Minebea TMD-100 User manual

III
Digital indicator for tension meter
TMD-100
Instruction Manual
EN294-1181-D


I
Introduction
Thank you very much for your purchasing the digital Indicator for tension meter, model TMD-100.
This manual explains the handling procedures and the notes when it would be used. Make use of it
properly after reading through the manual carefully.
Be sure to deliver the manual to the end user. Moreover, the end user should keep the manual at
hand after reading it over.
Marks and arrangements used in this manual
The following marks are attached to the explanation on the matters that indicate “Don’t do this.”,
“Take care” and “For reference”.
Be sure to read these items where these marks are attached
Warning
Caution
Warning may cause injury or accident that may harm to the operator.
Don’t do these things described here.
It is the description when danger to which the user owes injury is assumed in your handling mistaken, and When the occurrence only of the
physical loss or damage is assumed.
Caution during operation and working.
Be sure to read the item to prevent malfunctions.

II
For safe operation
Be sure to read this instruction manual before use.
1. Installation place
Caution
• Environmental temperature : -10 ℃ to 50℃
• Environmental humidity : Less than 85 %R.H. (Non condensing)
Warning
① Location where installation is not allowed.
• Do not locate the instrument in direct and/or high temperature area.
• Do not use the instrument in a high humid area.
• Do not install the instrument where there are vibrations and shocks.
• Do not use the instrument where there is excess of dusts and fine particles.
• Do not use the instrument where there are corrosive gas and salt and like that.
• Do not install the instrument where there is rapid change of temperature and humidity.
• Do not install the instrument near the devices that are magnetized or generate an electromagnetic field.
• Do not install the instrument where the instrument may be affected by radioactivity or radial rays.
• Avoid the location where chemical reaction may take place such as in a laboratory, or like that.
Use the instrument where the temperature/humidity specifies with the range as follows :
Don’t locate the instrument on the places as follows :
It may cause an unexpected faulty in the instrument.

III
② Installation
Caution
Each dimensions of the instrument and required dimensions for the environmental spaces are as follows :
When installing the instrument, install as referring to the following figures and secure the space around the instrument.
Unit : mm
Front Side
Rear Panel cutout size
200
150
1.6 200
188
145
189
0
+1
146
0
+1

IV
2. Power supply
Warning
Warning
Caution
3. Application note
Caution
Caution
Caution
Caution
Caution
Be sure to check that the power supply is off in connecting each cable. If the work is done while the power is on, there may have the case that
electric shock to the operator or even may have damage to the instrument.
Before supplying the power, check that the indication of power supply voltage/specifications for the instrument and the power going to supply
should be the same. If they are not equal, contact with us.
If you use the instrument without checking them, it may cause a damage in the instrument or electric shock to the operator.
Please consult Co. about the report when not matching. There is danger of causing damage and the electric shock of this instrument when
using it like the uncertainty.
Earth wire should be grounded securely.
When earth wire is not connected, it may cause a malfunction of the instrument or electric shock to the operator.
Before using a new instrument or exchanging the strain gage applied transducer for a new one, be sure to make calibration.
If calibration will not be made, the correct measuring results may not be obtained nor which may cause malfunction in the instrument and there
may exist damage in peripheral equipments. Besides, even though calibration has been made, there may occur the similar case when the results
are not correct, so make calibration, again.
In case of using the instrument, check that the connections are executed properly. If not connected properly, the correct measuring result will
not be obtained, nor it may cause malfunctions of the instrument, damage to the peripheral equipments or even more serious accidents.
When change of setting is made carelessly on the instrument during measurement, correct measured results may not be obtained and it may
cause malfunction in the instrument and even have the possibility of damage in peripheral instruments.
Don’t shock the instrument such as throwing something on it.
If neglected, it may cause destruction of the parts and damage to the electrical circuits.
Don’t remove the cover of this instrument, don’t peel off the panel sheet, and additionally, take this instrument apart.
It may cause a damage on the case and panel sheet, and ruins the environmental factor and operativeness.
At the time of shipment from the factory, the instrument has been plated with a clear sheet on the panel sheet for protective purpose.
In case of application, use the instrument after removing the clear sheet first.
Don’t push the panel sheet on the instrument with the excessive strong force nor push it with sharp edge object such as a driver. If neglected,
it may cause a damage to the panel switch and even have the possibility of damage to resist to environments or operational performance.

