manuals.online logo
Brands
  1. Home
  2. •
  3. Brands
  4. •
  5. Kimo
  6. •
  7. Data Logger
  8. •
  9. Kimo KISTOCK KT 320 User manual

Kimo KISTOCK KT 320 User manual

Other manuals for KISTOCK KT 320

2

This manual suits for next models

5

Other Kimo Data Logger manuals

Kimo 300 ST Manual

Kimo

Kimo 300 ST Manual

Kimo KISTOCK 320 Series User manual

Kimo

Kimo KISTOCK 320 Series User manual

Kimo KT 220 User manual

Kimo

Kimo KT 220 User manual

Kimo KISTOCK KT 320 User manual

Kimo

Kimo KISTOCK KT 320 User manual

Kimo KNT 320 User manual

Kimo

Kimo KNT 320 User manual

Kimo KISTOCK KT 320 User manual

Kimo

Kimo KISTOCK KT 320 User manual

Kimo KNT-310 User manual

Kimo

Kimo KNT-310 User manual

Kimo KT 120 User manual

Kimo

Kimo KT 120 User manual

Popular Data Logger manuals by other brands

UGCS SkyHub user manual

UGCS

UGCS SkyHub user manual

Avisaro 4.0 Series user manual

Avisaro

Avisaro 4.0 Series user manual

DDB C3 manual

DDB

DDB C3 manual

TESTO 176 instruction manual

TESTO

TESTO 176 instruction manual

HBM SOMAT XR CX22B-R operating manual

HBM

HBM SOMAT XR CX22B-R operating manual

Agtrix 380i manual

Agtrix

Agtrix 380i manual

Beijing ART Tech PCI2320 user manual

Beijing ART Tech

Beijing ART Tech PCI2320 user manual

Athena GET SL1 user manual

Athena

Athena GET SL1 user manual

Motec Systems C185 user manual

Motec Systems

Motec Systems C185 user manual

Beijing ART Tech USB2832 user manual

Beijing ART Tech

Beijing ART Tech USB2832 user manual

HT Italia XL421 user manual

HT Italia

HT Italia XL421 user manual

Racelogic VBOX LapTimer quick start guide

Racelogic

Racelogic VBOX LapTimer quick start guide

Edu-Logger 900-240 Guide

Edu-Logger

Edu-Logger 900-240 Guide

Elma DT173 manual

Elma

Elma DT173 manual

MadgeTech PRTemp140 Product user guide

MadgeTech

MadgeTech PRTemp140 Product user guide

Comet U0121M user guide

Comet

Comet U0121M user guide

pico Technology PT-104 user guide

pico Technology

pico Technology PT-104 user guide

Sensus DIAVASO user manual

Sensus

Sensus DIAVASO user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.

