Kaelus iVA0727A User manual

2017 | R99-0124RevC
This document contains information which is condential and the property of Kaelus, and which is not to be communicated to any person or
company, or used in any way without the previous authorization of Kaelus.
iVA A-SERIES OPERATING MANUAL
Cable And Network Analyzer System (iOS Version)

Operating Manual | Cable And Network Analyzer System (iOS Version)
2
Document Issue Description of Document
Change Date Authorization
Rev A Creation of document by
Eng (iOS) 29/04/16 R.H
Rev B
IR11200, IR11141 Section
remove and in case
charging warning.sdsds
30/04/17
Rev C Implementation of new
Kaelus format. 1/12/2017 LJ

3
© 2017 Innite Electronics, Inc., All Rights Reserved.
Precautions
This series of instrument is capable of producing up to 10 milliwatts of radio frequency (RF) power in the 600 MHz to 2750
MHz frequency band. Users are reminded that proper precautions must be taken to minimize exposure to these RF elds
to the recommended limits.
!WARNING: R F H A Z A R D
The iVA Series Cable and Network Analyzer is tted with a non user replaceable lithium-ion battery(Li-ion). Rechargeable
Li-ion batteries contain organic solvents and reactive materials. Erroneous handling of Li-ion batteries may result in heat
generation, explosion or re, possibly leading to injury.
Observe correct handling and disposal of Li-ion batteries.
!WARNING: B A T T E R Y
The instrument uses similar frequencies to several cellular telephone systems. Kaelus and its agents or distributors
accept no responsibility for interference to existing communications services as a result of improper operation or misuse.
If the equipment is used in a manner not specied by the manufacturer, the protection provided by the equipment may be
impaired.
DISCLAIMER: I N T E R F E R E N C E
This equipment should be protected from the ingress of liquids and precipitation such as rain and snow. The equipment
must be operated within the environmental conditions specied in Chapter 1.4. This equipment is not approved for use in
hazardous atmospheres, locations or medical applications.
GENERAL CONDITIONS OF USE
Do not recharge batteries while inside an accessory or transport case. Do not charge in a moving vehicle. Do not
operate or charge the equipment while inside an accessory or transport case.
!WARNING: C H A R G I N G
UNPACKING
Ensure that the total number of boxes of equipment is checked o against the delivery documentation. At the same
time inspect for any obvious transit damage. Should any damage be noted, notify the carrier immediately to le a transit
damage claim. Do not discard any packing material until notied by the carrier or Kaelus. Carefully unpack all containers
and check that all items listed on the delivery documentation. Please notify Kaelus of any damaged or missing items from
the shipment.

Operating Manual | Cable And Network Analyzer System (iOS Version)
4
Table of Contents
1. Operating Instructions 8
1.1 Introduction 8
1.2 Functional Description 8
1.3 Features 8
1.4 Specications 8
1.5 Construction and Features 8
1.6 Installing Software 10
1.6.1 Installing software on your iPad 10
1.7 Charging 11
1.7.1 Charging Constraints 11
1.7.2 Charging After the Battery has Been Run Flat 11
1.7.3 Charging Time 11
1.7.4 Battery Life 11
1.7.5 Safety 11
1.8 Switching on Your iVA 12
1.9 Bluetooth Connection Via Settings Menu 12
1.10 Connecting To Your iOS iPad (Application Information) 13
1.10.1 Selecting Your iVA in the iOS Application Device Menu 14
1.11 Navigation Menu 16
1.12 Traces Menu 17
1.12.1 Traces Menu Options 18
1.12.2 Trace Types 19
1.12.3 Trace Units 19
1.13 Channel Menu 20
1.13.1 Channel Menu Options 20
1.13.2 Subplots 21
1.14 Trace Zoom Functions. 22
1.14.1 Trace Autoscale 24
1.14.2 Trace Rest Scale 24
1.15 Return Loss Menu (iOS App Control) 24

