Kimo KNT 320 User manual

KNT 320 data collector
Data storage and recovery
for KISTOCK 2015 dataloggers


Table of contents
1 Safety...................................................................................................................................................4
1.1 Precautions for use.......................................................................................................................4
1.2 Symbols used...............................................................................................................................4
2 Device presentation..............................................................................................................................4
2.1 Use................................................................................................................................................4
2.2 Device description.........................................................................................................................4
2.3 Description of keys........................................................................................................................4
2.4 Connections..................................................................................................................................5
3 echnical features................................................................................................................................5
3.1 Data collector................................................................................................................................5
3.2 Housing.........................................................................................................................................5
3.3 Guarantee.....................................................................................................................................5
4 Device use............................................................................................................................................6
4.1 Functioning principle.....................................................................................................................6
4.2 Display..........................................................................................................................................6
4.3 Switch on/off the data collector.....................................................................................................6
4.4 Auto-off..........................................................................................................................................6
4.5 Display used storage capacity......................................................................................................6
4.6 Kistock – data collector interface..................................................................................................7
4.6.1 Download the Kistock data on the data collector..................................................................7
4.6.2 Reconfigure the Kistock and start an other measurement dataset.......................................8
4.6.3 Summary of functions..........................................................................................................10
4.7 Data collector – PC interface with Kilog 2015 software..............................................................11
4.7.1 Download data collector on PC...........................................................................................11
4.7.2 Display data collector state, delete memory and set date and time....................................11
4.7.3 Update the data collector.....................................................................................................12
4.8 Data collector – PC interface without Kilog 2015 software.........................................................13
5 Maintenance.......................................................................................................................................14
5.1 Recharge the battery..................................................................................................................14
5.2 Device cleaning...........................................................................................................................14
6 Accessories........................................................................................................................................14
7 roubleshooting..................................................................................................................................14

1 Safety
1.1 Precautions for use
Please always use the device in accordance with its intended use and within parameters described in the technical features
in order not to compromise the protection ensured by the device.
1.2 Symbols used
For your safety and in order to avoid any damage of the device, please follow the procedure
described in this user manual and read carefully the notes preceded by the following symbol:
The following symbol will also be used in this user manual:
Please read carefully the information notes indicated after this symbol.
2 Device presentation
2.1 Use
The KNT 320 data collector allows to download and store the recorded data from Kistock 2015 devices in order to
download them on a computer. The KNT 320 also allows to send a new configuration and start an other measurement
dataset without using the Kilog 2015 software.
•Compatible with Kistock 2015 range devices (except class 20)
•Up to 20 000 000 points stored
•Possibility to send a new configuration and start an other measurement dataset
•Integrated male U B connector
•Fast data transfer ( minute maximum)
•Rechargeable battery by USB connection on PC
2.2 Device description
2.3 Description of keys
OK key: allows to validate functions.
ON/OFF – election key: press once to start the device / press for 5 seconds to switch it off.
Press successively to scroll functions.
4Device presentation
“ON/OFF – election” key “OK” key
Display
Unlosabe PC USB connection cap

Dimensions
2.4 Connections
3 Technical features
3.1 Data collector
torage capacity Up to 20 000 000 points
peed of data transfer From the Kistock to the data collector: 400 values/second
From the data collector to the PC: 000 values/second
Operating temperature From 0 to +50°C
torage temperature From -20 to +70°C
Charging temperature From 0 to +45°C
Autonomy According to downloading frequency
Compatible software Kilog 20 5
3.2 Housin
Dimensions 00 x 42.5 x 5.9 mm
Weight 53 g
Display line LCD screen, 28 x 28 pixels
Dimensions: 32 x 25.5 mm
Control OK key
ON/OFF – Selection key
Material Compatible with food industry environment
ABS housing
Protection IP65
Communication
Data collector – PC: integrated A male USB connector
Data collector – Kistock: female USB / male micro-USB adapter
(supplied with the device, reference: KNT-A)
Battery power supply lithium polymer battery
Charging time from about 7 minutes (50%) to 2 hours ( 00%)
Environment Air and neural gases
3.3 Guarantee
KNT 320 data collector have -year guarantee for any manufacturing defect (return to our After-sales service required).
Technical features 5
Male USB connector Unlosable cap
Data collector – PC connection
Data collector – Kistock connection
Female USB / male micro-USB adapter, supplied with device
(reference: KNT-A)
Side view
Front view
Top view