V
Revision history
Date Manual No. Details of revision
May. 2006 DRW. NO.EN294-1181 First edition Ver.1.000
Oct. 2010 DRW. NO.EN294-1181A
Due to ECN. No, FN10-02140
-Change-
Minebea logo is changed.
May.2012 DRW. NO.EN294-1181B
Due to ECN. No, FN10-02140D
-Change-
Minebea logo is changed.
Jun.2016 DRW. NO.EN294-1181C
Due to ECN No, FN16-02057
-Deletion-
Delete 'Minebea Co., Ltd. Measuring Components Business Unit' from the
front cover.
Due to ECN No,FN10-02140-E
-Change-
Minebea logo is changed.
Aug.2017 DRW. NO.EN294-1181D Due to ECN FN17-02017
・Delete the company name in the contents.

VI
Contents
INTRODUCTION ..................................................................................................................................................................................... I
MARKS AND ARRANGEMENTS USED IN THIS MANUAL .................................................................................................... I
FOR SAFE OPERATION ..................................................................................................................................................................... II
1. INSTALLATION PLACE ...................................................................................................................................................................... II
2. POWER SUPPLY ............................................................................................................................................................................... IV
3. APPLICATION NOTE ........................................................................................................................................................................ IV
REVISION HISTORY .............................................................................................................................................................................V
1. NAME AND FUNCTION OF EACH POINT ......................................................................................................................... 1
1-1. FRONT PANEL ................................................................................................................................................................................... 1
1-2. REAR PANEL ..................................................................................................................................................................................... 2
2. CONNECTING METHOD ........................................................................................................................................................... 3
2-1. NOTES FOR CONNECTING WIRES ................................................................................................................................................. 3
2-2. CONNECTION WITH LOAD CELL .................................................................................................................................................... 3
2-3. CONNECTION OF ANALOG OUTPUT ............................................................................................................................................ 6
2-4. CONNECTION OF POWER SUPPLY AND EARTH ......................................................................................................................... 8
2-5. CONNECTION OF EXTERNAL CONTROL INPUT .......................................................................................................................... 9
2-6. CONNECTION OF EXTERNAL CONTROL OUTPUT ................................................................................................................... 10
3. OPERATION ................................................................................................................................................................................. 11
3-1. CHANGE OF MODE ......................................................................................................................................................................... 11
4. CALIBRATION ............................................................................................................................................................................. 13
4-1. KIND OF CALIBRATION METHOD ................................................................................................................................................. 13
4-2. SETTING ITEMS ............................................................................................................................................................................... 14
4-3. CALIBRATION PROCEDURE OF DIGITAL SPAN CALIBRATION MODE (CCAL CALIBRATION MODE) .......................... 15
4-4. CALIBRATION PROCEDURE OF DIGITAL ZERO AND SPAN CALIBRATION MODE (ACAL CALIBRATION MODE) ...... 20
4-5. CALIBRATION PROCEDURE OF ACTUAL ZERO/SPAN (LCAL CALIBRATION MODE) .................................................... 25
4-6. CALIBRATION PROCEDURE OF DIGITAL SPAN CALIBRATION MODE (SPAN2 CALIBRATION MODE) ......................... 30
4-7. FINE ADJUSTMENT OF ZERO ..................................................................................................................................................... 34
4-8. FINE ADJUSTMENT OF SPAN .................................................................................................................................................... 36
4-9. CALIBRATION ONLY OF ZERO POINT ...................................................................................................................................... 38
4-10. CALIBRATION LOCK .................................................................................................................................................................. 39
5. FUNCTION MODE ..................................................................................................................................................................... 40
5-1. METHOD OF SETTING FUNCTION MODE ................................................................................................................................... 40

