Operating instructions
Electronic pressure gauge
DMG S2601, DMG S2610
Translation from the German
© 2008 – 2010 Elster GmbH
Electronic pressure gauge ..............................1
DMG S2601, DMG S2610 .................................1
Contents ........................................................ 1
Safety .............................................................1
Checking the usage ........................................2
Type code ......................................................... ..2
Part designations ................................................2
Type label ...........................................................2
Commissioning ..................................................2
Segment displays ................................................ 2
Connecting the measuring hoses ........................ 2
Selecting the operating mode ..............................2
ZERO (zero point calibration) operating mode .....3
UNIT (unit selection) operating mode ..................3
AUTO (measured value resolution) operating
mode ................................................................3
Min./Max. measured value ..................................3
Hold function .......................................................3
Display lighting ....................................................3
Switching off .........................................................3
Maintenance ....................................................3
Changing the batteries...................................... 3
Accessories .....................................................4
Assistance in the event of malfunction............4
Technical data ....................... .........................4
Declaration of conformity ..............................4
Please read and keep in a safe place
Please read through these instructions
carefully before installing or operating. Follow-ing the
installation, pass the instructions on to the operator.
These instructions can also be found at
www.docuthek.com.
Explanation of symbols
●, 1 , 2 , 3 ...= Action
= Instruction
Liability
We will not be held liable for damages resulting from
non-observance of the instructions and non-compliant
use.
Safety instructions
Information that is relevant for safety is indicated in the
instructions as follows:
DANGER
Indicates potentially fatal situations.
WARNING
Indicates possible danger to life and limb.
!CAUTION
Indicates possible material damage.
All interventions may only be carried out by qualified
gas technicians. Electrical interventions may only be
carried out by qualified electricians.
Conversion, spare parts
All technical changes are prohibited. Only use OEM
spare parts.
Transport
On receipt of the product, check that the delivery is
complete (see Part designations). Report any trans-port
damage immediately.
Storage
Store the product in a clean, dry environment, protected
from shocks. Ambient temperature: see Technical data.
Disposal
Components, particularly batteries, are to be dis-posed
of separately.
On request, old units may be returned carriage paid to
Elster GmbH in accordance with the relevant waste
legislation requirements (for Germany: KrW-/AbfG [Act
for Promoting Closed Substance Cycle Waste
Management and Ensuring Environmentally Compat-
ible Waste Disposal]).
Contents
Safety

ADD FURNACE CO.,LTD.
44 ซอยบรมราชชนนี 
โทร: 02-888-3472 โทร: ออกแบบ:08-08-170-170 แฟกซ์: 02-888-3258
https://www.add-furnace.com E-mail: sales@add-furnace.com
S601, S610
For measuring negative pressure, positive pressure, differential
pressure and relative pressure of gaseous media in non-
explosive areas.
This function is only guaranteed when used within the specified
limits –see [Technical data –p. 4]. Any other use is considered
as non-compliant.
! CAUTION
Please observe the following to ensure that the
DMG is not damaged:
–Do not use the DMG in potentially explosive areas, or on people
or animals.
–Note the max. medium and ambient temperatures, see
[Technical data –p. 4].
–Protect the DMG against shocks and moisture.
Type code
Code Description
DM Electronic pressure gauge
2601 Measuring range
2610 0 to ±150 mbar (±2.18 psi)
0 to ±1000 mbar (±14.5 psi
Part designations
1 Measuring instrument including batteries and
2 Magnet
3 Display
4Connections
5 On/Off button
6 Menu button/unit selection
7 Program selection button
8 Pause button
9 Min./Max. button
Type label
Model, serial number and measuring range: see
type label.
Segment displays
1 Switching on the DMG: Press the On/Off button
for approx. 1 s.
▷▷All the available display segments are displayed
for approx. 2 s. The segment display time can
be extended for checking by keeping the On/Off
button pressed.
▷The serial number, e.g. 0123, and the firmware version,
e.g. 1.0.12, are displayed for approx. 2 s.
▷The device type, e.g. 2601, and measuring range, e.g.
150 hPa, are displayed for approx. 2 s.
Connecting the measuring hoses
▷▷The measuring hoses are available as a connec-
tion set, see [Accessories –p. 4].
1 Slide the measuring hoses onto ports P1+/P2-,
depending on the appropriate field of application,
and secure to the pressure test points.
Positive pressure: port P1+.
Negative pressure: port P2-.
Differential pressure: higher pressure at P1+,
lower pressure at P2-.
▷The unit is ready for operation.
Selecting the operating mode
▷It is possible to change between the different
operating modes using the program selection
button
Checking the usage
Commissioning