4 Device use
4.1 Functionin principle
4.2 Display
4.3 Switch on/off the data collector
➢To switch on the data collector, press key
The data collector switches on: it displays all the digits, then the home screen which indicates “on”.
➢To switch off the data collector, press key for 5 seconds
The data collector displays “off” and switches off.
4.4 Auto-off
To save battery power, after 5 minutes without activity, the data collector automatically switches off.
4.5 Display used stora e capacity
6Device use
Store the Kistock datasets
on the data collector
Send a new configuration on the Kistock devices
and start an other dataset
When the icon appears: data transfer is in progress
REC
Flashing slowly: data collector reached etween 70 and 80% of its storage capacity
Flashing quickly: data collector reached etween 80 and 90% of its storage capacity
Constant: data collector reached etween 90 and 100% of its storage capacity
FULL
Constant: indicates that the attery has to e recharged
Flashing: charging attery
Disappears when the attery is charged
BAT
Indicates that the data transfer is done
END
Transfer datasets stored on
the data collector
to the PC

If the dataset is in progress:
The Kistock displays *
The data collector displays ”Stop YES”.
Confirm dataset stop by pressing OK key (to cancel, press elect.,
“ top NO” is displayed then press OK).
The data collector displays the data transfer progression: 0-100%.
The dataset recorded by the Kistock is stored on the data
collector.
4.6 Kistock – data collector interface
4.6.1 Download the Kistoc data on the data collector
Procedure to download data:
When data are transferred from the Kistock to the data collector, these measurement data are copied. Only the
sending of a new configuration with the data collector or the Kilog 20 5 software allows to start an other dataset
(see procedure described page 9).
Device use 7
REC
If the data collector does not detect any point to download:
The data collector displays “No Pts” error message.
This message indicates that the dataset did not start already or that
the first point is not recorded yet.
Check that the dataset started:
If the Kistock is still waiting for dataset to start, it displays
flashing icon*.
Press the data collector OK key to start the dataset.
The data collector displays “Start Yes”.
Press OK to confirm (to cancel, press elect., “ top no” displays
then press OK).
If the dataset is finished:
The Kistock displays *
The data collector displays the data transfer progression: 0-100%
The dataset recorded by the Kistock is stored on the data
collector.
1- Plug the data
collector on the
Kistock
2- witch on the
data collector.
“On” displays.
3-a. tore the dataset on
the data collector
END
3-b. “ top YE ”
top or continue the
dataset
3-c. “No Pts”
No points detected
REC
1- witch on the data
collector. “On” is displayed.
2- Display used
storage capacity
Or wait
5 seconds
3- “On”
Back to home screen.
* only for models with display

4.6.2 Reconfigure the Kistoc and start an other measurement dataset
Once the measurement dataset is transferred to the data collector (see the downloading procedure page 7), it is possible to
start an other dataset directly from the data collector.
Two recording configurations can be sent:
•“Loop”: values are recorded continuously and when the storage capacity is full, the first recorded values are
overwritten by the new ones.
OR
•“Total memory”: values are recorded until the storage capacity of the device is reached.
The configuration which is sent depends on the Kistock last configuration
Initial configuration: top type Configuration sent by the data collector: top type
Date/time
Total memory
Duration
Records count
Total memory
Loop Loop
Other configuration parameters (recording intervals, stop by button, etc.) remain unchanged.
If a configuration different from “Loop” or “Total memory” recording mode is required, please use the Kilog 2015
software: for more details about device configuration via software, please see the Kilog 2015 software user
manual.
8Device use