VII
5-2. FUNCTION OF FUNCTION DATA .................................................................................................................................................. 41
5-3. C FUNCTION MODE ........................................................................................................................................................................ 46
6. VARIOUS FUNCTIONS BY FUNCTION DATA .............................................................................................................. 47
6-1. EFFECTIVE RANGE OF ZERO SET ............................................................................................................................................. 47
6-2. SETTING OF DECIMAL POINT DISPLAY POSITION .................................................................................................................. 47
6-3. A/D SAMPLING FREQUENCY ...................................................................................................................................................... 47
6-4. DISPLAY FREQUENCY ................................................................................................................................................................... 47
6-5. DIGITAL FILTER .............................................................................................................................................................................. 47
6-6. SELECTION OF OPERATIONAL INPUT FOR DS/WS DOWN ............................................................................................... 48
6-7. KEY LOCK ........................................................................................................................................................................................ 49
6-8. STABILIZATION FILTER ................................................................................................................................................................. 49
6-9. SETTING OF ANALOG OBJECT ................................................................................................................................................... 51
6-10. SETTING OF ANALOG OUTPUT TYPE ...................................................................................................................................... 51
6-11. SETTING OF EFFECTIVE AND INVALID FOR THE DISPLAY OF ADD AND DIFF ............................................................ 51
6-12. SETTING OF OUTPUT RATIO AT BOTH DS AND WS DOWN ............................................................................................ 51
6-13. SCALING OF ANALOG OUTPUT ................................................................................................................................................ 51
6-14. SETTING OF BRIDGE POWER SUPPLY VOLTAGE .................................................................................................................. 51
6-15. SETTING OF CHECK VALUE ....................................................................................................................................................... 51
6-16. SETTING OF PERMISSION/PROHIBITION OF CALIBRATION ............................................................................................... 53
6-17. ZERO CLEARNESS ..................................................................................................................................................................... 54
6-18. MEMORY CLEARNESS ................................................................................................................................................................. 54
6-19. DISPLAY COEFFICIENT ............................................................................................................................................................... 54
7. RECORD PLACE OF SET DATA ........................................................................................................................................ 55
7-1. DATA RECORDED IN EEPROM .................................................................................................................................................. 55
7-2. DATA RECORDED WITH RAM ..................................................................................................................................................... 55
8. CHECK MODE ............................................................................................................................................................................ 56
8-1. METHOD OF SETTING CHECK MODE ......................................................................................................................................... 56
8-2. CONFIRMATION OF ROM VERSION ........................................................................................................................................... 57
8-3. CONFIRMATION OF OPTIONAL CONNECTION .......................................................................................................................... 58
8-4. CONFIRMATION OF BRIDGE POWER SUPPLY VOLTAGE ........................................................................................................ 58
8-5. CONFIRMATION OF EXTERNAL CONTROL INPUT .................................................................................................................... 59
8-6. CONFIRMATION OF CONTACT OUTPUT .................................................................................................................................... 59
8-7. CONFIRMATION OF ANALOG OUTPUT ....................................................................................................................................... 60
9. ANALOG OUTPUT ................................................................................................................................................................... 61
9-1. CHANGING METHOD OF ANALOG OUTPUT .............................................................................................................................. 61
9-2. SCALING OF ANALOG OUTPUT ................................................................................................................................................... 63
9-3. FINE CONTROL 1 OF ANALOG OUTPUT .................................................................................................................................... 65

VIII
9-4. FINE CONTROL 2 OF ANALOG OUTPUT .................................................................................................................................... 67
10. MONITOR MODE ..................................................................................................................................................................... 69
10-1. METHOD OF SETTING MONITOR MODE .................................................................................................................................. 69
11. TROUBLESHOOTING ............................................................................................................................................................ 70
11-1. TROUBLESHOOTING EXECUTION ............................................................................................................................................. 70
12. SPECIFICATION ...................................................................................................................................................................... 78
12-1. ANALOG SPECIFICATION ........................................................................................................................................................... 78
12-2. DIGITAL SPECIFICATION ............................................................................................................................................................ 78
12-3. INTERFACE .................................................................................................................................................................................... 78
12-4. GENERAL SPECIFICATION ......................................................................................................................................................... 78
12-5. STANDARD SHIPMENT SPECIFICATION .................................................................................................................................. 79
12-6. ACCESSORY ................................................................................................................................................................................. 79
13. ERROR INDICATION .............................................................................................................................................................. 80
14. GUARANTEE ............................................................................................................................................................................. 81
14-1. GUARANTEE .................................................................................................................................................................................. 81
14-2. REPAIR ........................................................................................................................................................................................... 81
15. DISPLAY CHARACTER PATTERN .................................................................................................................................. 82
16. FUNCTION SETTING TABLE ............................................................................................................................................. 83
17. C FUNCTION DATA TABLE ............................................................................................................................................... 84
18. H FUNCTION MODE .............................................................................................................................................................. 85
18-1. METHOD OF SETTING H FUNCTION MODE ............................................................................................................................ 85
19. APPENDIX .................................................................................................................................................................................. 87

1
1.Name and function of each point
1-1. Front panel
(1) Load display section
Each load value of WS, DS, ADD, and DIFF is displayed.
(2) Status display section
HEALTHY After the power supply is turned on, it lights with various error being not occurred.
CHECK It lights on when the key and the key is pressed together, and CHECK ON.
WS DOWN It lights on when the key and the key is pressed together, or when the
equivalent signal from the outside is input.
DS DOWN It lights on when the key and the key is pressed together, or when the
equivalent signal from the outside is input
(3) Key
By pressing this key with the key, a present load value is memorized as a zero point, and the
display is adjusted to zero. Or, when the value is set, the set digit is carried.
(4) Key
By pressing this key with the key, the CHECK value is turned on and off. Or, when the value is set,
the digit lowers the set digit.
(5) Key
When the trouble is occurs on the WS side load cell, the WS false signal set beforehand is generated
from the inside of the indicator by pressing this key with the key. Or, when the value is set, one set
digit is lowered.
② Status display
section
① Load display section
⑨ Monitor terminal
③ Key
④ Key
⑤ Key
⑧ Key
⑦ Key
⑥ Key