USER MANUAL
CLASS 320 KISTOCK
KT 320 / KCC 320 / KP 320-321
KPA 320 / KTT 320
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
Table of contents
1 Safety instructions........................................................................................................................................................4
1.1 Precautions for use...............................................................................................................................................4
1.2 Symbols used....................................................................................................................................................... 4
1.3 Directive 2014/53/EU........................................................................................................................................... 4
2 Presentation of t e device............................................................................................................................................5
2.1 Use...................................................................................................................................................................... 5
2.2 Applications......................................................................................................................................................... 5
2.3 References...........................................................................................................................................................5
2.4 Description of t e device......................................................................................................................................6
2.5 Description of keys...............................................................................................................................................6
2.6 Description of LEDs..............................................................................................................................................6
2.7 Connections.........................................................................................................................................................6
2.8 Mounting.............................................................................................................................................................6
3 Tec nical features......................................................................................................................................................... 7
3.1 Tec nical features of t e devices...........................................................................................................................7
3.2 Programmed units................................................................................................................................................9
3.3 Free units............................................................................................................................................................. 9
3.4 Features of t e ousing........................................................................................................................................9
3.5 Features of optional probes................................................................................................................................10
3.6 Dimensions (in mm)............................................................................................................................................11
3.6.1 Devices...................................................................................................................................................... 11
3.6.2 Wall mount (in option)................................................................................................................................11
3.7 Guarantee period............................................................................................................................................... 11
4 Use of t e device....................................................................................................................................................... 12
4.1 Display...............................................................................................................................................................12
4.2 Function of LEDs................................................................................................................................................12
4.3 Function of keys.................................................................................................................................................13
4.3.1 Groups organisation...................................................................................................................................15
4.3.2 Measurements scroll...................................................................................................................................15
4.4 PC communication.............................................................................................................................................16
4.5 Conguration, datalogger download and data processing wit t e KILOG software..............................................16
5 Wireless connection function......................................................................................................................................17
6 Maintenance..............................................................................................................................................................17
6.1 Replace t e batteries..........................................................................................................................................17
6.2 Device cleaning..................................................................................................................................................17
6.3 Safety lock wall mount wit padlock...................................................................................................................17
7 Calibration................................................................................................................................................................. 18
