Beko METPOINT OCV User guide

EN
-
english
Instructions for installation and operation
Measurement accessories
METPOINT
®
OCV
for the measurement in systems up to 40 bar

2 Measurement accessories METPOINT® OCV 40 bar
Dear customer,
Thank you for deciding in favour of the METPOINT
®
OCV measuring device with sampling probe.
Please read these installation- and operating instructions carefully before mounting and starting up the
METPOINT
®
OCV measuring device, and follow our directions. Perfect functioning of the METPOINT
®
OCV, and thus exact measurement and monitoring of the residual oil content in compressed air, can
only be guaranteed when the provisions and notes stipulated here are strictly adhered to.
These instructions for installation and operation are a supplement to
the installation and operating instructions METPOINT
®
OCV for the
installation and operation of the measurement accessories to be applied
in systems with an operating pressure of 40 bar.
The installation and operating instructions of the integrated
components are not substituted by these, and they shall still be applied
in their entirety. They should always be available during the operation of
METPOINT
®
OCV in systems with an operating pressure of 40 bar.

Contents
Measurement accessories METPOINT® OCV 40 bar 3
1 Contents
1
Contents ......................................................................................................................................... 3
2
Operating conditions ...................................................................................................................... 4
2.1
Operating conditions METPOINT
®
OCV ........................................................................................ 4
2.2
Operating conditions ball valve 3/8“ for METPOINT
®
OCV............................................................ 4
2.3
Operating conditions pressure reducer for METPOINT
®
OCV....................................................... 4
3
General advice ............................................................................................................................... 5
4
Safety instructions .......................................................................................................................... 6
4.1
Safety labelling ............................................................................................................................... 6
4.2
Signal words in accordance with ANSI........................................................................................... 6
4.3
General safety instructions............................................................................................................. 7
4.4
Special advice for plants under pressure in accordance with the 97/23/EC Pressure Equipment
Directive.......................................................................................................................................... 8
4.5
Special safety instructions .............................................................................................................. 9
5
Field of application and proper use of the METPOINT
®
OCV 40 bar measurement accessories 10
6
Technical data .............................................................................................................................. 11
7
Transport and set-up.................................................................................................................... 12
8
Operating principle of the METPOINT
®
OCV 40 bar .................................................................... 13
9
Description structure and components......................................................................................... 14
10
Details pressure reducer .............................................................................................................. 15
11
Installation .................................................................................................................................... 16
11.1
Mechanical installation ................................................................................................................. 16
11.2
Electrical installation / connection to the electric mains ............................................................... 16
12
Start-up......................................................................................................................................... 16
12.1
Initial start-up................................................................................................................................ 16
12.2
Restart.......................................................................................................................................... 16
13
Removal from service .................................................................................................................. 17
14
Maintenance................................................................................................................................. 18
14.1
Care.............................................................................................................................................. 18
14.2
Calibration .................................................................................................................................... 18
15
Appendix....................................................................................................................................... 19

Operating conditions
4 Measurement accessories METPOINT® OCV 40 bar
2 Operating conditions
2.1 Operating conditions METPOINT
®
OCV
Max. permissible operating overpressure PS: 16 bar g
Working pressure compressed air: 7 bar g
Min. / max. permissible operating temperature
of the sensor unit: +5°C / +55°C
Min. / max. permissible operating temperature
of the evaluation unit: +5°C / +50°C
Min./max. ambient temperature: +5°C / +45°C
Supply voltage: 230 VAC 50 Hz
See also the separate instructions for installation and operation of the METPOINT
®
OCV.
2.2 Operating conditions ball valve 3/8“ for METPOINT
®
OCV
Max. permissible operating overpressure PS: 40 bar g
Working pressure compressed air: 38 bar g
Min. / max. permissible operating temperature: -30°C / +60°C
Min./max. ambient temperature: +5°C / +45°C
2.3 Operating conditions pressure reducer for METPOINT
®
OCV
Max. permissible operating overpressure PS: 40 bar g
Max. inlet pressure compressed air: 40 bar g
Max. outlet pressure compressed air: 10 bar g
Min. / max. permissible operating temperature: -30°C / +60°C
Min./max. ambient temperature: +5°C / +45°C
See also Chapter "Details pressure reducer", page 15 and Chapter "Appendix" (pressure governor),
page 19.