2
(6) Key
When the trouble is occurs on the DS side load cell, the DS false signal set beforehand is generated
from the inside of the indicator by pressing this key with key. Or, when the value is set, one value
of the set digit is lowered.
(7) Key
Used When shifting to the function mode.
(8) Key
Used for registering set value at the time of various kind of setting.
Also, the shift function of each key becomes effective by pushing each key while pushing this key.
(9) Monitor terminal
The analog output set beforehand can be monitored.
1-2. Rear panel
(1) Terminal block 1
The analog output is connected.
(2) Terminal block 2
The optional analog output is mounted.
(3) Fuse holder
The fuse of the specified capacity is mounted.
(4) Terminal block 3
A specified power supply voltage and grounding wire are connected.
(5) Terminal block 4
Various strain gauge type transducers, such as load cell, external control input for "ZERO", "WS
DOWN", "DS DOWN", "D S&W S DOWN", "LOCK" and external control output for "WS DOWN", "DS
DOWN", "HEALTHY", "ERROR", and "RUN" are connected.
① Terminal block 1
② Terminal block 2
③ Fuse holder
⑤ Terminal block 4
④ Terminal block 3

3
2.Connecting method
2-1. Notes for connecting wires
• When you connect wires, please proceed after turning off the power supply without fail.
• Please do not connect AC power supply until the installation is completed. There is no switch that switches
power supply ON/OFF in the main body.
• Please keep the cable connected with the main body away from the noise source, such as power supply line
and I/O for the control, etc. as much as possible.
• Please set the conduit wiring to be exclusive wiring, and avoid sharing with other lines.
• Please make sure to connect grounding wire, and the earth is to be the D class with single earth. Please do
not share with the earth of the power supply system.
• Please apply the shielded cable for connecting with the external control input and contact output, and the
shield is to be connected with the shield connector of the terminal, or main body.
2-2. Connection with load cell
(1) Connection of load cell and TMD-100
- When the tension force is measured with tension type or tension/compression type load cell, please connect the “green” with terminal No.
B1 or B2, and the “Blue” with terminal No. D1 or D2 respectively.
- As there may be the case that the wiring color is different form the standard, please confirm it with the inspection data sheet of the load
cell before connecting.
- When the length of CAB-502 is more than 30 m totally, there may have the case the accuracy becomes out of warranty because the
resistance of cable makes the input voltage of the instrument decrease.
- When the length of CAB-501 is more than 100 m totally, there may have the case that the remote sensing function will not work enough,
and the accuracy becomes out of warranty because of the resistance of cable.
RED
WHT
RED
WHT
SHORT BAR
SHORT BAR
GRN
BLU
GRN
BLU
SHIELD
SHIELD
WS side load cell DS side load cell
- Please refer the paragraph
19. Appendix – “List of
Terminal block” as for the
details.

4
(2) Connection of load cell, junction box B-304 for extension, and TMD-100
i) When CAB-502 (4-cores cable) is used
Internal terminal connection of B-304
SHORT BAR
SHORT BAR
GRN
BLU
GRN
BLU
SHIELD
SHIELD
RED
WHT
RED
WHT
WS side load cell DS side load cell
B-304
Junction box
B
-
304
Junction box
- Please refer the paragraph 19.
Appendix – “List of Terminal
block” as for the details.
Suitable crimp type terminal lugs
: 1.25 or 2-4
RED BLU GRN
to load cell
Terminal pitch 9.5 mm
YEL(Shield)WHT
RED BLU GRNWHT
YEL(Shield)
t
o TMD-100

5
ii) When CAB-501 (6-cores cable) is used
Internal terminal connection of B-304
Suitable crimp type terminal lugs
: 1.25 or 2-4
RED BLU GRN
to load cell
Terminal pitch 9.5 mm
YEL(Shield)WHT
YEL(Shield)
to TMD-100
RED BLU GRN
WHT
ORG BLK
GRN
BLU
GRN
BLU
SHIELD
SHIELD
ORG
BLK
RED
WHT
ORG
BLK
RED
WHT
WS side load cell DS side load cell
B-304
Junction box
B
-
304
Junction box
- Please refer the paragraph 19.
Appendix – “List of Terminal
block” as for the details.