5
© 2017 Innite Electronics, Inc., All Rights Reserved.
1.16 Cable Loss Measurement 25
1.17 Transmission Measurement 25
1.18 Distance to Fault 26
1.19 Spectrum Monitor Measurement 27
1.19.1 Setting Sweep Start and Stop Frequency 27
1.19.2 Setting Markers 28
1.20 Stimulus Mode 30
1.20.1 Stimulus Mode Options 31
1.21 Setting Test Limits 31
1.21.1 Test Limit Options 32
1.21.2 Minimum Limit 33
1.21.3 Maximum Limit 34
1.21.4 Ripple Limit 35
1.21.5 Tolerance Limit 36
1.22 Generating Reports 37
1.22.1 Site Information Menu 37
1.22.2 Entering Contractor Details and Logo 38
1.22.3 Adding Photo’s 39
1.22.4 Geotagging 40
1.22.5 Retrieving CSV data 40
1.22.6 Tags 40
1.22.7 Editing Reports 42
1.23 State Files 44
1.24 Settings 45
1.24.1 General Menu 45
1.25 Cleaning the Equipment 46
1.26 Turning o the IVA 46
2. Troubleshooting Guide 47
3. Ce Declaration of Conformity 49
4. Bluetooth Product Listing 50
5. End of Life Statement 51
6. Contacts 51

Operating Manual | Cable And Network Analyzer System (iOS Version)
6
List of Figures
Figure 1 iVA Equipment View 9
Figure 2 Install Screen from the App Store. 10
Figure 3 Switching on the iVA 12
Figure 4 Select the iVA’s to connect to by Serial number 12
Figure 6 iOS iVA Application icon 13
Figure 5 Allowing the iVA app via Bluetooth 13
Figure 7 iOS Application, Devices Menu 14
Figure 8 Application, Navigation Menu 16
Figure 9 Add and remove traces menu DTF, iOS application 17
Figure 10 Application, Channel Menu 20
Figure 11 Subplot Display 21
Figure 12 Application X axis Zoom 22
Figure 13 Application Y Axis Zoom 22
Figure 14 Application X and Y Zoom Together 23
Figure 15 Double tap and Slide Zoom 23
Figure 16 Return Loss Menu (iOS Application Control) 24
Figure 17 Transmission Calibration Menu 26
Figure 18 Distance to Fault Display, iOS Application 26
Figure 19 Spectrum Monitor, Override Stimulus 27
Figure 20 Marker Setting 28
Figure 21 Marker Focus 28
Figure 22 Trace Hold (Max-Min Menu) 30
Figure 23 Stimulus Mode 30
Figure 24 Limits Menu 31
Figure 25 Minimum Spec Limit 33
Figure 26 Max Limit Drawn on a Return Loss Trace 34
Figure 27 Ripple Limit on a Trace 35
Figure 28 Tolerance Spec Limit 36
Figure 29 Report Site Information 37
Figure 30 Merging External Reports 37

7
© 2017 Innite Electronics, Inc., All Rights Reserved.
Figure 31 Re-ordering Reports 38
Figure 32 Contractor Details Menu 38
Figure 33 Record and Photo Menu 39
Figure 34 Tag menu 40
Figure 35 Plot expanded view (Reports) 42
Figure 36 Edit Plot View 43
Figure 37 Remove a Plot 44
Figure 38 Settings General Menu 45
Figure 39 Settings Units 45

Operating Manual | Cable And Network Analyzer System (iOS Version)
8
1. Operating Instructions
1.1 Introduction
This manual describes the operation of the iVA (Cable and Network Analyzer) developed by Kaelus.
The iVA Cable and Network Analyzer enables users to measure Through Loss/VSWR/Return Loss and the location of the
VSWR/Return Loss faults in their RF infrastructure. The Bluetooth interface allows unprecedented measurement exibility
and opens up new & exciting possibilities in sweep testing.
The iVA is a rugged remotely controlled battery operated module that can be used with any Bluetooth enabled iOS iPAD.
1.2 Functional Description
The iVA can be connected to by an iPAD via an application available in the APP Store. The iVA is connected via the iPADs
Bluetooth interface.
1.3 Features
• Rugged and reliable; designed with tower climbers in mind
• Accurate measurement of swept VSWR/Return loss and Distance to Fault (DTF) in the RF Path.
• Simplies site certication sweep testing dramatically reducing test time on site.
1.4 Specications
Please refer to the Datasheet on our webpage
1.5 Construction and Features
The external details of the iVA test set are shown in Figure 1.