Genera advice
Measurement accessories METPOINT® OCV 40 bar 5
3 General advice
Check, prior to reading the operating instructions, whether or not these instructions
correspond to the device.
Read these installation and operating instructions carefully prior to any intervention
regarding the METPOINT
®
OCV in the version for the measurement in systems with an
operating pressure of 40 bar.
The operating instructions must be accessible at all times at the place of application of the
device.
To ensure safe operation, the device must only be operated and maintained in accordance
with the indications in the operating instructions. In addition, the national and operational
statutory provisions and safety regulations, as well as the accident prevention regulations
required for the respective case of application, need to be observed during employment.
This applies accordingly when accessories are used.
Non-observance of the installation and operating instructions involves risks for persons and
systems.
System performance tests, installation, setting and service measures must only be carried
out by authorised skilled personnel.
In the event of improper use of the METPOINT
®
OCV in the version for the measurement in
systems with an operating pressure of 40 bar, or in the event of non-observance of these
instructions, warranty and liability claims are excluded.
If you have any queries regarding these instructions, please contact BEKO
TECHNOLOGIES GMBH.
This symbol stands for actions which can also be implemented by the operators, provided
that they are skilled and authorised accordingly.

Safety instructions
6 Measurement accessories METPOINT® OCV 40 bar
4 Safety instructions
In addition to these operating instructions, it is imperative to adhere to the relevant accident prevention
regulations, the generally accepted safety rules, the EU directives and country-specific standards and
provisions.
4.1 Safety labelling
Safety instructions whose non-observance represents a threat to life and physical condition or can lead
to property damage are labelled in accordance with DIN 4844 and ANSI Z535.
Safety pictograms in accordance with DIN 4844
General danger symbol
Compressed gas
Note
Observe operating instructions
Disconnect from voltage
4.2 Signal words in accordance with ANSI
Danger! Imminent hazard
Serious injury or death
Warning! Potential hazard
Serious injury or death
Caution! Imminent hazard
Injury or property damage
Notice! Potential hazard
Injury or property damage
Important! Additional advice, info, hints
No danger

Safety instructions
Measurement accessories METPOINT® OCV 40 bar 7
4.3 General safety instructions
DANGER!
Compressed gas!
Risk of serious injury or death through contact with quickly or suddenly escaping
compressed gas or through bursting plant components.
It is the obligation of the operator to ensure that the connected pressure generator is
protected against the exceeding of the maximum operating overpressure and of the
temperature limits at the METPOINT
®
OCV in the version for the measurement in systems
with an operating pressure of 40 bar.
Make sure that the system is pressureless prior to integrating the METPOINT
®
OCV in the
version for the measurement in systems with an operating pressure of 40 bar into the
compressed gas system!
Any installation works must only be carried out on the pressureless system. Only use
fittings and connecting elements that are approved for this application. It is of vital
importance to observe the specifications of the respective manufacturer.
Unpressurise the plant prior to any maintenance or repair works!
Check all connections prior to the initial start-up and retighten them if necessary.
Do not exceed the max. permissible operating pressure (see type plate).
The max. permissible operating overpressure is specified on the type plate and in the
technical data (see "Technical data" section).
Sudden impact through pressure build-up can cause damage and escape of compressed
gas.
Avoid a sudden pressure build-up by operating the valves slowly.
Maintenance, inspection and installation works must be carried out by authorised and
skilled personnel only. Prior to undertaking any measures on the METPOINT
®
OCV in the
version for the measurement in systems with an operating pressure of 40 bar, the skilled
personnel shall read up on the device by carefully studying the operating instructions. The
operator is responsible for the adherence to these provisions. The respective directives in
force apply to the qualification and expertise of the skilled personnel.
Never make structural modifications to the plant!
Use genuine spare parts and accessories only!
The general safety and accident prevention regulations apply!
DANGER!
Supply voltage!
The operation and maintenance of electrically-powered machines and devices must only be
undertaken by correspondingly qualified and authorised personnel. Prior to carrying out
service measures, no matter which type, the following notes must be considered:
Make sure that no parts of the system are energised and that the METPOINT
®
OCV cannot
be connected with the electric supply mains during service measures.
WARNING!
Unauthorised interventions!
Any modification of the components that has not been checked and approved by the
manufacturer is potentially dangerous and will void any warranty and liability on the
manufacturer's part.

