Kewtech KEW 2200 User manual

kewtechcorp.com
KEW 2200
Instruction Manual
COM V/�
�
�
HOLD
OFF
A
V
V
600V
1000A
CAT. II 600V
CAT. III 300V
CAT. III
digital clamp meter
KEW 2200
Current Sensor
Trigger
Function Switch
V/ΩTerminal
COM Terminal
LED
Data Hold Key
Mode Key
Lanyard
kewtechcorp.com

kewtechcorp.com
1. Specification
Accuracy guaranteed
100% or less of range
Temperature 23 ± 5°C
Humidity 45 - 75%
AC A Auto Range
Range Accuracy
40A 0.00, 0.03-41.99A ±1.4 %rdg±6dgt
(50/60Hz)
±1.6 %rdg±6dgt
(45-65Hz)
400A 32.0-419.9A
1000A 320-1049A
Input protective current AC 1200A
AC V Auto Range
Range Accuracy
4V
0.000, 0.005-4.199V
±1.8 %rdg±7dgt
(45-65Hz)
±2.3 %rdg±8dgt
(65-500Hz)
40V 3.20-41.99V
400V 32.0-419.9V
600V 320-629V
DC V Auto Range
Range Accuracy
400mV ±0.0 -±419.9mV *1
4V ±0.320-±4.199V
±1.0%rdg±3dgt
40V ±3.20-±41.99V
400V ±32.0 -±419.9V
600V ±320-±629V
*1 : Accuracy is not guaranteed
AC V/DC V Input impedance
>100MΩ (400mV Range)
11MΩ (4V Range)
10MΩ (40/400/600V Range)
Barrier of Instrument body
Ca
p
Barrier of Test Leads
WARNING
Read through and understand the
instructions contained in this manual
before using the instrument.

kewtechcorp.com
Resistance Auto Range / Continuity
Range Accuracy
400Ω 0.0-419.9Ω
±2.0%rdg±4dgt
4kΩ 0.320-4.199 kΩ
40kΩ 3.20 -41.99 kΩ
400kΩ 32.0-419.9 kΩ
4MΩ 0.320-4.199 MΩ ±4.0%rdg±4dgt
40MΩ 3.20-41.99 MΩ ±8.0%rdg±4dgt
Cont. 0.0 -419.9Ω Buzzer threshold value
50±30Ω
Open-loop voltage
<3.4V (400Ω / Cont Range)
0.7V typ (4kΩ Range)
0.47V typ (40k - 40MΩ Range)
Input protective voltage
AC/DC 600V 10 sec
Measuring method
Dual integration
Over-range indication
OL
Measurement cycle
2.5 times per second
Applicable Standards
• IEC/EN 61010-1/ 61010-2-032/ 61010-031 /
61010-2-030
Pollution degree 2
Indoor use
Altitude up to 2000m
Current measurement section CAT.III 600V
Voltage measurement section CAT.II 600V /
CAT.III 300V
EN61326 (EMC)
In the radio-frequency electromagnetic field of
3V/m, accuracy is within five times the rated
accuracy.
EN50581 (RoHS)

kewtechcorp.com
2. Other Function
Data Hold
Press the Data Hold Key to freeze the reading.
Press the Data Hold Key again to release the
display.
Indicated“ ”on LCD
Low battery indication
Indicated “
” on LCD at 2.3±0.15V or less
Sleep Function
Automatically powers off after about 10min
of inactivity. To disable the sleep function,
power the instrument on with the Data Hold Key
pressed.
(indicated " " for about 2 seconds on LCD)
Withstand voltage
AC 5320Vrms 5sec between current sensor
and enclosure
AC 3540Vrms 5sec between circuit and enclosure
Insulation resistance
>100MΩ /1000V between enclosure and
electrical circuit
Operating temperature and humidity range
0 to 40°C 85%RH or less (no condensation)
Storage temperature and humidity range
-20 to 60°C 85%RH or less (no condensation)
Power source
DC3V×2 / R03/LR03 (AAA) ×2
Current consumption
< 3mA
Battery life
Approx. 350 hours (AC A, continuous, no load)
Dimension, Weight
190(L)×68(W)×20(D)mm
approx. 120g (including batteries)
Accessories
Test leads Model 7107A 1set
Battery R03(AAA) 2pcs
Instruction manual 1pce
Carrying case Model 9160
1pce