9
© 2017 Innite Electronics, Inc., All Rights Reserved.
1 Power On/O Button
Power on turns on the iVA unit. A short touch will turn on the iVA, indicated by a Green LED (See item 3). A long touch
until the Tx Rx activity LEDS ash then release will turn the iVA o.
2 Bluetooth Activity LED
Indicates when Bluetooth signaling is active
• Solid Blue indicates connected
• Flashing indicates disconnected
3 Power On/ Charge LED’s
Solid orange indicates charging. Green indicates switched on.
4 Tx, Rx Activity LED’s
Indicates when Transmit (Red led) or Receive (Green led) is active.
5 RF Input /Output Port
N Type connector.
6 USB Port
USB Port covered by a protective splashproof cover. Charging the iVA battery is done through the USB port. Connection
to your charger is done through the USB port.
7 Protective Cover
Protective cover for the RF Input Output connector.
8 Lifting Lug
Provides a fastening point for a lanyard or line to help prevent the unit being dropped, and support the measurement head
near the cable end or measurement point.
9 Protective Cover Retaining Hook
Provides a fastening point to hook on the protective cover to keep it out of the way while connecting and making
measurements.
6
8
41
2
3
597
Figure 1 iVA Equipment View

Operating Manual | Cable And Network Analyzer System (iOS Version)
10
1.6 Installing Software
1.6.1 Installing software on your iPad
1.6.1.1 App Store Releases
Typically the stand alone iOS installable le will be released via the App Store
1.6.1.2 App Store releases
Go to the App store and search for Kaelus iVA and press install.
1.6.1.3 Minimum System Requirements
iOS 8.0 and iPhone 4
Figure 2 Install Screen from the App Store.

11
© 2017 Innite Electronics, Inc., All Rights Reserved.
1.7 Charging
The iVA has a non user replaceable 3.6V Li-ion battery, that is charged via the USB port.
1.7.1 Charging Constraints
There is a variety of charging currents delivered from USB Ports. Ideally your USB Port should be able to deliver 500mA
of current or more. 500mA is sucient for intermittent use at maximum operation current and charging, you may nd that
heavy continuous use will not allow you to take on extra charging with a minimal current USB port.
While long USB cables are handy for moving the iVA into harder to reach places. Low quality USB cables can reduce the
power available to charge the iVA.
The iVA will self-limit at 1.2A, and is designed to charge from a number of dierent sources. A heavy use pattern may
require the use of the supplied USB Charger.
The supplied USB charger and small USB cable is recommended for charging.
1.7.2 Charging After the Battery has Been Run Flat
The iVA has a number of protection systems to reduce the chance of battery damage from complete discharge. The iVA
hardware will prevent higher current charging of a discharged battery. If discharged, it is necessary to charge the iVA while
switched o initially for at least 1 Hour. Note : That if the iVA has an adequate charge level remaining, normal and higher
charge rates can occur and normal use can commence immediately once connected for charging, lower current USB
charge sources may not support continuous heavy usage.
If the iVA is not going to be used for a signicant time and has been run at, it is recommended to give the iVA at least an
hours charge before storing. Li ion batteries can self-discharge at a rate of approximately 1% per month, if not used for
many months a periodic recharge may be needed.
1.7.3 Charging Time
Charging time will depend on the USB port current capacity. A new iVA, switched o for charging with the supplied a 5V
1.2A charger can reach fully charged in approximately 3.5 Hrs.
1.7.4 Battery Life
Fully charged with continuous return loss sweeps and under Bluetooth control the iVA will last typically 3 Hrs of use. With
connect and disconnect time, rigging and installation work time a nominal 8Hr work shift is readily achieved.
1.7.5 Safety
Any attempt to replace the internal battery could damage the unit. Care should be taken to not expose the battery to high
temperatures and to not overcharge the battery.