6
2-3. Connection of analog output
This instrument can select voltage output/current output by the setting with each channel.
Please refer the paragraph 9-1. Changing method of analog output for setting method.
(1) Specifications of current output
Specification Content
Output DC4 mA to 20 mA
Load resistance 510 ohm or less
Non-linearity 0.05 %F.S.
Resolution About 1/12 000
Over range "-OL" display : approx. DC0 mA
"OL" display : approx. DC24 mA
Output frequency Synchronized with A/D sampling speed.
- The analog output of this instrument executes rewriting the output synchronizing with A/D sampling speed.
- After turning on the power supply, the analog output has the output fluctuation element.
To use it with stable condition, please use it about one hour after turning on the power supply.
- +
A
Shield
Load resistance 510 ohm or less
- Please refer the paragraph 19.
Appendix – “List of Terminal
block” as for the details.

7
(2) Specifications of voltage output
Specification Content
Output Selection from DC0 V to 10 V, DC0 V to 5 V, DC0 V to 1 V, DC±10 V, DC±5 V, and DC±1 V
Load resistance 5 k-ohm or more
Non-linearity 0.05 %F.S.
Resolution About 1/12 000
Over range "-OL" display : DC-2.5 V, DC-1.25 V, DC-0.2 V, DC-14.0 V, DC-7.5 V, and DC-1.2 V
"OL" display : DC14.0 V, DC7.5 V, DC1.2 V, DC14.0 V, DC7.5 V and DC1.2 V
Output frequency Synchronized with A/D sampling speed.
- The analog output of this instrument executes rewriting the output synchronizing with A/D sampling speed.
- After turning on the power supply, the analog output has the output fluctuation element.
To use it with stable condition, please use it about one hour after turning on the power supply.
V
-
Shield
Load resistance 5 k-ohm or more
- Please refer the paragraph 19.
Appendix – “List of Terminal block”
as for the details.

8
2-4. Connection of power supply and earth
Please connect the power supply and the earth as shown in the figure below. Please give the earth as D
class single earth.
Power supply voltage: AC100 V to AC240 V(permissible variable range AC85 V to AC264 V)
Power supply frequency: 50/60 Hz
- Connections with power supply and the earth is made securely as showing in the figure, and also within the rated capacity of an instrument.
- Grounding is to be the D class with single earth.
There is a possibility of causing an unexpected malfunction by receiving the influence of the noise from other equipment.
~
AC100 V to AC240 V

9
2-5. Connection of external control input
Each terminal of the external control input of "ZERO", "WS DOWN", "DS DOWN", "DS&WS DOWN", and
"LOCK" is connected with "COM.1" by contact output or open collector as shown in the figure below.
Please connect the external control input surely as shown in the above drawing.
If neglected it causes an unexpected breakdown and the malfunction.
Please use the shielded cable for the connection of external control input, and connect the shield with the terminal E1 or E2 of this instrument.
If not connected, there is a possibility of causing the malfunction by the influence of the external noise, etc.
- Please refer the paragraph 19.
Appendix – “List of Terminal
block” as for the details.
Or,

10
2-6. Connection of external control output
Please connect the external control output for "WS DOWN", "DS DOWN", "HEALTHY", "ERROR", and
"RUN" within the range of the following specification with "COM.2" as shown in the figure below.
Specification Content
Composition of contact point Five points (Only the ERROR is b contact point, and the other is a contact point)
Rated load AC250 V or DC30 V
Ratings energizing electric current 1 A (load resistance)
Equivalent circuit
- Please connect the serge preventive element suitable for the characteristic of the connected external load for this protection of contact
point. It causes an unexpected breakdown and the malfunction like the loss and welding etc. of this contact point.
~
Or, Or,
Load Serge preventive element
HEALTHY、DS DOWN、RUN、WS DOWN
COM.2
COM.2
ERROR
- Please refer the paragraph 19.
Appendix – “List of Terminal
block” as for the details
Other Minebea Measuring Instrument manuals

Minebea
Minebea CSD-701C User manual

Minebea
Minebea CSD-801 User manual

Minebea
Minebea CSD-904-EX User manual

Minebea
Minebea MIS-202-PT User manual

Minebea
Minebea CSD-892B-07-25 User manual

Minebea
Minebea CSD-709 User manual

Minebea
Minebea CSD-892B-73 User manual

Minebea
Minebea CSD-581-74 User manual

Minebea
Minebea CSD-891B User manual

Minebea
Minebea CSD-903-73 User manual