7.1 KCC 320: perform a CO2 measurement verication..............................................................................................18
7.2 KP 320 – KP 321: perform an auto-zero..............................................................................................................18
8 Accessories................................................................................................................................................................ 19
9 Troubles ooting..........................................................................................................................................................19
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
1 Safety instructions
1.1 Precautions for use
Please always use t e device in accordance wit its intended use and wit in parameters described in t e tec nical features
in order not to compromise t e protection ensured by t e device.
1.2 Symbols used
For your safety and in order to avoid any damage of t e device, please follow t e procedure
described in t is user manual and read carefully t e notes preceded by t e following symbol:
T e following symbol will also be used in t is user manual:
Please read carefully t e information notes indicated after t is symbol.
1.3 Directive 2014/53/EU
Hereby, Sauermann Industrie SAS declares t at t e radio equipment type Kistock 320 is in compliance wit Directive
2014/53/EU. T e full text of t e EU declaration of conformity is available at t e following internet address:
www.sauermanngroup.co m
4
Safety instructions
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
2 Presentation of the device
2.1 Use
T e KISTOCK class 320 dataloggers allow t e measurement of several parameters:
•KT 320: internal measurement of temperature wit two universal inputs for probe
•KCC 320: internal measurement of temperature, umidity, atmosp eric pressure and CO2
•KP 320 – KP 321: internal measurement of differential pressure wit two measuring ranges
•KPA 320: internal measurement of temperature, ygrometry and atmosp eric pressure
•KTT 320: model wit four t ermocouple inputs
Communication between device and PC is carried out wit a USB cable wit a micro-USB female connector.
T e low-energy wireless connection (possibility to deactivate t is function) allows to communicate wit smartp ones and
tablets, working wit Android and IOS.
2.2 pplications
T e KISTOCK dataloggers are ideal for different parameters surveillance (temperature, ygrometry, lig t, current, voltage,
impulsion, relative pressure…). T ey ensure t e traceability in t e food industry environment as well as t ey validate t e
proper functioning of industrial installations.
2.3 References
Device
reference Display
Internal sensors External sensors
arameters Number of
recording points
Number Type Numbe
rType
KT 320
Yes
1 Temperature 2 Inputs for SMART
PLUG* probes
Temperature, ygrometry,
current, voltage, impulsion
2 000 000
KCC 320 4
Temperature, ygrometry,
atmosp eric pressure,
CO2
-
Temperature, ygrometry,
atmosp eric pressure, CO2
KP 320
KP 321 1 Differential pressure Differential pressure
KPA 320 3 Temperature, ygrometry,
atmosp eric pressure
Temperature, ygrometry,
atmosp eric pressure
KTT 320 - 4
Inputs for
t ermocouple
probes
Temperature
* Input which allows to plug different compatible SMART PLUG probes see optional probes and cables page 10.
resentation of the device
5
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
2.4 Description of the device
2.5 Description of keys
OK key: allows to start or stop t e dataset or c ange of scrolling group, see page 13.
Selection key: allows t e functions scroll, see page 13.
2.6 Description of LEDs
2.7 Connections
T e communication between t e device and t e computer is carried out via an USB cable and wit t e female micro-USB
connector.
2.8 Mounting
T e class 320 KISTOCK as magnetic mountings, so you can x it easily.
6
resentation of the device
“Selection” key “OK” key
Micro-USB connector
KT 320: 2 mini-DIN
connections
KCC 320 and
K A 320
K 320 and K 321: 2
pressure connections
KTT 320: 4 mini-t ermocouple
connections
Alarm LED Operating LED
Alarm LED Operating LED
Magnetic mountings
Display
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
3 Technical features
3.1 Technical features of the devices
KT 320 KTT 320
Units displayed
°C, °F, °Ctd, °Ftd,%RH, mV, V, mA, A
Programmed and free units are also
available1 (see table page 9)
°C, °F
Resolution 0.1°C, 0.1°F, 0.1%RH, 1 mV, 0.001 V,
0.001 mA, 0.1 A 0.1°C, 0.1°F
External input Female micro-USB connector
Input for probe 2 SMART PLUG2 inputs 4 inputs for t ermocouple probes
(K, J, T, N, S)
Internal sensor Temperature -
Type of sensor CTN T ermocouple
Measuring range Measuring range of t e internal sensor3:
From -40 to +70°C
K: from -200 to +1300°C
J: from -100 to +750°C
T: from -200 to +400°C
N: from -200 to +1300°C
S: from 0 to 1760°C
Accuracy4±0.4°C from -20 to 70°C
±0.8°C below -20°C
K, J, T, N: ±0.4°C from 0 to 1300°C
±(0.3% of t e reading +0.4°C) below 0°C
S: ±0.6°C
Setpoints alarm 2 setpoint alarms on eac c annel
Frequency of
measurements From 1 second to 24 ours
Operating temperature From -40 to +70°C From -20 to 70°C
Storage temperature From -40 to +85°C
Battery life 7 years5
European directives 2011/65EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU
1
Some units are available only wit optional probes.
2
Input w ic allows to plug different SMART PLUG compatible probes: see optional probes and cables page 10.
3
Ot er measuring ranges are available according to t e connected probe: see optional probes and cables page 10.
4
All accuracies indicated in t is document were stated in laboratory conditions and can be guaranteed for measurement carried out in t e same conditions, or carried out wit
calibration
compensation.
5 On t e basis of 1 measurement eac 15 minutes at 25°C.
Technical features
7
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
KCC 320 K A 320
Units displayed °C, °F, %RH, Pa, ppm °C, °F, %RH, Pa
Resolution 0.1°C, 1 ppm, 0.1%RH, 1 Pa 0.1°C, 0.1%RH, 1 Pa
External input Micro-USB female connector
Input for probe - -
Internal sensor Hygrometry, temperature, atmosp eric pressure,
CO2
