Keison KANE3100 User manual

KANE3100
PRESSURE METER

GENERAL OPERATION
When the instrument is first switched on
using the ON/OFF key, the top line of the
display shows in sequence the date, the time,
the model version number, the battery status
and the pressure units. The bottom line of
the display shows a countdown number that
starts at 10 and reduces to zero. At zero the
meter display two lines of data.
The display shows 2 lines of data and all
data can be printed via an optional infrared
printer. The printed data can be 'live' data,
'frozen’ data or ‘stored’ data. 255 sets of
tests can be stored in non-volatile memory.
GENERAL OPERATION
When the instrument is first switched on
using the ON/OFF key, the top line of the
display shows in sequence the date, the time,
the model version number, the battery status
and the pressure units. The bottom line of
the display shows a countdown number that
starts at 10 and reduces to zero. At zero the
meter display two lines of data.
The display shows 2 lines of data and all
data can be printed via an optional infrared
printer. The printed data can be 'live' data,
'frozen data' or stored data. 255 sets of tests
can be stored in non-volatile memory.

Two lines of 20 characters can be added to
the header of printouts.
The meters are controlled using 4 buttons
The four buttons are:
PRINT/BACKLIGHT
A quick press to PRINT
Press and hold until there is a beep to toggle
the BACKLIGHT
SCROLL/MENU
Press and hold for MENU.
A quick press until there is a beep to
SCROLL the bottom line
Two lines of 20 characters can be added to
the header of printouts.
The meters are controlled using 4 buttons
The four buttons are:
PRINT/BACKLIGHT
A quick press to PRINT
Press and hold until there is a beep to toggle
the BACKLIGHT
SCROLL/MENU
Press and hold for MENU.
A quick press until there is a beep to
SCROLL the bottom line

HOLD/ZERO
A quick press to HOLD the reading and the
display flashes.
Press and hold until there is a beep to ZERO
the reading
ON/OFF
Press for ON or OFF.
The meter counts down from 10 after OFF is
pressed
Press HOLD/ZERO to cancel OFF .
The buttons with UP, DOWN and ENTER
arrows also change settings such as date,
time, and other menu items when MENU
mode has been selected.
HOLD/ZERO
A quick press to HOLD the reading and the
display flashes.
Press and hold until there is a beep to ZERO
the reading
ON/OFF
Press for ON or OFF.
The meter counts down from 10 after OFF is
pressed
Press HOLD/ZERO to cancel OFF .
The buttons with UP, DOWN and ENTER
arrows also change settings such as date,
time, and other menu items when MENU
mode has been selected.

1. BEFORE USING THE
METER FOR THE FIRST TIME:
Unscrew the bottom two screws, remove the
cover and fit a new battery in the battery
compartment. Take great care to ensure
that the battery is fitted with the correct
battery polarity. Then replace the battery
cover and screws, taking care not to over
tighten them. Always check that the meter is
working correctly after replacing the battery.
Set the meter’s correct time, date and units.
These are then stored when the meter is
switched off.
1. BEFORE USING THE
METER FOR THE FIRST TIME:
Unscrew the bottom two screws, remove the
cover and fit a new battery in the battery
compartment. Take great care to ensure
that the battery is fitted with the correct
battery polarity. Then replace the battery
cover and screws, taking care not to over
tighten them. Always check that the meter is
working correctly after replacing the battery.
Set the meter’s correct time, date and units.
These are then stored when the meter is
switched off.

2. BEFORE USING THE
METER EVERY TIME:
After switch on, check that date and time are
correct and battery power is sufficient and
the displayed units are the ones required.
SAFETY WARNING
Never connect to a pressure source where
you are not sure what the maximum
pressure might be. Always ensure that the
meter you are using is correctly rated for the
pressure that you intend to measure.
Excessive pressure (> 5 times nominal
range) can permanently damage the meter’s
pressure sensor.
2. BEFORE USING THE
METER EVERY TIME:
After switch on, check that date and time are
correct and battery power is sufficient and
the displayed units are the ones required.
SAFETY WARNING
Never connect to a pressure source where
you are not sure what the maximum
pressure might be. Always ensure that the
meter you are using is correctly rated for the
pressure that you intend to measure.
Excessive pressure (> 5 times nominal
range) can permanently damage the meter’s
pressure sensor.