Operating Manual | Cable And Network Analyzer System (iOS Version)
12
1.8 Switching on Your iVA
A momentary touch on the Power switch on the iVA. Note: the pre–use charging instructions in Section 1.7. Turn on
the iVA prior to starting the iVA application. If you have recently installed or updated the iVA Application, ensure the iVA
is charged and will be operating from a reliable power source. After initial installation or application update, it may be
necessary for the application to update the iVA rmware and download the factory calibration data. It is necessary that the
iVA and iPAD remain on until these processes have completed.
1.9 Bluetooth Connection Via Settings Menu
In order for the iVA App to connect to the iVA, a Bluetooth connection must rst be established via the settings menu.
Make note of the serial number of your iVA and locate it in the iPAD’s Bluetooth devices menu and connect. You will then
need to allow the Bluetooth device to connect to the iVA Application. This will need to be repeated for all of the iVA’s in
use. iOS restricts the number of iVA’s that can be connected to 4 (As of Version 9.0.2) .
Figure 3 Switching on the iVA
Figure 4 Select the iVA’s to connect to by Serial number

13
© 2017 Innite Electronics, Inc., All Rights Reserved.
1.10 Connecting To Your iOS iPad (Application Information)
Once you have installed the iOS Application there will be an icon in your applications list. Click on it to start the program.
Figure 5 Allowing the iVA app via Bluetooth
If the Kaelus iVA application icon is not visible on your iPad screen, check in the Apps menu. Typically the Program will
start on the devices menu screen to allow you to rst connect to the iVA, when the iVA is rst switched on the Blue LED
will be ashing, indicating that the iVA is not yet paired to a iPad.
Figure 6 iOS iVA Application icon

Operating Manual | Cable And Network Analyzer System (iOS Version)
14
1.10.1 Selecting Your iVA in the iOS Application Device Menu
Ensure the Bluetooth is enabled on your iOS device, the Bluetooth switch can be found in the iOS/Apple Settings menu.
1.10.1.1 Devices Menu
1.10.1.1.1 Show Details
Displays iVA’s device information and iVA rmware version.
1.10.1.1.2 Connect Icons
The icon will rotate while establishing a connection to the selected iVA. The iOS iPAD/device. Connection to your iVA can
be conrmed in several ways. On the iPAD/device App the icon will stop rotating and be a solid/brighter Yellow. On the
iVA the Bluetooth LED will change from ashing to continuous ON. The identify function, see item 3 below can also be
used.
1.10.1.1.3 Identify
The identify function key allows you to get a visual response from the iVA/s you are connected to in your device list
pressing will cause the blue Bluetooth, red tx and green rx leds to ash for approximately 6 seconds. This allows
you to view which is the physical device on your list.
Figure 7 iOS Application, Devices Menu
1.10.1.1.1
1.10.1.1.2
1.10.1.1.3
1.10.1.1.4
1.10.1.1.5
1.10.1.1.6
1.10.1.1.7

