LAIA Cute Thermal User manual

Copyright © 2020 Laia · www.laiatech.com All rights reserved
Contents
OVERVIEW .............................................................................................. 2
SPECIFICATIONS ........................................................................................ 3
CONNECTION MODE ................................................................................... 4
BACKGROUND SOFTWARE ............................................................................. 5
4.1 INTERFACE....................................................................................... 5
4.2 DRIVER INSTALLATION ............................................................................ 5
4.3 SOFTWARE APPLICATION .......................................................................... 9
4.3.1 Real-time image..................................................................... 9
4.3.2 Temp measurement .............................................................. 11
4.3.3 Zone temp measurement ........................................................ 12
4.3.4 Color palette ...................................................................... 12
4.3.5 Alarm .............................................................................. 13
4.3.6 Snapshot .......................................................................... 13
4.4 TEMP MEASUREMENT .......................................................................... 13
4.4.1 Distance ........................................................................... 13
4.4.2 Calibration......................................................................... 14
4.4.2.1 No blackbody mode................................................. 14
4.4.2.2 Blackbody mode .................................................... 14
DEPLOYMENT PRECAUTIONS ........................................................................ 15

Copyright © 2020 Laia · www.laiatech.com All rights reserved
Overview
CUTE THERMAL is a high-precision thermal imaging, which can measure the temp of target
object online in real time, output thermal image video and check the over-temp condition. Going
with different matching platform software, it can be suitable for different usage modes (such as
power device temp measurement, fire alarm, human body temp measurement and screening).
This document only introduces the usage modes for human body temp measurement and
screening.
CUTE THERMAL employs USB supply power and transmit data are completed through one
USB line, realizing convenient and rapid deployment.
Based on on-site deployment of clients, CUTE THERMAL can carry out temp compensation
varying with environment changes voluntarily without continuous blackbody calibration and
control the error within the range of ±0.6°C (±1.08°F).

Copyright © 2020 Laia · www.laiatech.com All rights reserved
Specifications
CUTE THERMAL specifications are as follows:
Parameters Index
Infrared
thermal
imaging
Resolution 320x240
Response wave band
8-14um
Frame rate 9Hz
NETD 70mK@25°C (77°F)
Field angle 34.4 in horizontal, 25.8 in vertical
Lens 6.5mm
Measurement range -10°C - 300°C (14°C-572°C)
Measurement
accuracy
For human body, the temp compensation algorithm can reach
±0.6°C (±1.08°F)
Measurement Temp measuring area and center point can be set
Color palette Whitehot, Rainbow, Iron, Tyrian.
General
Interface
Power supply and data transmission through standard Micro USB
2.0
Language English
Operating temp
-20°C (-4°F) ~ +60°C (+140°F) (for the requirement of accurate temp
measurement of human body, it is recommended to use at ambient
temp of 10°C (50°F) ~ 30°C (+86°C))
Storage temp -40°C (-40°F)- +85°C (+185°F)
Waterproof and
dustproof IP54
Size 129mm*73mm*61mm (L*W*H)
Net weight 295g
Picture storage BMP, JPG.
Installation Standard tripod or pan-tilt hoisting is adopted.
Software
Temp display
High temp tracking in measurement area can be set, and fixed-point
measurement of center temp can be carried out
Alarm
Available for alarm over the set threshold temp (high and low
temp), can sound alarm, snapshot alarm photos and store
simultaneously.
Frame temp
measurement
Temp measurement is set according to different entry and exit
personnel channels
Temp compensation
The staff can set up temp compensation according to the
environments
Photograph Manually under opening, automatically under alarming
Internet cloud
upload Customized according to cloud requirements

Copyright © 2020 Laia · www.laiatech.com All rights reserved
Background software
3.1 Interface
This system only available for Win10 OS, and the background software interface is as follows:
3.2 Driver installation
After connecting the computer to the CUTE THERMAL camera with Micro USB 2.0 cable, open the
"Device Manager", and locate the "Universal Serial Bus devices" or “other devices” or "PIR324
Thermal Camera" with exclamation mark in "USB"

Copyright © 2020 Laia · www.laiatech.com All rights reserved
Double click and click "Update Driver", browse and find the driver directory. The exclamation
mark disappears after installation completed
Double click "DYTiRViewer.exe" to operate the software.
Please refer to the following steps, if you are prompted that the driver cannot be installed due to
driver hash value:
Open the "Settings" and click “Update and Security”:

Copyright © 2020 Laia · www.laiatech.com All rights reserved
3.3 Software application
Double click DYTirViewer.exe to run the program. If the program cannot be run after the driver
installation is completed, please confirm whether the last supported "Microsoft VC++
Redistributable" has been installed. If not, please search and download the corresponding version
online, and the program will run normally after installation.
3.3.1 Real-time image
Select the camera in red box in the figure below, click the "Play", and the current image of the
camera will be displayed on the right. Click "Stop" to stop displaying the real-time image. Click
"Photo" to select "Folder" and save the image.

