5(8126 X81) Precision Air Pressure Sensor Operating Instructions
Operating Instructions
Precision Air Pressure Sensor (8126 X81)
2.0 Putting into operation
2.1 Choice of installation place
As a barometer needs a free access to the ambient outer air
it has to be installed outdoors without influence of pressure
oscillations caused by movements of the vessel. Therefore
the sensitive barometer should be installed on a place which
is located at a wind protected place. In general the housing is
splash-water tight when mounted correctly. Therefore the air
inlet of the housing has to show downwards in order to prevent
the penetration of sea and rain water.
As barometric pressure values world-wide are based on
measurements above sea level the altitude of the barometer
should be noted for later compensations. Otherwise wrong
measurements of approximately 1 hPa per 8 m height have
to be considered.
In case of the instrument has to be installed indoors due to
special requirements a pressure link to the outer atmosphere
has to be carried out by means of a flexible rubber pipe or
a similar material. Otherwise overpressure cause by air-
conditioners will falsify the measurements.
2.2 Mounting
The mounting must be carried out on a flat surface with four
stainless steel hexagonal screws of the size M8.
2.3 Electrical connection
The electrical connection (drawing see right site) must be
carried out in the external junction box according to the wir-
ing diagram of the entire system. Important note: To prevent
electrical hazards for installation staff and damages due to
short circuits, it is absolutely necessary to switch off the main
power supply of the system until the installation is finished.
Attention: wrong connection may cause a destruction of this
and other connected components !
2.4 Cable entries
The common signal and power cable will be linked inside the
housing by means of a cable gland according to VG standards.
The selection of the right cable inlet will be carried out accord-
ing to the project specific bloc diagram respectively the wiring
diagram. The corresponding positions are clearly illustrated in
the discussed drawings.As only one item is present a special
identification procedure is not required.
For various outer diameters of the installation cables different
sized cable cones will be delivered together with the item.
2.5 Setting into operation
After having finished the electrical and mechanical installation
work of this item as mentioned in this manual and other related
instructions the wiring should be checked once again before
switching on the mains for the whole measuring system.
No further operating handling is required.
When the power is turned on, the sensor automatically starts
sending data.After a settling time of approx. 20 sec the sensor
provides specification-compliant measured values.
2.6 Performance check
Readings of barometric values can be obtained from the
connected pressure indicators or from external indicator con-
soles. As suitable performance check the read data can be
compared with these of other stations in the neighbourhood.
We recommend to ask a bureau of the local weather service
which should be located in harbours for precise data.
To be able to compare data all pressure values must
be related to the sea level.
Electrical connection
Data
Data
GNWTRDBU
Output A Green
RS485 B White
Air pressure
RS485 half-duplex
11...28 V DC
(24 V nominal)