15
© 2017 Innite Electronics, Inc., All Rights Reserved.
1.10.1.1.4 Battery Charge Indicator
The Battery charge indicator, indicates the battery charge level and whether the iVA is charging via the USB port.
Charging
50% Charge remaining
Battery Low
Battery Very Low
1.10.1.1.5 Progress Bar
Shows the progress of the transfers to and from the iVA. This could be factory calibration data or rmware (ensures your
iVA rmware is compatible with the application version you are using), and will occur on the rst use of the iVA or after
an update.
1.10.1.1.6 Enter Calibration Menu
Calibration saved Requires calibration
When the user enters any test mode (Return Loss, Cable Loss, DTF), the calibration status will be displayed and
color coded as described below. The accuracy of the iVA is quite good directly from the factory. However, it is highly
recommended that an Open-Short-Load (OSL) calibration be performed to maximize data accuracy. The Calibration
status will change depending on how much time has passed since the calibration was performed and if the temperature
has changed by more than 8 degrees since the last calibration. The table below indicates the recommended uses for the
dierent calibration states.
CAL Performed CAL Status Status Shown on Plots
and Reports
Acceptable for Use as
Follows
OSL Calibration
Last Calibrated Less than
12 Hours Prior Last Calibrated Time Prior Reportable Measurements
Last Calibrated Between 12-
36 Hours Prior
Last Calibrated =
Uncalibrated Fault nding and repairs
Greater than 36 Hours Last Calibrated =
Uncalibrated
General quick check and
repairs at a given
measurement plane
Last than ±8° C Temp
Change Last Calibrated Time Prior Reportable Measurements
Greater than ±8° C Last Calibrated =
Uncalibrated Fault nding and repairs
From the table we can see that once the iVA has been given an OSL Calibration, the iVA calibration indicator will remain
green for 12 Hours providing there has not been more than an 8 Degrees temperature change. The iVA can continue to be
used for fault nding and repairs and then be re-calibrated for the site report.
1.10.1.1.7 Pull Down Menu
Selects what Port number is associated with the connected iVA

Operating Manual | Cable And Network Analyzer System (iOS Version)
16
1.11 Navigation Menu
1.11.1.1.1 Navigation Side Panel Menu
Touch to open and touch outside the panel to close.
1.11.1.1.2 Return Loss. See Section 1.15
1.11.1.1.3 CableLoss. See Section 1.13
1.11.1.1.4 DTF (Distance To Fault) See Section 1.18
1.11.1.1.5 Spectrum Monitor. See Section 1.19
1.11.1.1.6 Transmission. (Optional Feature) See Section 1.17
1.11.1.1.7 Add New Mode
This allows custom modes to be created for specic test scenarios where needed.
1.11.1.1.8 Stimulus Mode. See Section 1.20
1.11.1.1.1
1.11.1.1.2
1.11.1.1.3
1.11.1.1.4
Figure 8 Application, Navigation Menu
1.11.1.1.5
1.11.1.1.6
1.11.1.1.7
1.11.1.1.8
1.11.1.1.9
1.11.1.1.10
1.11.1.1.11
1.11.1.1.12
1.11.1.1.13
1.11.1.1.14

17
© 2017 Innite Electronics, Inc., All Rights Reserved.
1.11.1.1.9 Devices. See Section 1.10.1
The devices menu allows you to scan and select the iVA device you want to use for your measurements.
Note: Your device/s must be connected rst, in the Bluetooth settings before items 1.11.1.1.2 to 1.11.1.1.4 above and
item 1.11.1.1.6 are able to operate. See 1.10.1.1 for more detail.
1.11.1.1.10 Calibrate Menu
1.11.1.1.11 Limits. Refer to Section 1.20
1.11.1.1.12 Report. See Section 1.22
1.11.1.1.13 Settings
1.11.1.1.14 Notications
1.12 Traces Menu
1.12.1.1
1.12.1.4
1.12.1.6
1.12.1.3
1.12.1.2
1.12.1.5
1.12.1.10
1.12.1.9
1.12.1.11
1.12.1.14
1.12.1.7
Figure 9 Add and remove traces menu DTF, iOS application
1.12.2
1.12.1.13