Copyright © 2020 Laia · www.laiatech.com All rights reserved
3.3.2 Temp measurement
CUTE THERMAL infrared thermal imager needs 10min of preheating after start-up. After 10min,
the thermal imager can output accurate temp values stably. View the start-up time at the top right
of the software. Temp fluctuation is normal within 0.5°C (0.9°F). In context of human body temp
measurement, because the high accuracy is required, the environment needs to be stable,
avoiding air opening, outdoor and constantly changing occasions, such as strong wind, air
conditioning tuyere, fan and hot water surrounded, otherwise the measurement may be
inaccurate.
The upper-left shows the current maximum temperature in the whole image, and if add temp zone,
then the maximum temp only show the current zone area highest temp.
Press the upper right corner configuration icon, there will be a temperature unit transformation,
press the unit and OK, then all the temperature unit will be transformed.

Copyright © 2020 Laia · www.laiatech.com All rights reserved
3.3.3 Zone temp measurement
If you do not select area to be measured, the location and temp of the max temp in entire screen
will be displayed. When you select the zone, only the max temp in the selected area will be
displayed.
On the real-time displayed image, you can select to add a temp measurement zone on the left to
select area to measured, and you can select up to 3 areas for temp measurement. Click "Clear" will
clear all temp measuring zones displayed in the screen. There are 1,2,3 for the label of the zone to
differential different zones.
3.3.4 Color palette
You can click "Color" at the red circle in the figure below to display the thermal image.

Copyright © 2020 Laia · www.laiatech.com All rights reserved
The optional color palettes are as follows:
•Rainbow
•Iron
•Tyrian
•Whitehot
3.3.5 Alarm
Available for image alarms and sound alarms, and automatic saving of snapshot when alarms occur.
When the temp exceeds the threshold, the area temp measuring box will turn red to give an alarm.
Click on the ellipsis following the word "Voice Alarm" to select different sounds and intervals for
sound production, and click on the ellipsis following the word "Alarm Photo" to select the directory
and interval for automatic snapshot.
3.3.6 Snapshot
If "Alarm Photo" is checked, the snapshot will be displayed on rightmost side of the software and
the snapshot time will be displayed. Click this picture to view with Win10 default software.
3.4 Temp measurement
3.4.1 Distance
The measurement accuracy of the infrared thermal imager is affected by the distance to the

Copyright © 2020 Laia · www.laiatech.com All rights reserved
targeted object. The closer the distance is, the stronger the thermal radiation received by the
sensor is, and the higher the temp displayed is. It is recommended to measure within 2-3 meters.
3.4.2 Calibration
Calibration is in two modes:
•No blackbody mode
•Blackbody mode
3.4.2.1 No blackbody mode
If there is no blackbody, the thermal imager needs to be calibrated as follows:
Enable the thermal imaging thermometer, after stable operation, first measure the forehead temp
of a normal human body with a medical infrared thermometer indicated as T1, then measure the
temp of the human body with thermal imaging at a fixed distance of 2-3 meters from the camera
indicated as T2, and enter the compensated temp value T=T1-T2 in the temp adjustment box to
start normal temp measurement.
3.4.2.2 Blackbody mode
Blackbody should be placed at a position, distance and height close to the measured head as far
as possible to ensure temp measurement accuracy. Check "Blackbody Mode" and click "Select
Blackbody Position", then the mouse will become the crosshair on the video. Align the crosshair
with the center of the blackbody and click again. Enter the temp value set by the blackbody in pop-
up input box, click "OK" and start measuring.
Please try to use frame to measure and avoid blackbody, otherwise the temp set by blackbody is
higher than that of human body, which may cause the temp value of human body not to be
displayed.
The "Temp Calibration Value" in blackbody mode cannot be entered manually. The value in the
box is the one currently compensated. The historical temp measurement curve of boldface will be
displayed below. If this curve rises or falls sharply, it is necessary to check whether the blackbody

Copyright © 2020 Laia · www.laiatech.com All rights reserved
is blocked or the camera and blackbody are moved.
Deployment Precautions
1. Environmental changes have a great influence on the temp measurement accuracy of infrared
thermal imaging. Try to choose a stable room temp environment, and do not deploy thermal
imaging thermometer beside tuyere, air conditioner direct blowing position, fan direct
blowing position, heating device, etc.
2. Behind the position where the lens is located, try not to place other high-temp device higher
than the body temp, such as heating air conditioner, water boiling machine, high-power
incandescent lamp and other high-temp objects, to prevent false alarm.
3. If the high-temp objects cannot be excluded in the picture, a frame area can be drawn on the
software video, so that only the temp in the area frame will be measured to avoid other
interferences.
4. It is suggested to correct the temp every day or every few days according to the method of
3.4.2.1 to maintain the accuracy.
5. When the thermal camera is turned on for 10min, the internal heat balance of the machine
can be reached. At this time, the temp measurement value is the most accurate, so it is
recommended to measure 10min after turning on the camera.
6. The placement of blackbody will directly affect the measurement accuracy, so it should be
placed at a position, distance and height close to the measured head as far as possible to
ensure the measurement accuracy.
7. If the temp measuring software is interrupted due to the disconnection of the network cable
and the instability of WINDOWS system in intermediate process, if the interruption time is less
than 5min, the measurement can be continued after starting up; if more than 20min, and the
machine cools down, it should be continued 10min later after start-up.
8. Try to ensure one person appearing in the measuring area at a time when selecting position,
to prevent multiple people from entering simultaneously to give false alarm.
9. Some non-infected fever patients may have an alarm of ultra-high temp, such as after drinking,
exercising violently, after drinking boiled water, local facial inflammation, long-term sun
exposure, just coming out of a sultry environment, etc. At this time, the tested person can be
allowed to being calm down for a while or cool, then measure and screen.