3. THE MENU FUNCTIONS
The MENU structure comprises main menu
options. STORE, SETUP and PRESSURE
have a sub-menu structure. SERVICE is for
use by an approved service organisation
only.
MENU Press MENU and use the
UP button to select the
following function for
change:
STORE
SETUP
PRESSURE
SERVICE
EXIT
3. THE MENU FUNCTIONS
The MENU structure comprises main menu
options. STORE, SETUP and PRESSURE
have a sub-menu structure. SERVICE is for
use by an approved service organisation
only.
MENU Press MENU and use the
UP button to select the
following function for
change:
STORE
SETUP
PRESSURE
SERVICE
EXIT

Press ENTER to select a main menu
function
When you have selected the function to
change, press ENTER to select.
Repeat this to scroll through the menu and
select (using the ENTER key) and change
(using the UP / DOWN keys) the function.
The final, logical ENTER returns you to the
main menu display.
To exit the Menu function the final logical
ENTER is pressed. No changes are made if
the MENU function is aborted by pressing
ON/OFF.
Press ENTER to select a main menu
function
When you have selected the function to
change, press ENTER to select.
Repeat this to scroll through the menu and
select (using the ENTER key) and change
(using the UP / DOWN keys) the function.
The final, logical ENTER returns you to the
main menu display.
To exit the Menu function the final logical
ENTER is pressed. No changes are made if
the MENU function is aborted by pressing
ON/OFF.

PRESSURE
SMOOTH
OFF = Normal
ON = Readings filtered to
damp out fluctuations
RESOLVE
HIGH = maximum number
of decimal places
LOW = one less decimal
place displayed
PS UNITS
mBAR, inH2O, hPa,
mmHg, PSI, kPa, Pa,
mmH2O
PRESSURE
SMOOTH
OFF = Normal
ON = Readings filtered to
damp out fluctuations
RESOLVE
HIGH = maximum number
of decimal places
LOW = one less decimal
place displayed
PS UNITS
mBAR, inH2O, hPa,
mmHg, PSI, kPa, Pa,
mmH2O

SETUP 1
2
3
4
5
6
7
8
Temperature units -
o
F
or
o
C
AUTO-OFF adjusts the auto
power off time.
A setting of 00 sets the
meter for manual switch off
only.
Time – Uses “Military” time
as standard:
7am = 07:00, 7pm = 19:00
Date user defined format
Header – 2 lines each of up
to 20 characters that appear
on the printout.
Use ENTER and UP/DOWN.
Language
The display’s contrast.
EXIT
SETUP 1
2
3
4
5
6
7
8
Temperature units -
o
F
or
o
C
AUTO-OFF adjusts the auto
power off time.
A setting of 00 sets the
meter for manual switch off
only.
Time – Uses “Military” time
as standard:
7am = 07:00, 7pm = 19:00
Date user defined format
Header – 2 lines each of up
to 20 characters that appear
on the printout.
Use ENTER and UP/DOWN.
Language
The display’s contrast.
EXIT

STORE
VIEW:
The test number appears on
the top line.
Use UP or DOWN to
change test number.
The bottom line shows the
measured value.
Use ENTER to exit this
mode.
LOG :
Saves a set of results when
ENTER is pressed. Storage
capacity is 255 sets of
results
STORE
VIEW:
The test number appears on
the top line.
Use UP or DOWN to
change test number.
The bottom line shows the
measured value.
Use ENTER to exit this
mode.
LOG :
Saves a set of results when
ENTER is pressed. Storage
capacity is 255 sets of
results