Operating Manual | Cable And Network Analyzer System (iOS Version)
18
1.12.1 Traces Menu Options
1.12.1.1 Remove Trace
The highlighted trace will be removed.
1.12.1.2 Shows the Trace Number.
This shows the trace number for the current parameter menu, other traces could be Tr2, Tr1 and so on.
1.12.1.3 Add an Additional Trace
Adds a trace to the current display channel.
1.12.1.4 DTF Range
Sets the maximum measurement range and trace X Scale on the DTF graph.
1.12.1.5 Cable Type
There are a number of pre-congured cable types in a pull down menu. This populates items 1.12.1.6, and 1.12.1.7
automatically, if your desired cable is not displayed, select custom and enter items 1.12.1.6 and 1.12.1.7 from your
cable datasheet.
1.12.1.6 Velocity Factor
The relative speed of the radio waves in the desired cable compared to air, being 1.00, Incorrect velocity factor entry can
introduce errors in the distance to fault measurements.
1.12.1.7 Cable Loss Per Unit Length
Length Units can be set in the settings menu. Enter the cable loss per meter.
1.12.1.8 Annotations Switch
If the Annotations switch is on, the entered Limits are overlaid on the display, individual user entered spec limits can be
applied to this trace by switching them on (See item 9).
1.12.1.9 Apply Limits Switches
Spec Limits entered ii the Limits menu can be applied to the selected trace, via an ON/OFF switch in this sub-menu.
1.12.1.10 Open Limits Menu
This will take you directly to the limits edit menu See Section 1.20
1.12.1.11 Override Stimulus Menu
Switch this on to set specic frequency stimulus for this trace.
1.12.1.12 Trace memory
Refer to Figure 21, The Trace memory function will store the currently focused trace into memory by pressing , the
memory icon is a toggle function, a solid symbol indicates the memory has contents already. Memory traces are
indicated by a dotted trace line See Figure 21. If a solid symbol is indicated, pressing it will erase the current memory,
pressing will move the current focused trace into that memory.
1.12.1.13 Trace Group
Traces in the same group are executed in order of trace number. If you have multiple iVA’s connected and want traces to
execute at a similar time, the traces can be allocated to a separate group or groups.

19
© 2017 Innite Electronics, Inc., All Rights Reserved.
1.12.1.14 Trace Math (Optional Feature)
The Trace math menu can from the currently highlighted trace. Add or subtract another displayed trace, or add or subtract
a previous memory trace of the highlighted trace See Section 1.12.1.12.
Note : The traces will need to be the same type
1.12.2 Trace Types
Dierent trace types can be included on the same measurement channel allowing for quick comparisons on the same
display.
Trace types available include .
1.12.2.1 Return Loss
See Section 1.15
1.12.2.2 Cable Loss
See Section 1.16
1.12.2.3 DTF
See Section 1.18
1.12.2.4 Spectrum
See Section 1.19
1.12.2.5 Transmission
See Section Section 1.17
1.12.3 Trace Units
The trace units available will depend on the trace type selected above.
1.12.3.1 dBm
Decibels with respect to a 1 mW in 50 Ohms
1.12.3.2 mW
milliwatts in 50 Ohms
1.12.3.3 dB
dB return loss or Cable loss
1.12.3.4 VSWR
Voltage Standing Wave ratio
1.12.3.5 Mag
Magnitude
1.12.3.6 Real
The real component of the complex characteristic.
1.12.3.7 Imag
The imaginary component of the complex characteristic.
1.12.3.8 Degrees
The degrees of Phase shift.

Operating Manual | Cable And Network Analyzer System (iOS Version)
20
1.13 Channel Menu
Channel frequency stimulus is used by default on traces, See Figure 9 to Override
1. Channel menu
2. Run
Selects between trace running and stopped
3. Trace Modules
3.1 Single sweep
The highlighted trace will be swept once
3.2 Continuous Sweep
The highlighted trace will be swept continuously
3.3 All Single
All traces will be swept once.
3.4 All Continuous
All traces will be swept/repeated Continuously.
4. Add and remove traces. See Section 1.12
5. Stimulus start frequency
6. Stimulus stop frequency
7. Number of points on plot
1
4
10
9
8
5
6
7
23
Figure 10 Application, Channel Menu
11
1.13.1 Channel Menu Options
Other manuals for iVA0727A
1
This manual suits for next models
3
Table of contents
Other Kaelus Measuring Instrument manuals