
g) Calculate|(a1-a2)- (b1-b2) |= e ;
f) If e>1.5mm, the instrument accuracy is out of tolerance, an it nee s repair.
5.3. Self-calibration adjustment
See as following figure, there are two self-calibration apertures on the instrument, the self-calibration aperture A is
correspon ing with the a justment of fore an aft irection (the checke error in item 2 accuracy self calibration), the
self-calibration aperture B is correspon ing with the a justment of left-right irection (the checke error in item 1
accuracy self calibration).
1) When a justing, use the
subtense 3mm hexagon spanner.
2) The a justment of both irections will influence each other. When making fine a justment in the left-right irection,
there may be changes in fore an aft irection. Also, when making fine a justment in fore an aft irection, there may be
changes in left-right irection. Accor ingly, please repeat the a justment of both irections when making fine
a justment.
3) The a justing screw of self-calibration shoul not excee 4 circles (clockwise or anti-clockwise irection).
4) If it can not calibrate the instrument accuracy through the self-calibration aperture a justment, please contact the
ealer for servicing.
6. Demonstration
Self-calibration aperture A
Self-calibration aperture B