Safety instructions
8 Measurement accessories METPOINT® OCV 40 bar
In the event of fire, do not use water to extinguish it (neither in the vicinity of the system nor
directed directly at the system).
4.4 Special advice for plants under pressure in accordance with the 97/23/EC Pressure
Equipment Directive
The proper use of the measuring device is the basic requirement for safe operation. Therefore, the
user must proceed as follows:
•The METPOINT
®
OCV in the version for the measurement in systems with an operating
pressure of 40 bar must only be employed within the pressure and temperature range limits
indicated by the manufacturer on the type plate.
•No welding measures must be undertaken at the housing parts of the devices.
•The maximum operating pressure indicated by the manufacturer on the type plate must not
be exceeded. It is the user's responsibility to install the appropriate safety and control
devices.
•The documents related to the METPOINT
®
OCV in the version for the measurement in
systems with an operating pressure of 40 bar (manual, operating instructions,
manufacturer's declaration etc.) must be kept safe for future reference.
•No objects whatsoever must be installed at or placed on the METPOINT
®
OCV in the
version for the measurement in systems with an operating pressure of 40 bar and the
connecting lines.
DANGER!
High pressure!
Risk of serious injury or death through contact with quickly or suddenly escaping
compressed air or through bursting plant components.
Never make structural modifications to the plant!
Use genuine spare parts and accessories only!
WARNING!
Unauthorised intervention!
Unauthorised interventions may endanger persons and plants and lead to malfunction.
Unauthorised interventions, modification and abuse of the pressure devices are prohibited.
User of the devices must observe the local and national pressure equipment regulations in
the country of installation.

Safety instructions
Measurement accessories METPOINT® OCV 40 bar 9
4.5 Special safety instructions
IMPORTANT!
Compressed air!
Exceeding the maximum pressure can lead to damaging the plant.
Observe the indications regarding the maximum pressure on the type plate!
IMPORTANT!
Operational reliability at risk!
Incorrect installation can compromise the operational reliability and affect service measures.
Do not exceed the permissible volume flow and working pressure!
Do not operate below the working pressure!
NOTE!
Service measures!
Any service measures on the METPOINT
®
OCV in the version for the measurement in
systems with an operating pressure of 40 bar must only be carried out on the de-activated,
pressureless and de-energised METPOINT
®
OCV.

Fie d of app ication and proper use of the METPOINT® OCV 40 bar
measurement accessories
10 Measurement accessories METPOINT® OCV 40 bar
5 Field of application and proper use of the METPOINT
®
OCV 40 bar
measurement accessories
The METPOINT
®
OCV measurement accessories serve to expand the operating pressure range of the
METPOINT
®
OCV to 40 bar. With this, the stationary measurement and monitoring of the vaporous
and gaseous residual oil content of the compressed air (ISO 8573:2001) in systems with an operating
pressure of40 bar is possible. Any other use is considered improper.
The manufacturer and the marketer shall not be liable for the consequences of improper use; the user
and operator alone is responsible for all risks that may arise.
The proper employment and use of the METPOINT
®
OCV according to directions requires the exact
compliance with the installation instructions, in particular in respect of:
•The place of installation, installation conditions
•The voltage supply and frequency
•Pressure, temperature and humidity of the inlet air
•Ambient temperature
•The measuring device is supplied as accepted in the factory. The user only needs to establish
the connections with the supply mains and interconnect the components, which is described in
the following chapters.
WARNING!
Employment for the prevention of dangerous conditions!
A device of the METPOINT
®
OCV series must not be employed as the sole means of
preventing dangerous conditions at machines and plants.
Machines and plants must be constructed in such a manner that defective conditions do not
lead to a dangerous situation for persons or plants.
WARNING!
Improper use!
The METPOINT
®
OCV measuring devices and the accessories offer state-of-the-art
technology and are operationally safe. Residual risks may emanate from the measuring
device and the accessories when improperly installed, employed, or operated by untrained
personnel.
The METPOINT
®
OCV measuring devices and their accessories must only be used by
qualified and authorised personnel and exclusively in accordance with the technical data.