kewtechcorp.com
Correct Incorrect
digital clamp meter
KEW 2200
digital clamp meter
KEW 2200
Press the trigger to open the Current Sensor
and clamp the single conductor (Dia. 33mm
max.) under test.
NOTE
Measurement accuracy is guaranteed when the
measured object is placed at the centre (zone A)
of the Current Sensor. In zone B, 4% of tolerance
should be added to the specified accuracy. In
zone C, measured values should be considered as
reference values (Accuracy is not guaranteed).
3. AC A Measurement
DANGER
Never measure current while the test leads
are inserted into the V/Ωand/or COM
Terminals.
Correct Incorrect
C
C
B Center (zone A)
A
B
Centre (zone A)

kewtechcorp.com
4. AC V/DC V Measurement
DANGER
600V
Never make measurement on a circuit in
which voltage over 600V exists.
Black test lead
A
C
V
Red test lead
DC
V
Black(-) Red(+)
digital clamp meter
KEW 2200
digital clamp meter
KEW 2200
AC V DC V

kewtechcorp.com
4. AC V/DC V Measurement
DANGER
600V
Never make measurement on a circuit in
which voltage over 600V exists.
5.
Resistance (Continuity) Measurement
WARNING
Never use the instrument on an
energized circuit.
Beeps when less than 50±30Ω.
Test lead
Test lead
Resistance Continuity
LCD
/ Press
♪
digital clamp meter
KEW 2200
digital clamp meter
KEW 2200
Press

kewtechcorp.com
WARNING
• Read through and understand the
instructions contained in this manual before
using the instrument.
• Keep the manual at hand to enable quick
reference whenever necessary.
• The instrument is to be used only in its
intended applications.
• Understand and follow all the safety
instructions contained in the manual.
• It is essential that the above instructions
are adhered to.
• Failure to follow the above instructions
may impair the protection provided by the
instrument and test leads, and may cause
injury, instrument damage and/or damage
to equipment under test.
The symbol indicated on the instrument
means that the user must refer to the related
parts in the manual for safe operation of
the instrument. It is essential to read the
instructions wherever the symbol appears
in the manual.
DANGER is reserved for conditions and
actions that are likely to cause serious or
fatal injury.
WARNING is reserved for conditions and
actions that can cause serious or fatal
injury.
CAUTION is reserved for conditions
and actions that can cause injury or
instrument damage.
6. Safety Warnings
This instrument has been designed,
manufactured and tested according to IEC
61010: Safety requirements for Electronic
Measuring apparatus, and delivered in the best
condition after having passed final inspection.
This instruction manual contains warnings and
safety rules which must be observed by the
user to ensure safe operation of the instrument
and retain it in a safe condition. Therefore,
read through these operating instructions
before using the instrument.

kewtechcorp.com
DANGER
• Never make measurement on a circuit in
which voltage over AC/DC600V exists.
• Do not attempt to make measurement in the
presence of flammable gasses. Otherwise,
the use of the instrument may cause sparking,
which can lead to an explosion.
• Never attempt to use the instrument if its
surface or your hand is wet.
Measurement Category
CAT.II
Primary electrical circuits of equipment connected
to an AC electrical outlet by a power cord.
CAT.III
Primary electrical circuits of the equipment
connected directly to the distribution panel, and
feeders from the distribution panel to outlets.
CAT.IV
The circuit from the service drop to the service
entrance, and to the power meter and primary
over current protection device (distribution panel).
Current measurement section of this
instrument is designed for CAT.III 600V and
Voltage measurement section is for CAT.III
300V / CAT.II 600V respectively.
Test leads 7107A with the Cap is designed
for CAT.IV 600V / CAT.III 1000V and without
the Cap is for CAT.II 1000V.
Marks listed below are used on this instrument.
User must refer to the manual.
Instrument with double or reinforced insulation
Indicates that this instrument can clamp on
bare conductors when measuring a voltage
corresponding to the applicable measurement
category, which is marked next to this symbol.
AC
DC
Ground (Earth)
This instrument is subject to WEEE Directive
(2002/96/EC). Please contact our dealer near
you at disposal.

