Kilotech KWD 1000 Series User manual

Page 1 of 8
KILOTECH
Washdown Weighing Scale Service manual # 2004(WB10D)
(WB10D.PCB, FAT60K1-40B.HEX)
Model: KWD 1000 Series
FEATURES:
•Washdown scale to meet IP65 requirements
•6Key keypad
•1” black LCD display with backlight
•1/3000 External resolution
•Auto zeroing and tare function
•Kg/lb unit weight conversion
•Overload protected load cell
•Low power indication
•Rechargeable battery (DC 6V / 4.5A) or 120VAC 60 hz. power.
Before using for the first time:
•Place the scale on a stable surface.
•Adjust four feet to center the leveling bubble on the scale.
•Avoid hot direct sunlight. Do not place scale near ventilation system.
Power supply:
•Power switch: it is located on the underside of the indicator.
•Scale can be plugged directly into wall outlet for use or for charging.
•Power consumption: about DC 20mA, with backlight about DC 40mA.
•Indicator will blank-out and low battery icon flashes to indicate the battery needs recharging. In order to
prolong battery life, recharge the battery every day.
Note:
•When recharging, do not share power with other equipment and avoid electromagnetic interference from other
devices.
•If the scale will not be used forlong period of time, the scale must be recharged (internal battery) every 3 months.
Recharge scale before using.
•Red, which means charging and green means charging is finished.

Page 2 of 8
Available capacity and division:
The KWD 1000 is available in 3 capacities. Ask your dealer for more specifications
Resolution Capacity Division
1/3000 60 kg 20 g
1/3000 150 kg 50 g
1/3000 300 kg 100 g
Keypad Function:
1. ZERO (+): a)Press this key to return the display to “0” reading.
b)Press this key to deduct the empty container weight.
c)When performing as setting functions: Use this key to increase digits (0-9)
to your required value during calibration.
2. kg/lb : Press this key to change unit weight (kg /lb)
3. HI/LO : a)Preset weight key to check weighing function.
-Method 1:
-Place the desired weight value on platform.
-Preset HI/LO key to activate the preset value.
-When a weight is placed on platform and meets or exceed the preset value
set, a beep sound will occur to notify the operator.
-To clear preset alarm function, press HI/LO key again when the display
reads ‘0’.
-Method 2:
-Press HI/LO key to display LO then key-in lower weight,
-Press HI/LO key again when displays HI then key-in high weight value,
-When finished press ENTER .
-When loading reaches the pre-set range, it will be a beep sound.
-To clear preset alarm function, press HI/LO key again when the display
reads ‘0’.
b)When performing a setting function: use as a set up key.
4. NET/GROSS : a) When using TARE key to convert net weight or gross weight function.
b) Or perform as setting function: use this key as right → movement key
during calibration.
5. TARE : To deduct the weight of the container.

Page 3 of 8
1. Ensure that there is no weight on the scale platform.
2. Place an empty container on the scale platform
3. Press the TARE key to deduct the weight of the empty container.
4. Place a weight in the container to obtain the net weight of the item.
•To clear the tare function, simply press the TARE key when platform is empty.
•The Tare indicator light will be ON when TARE is activated.
•The Tare range is from capacity 0kg to the max. capacity of the scale.
6. ENTER : When finished Hi/Lo setting, press ENTER key to confirm.
Backlight function setting:
1. Press TARE for 3 seconds. Release when displays 0 . Backlight function is OFF. (originate
setting)
2. Press HI/LO for 3 seconds. Release when displays 1 . Backlight function is ON continually.
3. Press MODE for 3 seconds. Release when displays 2 . Backlight willAUTO-ON when a load is
on the platform and AUTO-OFF when the platform is empty.
Simple Calibration
1. Find the J8 jumper to switch to ‘ON’. Press HI/LO key and turn on the power at same time.
After 10 seconds the display shows P6 , release all keys.
2. Display will count backwards about 20 seconds. Scale will display calibration value. If value
needs to be changed, press NW/GW →(right movement key) and ZERO + (increase key) to
adjust the value to match your weighing standard’s value.
3. After adjusting the value, place the weighing standard on scale platform. When stable, press
HI/LO to displays P8 . About 5 seconds later there will be “BEEP” sound or indicator lights
on/off which means the calibration is completed. The scale should display same value as the
weighing standard. If there is a difference, press HI/LO key again until the difference has
been removed.
4. Power OFF and ON to reset..
Special Attentions:

Page 4 of 8
zUse maximum (full) capacity weighing standard to calibrate to avoid any difference.
Minimum is 1/6.
zDo not use weighing standards beyond maximum capacity of scale.
zKeep the scale stable during calibration.
Installations & Attentions
zUse correct power supplied, connect ground line and do not share the power with other heavy
duty equipment.
zDO NOT leave any objects on the scale for a long period of time.
zMake sure the display shows “0” before operation.
zDO NOT remove any parts from the scale while the power is on.
Insufficient Power Warning
When operating using the battery, if insufficient power level is reached, battery indicator will
appear and scale will no longer operate. Scale battery needs to be recharged.
Charging Instruction
zPlease use correct charger specification and voltage to prevent any damage.
Adapter charging specification is 6V/1A. If the charging light turns red, this means scale is
recharging. If the charging light turns green, this means scale recharging is finished. Unplug
the power cord from the wall receptacle first then unplug from the scale adapter.
zTo prolong battery life, recharge the battery whenever available or at the end of every day.
Recharge every week when operated over 48 hours. Charge 8 hours for about 80 % full.
zWhen using (6V/10A) battery, scale can be operated for 100 hours,; if 6V/4.5A(standard) can
be operated for 50 hours.
zIf the battery is unable to charge, the adapter may be damaged.
zWhen hen the battery is to low, the battery icon will appear and the screen will blank out.
When charging battery, the red charging light should come on. If the red light does not come
on, unplug and have the scale serviced.
When no display
zPlease check whether the scale is properly connected and battery fully charged..
zCheck whether the main board is damaged.
Weight calibration inner set up basic functions’chart
Displayed Function Description Set up Functions

Page 5 of 8
PA 0 Decimal or Hexadecimal numeration Hexadecimal
PP 0 Unit weight when turn on power 0:kg . 1: lb (validity: L5=0)
PL 10 Display times at per second. Range:00-99 times to display per second.(adapt A4)
PH 0 Stability Range:00-99 to select difference of stability(adaptA4)
P0 0 Kg/lb or g/oz unit weight conversion 0: kg/lb conversion
P1 0 If nearby zero unstable, value around zero
+/-? division to display zero +/- 00..99 range to display 0
P3 0 To warn or not when over max. capacity 9
divisions To warn when over
P9 1 RS232 baud rate 0:2400,1:4800,2:9600,3:19200,4:28800,5:57600.
L0 1 Select Hi/Lo buzzer range 0: outside range with buzzer, around ‘0’ no buzzer
1: inside range with buzzer
L5 Select keypad function to be stored 0: “mode” change is not stored
1: “mode” change is stored.
L- Min. graduation selection 0:one range graduation, 1:two range graduation
H1 Weight unit selection 00: kg/lb, 01:g/oz, 02:g/lb, 03:kg/kg, 04:g/g
H3 Select weight unit change is allowed or
Not. 0: mode change is allowed , 1: mode change is not
allowed.
HL 06 Zero tracing+/- ? division and ? seconds
Selection 01: 0.6 division/1 sec.
02: 0.5 division/2 sec, 03: 0.5 division/3 sec.
04: 0.5 division/4 sec., 05: 1.0 division/1 sec.
06: 1.0 division/2 sec.(setup), 07: 1.0 division/3 sec.
08: 1.0 division/4 sec., 09: 1.5 division/1 sec.
10: 1.5 division/2 sec.,11: 1.5 division/3 sec.,
12: 1.5 division/4 sec.
A3 RS232 transmit method 00: continuous sending (18times/sec.),01:press “*”key t
o
send one data,99:no use.
A4 Improve stability range 00:stability full range,90:more than 90% max. capacity,
99:more than 99% max. capacity
A7 Around 0kg +/- ? division with buzzer While setup Hi/Lo around 0kg +/- ? divisions without
buzzer(available at L0=0)
P4 Maximum capacity set up 10 … 999999(better to make 1/2 load cell capacity)
P5 Decimal point set up 0 … 0.00000
P6 Choice of min. graduation and will auto
Warm up to catch zero value. 01, 02, 05, 10, 20, 50
P7 Calibration with weighing standard With max. capacity ‘kg’to calibrate is better
Or less than 1/6 capacity.
How to set up the functions:
1. Find the J8 jumper to switch ‘ON’. Press ZERO key and turn on power at same time, when
displays PA 0 release the ZERO key.
2. When displays PA 1 press ZERO key to choose decimal or hexadecimal system then press
HI/LO key. (0 is for hexadecimal(under 600kg), 1 is for decimal(over 600kg.)
3. When displays PP 0 press ZERO key to choose unit weight after turn on then press HI/LO
key. (0 is for kg , 1 is for lb ).(P.S. available only at L5=0)
4. When displays PL 10 press NW/GW→& ZERO to select display times at per second,
range 00..99 ,then press HI/LO . (this function will adapt to A4 function).
5. When displays PH 05 press NW/GW→& ZERO to select stability difference, range 00..99
then press HI/LO . (this function will adapt toA4 function.)