ADD FURNACE CO.,LTD.
44 ซอยบรมราชชนนี 
โทร: 02-888-3472 โทร: ออกแบบ:08-08-170-170 แฟกซ์: 02-888-3258
https://www.add-furnace.com E-mail: sales@add-furnace.com
ZERO (zero point calibration) operating mode
▷In ZERO operating mode, the zero point can be
set at any time and as often as required.
1Press menu button
UNIT (unit selection) operating mode
1Select the unit according to the measuring
mode using the menu button
AUTO (measured value resolution) operating
mode
▷Availability of this operating mode is dependent
on the measuring ranges.
▷In AUTO operating mode, the best possible
display of the measured value is preset
automatically. The resolution is thus adjusted or
reduced as the measured value increases.
▷The desired measured value display can be
preset manually using the menu button , so that
irrelevant digits after the decimal point, for example,
can be masked.
Auto: automatic measuring range selection,
Lo (Low): low measuring range (higher resolution),
HI (High): high measuring range (low resolution).
Min./Max. measured value
1 Press min./max. button .
▷The display shows the minimum measured
pressure value.
2 .Press min./max. button again.
▷The display shows the maximum measured
pressure value.
3 . To delete the min./max. measured value, press
the menu button .
▷The current measured pressure value is displayed again.
Hold function
▷In all operating modes, the measured value can be held
by pressing the pause button ӀӀ .
▷appears in the top left corner of the display.
▷By pressing the pause button again, the measured
value is released.
Display lighting
▷The display lighting can be switched on and off
by briefly pressing the On/Off button .
Switching off
▷Press the On/Off button for approx. 2seconds.
▷Automatic time-dependent switch-off is currently
not implemented.
▷It is recommended that you have the instrument
checked regularly (< 24 months) and that it be
recalibrated regularly if used industrially.
Changing the batteries
▷Approved batteries, see [Technical data –p. 4].
▷The charge status is displayed in the top right
corner of the display.
▷Always change both batteries at the same time.
3 Insert new batteries.
4 Ensure that these are connected with the correct
polarity.
5 Place the cover back on the battery compartment
to close.
6 Dispose of used batteries separately.
Maintenance