Procedure to send a new configuration:
Once the dataset is transferred to the data collector (see downloading procedure described page 7, screens to 3):
When a new configuration is sent on the Kistock with the data collector, two types of start can be sent: start by PC
or start by button. The type of start which is sent depends on an immediate start or not of the dataset with the
data collector.
Device use 9
No new configuration is sent:
The Kistock displays *.
The data collector displays “On”.
No new configuration has been sent on the Kistock.
To start a new measurement dataset, reconfigure the device
with the data collector, following the opposite procedure, or
with the Kilog 2015 software (please see the Kilog 20 5 software
user manual).
A new configuration is sent:
The Kistock displays flashing icon*.
Device is waiting for start.
The data collector displays “On”.
A new configuration has been sent but
the dataset did not start.
The Kistock is automatically configured with
a type of start by button: the new dataset
can be started by pressing the OK key of the
Kistock.
An other measurement dataset starts:
The data collector displays the new
configuration transfer progression: 0-100%
then displays “On”.
The Kistock displays *
A new configuration has been sent and
an other measurement dataset starts.
The Kistock is automatically configured
with type of start by PC.
3-a. Dataset is
stored on the data
collector
4- “Conf YE ”
end a new
configuration
5- “ tart YE ”
tart an other
dataset
REC
REC
END
5- “ tart NO”
Do not start an
other dataset
4- “Conf NO”
Do not send a new
configuration
6- An other
measurement
dataset starts
6- “On”
Back to
home screen
6- “On”
Back to
home screen
* only for models with display

4.6.3 Summary of functions
0 Device use
Dataset is in progress
“Stop YES” is displayed.
Allows to stop the dataset.
Plug the data collector
on the Kistock
witch on the data
collector.
“On” is displayed.
Dataset is finished
Transfer progression is
displayed: 0-100/% .
Dataset is stored on the data
collector.
“Stop NO” is displayed.
Allows not to stop the
dataset.
The data collector does not detect
any point to download. The error
message “No Pts” is displayed. The
dataset did not start already or the first
point has not been recorded yet.
“Conf YES” is displayed.
Allows to send a new
configuration on the Kistock.
“Start NO” is
displayed
Allows not to start
the measurement
dataset.
“Conf NO” is displayed.
Allows not to send a new
configuration.
“Start YES” is displayed.
Allows to start the
measurement dataset.
An other
measurement
dataset starts.
0-100/% is
displayed.
“Start NO” is displayed
Allows not to start the
measurement dataset.
“Start YES” is displayed.
Allows to start the
measurement dataset.
The measurement
dataset starts.

4.7 Data collector – PC interface with Kilo 2015 software
4.7.1 Download data collector on PC
➢Plug the data collector on a USB port of the computer.
➢Launch the Kilog 2015 software by double clicking on the icon.
➢Click on “Download” button, or go to “Datalogger” menu and select “Downloading”.
The following window opens.
➢Tick the boxes of datasets to transfer.
➢If necessary, define the files saving location by clicking on
“Modify”. ( )
➢Select the required option about “Management of existing
files” (2).
➢To delete automatically the datasets which are transferred
from the data collector to the computer, tick the
corresponding box in “Options” (3).
➢Click on “Validate”.
The measurement datasets are displayed and
automatically saved in the defined location.
4.7.2 Display data collector state, delete memory and set date and time
➢Plug the data collector on a USB port of the computer
➢Launch the Kilog 2015 software by double clicking on the icon.
➢Click on “Configure” button, or go to “Datalogger” menu and select “Configuration”
The “Device state” (1) window opens.
The device name, serial num er and version num er are
indicated on the top of the window.
The num er of recorded datasets and the used storage
capacity are displayed.
➢Click on “Clear memory” (2)
A diagram summarises the memory state and a list of the
recorded datasets is displayed.
➢Select the datasets to delete by ticking the corresponding
box then click on “Validate”
A confirmation message is displayed: select Yes to confirm
or No to cancel.
Device use
( )
(2)
(3)
( )
(2)

