Range
Resolution
Accuracy
±1 % FS
Temp . Range
Temp.
Resolution
Temp.
Accuracy
TDS Constant
0.5
Power LR44 x 4
Battery Life > 100 hours
Ambient
Temperature
Range
Waterproof
IP67
Dimensions
163 x 43 x 20 mm
Weight
71 g
Parameters
COND110N /
TDS110N
COND115N /
TDS115N
COND111N /
TDS111N
COND / Temp
TDS / Temp
COND / Temp
TDS / Temp
COND / Temp
TDS / Temp
10 uS/cm (Cond)
10 mg/L (TDS)
0.1 °C or ( ≤ 99.9 °F, 0.2 °F; > 99.9 °F, 1 °F )
0.5 uS/cm (Cond)
0.3 mg/L (TDS)
0.10 mS/cm (Cond)
0.05 g/L (TDS)
±1 % F.S.( ≥ 10 mS/cm
or ≥ 5.00 g/L ±2 % F.S.)
0~2000 uS/cm (Cond)
0~2000 mg/L (TDS)
0.0~100.0 uS/cm (Cond)
0.0~50.0 mg/L (TDS)
0.00~20.00 mS/cm (Cond)
0.00~10.00 g/L (TDS)
Warranty
Jenco warrants this product to be free from significant deviations in
material and workmanship for a period of 1 year from date of purchase. If
repair or adjustment is necessary and has not been the result of abuse or
misuse, within the year period, please return-freight-prepaid and the
correction of the defect will be made free of charge. If you purchased the
item from our Jenco distributors and it is under warranty, please contact
them to notify us of the situation. Jenco Service Department alone will
determine if the product problem is due to deviations or customer misuse.
※Out-of-warranty products will be repaired on a charge basis.
About Jenco
Jenco is a family-operated manufacturer with 45 years of specialization in
designing and producing water quality instruments. For more information
about the Wand tester or other Jenco products, please visit
https://jencoi.com
Calibrate
1. Remove the electrode cap of the instrument and rinse the
electrode & ATC assembly in distilled water.
2. Immerse electrode assembly into the standard solution.
Use the following table to determine the right standard
solutions for your tester.
3. Note the temperature displayed is the temperature of the
standard solution.
4. Press the “Cal” button for 2 seconds to enter the calibration
mode. Conductivity or TDS value of the standard solution
and “Cal” indicator is displayed.
5. Calibration data will automatically be stored when the
measurement values are stable.
6. Calibration is complete.
Measure
1. From “Measure Mode”, dip the electrode end of the tester
into the test solution.
Hold Data
1. When the conductivity or TDS reading is stable, press “Hold”
key once to hold and lock the display to the latest
measurement values.
2. Press “Hold” key again to release and unlock the display and
the tester returns to “Measure Mode”. The unit is now ready
for another measurement.
Specifications
Error Displays and Troubleshooting
[ Bring solution to a lower
temperature ]
[ Bring solution to a higher
temperature ]
[ Tester:Recalibrate with standard
solutions and take the measurement
again. If the readings are out of
range, contact vendor ]
[ Bring solution to a lower
temperature ]
Correction of slope beyond 50%
[ Calibrate with new standard
solutions or replace the tester ]
COND110N or TDS110N
COND115N or TDS115N
COND111N or TDS111N