Page 6 of 8
6. When displays P0 0 press ZERO to select ‘kg/lb’, ‘kg/kg’as unit, then press HI/LO
(0: kg/lb, 1: kg/kg)
7. When displays P1 0 press ZERO key to choose nearby zero ±? division to display ‘0’ then
press HI/LO ( 0: no function, 01..99 ±? division )
8. When displays P3 0 press ZERO key to choose when over max. capacity 9 divisions to warn
or not then HI/LO key. (0 is to warn when over, 1 is not to warn.)
9. When displays P9 1 press ZERO key to select RS232 baud rate transmitting.
(0:2400, 1:4800, 2:9600, 3:19200, 4:28800, 5:57600.) then press HI/LO key.
10. When displays L0 1 press ZERO to select Hi/Lo buzzer range, then press HI/LO .
0: outside range with buzzer, but around 0kg no buzzer.
1: inside range with buzzer
11. When displays L5 1 press ZERO to select key function is stored or not when switch on. Then
press HI/LO (0: Mode function is not stored, 1: Mode is stored when turn on power)
12. When displays L- 0 press ZERO to select min. graduation for one or two range then press
HI/LO . (0: one range , 1: two range.)
13. When displays H1 00 press ZERO to select kg/lb or g/oz as unit, then press HI/LO
0: kg/lb as unit, 1: g/oz as unit (use at 120g~600g, resolution: 1/12000)
02: g/lb as unit (use at 1200g~6000g)
03: kg/kg as unit, 04: g/g as unit.
14. When displays H3 1 press ZERO to select “unit” change is allowed or not. Press HI/LO (0:
weight unit change is not allowed, 1: weight unit change is allowed.)
15. When displays HL 06 press ZERO to select with zero-tracing function and time. Then press
HI/LO .
*02: 0.5 division/2 sec, 03: 0.5 division/3 sec.
*04: 0.5 division/4 sec., 05: 1.0 division/1 sec.
*06: 1.0 division/2 sec., 07: 1.0 division/3 sec.
*08: 1.0 division/4 sec., 09: 1.5 division/1 sec.
*10: 1.5 division/2 sec.,(originate setup)
*11: 1.5 division/3 sec., 12: 1.5 division/4 sec.
16. When displays A3 press ZERO to select transmitting method through RS232, then press
HI/LO . (00:continuous sending 18times/sec., 01: press “*” key to send one data, 99: no use
function.)
17. When displays A4 press ZERO to select stability range, then press HI/LO .
(00: at full range, 90: more than 90% of max. capacity, 99: more than 99% of max. capacity).
18. When displays A7 04 press ZERO to select nearby 0kg +/- ? division with buzzer. Then press
HI/LO. (when at Hi/Lo mode nearby 0kg less than +/- ? division with buzzer.
19. When displays P4 , which means all setup finished. Switch off code and turn on and off scale
again.