Technica data
Measurement accessories METPOINT® OCV 40 bar 11
6 Technical data
Name
Measurement accessories METPOINT® OCV
for the measurement in systems with an
operating pressure of up to 40 bar
Manufacturer BEKO TECHNOLOGIES GMBH Germany,
41468 Neuss, Im Taubental 7
Medium
Compressed air, free from aggressive, corrosive,
caustic, toxic, flammable, and fire-accelerating
constituents
Connection sensor unit G 3/8“ internal thread
Connection measuring probe 3/8“ external thread
Connection ball valve G 3/8“ internal thread
Connection pressure reducer ¼ NPT internal thread
Max. permissible operating
overpressure
METPOINT
®
OCV 16 bar(g)
Ball valve for METPOINT
®
OCV 40 bar(g)
Pressure reducer for METPOINT
®
OCV 40 bar(g)
Operating overpressure METPOINT
®
OCV 7 bar(g)
Operating overpressure ball valve 38 bar(g)
Operating overpressure pressure
reducer
Primary pressure 38 bar(g)
Secondary pressure 7 bar(g)
Min./max. permissible operating
temperature +5 °C / +50 °C
Min./max. ambient temperature +5°C / +45°C
Supply voltage METPOINT
®
OCV 230 V AC 50 Hz

Transport and set-up
12 Measurement accessories METPOINT® OCV 40 bar
7 Transport and set-up
IMPORTANT!
Danger through damaged components!
Under no circumstances should you start up damaged components. Defective
components can impair the operational reliability and cause further damage.
The function and service life of the METPOINT
®
OCV in the version for the measurement in systems
with an operating pressure of 40 bar depends on the conditions at the place of installation. This place
must meet the following requirements:
1. Weather-proof and frost-free installation within a building.
2. The ambient temperature must not drop below or exceed the indications on the type plate.
3. The installation conditions in the separate installation and operating instructions need to be
observed and followed.

Operating princip e of the METPOINT® OCV 40 bar
Measurement accessories METPOINT® OCV 40 bar 13
8 Operating principle of the METPOINT
®
OCV 40 bar
The METPOINT
®
OCV in the version for an operating pressure of 40 bar was developed for the
detection of vaporous and gaseous hydrocarbons in compressed air that is free from aggressive,
caustic, toxic, flammable, and fire-accelerating constituents.
To determine the residual oil content in systems with an operating pressure of up to 40 bar(g), a partial
volume flow can be taken from the system, and the system pressure reduced to 7 bar(g). With the
reduced operating pressure, the medium flows through the METPOINT
®
OCV.
The measuring probe of the METPOINT
®
OCV is integrated into the pipework system of the customer.
The provided ball valve allows for the separation of the METPOINT
®
OCV from the system for
maintenance and calibration, without needing to interrupt the compressed-air supply. For the reduction
of the operating pressure, a special pressure reducer is installed.
For the tubing from the pressure reducer to the METPOINT
®
OCV, a stainless steel pipe Ø 6 x 1 mm is
used. The length of the pipework must not exceed 1000 mm.
The medium travels at approximately 2 l/min through the METPOINT
®
OCV, which means that the
"loss" of compressed air through the measurement in the withdrawn partial flow is negligible.

Description structure and components
14 Measurement accessories METPOINT® OCV 40 bar
9 Description structure and components
1 - Sensor unit METPOINT
®
OCV (component part METPOINT
®
OCV)
2 - Pressure reducer (measurement accessories)
3 -
Connection between the pressure reducer and the ball valve, 1m stainless steel
pipe 6 x 1 mm
(measurement accessories)
4 - Ball valve 3/8“ (measurement accessories)
5 - Sampling probe (component part METPOINT
®
OCV)

Detai s pressure reducer
Measurement accessories METPOINT® OCV 40 bar 15
10 Details pressure reducer
The pressure reducer is intended for a maximum primary pressure of 40 bar(g). The secondary
pressure is adjusted to 7 bar(g) and can be read at the manometer.
No further settings are required on the part of the customer.

