Remote Control Box RC01-TOS/RC02-TOS enables a user to control the starting
and stopping operations of Kikusui’s Withstanding Voltage or Insulation Resistance
Tester from a remote location.
You can use this remote control box by simply connecting it to the REMOTE termi-
nal on the panel of the Tester.
RC01-TOS is equipped with one START switch. RC02-TOS is equipped with two
START switches, both of which you have to press simultaneously to be able to start
testing.
• The signals sent from the Remote Control Box are capable of turning ON/
OFF a high voltage. Therefore, certain situations may create extremely
hazardous conditions. Make sure not to generate high voltages carelessly
and provide adequate safety measures to prevent anyone from touching
the device under test, high voltage test lead, probe, and output terminals
when high voltage is generated.
Never use the Remote Control Box unless the above measures are ade-
quately provided.
• RC01-TOS/RC02-TOS is designed so that it can be used with the Tester. Before
using the Remote Control Box, read WARNING and CAUTION descriptions in
the Tester operation manual thoroughly.
Functions
RC01-TOS/RC02-TOS
OPERATION MANUAL
REMOTE CONTROL BOX
WARNING
NOTE
OPERATE switch
The START switch will be effective only when this
switch is ON. When this switch is OFF, a forced
stop is activated.
START switch
A test can be started by pressing this switch when
the OPERATE switch is ON and the Tester is in the
READY status.
STOP switch
This switch is used to cut off the output voltage or to
cancel the FAIL status. The switch provides a func-
tion similar to the STOP switch on a Tester.