AUTO STO :
Automatically stores readings at a preset
interval. Press ENTER then select YES or
NO. If YES set the desired time interval
between 2 and 99 seconds and then press
ENTER to activate and exit.
DEL ALL:
Clears all the memory when YES and
ENTER pressed.
EXIT
AUTO STO :
Automatically stores readings at a preset
interval. Press ENTER then select YES or
NO. If YES set the desired time interval
between 2 and 99 seconds and then press
ENTER to activate and exit.
DEL ALL:
Clears all the memory when YES and
ENTER pressed.
EXIT

BOTTOM
LINE
OPTIONS
Max Pressure
Min Pressure
Time
Date
Pressure units
Battery Status
Instrument Internal Temp
BOTTOM
LINE
OPTIONS
Max Pressure
Min Pressure
Time
Date
Pressure units
Battery Status
Instrument Internal Temp

Max Pressure is displayed as :
P 1.3
¯1.9
Min Pressure is displayed as :
P 1.3
_ 1.1
BOTTOM
LINE
OPTIONS
BOTTOM
LINE
OPTIONS
Max Pressure is displayed as :
P 1.3
¯1.9
Min Pressure is displayed as :
P 1.3
_1.1

4. MEASURING
Make sure you do not exceed the meter’s
operating specifications.
•Do not exceed the meter’s internal
temperature operating range
•Do not put the meter on a hot surface
When taking critical draft measurements
always re-zero the meter in the position you
are taking the readings and hold the meter
still during the test.
4. MEASURING
Make sure you do not exceed the meter’s
operating specifications.
•Do not exceed the meter’s internal
temperature operating range
•Do not put the meter on a hot surface
When taking critical draft measurements
always re-zero the meter in the position you
are taking the readings and hold the meter
still during the test.

Port 1 Differential Port
IR Printer Output
Always use the top left hand port (Port 1) for
taking single channel measurements (pressure
or vacuum). Only use the right hand port for
differential measurements
.
Port 1 Differential Port
IR Printer Output
Always use the top left hand port (Port 1) for
taking single channel measurements (pressure
or vacuum). Only use the right hand port for
differential measurements.

If the pressure being measured exceeds the
meter’s design range the display will show
OR
for “over-range”
When taking draft readings at very low
pressure or draft levels, for maximum
accuracy, re-zero the meter in the
orientation that it is being used. This
eliminated gravity effects on the very
sensitive pressure transducer. It is also
recommend that the meter is switched on
for at least five minutes and then re-
zeroed before taking such sensitive
measurements.
If the pressure being measured exceeds the
meter’s design range the display will show
OR
for “over-range”
When taking draft readings at very low
pressure or draft levels, for maximum
accuracy, re-zero the meter in the
orientation that it is being used. This
eliminated gravity effects on the very
sensitive pressure transducer. It is also
recommend that the meter is switched on
for at least five minutes and then re-
zeroed before taking such sensitive
measurements.

EXAMPLE PRINTOUTS
The standard
printout is:
The MAX and
MIN readings
are those
observed since
last switch on
EXAMPLE PRINTOUTS
The standard
printout is:
The MAX and
MIN readings
are those
observed since
last switch on

NOTE:
Printouts of
stored readings
will also include
the TEST NO.
below the
header.
NOTE:
Printouts of
stored readings
will also include
the TEST NO.
below the
header.

5. METER ANNUAL
RECALIBRATION AND
SERVICE
The meter should be re-calibrated and
serviced annually.
Local regulations may require more frequent
re-calibration.
5. METER ANNUAL
RECALIBRATION AND
SERVICE
The meter should be re-calibrated and
serviced annually.
Local regulations may require more frequent
re-calibration.
This manual suits for next models
5
Table of contents
Other Keison Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

janitza
janitza UMG 508 Handbook

MSA
MSA Ultima X Series Installation & maintenance instructions

Mitsubishi Electric
Mitsubishi Electric EcoMonitorPro EMU2-RD3-F instruction manual

Xylem
Xylem SENSUS PolluFlow quick guide

Sartorius
Sartorius LMA200PM operating instructions

NPI
NPI VA-10M operating instructions