Insta ation
16 Measurement accessories METPOINT® OCV 40 bar
11 Installation
11.1 Mechanical installation
The installation of the METPOINT
®
OCV and of the 3/8“ ball valve is carried out in accordance with the
specifications in the separate instructions for the installation and operation of the METPOINT
®
OCV.
The sensor unit of the METPOINT
®
OCV must be fixed on the part of the customer.
The pressure reducer is firmly connected with the inlet into the sensor unit and is fixed in the factory
using a holding plate.
The connection between the pressure reducer and the ball valve can be flexibly designed. For this
purpose, the supplied stainless steel pipe 6 x 1 mm and fittings must be used.
The installation can be implemented by the customer, but we recommend having the device installed
and started up by our service crew.
11.2 Electrical installation / connection to the electric mains
The electrical installation of the METPOINT
®
OCV is carried out in accordance with the specifications
in the separate instructions for the installation and operation of the METPOINT
®
OCV.
12 Start-up
12.1 Initial start-up
Please proceed according to the description in the separate instructions for installation and operation in
order to start up the METPOINT
®
OCV.
12.2 Restart
When the METPOINT
®
OCV is started up after a long rest period, please proceed as if you were
starting up the device for the first time.
Check, above all after service measures or repair, whether or not the components are tubed
completely and electrically connected. If this is not the case, carry out the installation properly.
Check all of the pipe joints for tightness.

Remova from service
Measurement accessories METPOINT® OCV 40 bar 17
13 Removal from service
DANGER!
High pressure!
Risk of serious injury or death through contact with quickly or suddenly escaping
compressed air or through bursting plant components.
Unpressurise the plant prior to any maintenance or repair works!
DANGER!
Supply voltage!
Contact with non-insulated parts carrying supply voltage involves the risk of an electric
shock resulting in injury and death.
The connection to the electric mains and the corresponding protective equipment must
comply with the legal provisions in force at the place of installation of the METPOINT
®
OCV;
the connection and installation must be carried out by correspondingly qualified skilled
persons.
NOTE!
No repair and maintenance measures under supply voltage!
De-energise all connector cables prior to repair or maintenance!
The repair and maintenance measures must only be carried out by correspondingly
qualified personnel and in accordance with the legal stipulations in force!
The general safety regulations in accordance with UVV and VDE shall apply, as well as all
relevant safety regulations of the country where the devices are operated.
.
Switch off the compressor.
Slowly decompress the system.
Disconnect the device from the power supply.

Maintenance
18 Measurement accessories METPOINT® OCV 40 bar
14 Maintenance
The METPOINT
®
OCV in the version for an operating pressure of 40 bar must be subjected to regular
inspections and service measures through authorised skilled personnel.
For the maintenance of the METPOINT
®
OCV, the specifications in the separate installation and
operating instructions apply.
Use original BEKO spare parts only.
14.1 Care
NOTE!
Aggressive cleaning agents and solvents must not be used!
14.2 Calibration
For the calibration of the METPOINT
®
OCV, the specifications in the separate installation and
operating instructions apply.

Appendix
Measurement accessories METPOINT® OCV 40 bar 19
15 Appendix

Appendix
20 Measurement accessories METPOINT® OCV 40 bar
Other manuals for METPOINT OCV
2
Table of contents
Other Beko Measuring Instrument manuals

Beko
Beko METPOINT FLM SF13 User manual

Beko
Beko METPOINT OCV User manual

Beko
Beko FLM SF53 User manual

Beko
Beko BEKOMAT 12 KW User guide

Beko
Beko DP 109 User guide

Beko
Beko METPOINT BDL compact User manual

Beko
Beko METPOINT User manual

Beko
Beko FLM SF53 Owner's manual

Beko
Beko METPOINT BDL portable User guide
Popular Measuring Instrument manuals by other brands

Master Meter
Master Meter eLinx quick start guide

JensPrima
JensPrima innoCon 6501P Operation manual

Vision & Control
Vision & Control TOB42/11.0-165-F10-WN Instructions for use

Nikon
Nikon Laser 800 - Monarch Laser 800 Rangefinder instruction manual

LEGRAND
LEGRAND IME NEMO 96 EC installation manual

ICP DAS USA
ICP DAS USA PM-3133-RCT Series user manual