ADD FURNACE CO.,LTD.
44 ซอยบรมราชชนนี 
โทร: 02-888-3472 โทร: ออกแบบ:08-08-170-170 แฟกซ์: 02-888-3258
https://www.add-furnace.com E-mail: sales@add-furnace.com
▷Connection set:
2 x blue plastic hoses,3 x 0.5 mm,
2 x silicone hoses, 5 x 2 mm,
2 x hose connectors D3/D5.
▷Order No.: 03352147.
Fault
Cause
Remedy
The DMG will not switch on.
The battery charge is too low.
Change batteries.
Pressure is not displayed.
The sensor is defective.
Return unit to the manufacturer.
Dimensions including protective sleeve W x H x D:
66 x 37 x 143 mm (2.7 x 1.4 x 5.6").
Weight: approx. 250 g (8.8 oz).
Housing: plastic, blue,
protective sleeve: plasticized PVC.
Display: transflective.
Response delay: 90% (T 90) of the measured
value will appear on the display within 10 s.
Ambient and medium temperatures:
0 to +40°C/+32 to +104°F, storage temperature:
-20 to +50°C/-4 to +122°F.
Approved batteries: 2 x 1.5 V (type miniature
LR6/AA).
Power supply: rated voltage via power adapter
(USB).
DMG 2601 measuring range:
rated 150 mbar (2.2 psi),
max. 180 mbar (2.6 psi),
max. positive pressure: 1.35 bar (19.6 psi).
DMG 2610 measuring range:
rated 1000 mbar (14.5 psi), max. 1500 mbar
(21.8 psi), max. positive pressure: 3 bar.
Resolution: DMG 2601: 0.01 (< 19.99); 0.1 (> 20),
DMG 2610: 0.1 (< 199.9); 1 (> 200).
Accuracy [% of measured value]:
DMG 2601:
1 ±1 digit (< 130 mbar/1.9 psi);
1.5 ±1 digit (> 130 mbar/1.9 psi).
DMG 2610:
1 ±1 digit (< 1000 mbar/14.5 psi);
1.5 ±1 digit (> 1000 mbar/14.5 psi).
Enclosure: IP 40 EN 60529
We hereby declare that the products DMG 2601
and 2610 comply with the requirements of the listed
Directives and Standards.
Directives:
–2004/108/ECStandards:
–EN 55022
–EN 61000-6-1
–EN 61000-4-2
–EN 61000-4-3
The production is subject to a Quality System pursu-
ant to DIN EN ISO 9001.
Elster GmbH
Scan of the Declaration of conformity (D, GB)–see
www.docuthek.com
?
!
•
?
!
•
?
!
•
Accessories
Assistance in the event of
malfunction
Technical data
Declaration of conformity

This manual suits for next models

1

Popular Measuring Instrument manuals by other brands

Praktica LRF-7 manual

Praktica

Praktica LRF-7 manual

DAKOTA ULTRASONICS ZX-5 DL Operation manual

DAKOTA ULTRASONICS

DAKOTA ULTRASONICS ZX-5 DL Operation manual

AQUATERR INSTRUMENTS M-300 user manual

AQUATERR INSTRUMENTS

AQUATERR INSTRUMENTS M-300 user manual

Kusam-meco KM 2009 Operation manual

Kusam-meco

Kusam-meco KM 2009 Operation manual

Triplett RHT05 user manual

Triplett

Triplett RHT05 user manual

Bacharach PUR-CHEK PRO manual

Bacharach

Bacharach PUR-CHEK PRO manual

ETM ETM770 user manual

ETM

ETM ETM770 user manual

General tools RHMG700DL user manual

General tools

General tools RHMG700DL user manual

Teledyne Lecroy HVD3102 Operator's manual

Teledyne Lecroy

Teledyne Lecroy HVD3102 Operator's manual

Delta Electronics Platinum Temperature Sensors DVP04PT-S instruction sheet

Delta Electronics

Delta Electronics Platinum Temperature Sensors DVP04PT-S instruction sheet

Roland R-09 Brief instructions

Roland

Roland R-09 Brief instructions

Onicon D-1401 Installation and operation guide

Onicon

Onicon D-1401 Installation and operation guide

Water I.D. PoolLab 2.0 user manual

Water I.D.

Water I.D. PoolLab 2.0 user manual

SIGLENT SSA3000X-R user manual

SIGLENT

SIGLENT SSA3000X-R user manual

CHARDER MEDICAL HM80M instruction manual

CHARDER MEDICAL

CHARDER MEDICAL HM80M instruction manual

KYORITSU KEW6205 instruction manual

KYORITSU

KYORITSU KEW6205 instruction manual

Laser Technology TruPulse 200X Connection guide

Laser Technology

Laser Technology TruPulse 200X Connection guide

Adeunis RF LoRaWAN TEMP user guide

Adeunis RF

Adeunis RF LoRaWAN TEMP user guide

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.