kewtechcorp.com
• Do not exceed the maximum allowable input of
any measuring range.
• Never open the Battery cover during a
measurement.
• To avoid electrical shock by touching the
equipment under test or its surroundings, be
sure to wear insulated protective gear.
• Never measure current while the test leads are
inserted into the input terminals.
• Barriers on the instrument body and the test
leads provide protection to keep your fingers
and hands from touching an object under test.
• Keep your fingers and hands behind the
barriers during measurement.
WARNING
• Never attempt to make measurement if any
abnormal conditions, such as broken case
and exposed metal parts are found on the
instrument or test leads.
• Verify proper operation on a known source
before use or taking action as a result of the
indication of the instrument.
•Firmly attach the Caps to the test leads
when performing measurements at CAT.
III or higher test environment.
•When the KEW2200 and the test leads
are combined and used together,
whichever is the lowest category &
voltage to earth specification applies.
• Do not rotate the Function Switch while the
test leads are being connected.
•
Do not install substitute parts or make any
modification to the instrument. For repair or
re-calibration, return the instrument to your
local distributor from where it was purchased.
CAUTION
• Use of this instrument is limited to domestic,
commercial and light industry applications.
•If equipment generating strong
electromagnetic Interference or strong
magnetic fields due to large currents exist
nearby, the instrument may not work correctly.
•Set the Function Switch to an appropriate
position before starting measurement.
•Firmly insert the test leads.
•The LCD may show some readings at the
ACV and the DCV ranges even while the test
leads are open. And, it may show some digits

kewtechcorp.com
7. Battery Replacement
CAUTION
• Do not mix old and new batteries.
• Install batteries in correct polarity as
indicated in the Battery Compartment.
(1) Set the Function Switch to "OFF" position.
(2) Unscrew and remove the Battery
Compartment Cover on the bottom of the
instrument.
(3) Replace the batteries observing correct
polarity. Use new two
R03/LR03 (AAA)
1.5V batteries.
(4) Install the Battery
Compartment and
tighten the screw.
WARNING
• Replace the batteries when a Low Battery
Voltage warning " " mark(< 2.3±0.15V)
is indicated on the LCD. Otherwise, precise
measurement cannot be made. Note that
when the battery is completely exhausted,
the LCD goes blank without showing " "
mark.
• Do not try to replace the batteries if the
surface of the instrument is wet.
• Disconnect the test leads from the object
under test and power off the instrument
before opening the Battery Compartment
Cover for Battery replacement.
instead of 0 when short-circuiting the test
leads. However, these phenomena doesn’t
affect measurement results.
•This instrument isn’t dust & waterproof.
•Keep away from dust and water.
•
Be sure to power off the instrument after use.
• When the instrument will not be in use for
a long period, place it in storage after
removing the batteries.
•
Do not expose the instrument to the direct sun,
high temperature and humidity or dewfall.
• Use a slightly damp cloth or neutral
detergent for cleaning the instrument. Do not
use abrasives or solvents.

kewtechcorp.com
Certificate
of Conformity &
Warranty Registration
This instrument has been calibrated
using equipment which has itself been
calibrated to standards traceable to
International Standards monitored by
BIPM (International Bureau of Weights
and Measures)
This certificate guarantees that the
product has been fully inspected and
conforms to all the relevant published
specifications.
Free Two Year Warranty
Kewtech's Two Year Warranty enhances
the customers' legal rights/ it covers all
manufacturing defects for a two year
period but Kewtech reserves the right to
exclude abuse or accidental damage.
To register your free guarantee simply go
to kewtechcorp.com - the link is on the
home page.
Re-calibration Service
Regular re-calibration is recommended
for this instrument. Kewtech recommends
that with normal use the instrument
is calibrated at least once in every 12
month interval.
When the instrument is due for
re-calibration return it to the address
below marked for the attention of the
Calibration Department.
Kewtech Corp Ltd
Shaw Wood Business Park
Shaw Wood Way
Doncaster DN2 5TB
t: 01302 761044
Table of contents
Other Kewtech Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Sekonic
Sekonic SPEEDMASTER L-858D Startup guide

Omega
Omega DTG-RTD100 Series user guide

AEMC
AEMC PowerPad IV quick start guide

EUTECH INSTRUMENTS
EUTECH INSTRUMENTS WATERPROOF CYBERSCAN DO 300 DISSOLVED OXYGEN... Manuel d'instructions

Leica Geosystems
Leica Geosystems TPS700 Series user manual

T&D
T&D TR41A user manual