➢Click on “Date and time” (3)
➢Set a date and time synchronisation with PC or customise
them, then click on “Validate”.
4.7.3 Update the data collector
It is possible to update the KNT 320 data collector by connecting it
on the Kilog 2015 software:
➢Click on “Update” (4).
•Carry out an automatic update
The automatic updating needs an internet connection. It can be
protected by a proxy server. In this case:
➢Tick the “Automatic” box.
➢Tick the “Use a proxy server” box.
➢Choose “Default proxy”: the KILOG software uses the
proxy parameters of the computer internet connection.
Or
➢Choose “Customised proxy”: indicate the Proxy address then the user name and password.
➢Click on “ tart update” to start the device updating.
The progress of the updating displays in the progression ar.
A message indicating the updating is finished displays at the end.
•Carry out a manual update
➢Tick the “Manual updating” box.
➢Click on “Browse” and take the updating file given by the KIMO after-sales service (.zip file type).
➢Click on “ tart update” to start the device updating.
The progress of the updating displays in the progression ar.
A message indicating the updating is finished is displayed at the end.
•Check the effective updating
To verify the effective updating, check the version number indicated in the device references of the device state window:
2 Device use
(3)
(4)

4.8 Data collector – PC interface without Kilo 2015 software
It is also possible to manage the files of the data collector without using the Kilog 2015 software.
In this case, the KNT 320 data collector works as a USB flash drive. It is recognised by the computer as a removable disk
with the name “KNT-320”.
➢Plug the data collector on a USB port of the computer.
The following auto-play window opens:
➢Click on “Open folder to view files”.
The folders are organised according to the serial num er of the devices used. The su -folders are settled y dates
of the measurement datasets. In these su -folders are stored the measurement datasets files: they are named as
the dataset has een defined.
➢Follow the same procedure as with a USB key to rename, copy, transfer or delete files.
➢For a data processing, the Kilog 2015 software is necessary. If the software is installed on your computer, double-
click on the required dataset file to open it.
Device use 3

5 Maintenance
5.1 Rechar e the battery
The KNT 320 data collector power supply is ensured by a rechargeable lithium polymer battery.
When the battery is low, icon is displayed.
➢To recharge the battery, plug the data collector on a USB port of your PC.
During the charging time on PC, icon is displayed and the data collector displays “USB”.
When the attery is totally recharged, icon disappears.
Charging time is about 7 minutes to charge 50% of the battery, about 2 hours to recharge it completely.
5.2 Device cleanin
Please avoid any aggressive solvent.
Please protect the device form any cleaning produce containing formalin.
6 Accessories
Accessories References Illustrations
1 female U B / male micro-U B adapter
Supplied with the device KNT-A
KILOG 2015 oftware
The Kilog 20 5 configuration and data processing
software allows to configure, save and process your
data in a very simple way.
Software only: Ref. KILOG-3-N
Complete set: software + USB cable,
Ref. KIC-3-N
Only the accessories supplied with the device must be used.
Only the KILOG 2015 software is compatible with the KNT 320 data collector
7 Troubleshooting
Problem Probable cause and possible solution
“Err” error message is displayed
The data collector can not communicate with the Kistock:
Check that the USB connector is correctly plugged.
Once the connection established, press OK to go back to home screen:
the data collector displays “On”.
The data collector does not
switch on
The battery must be recharged:
Plug the data collector on a USB port to recharge it (see § 5. ).
4 Trou leshooting
BAT
BAT
BAT


NTang – KNT 320 – 08/ 2/ 6 – RCS (24) Périgueux 349 282 095 Non-contractual document – We reserve the right to modify the characteristics of our products without prior notice.
Once returned to KIMO, required waste collection will be assured in the respect of the environment in accordance to guidelines relating to WEEE.
BE CAREFUL! Material damages can happen, so please apply the precautionary measures indicated.