KITO Harrington DHLT Manual

Dynamic Hoist Load Tester (DHLT)
Calibration Procedure
HARRINGTON HOISTS,INC.CONFIDENTIAL
1 of 4
General Use
EDOC0430
Rev. 01 10-5-2016
PURPOSE
To provide instructions for calibration of the DHLT.
SCOPE
(1) Includes the DHLT. Excludes the SHLT (Static Hoist Load Tester).
(2) Includes only calibration. Excludes setting the Pressure Switch and setting the Limit Switch.
(3) Calibration based on using Harrington 5T Electric Chain Hoist (either NER050L or ER050L).
INSTRUCTIONS
1. Collect Calibration Data
(a) Assemble 5 Ton Electric Chain Hoist and properly calibrated Force Measuring Device (Load
Cell or Dynamometer) to the 5 Ton link in the test stand (see Fig. 1).
(b) Connect power to test stand control panel.
(c) Connect pendant to test stand control panel.
(d) Connect cable between test stand control panel and hoist.
(e) Connect power to the hoist.
(f) Ensure the Accumulator Discharge Valve located between the accumulator and the reservoir
is closed (handle is perpendicular to the hoses).
(g) Press the On/Off pushbutton on the test stand control panel to turn on the DHLT. The DHLT
pump should begin operating and maintain system pressure between approx 1700-2300 psi.
For DHLT units shipped after 10/2010, the pump should begin operating and maintaining
system pressure between approx. 1700 – 2450 psi.
(h) Turn on the Dynamometer; assure that it is freely suspended from the Test Hoist with no
tension on it (other than the weight of the attached lower shackle). Press its zeroing button
and ensure the read-out indicates zero load.
Note: Do not press the zeroing button any time after testing starts.
(i) Use the Test Hoist to level the lower beam assembly; this minimizes any effects associated
with slant-lifting.
(j) Follow the calibration procedure for both the 5 ton and 10 ton links as follows. Using the
Pressure Control Valve (PCV), slowly adjust the cylinder pressure to increase the system
pressure. An increase of system pressure can be verified by the Cylinder Pressure Gauge
(CPG). See the last page of this document for a data recording chart.
Note: Small movements in the PCV correspond to large changes in the Dynamometer
reading, particularly on the 10 Ton link; avoid large or rapid adjustments to the PCV as
these can lead to pressure oscillations.
Note: During the following procedure, if you go above the target cylinder pressure while
adjusting “up”, reduce the cylinder pressure back down below the target value, and then
raise it up to the target value again. Similarly, if you go below the target cylinder
pressure while adjusting “down”, raise the cylinder pressure back above the target value,
and then lower it down to the target value again.

Dynamic Hoist Load Tester (DHLT)
Calibration Procedure
HARRINGTON HOISTS,INC.CONFIDENTIAL
2 of 4
General Use
EDOC0430
Rev. 01 10-5-2016
a. 5 Ton link:
i. Adjust cylinder pressure up to 200 psi and record dynamometer reading.
ii. Adjust cylinder pressure up from 200 psi to 1300 psi in 100 psi increments
recording the dynamometer reading at each increment.
iii. Adjust cylinder pressure just slightly above 1300 psi then back down to 1300 psi
and record dynamometer reading.
iv. Adjust cylinder pressure down from 1300 psi to 200 psi in 100 psi increments,
recording the dynamometer reading at each increment.
v. Average the two dynamometer readings for each cylinder pressure increment
recorded and record these results onto the DHLT Calibration Data Sheet.
vi. Adjust cylinder pressure to zero.
vii. Use the Test Hoist to lower the lower beam assembly to its fully lowered
position; this will remove the load from the Test Hoist.
viii. Remove the Test Hoist, dynamometer, and associated shackles from the DHLT’s
5 Ton link and attach them to the DHLT’s 10 Ton link.
b. 10 Ton Link
i. Repeat steps 1(h) and 1(i) above.
ii. Adjust cylinder pressure up to 200 psi and record dynamometer reading.
iii. Adjust cylinder pressure up from 200 psi to 700 psi in 100 psi increments,
recording the dynamometer reading at each increment.
iv. Adjust cylinder pressure just slightly above 700 psi, and back down to 700 psi,
and record dynamometer reading.
v. Adjust cylinder pressure down from 700 psi to 200 psi in 100 psi increments,
recording the dynamometer reading at each increment.
vi. Average the two dynamometer readings for each cylinder pressure increment
recorded, and record these results onto the DHLT Calibration Data Sheet.
vii. Use the Test Hoist to lower the lower beam assembly to its fully lowered
position; this will remove the load from the Test Hoist.
viii. Remove the Test Hoist, dynamometer, and associated shackles from the DHLT’s
10 Ton link.
2. Prepare Calibration Chart
(a) Enter the date and the DHLT serial number in the designated areas in the upper left-hand
area of the calibration spreadsheet (Request DHLT Cal rev00.xls from Harrington).
(b) Enter the “Average” values from the DHLT Calibration Data Sheet into the corresponding
yellow-colored cells of the calibration spreadsheet. This will cause the spreadsheet to (i)
generate two lines on the graph area of the spreadsheet, and (ii) enter values in the table
area (load vs. pressure) of the spreadsheet.

Dynamic Hoist Load Tester (DHLT)
Calibration Procedure
HARRINGTON HOISTS,INC.CONFIDENTIAL
3 of 4
General Use
EDOC0430
Rev. 01 10-5-2016
(c) Confirm that the two lines generated on the graph are linear, or close to linear (they will be
linear if their R2values are very close to 1). If not, investigate the cause, correct, and obtain
new test data. Recalibration may be required to obtain accurate values.
(d) Print the spreadsheet table (load vs. pressure) and attach it to the DHLT in a manner that is
visible to the operator.
Figure 1

Dynamic Hoist Load Tester (DHLT)
Calibration Procedure
HARRINGTON HOISTS,INC.CONFIDENTIAL
4 of 4
General Use
EDOC0430
Rev. 01 10-5-2016
DHLT CALIBRATION DATA SHEET
Date: ____________________________
Serial Number: ______________________________
Cylinder
Pressure
(psi)
Force (lbs) Measured by Load Cell
5 Ton Link 10 Ton Link
Up Down Average Up Down Average
200
300
400
500
600
700
800
900
1,000
1,100
1,200
1,300
Table of contents
Popular Test Equipment manuals by other brands

Leclanche
Leclanche AlfaBat Pro Start

Keysight Technologies
Keysight Technologies N1091 CA Series Getting started

barfield
barfield TT1200A User instruction manual

Lollipop
Lollipop Femometer manual

Tektronix
Tektronix 4 Series Declassification and security instructions

Symmetricom
Symmetricom TimeSource 2700 manual