Page 7 of 8
How to set up capacity, graduation & weighing standard calibration
After each set up, you may turn off and on the scale to operate (do not need to go through to the end) .
1. Find the unlock code J8 jumber on mainboard to let it ‘ON’, then press NW/GW→and turn on
power at same time.
2. When displays P4 let go the key then displays maximum capacity 30000
press NW/GW→ & ZERO+ keys to change the figures (0…99999) then press HI/LO to set up.
3.When displays P5 after one second then decimal point location 30.000 press →NW/GW &
ZERO+ keys to change decimal point location (0…0.0000) then press HI/LO.
4. When displays P6 then minimum graduation 05 press ZERO+ key to change the
Graduations (1…50) then press then press HI/LO
5. While display will show 99999 reduced backward to 00000, it will catch the zero point value
automatically and 20 seconds later, displays P7 then weighing standard’s value 30.000
press →NW/GW & ZERO+ keys to change the figures (10…99999) to adjust the value to
match weighing standard then place the weighing standard (use kg only) on scale & press
SET after stable.
6.When displays P 8 after 5 seconds later, there’ll be pi-pi sound or indicator lights off & on,
which means setting finished, 2 seconds later displays weighing standard’s value 30.000 then
switch the code to ‘Off’then turn off power and turn on the scale.
7. If there’s any difference, press HI/LO key to adjust till no difference.
Zero Point Value Displayed
1. When turn on scale, the zero value is showed before displayed 0kg, it must be less 710000.
2. Or you may press ZERO key 2 seconds, the zero point value will be displayed,
The value range is about 300000 … 500000. You may re-press ZERO key twice to restore the
displayed value.
3. Zero value range is best on 40000 ~ 470000, up to two multiple of max. capacity.
(ex. A max. capacity scale 30kg, load 30kg standard and press ZERO you may load another
30kg standard, there might be with difference (about 1 to 4 divisions) which is to prevent
improper loading to raise the zero value or unable to reach max. capacity situation.
4. The difference between No loading and full loading is at 1mV/V, best is at 200000.
R8=150 Ohm, R9, R10=18 K ohm, Y1=40 MHz,A/D / 2, 1 mV/V
Displayed How to get rid of the problems
PPP1 Means load cell signal (S+, S-) is incorrectly installed. Reverse connections.
PPP2 Means display resolution is bigger than 1/30010,. Adjust PLor PH value.
PPP3 Means that doing calibration, no weight or insufficient weight was applied to the
platform causing low load cell output voltage and incomplete calibration. Re-do
calibration process.
ERRA Means that during calibration, the weights used exceeded the value required. Re-do

Page 8 of 8
calibration process.
---OL--- Means over loading (OLat 9 divisions over max. capacity) or load cell is disconnect.
Display Value drifts….. Increase the value of PL and PH settings. This will increase stability but decrease
display speed.
No BEEPafter calibration The load cell signal (S+, S-) is incorrectly installed. Check and re-install.
Indicator lights blink If this happens during calibration, the load cell connections may be installed incorrectly.
No load on scale but not at zero.
Press TARE if in tare mode;
If not in tare mode, press ZERO if within zero range.
Change the P1 setting.
Setup memory disappears. May be caused by outside interference. Correct by doing one of the following.
1. 93LC46 In this IC 4 & 5 pin to weld a 4.7K resistor. Then re-setup max. capacity and graduation and
calibration.
2. The load cell connection wire is disconnected, use tape to protect it.
3. U13(93LC46) Pin 8 power connect C30 + end; U13 pin 5 connect to C30 – end.
4. Transformer side (AC7…12V) or charging hole to add a EMI filter core.
5. Replace memory IC 93LC46 (3V)
6. Add a stabilizer.