Hygrometry, temperature, atmosp eric pressure
Tolerated
overpressure - 1260 Pa
Type of sensor
Temperature and ygrometry: capacitive
Atmosp eric pressure: piezo-resistive
CO2: NDIR
Temperature and ygrometry: cpacitive
Atmosp eric pressure: piezo-resistive
Measuring range
Temperature: from -20 to 70°C
Hygrometry: from 0 to 100%RH
Atmosp eric pressure: from 800 to 1100 Pa
CO2: from 0 to 5000 ppm
Temperature: from -20 to 70°C
Hygrometry: from 0 to 100%RH
Atmosp eric pressure: from 800 to 1100 Pa
Accuracy*
Temperature: ±0.4°C from 0 to 50°C
±0.8°C below 0°C or above 50°C
Humidity**: ±2%RH
from 5 to 95%, 15 to 25°C
Atm. pressure: ±3 Pa
CO2: ±50 ppm ±3% of t e reading
Temperature: ±0.4°C from 0 to 50°C
±0.8°C below 0°C or above 50°C
Humidity**: ±2%RH
from 5 to 95%, 15 to 25°C
Atm. pressure: ±3 Pa
Setpoints alarm 2 setpoint alarms on eac c annel
Frequency of
measurements
From 1 minute to 24 ours
(15 sec in on-line mode) From 1 second to 24 ours
Operating
temperature From 0 to +50°C
Storage
temperature From -40 to +85°C
Battery life 3 years*** 7 years***
European
directives 2011/65EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU
*
All accuracies indicated in t is document were stated in laboratory conditions and can be guaranteed for measurement carried out in t e same conditions, or carried out wit
calibration compensation.
** Factory calibration uncertainty: ±0.88%RH. Temperature dependence: ±0.04 x (T-20) %RH (if T<15°C or T>25°C)
*** On t e basis of 1 measurement eac 15 minutes at 25°C.
8
Technical features
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
K 320 K 321
Units displayed Pa
Measuring range ±1000 Pa ±10000 Pa
Resolution 1 Pa
Accuracy* ±0.5% of t e reading ±3 Pa ±0.5% of t e reading ±30 Pa
Tolerated overpressure 21 000 Pa 69 000 Pa
External input Micro-USB female connector
Input for probe 2 pressure connections
Internal sensor Differential pressure
Setpoints alarm 2 setpoint alarms on eac c annel
Frequency of measurement From 1 second to 24 ours
Operating temperature From 5 to +50°C
Storage temperature From -40 to 85°C
Battery life 7 years**
European directives 2011/65EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU
* All accuracies indicated in t is document were stated in laboratory conditions and can be guaranteed for measurement carried out in t e same conditions, or carried out wit
calibration
compensation.
** On t e basis of 1 measurement eac 15 minutes at 25°C.
3.2 Programmed units
T e available programmed units for t e KT 320 and KTT 320 KISTOCK are t e following:
•m/s
•fpm
•m³/s
•°C
•°F
•%HR
•K
•PSI
•Pa
•mmH2O
•inWg
•kPa
•mmHg
•mbar
•g/Kg
•bar
•Pa
•daPa
•°Ctd
•°Ftd
•°Ctw
•°Ftw
•kj/kg
•mA
•A
•mV
•V
•Hz
•tr/
min
•rpm
•ppm
3.3 Free units
For t e free units creation, please see t e KILOG software user manual.
3.4 Features of the housing
Dimensions 110.2 x 79 x 35.4 mm
Weight KT 320, KCC 320, KP 320, KP 321: 206 g. KTT 320 and KPA 320: 200 g.
Display 2 lines LCD screen. Screen size: 49.5 x 45 mm
2 indication LEDS (red and green)
Control 1 OK key
1 Selection key
Material Compatible wit food industry environment
ABS ousing
rotection
IP65: KT 320, KP 320 and KP 321*
IP 54: KTT 320**
IP40: KCC 320 and KPA 320
C communication Micro-USB female connector
USB cable
Battery power supply 2 double AA lit ium 3.6 V batteries
Environmental conditions of use
Air and neutral gases
Hygrometry: en conditions de non-condensation
Altitude: 2000 m
* Wit t e pressure connectors plugged for KP 320 and KP 321.
** Wit all t e t ermocouple probes connected.
Technical features
9
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
3.5 Features of optional probes
Reference Description Measuring range
External or ambient t ermo- ygrometric probes
KITHA Interc angeable ygrometry and ambient temperature probe Hygrometry: from 0 to 100%HR
Temperature: from -20 to +70°C
KITHP-130 Remote interc angeable ygrometry and temperature probe
KITHI-150 Remote interc angeable ygrometry and temperature probe Hygrometry: from 0 to 100%HR
Temperature: from -40 to +180°C
General use or insertion Pt 100 temperature probes
KIRGA-50 / KIRGA-
150 IP65 immersion probe (50 or 150 mm) From -40 to +120°C
KIRAM-150 Ambient probe 150 mm
KIRPA-150 Penetration probe IP65
From -50 to +250°C
KIPI3-150-E IP68 penetration probe wit andle
KITI3-100-E IP68 penetration probe wit T- andle
KITBI3-100-E IP68 penetration probe wit corkscrew andle
KIRV-320 Velcro probe From -20 to +90°C
KICA-320 Smart adapter for Pt100 probe From -200 to +600°C according to
t e probe
Input current, voltage and impulsion cables
KICT Voltage input cable 0-5 V or 0-10 V
KICC Current input cable 0-20 mA or 4-20 mA
KICI Pulse input cable
Maximal voltage: 5 V
Type of input: TTL frequency counting
Maximal frequency: 10 kHz
Maximum number of recordable
points: 20000 points
Clamp-on ammeters
KIPID-50 Ammeter clamp from 0 to 50 A, frequency range from 40 to
5000 Hz From 0 to 50 AAC
KIPID-100 Ammeter clamp from 0 to 100 A, frequency range from 40 to
5000 Hz From 1 to 100 AAC
KIPID-200 Ammeter clamp from 0 to 200 A, frequency range from 40 to
5000 Hz From 1 to 200 AAC
KIPID-600 Ammeter clamp from 0 to 600 A, frequency range from 40 to
5000 Hz From 1 to 600 AAC
T ermocouple probes
All t e t ermocouple temperature probes for t e KTT 320 KISTOCK ave a class 1 sensitive element as per IEC 584-1, 2
and 3 standards.
For more details about t e available t ermocouple probes, please see t e ''T ermocouple probes'' datas eet.
For more details, please see the “Measuring probes for KT 320 KISTOCK” and “Thermocouple probes”
datasheets.
10
Technical features
All t e probes for t e KT 320 KISTOCK ave t e SMART LUG tec nology. An automatic
recognition and t e adjustment make t em 100% interc angeable.
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner
Connect a probe:
➢Open the mini-DIN connection cap on the bottom of the KISTOCK.
➢Connect the probe in such a way the mark on the probe is in front of the user.
3.6 Dimensions (in mm)
3.6.1 Devices
3.6.2 Wall mount (in option)
3.7 Guarantee period
KISTOCK dataloggers ave 1-year guarantee for any manufacturing defect (return to our After-sales service required).
Technical features
11
KT 320 KCC 320 / K A 320 K 320 / K 321
KTT 320
Mark
[email protected] www.digiparts.ch
Ihr Schweizer Industriepartner