector
Tip of Probe
Figure 3 (b) Waveform adjustment of probe
in matching box (closeup)
Screwdriver
Hole for waveform adjustment
Clamp
Output cable
Ground lead

6
High Voltage Probe HV-P60A
(4) Adjust the input range of the oscilloscope, and display the
waveform on the screen of the oscilloscope.
(5) The driver for the adjustment is inserted in the adjustment hole
of the Matching box (Refer to Figure 3(b)). The waveform of the
probe is adjusted to become the ideal waveform like “Figure 3(e)
Adequated” while seeing the waveform (following Figure 3(c)
and Figure 3(d)) displayed in the oscilloscope.
Figure 3(d) Undercompensated
Figure 3(c) Overcompensated
Figure 3(e) Adequated

7
High Voltage Probe HV-P60A
4. Measuring method
4.1 Installation and connection
(1) Connection and fixation of Probe body side
A. When Probe body is set up and used for the stand
* Refer to following 1) to 5) and Figure 4(a).
1) The banana chip terminal (green) for the playground in the
stand is inserted in the terminal GND of the bottom of
Probe body.
2) The output cable is passed from the hole plug on the stand
side, and it is drawn out through the hole in the probe
receiving part.
3) The output cable is connected with the bottom of Probe
body.
4) Probe body is inserted in the probe receiving part and is
tightened with two fixed screws to be fixed.
5) Please connect the terminal GND of the probe stand with
earth GND for safety.
Figure 4(a) Probe body HV-P60A and Probe stand SK-301
Probe body
HV-P60A
Probe stand
SK-301
Probe receiving part
Hole plug
* Hole to pass cable from output
of the bottom of Probe body
Ground lead
Fixed screws
GND Terminal of
Probe stand

8
High Voltage Probe HV-P60A
B. When Probe body is set up horizontally or downward and
used for the stand
* Refer to following 1) to 5) and Figure 4(a), (b) and (c).
1) The bottom of Probe body is fixed from the back of the
probe receiving part by removing the probe receiving part
from the stand, and using four screws of the attachment for
the stand (Refer to Figure 4(b) and Figure 4(c)).
2) The banana chip terminal (green) for the playground in the
stand is inserted in the terminal GND of the bottom of
Probe body.
3) The output cable is passed from the hole plug (Refer to last
page and Figure 4(a)) on the stand side, and it is drawn out
from the hole in the probe receiving part.
4) The output cable is connected with the bottom of Probe
body.
5) The probe receiving part is fixed to the stand with four
make-up screws. (Refer to Figure 4(b))
6) The position in which the stand is installed is chosen, and
the stand installation hole is fixed with the screw etc. (Refer
to four places and Figures 4(b))
Figure 4(b) Probe stand SK-301 and Probe receiving part
Probe stand
SK-301
Probe receiving part
Stand installation
Hole (four places)
Make-up screw (four places)
Note
When buying it, probe stand SK-301 has installed the probe receiving
part for HV-P30A. Replace the probe receiving part for HV-P60A and
replace the top and bottom. (Refer to the manual of SK-301 for details.)

9
High Voltage Probe HV-P60A
Figure 4(c) Screw hole position for installation of
Probe body on Probe receiving part
Probe receiving part (back)
Screw hole position for
installation of Probe
body (four places)

10
High Voltage Probe HV-P60A
(2) Connection of oscilloscope and object to be measured
(Refer to Figure 5.)
1) The power switch of the object to be measured is turned off.
2) The ground terminal of Probe body is connected with the earth
GND of the object to be measured and the terminal GND of the
oscilloscope.
Connect the terminal GND of the stand with the earth of the
object to be measured and the terminal GND of the oscilloscope
when connecting the ground terminal of Probe body with the
terminal GND of the stand by (1) on the page beforehand.
3) The connector of Probe body and INPUT of the oscilloscope is
connected with the output cable.
Connect the output cable with INPUT of the oscilloscope when
the connector of Probe body is connected with the output cable
beforehand by (1) on the page beforehand.
4) The earth lead of the attachment is touched to the measurement
part beforehand, and it is confirmed the voltage is not generated.
Caution!
Connect following 2), 3) and 5) with each terminal firmly so that
there is no shake.
Probe body
Figure 5 Example of connecting measurement
(object to be measured and oscilloscope)
Ground terminal
Object to be
measured
Oscilloscope
Matching box
Earth lead of the attachment
Output cable
Ground lead
Connector
Clamp
Tip
Measurement part
This manual suits for next models
1
Table of contents
Other Iwatsu Measuring Instrument manuals

Iwatsu
Iwatsu SS-530 User manual

Iwatsu
Iwatsu SS-540 User manual

Iwatsu
Iwatsu SY-8218 User manual

Iwatsu
Iwatsu SS-623S User manual

Iwatsu
Iwatsu SS-500 User manual

Iwatsu
Iwatsu SS-521 User manual

Iwatsu
Iwatsu SS-281A User manual

Iwatsu
Iwatsu SS-520 User manual

Iwatsu
Iwatsu SS-293S User manual

Iwatsu
Iwatsu SS-281A-H User manual