4. VerifyDAandADwhenthemaximumvoltageisupto90V:
Method:setthepowersupplyto100Vandthemaximumcurrentoutputto1A.
Setthemultimetertothevoltagerange.Combinethepowersupplyandthe
multimeterintotheloadinterfaceatthesametime.TurnonON/OFFswitchand
adjusttheknobtomakethevoltageonthemultimeterdisplay90V.Andpressthe
leftandrightkeystoadjusttheincreasedandreducedvalue.ThenpressENTER
toenterthenextcheck,orpressESCtoreturntothepreviouscheck.
5. CheckthezeropointofDAlowcurrent:
Method:setthepowersupplyto5Vandthemaximumcurrentoutputto1A.Set
themultimetertothecurrentrange.Connectthepowersupplyandthe
multimetertotheloadinterface.TurnontheON/OFFswitchandadjusttheknob
tomakethemultimetercurrentdisplaybetween1mAand2mA.Andpressthe
leftandrightkeystoadjusttheincreasedandreducedvalue.ThenpressENTER
toenterthenextcheck,orpressESCtoreturntothepreviouscheck.
6. CheckthezeropointofDAhighcurrent:
Method:setthepowersupplyto5Vandthemaximumcurrentoutputto1A.Set
themultimetertothecurrentrange.Connectthepowersupplyandthe
multimetertotheloadinterface.TurnonON/OFFswitchandadjusttheknobto
makethecurrentonthemultimeterdisplaybetween1mAand2mA.Andpress
theleftandrightkeystoadjusttheincreasedandreducedvalue.Thenpress
ENTERtoenterthenextcheck,orpressESCtoreturntothepreviouscheck.
7. VerifyDAandADwhenthemaximumvalueofthecurrentlowgearis3A:
Method:setthepowersupplyto5Vandthemaximumcurrentoutputto5A,set
themultimetertothecurrentrange,connectthepowersupplyandthe
multimetertotheloadinterface,turnontheON/OFFswitch,andadjusttheknob
tomakethecurrentonthemultimeterdisplay3A.Andpresstheleftandright
keystoadjusttheincreasedandreducedvalue.ThenpressENTERtoenterthe
nextcheck,orpressESCtoreturntothepreviouscheck.
8. VerifyDAandADwhenthemaximumvalueofthecurrentlowgearis9.9A:
Method:setthepowersupplyto5Aandthemaximumoutputto11A,setthe
multimetertothecurrentrange,connectthepowersupplyandthemultimeter
totheloadinterface,turnontheON/OFFswitch,andadjusttheknobtomake
thecurrentonthemultimeterdisplay9.9A.Youcanpresstheleftandrightkeys
toincreaseordecrease.ThenpressENTERtosaveandcompletetheverification,
orpressESCtoreturntothepreviousverification.