Installation Note N9020-90240 7
Installation Procedure
Installation Procedure
Analyzer Information
1. Connect a power cord to the analyzer and turn on the analyzer.
2. After the analyzer has completed turning on, press System, Show, System. Make note of the
following information from the Show System screen:
Product Number ________________
Serial Number ________________
Instrument S/W Revision ________________
3. Check for the presence of the options listed below in the Show System screen. Put a check mark or
“X” after each option listed below that appears in the Show System menu.
N9020A-B40 ______
N9020A-DP2 ______
N9020A-MPB ______
N9020A-BBA ______
N9020A-508 ______
N9020A-513 ______
N9020A-526 ______
4. Refer to the data in step 2 above. If the Product Number is not N9020A, do not proceed with the
installation of this kit. This kit is to be installed only on N9020A signal analyzers.
5. Refer to the data in step 2 above. If the Serial Number prefix (the first six characters of the serial
number) is not MY5233, SG5233, or US5233 or later, do not proceed with the installation of this
kit. This kit is to be installed only on N9020A signal analyzers with later serial numbers.
6. Refer to the data in step 3 above. If option N9020A-BBA is present, do not proceed with the
installation of this kit. MXA signal analyzers with Option BBA do not have sufficient room to
install the two additional boards required for analysis bandwidths of 85 MHz and wider.
7. Refer to the data in step 3 above. If Option N9020A-B40 is not already present, an additional
upgrade kit (refer to www.agilent.com/find/mxa_upgrades for details) will also need to be installed
to first upgrade the analysis bandwidth to 40 MHz. This 40 MHz analysis bandwidth upgrade
should be performed prior to installing this 40 MHz to 85 MHz analysis bandwidth upgrade kit.
8. Refer to the data in step 3 above. If Option N9020A-MPB is not already present and either
frequency range option N9020A-508, N9020A-513, or N9020A-526 is present, it will be necessary
to upgrade the analyzer to add option N9020A-MPB before installing this kit. Refer to