HND-P121/-123/-126/-127/-129
page 10 HND-P121-129 K04/1121
Supported functions:
7.6 Zero Displacement Sensor ('OFFS')
A zero displacement can be carried out for the measured value:
value displayed = value measured - offset
Standard setting: 'off' = 0.0°, i.e. no zero displacement will be carried out.
Together with the scale correction (see below) this factor is mainly used to
compensate for sensor deviations. Input is in the display unit.
7.7 Scale Correction Sensor ('SCAL')
The scale of the measuring can be influenced by this setting (factor is in %):
displayed value = measured value * (1+Scal/100)
Standard setting: 'off' =0.000, i.e. value is not corrected. Together with the zero
displacement (see above) this factor is mainly used to compensate for sensor
deviations.
7.8 Pressure Connection
2 (or 1) universal pressure connector for 6 x 1 mm (4 mm tube inner diameter) or
8 x 1 mm (6 mm tube inner diameter) plastic tubes.
7.8.1 Device type with absolute pressure (HND-P129)
Connect plastic tube to pressure port.
Code Name/Function Code Name/Function
0 Read measurement value 200 Read min display range
3 Read system state 201 Read max display range
6 Read min memory 202 Read display range - unit
7 Read max memory 204 Read display range – decimal point
12 Read ID number 208 Read # of channels
32 Read configuration flag
BitCorrectToSealevel:32
214 Read scale adjustment [%]
216 Read offset adjustment
160 Set configuration flag (see above) 220 Read altitude (only abs. press sensors)
174 Clear min memory 221 Set altitude (only abs. press sensors)
175 Clear max memory 222 Read power off time (Conf-P.oFF)
176 Read min measuring range 223 Set power off time (Conf-P.oFF)
177 Read max measuring range 240 Reset
178 Read measuring range – measuring unit 254 Program version
179 Read measuring range – decimal point
180 Read kind of measuring of sensor
199 Read kind of measuring of display