Lambrecht 252 User manual

Thermo-Hygrograph
Operating Instructions
Thermo-Hygrograph (252)
Enclosure: Operating Instructions “Drum Recorders“
Thermo-Hygrograph (252) OperatingInstructions
1 General
Thermo-Hygrographs are used to measure and record air
temperature and relative humidity (r.h.). The measured values
are recorded, using drum recorders driven by clockworks.
These rugged and reliable precision products have been
proven over decades of use in business, science and culture
around the world. They are suited for service in industry and
private environments. To take advantage of the high measur-
ing accuracy and to maintain them for many years in service,
please observe our following recommendations.
2 Place of installation
Install the instrument horizontally in a vibration-free location.
Depending on the model, the permissible ambient tempera-
ture varies between -35...+80 °C. The instrument should be
protected against direct heat radiation and conduction. For
outdoor installation, a protection screen against radiation and
precipitation is required.
3 Setting into operation
To open the instrument, gently push down the hood 2, pull
the locking knob 1 and swing up the hood. At the same time
hold the styrofoam cushion 22 (located under the cover) and
remove it after opening the instrument.
3.1 Removing the transport lock
Loosen the recording arm‘s transport lock 4 by turning it on the
lifting rod 7 by about 90°. Remove the cardboard wedge at the
switch-o lever 3 and save it. Carefully remove the protective
cap from the felt-tipped pen 5 and save it.
3.2 Mounting the humidity measuring
element
Take the humidity measuring element N out of its separate
packing. Put the measuring element‘s mount 8 provided with
two bore holes into the upper hooks. Put the lower mount 9
into the lower hook. At the same time lift the recording arm 10;
regenerate the humidity measuring element N as described
under 6. The humidity measuring element K is already mount-
ed and is maintenance-free.

Operating Instructions
Thermo-Hygrograph (252)
Enclosure: Operating Instructions “Drum Recorders“
Thermo-Hygrograph (252) OperatingInstructions
3.3 Inserting the battery
(quartz clockworks only)
Lift the recording arms using the switch-o lever 3. Pull o the
drum 12 straight and very carefully, making sure not to bend
it sideways. Then push aside the yoke spring and check the
polarity when inserting the battery (1.5 V, type Mignon/AA).
Adjust the desired recording period at the gear wheels. Then
remount everything carefully (don’t bend the recording arms).
3.4 Winding up the clockwork
(mechanical clockworks only)
The moving spring is stretched by turning the winding key 6
anti-clockwise.
3.5 Setting the drum to local time
The felt-tipped pens must be very close to the paper surface
(adjusted by using the switch-o lever 3). Turn the drum 12
counterclockwise until the pen matches the desired time line.
If you have turned the drum too far, turn it back beyond the
desired time line so that you can resume the adjustment from
the beginning. This way the backlash of the toothed gears will
be eliminated. Since a graph paper has been already inserted
in the factory, the instrument is ready for operation after closing
the hood. Attention: You must hear locking knob 1 clicking in
place! After taking the device to the desired place of instal-
lation, you can put the felt-tipped pen into writing position by
turning the switch-o lever 3.
3.6 Change of the recording period of me-
chanically reversible clockworks (Fig. 3)
When delivered, the recording period is 7 days.
In order to change the recording period to 1 day, open the
instrument’s housing 2, unscrew the winding key 6clockwise
and pull the drum completely o the axis.
The gear wheel in the lower part of the drum has to be pulled
off and is re-inserted vice versa. Afterwards re-assemble
everything.
4.0 Service and maintenance
4.1 Changing graph paper
Using the switch-o lever 3, take the pen o the paper. Loosen
the metal clasp 13 on the clockwork drum by pushing it up.
Remove the old graph paper and put on a new one. Both chart
ends must be below the metal clasp; the lower edge of the
chart must t evenly to the lower edge of the clockwork drum.
Rex the metal clasp. Set the time as described under 3.5.
4.2 Changing the felt-tipped pens
Carefully pull the used felt-tipped pen 5 from the recording arm
and attach a new pen. Remove the protective cap and save it.
4.3 Felt-tipped pen contact pressure
To adjust the correct contact pressure, rst put the recording
arm into writing position using the switch-o lever 3. Tilt the
device by about 30° to the front. If the felt-tipped pens are
adjusted correctly, they should now be 1...2 mm from the chart
surface. You may have to turn the screws 14 to set the tips to
the correct distance. Too much pressure (thick recording line)
generates too much friction between the felt-tipped pens and
the graph paper, leading to wrong measuring results. If the
pressure is too weak (thin recording line) there is a risk that
the recording will be interrupted.
4.4 Cleaning and care
The bearings should not be oiled. Every three to ve years,
mechanical clockworks should be maintained by a clock mak-
er. The rest of the device may be carefully cleaned, using a
soft brush or a damp cloth.
4.5 Putting out of service
If the device is temporarily put out of service, the felt-tipped
pens must be covered with the supplied caps to prevent them
from drying out.
4.6 Transport
Use the switch-o lever 3 to lift the recording arms and secure
them with the cardboard wedge (see 2.1); put the protective
caps onto the felt-tipped pens. The recording arm 10 of
the humidity measuring element must be secured with the
transport lock 4 so that the measuring element is unstressed
(recording arm tip at approx. 20% r.h.). For dispatch the hu-
midity measuring element (except the humidity measuring ele-
ment K) must be unhooked and put into the original ip pack.
During transport the quartz clockwork 15 should not be in
operation.
5.0 Clockworks (gear wheels or DIP switch)
5.1 Mechanical clockwork (Fig. 2)
Recording time 1 or 7 days (depending on model)
Temperature range -35...+65 °C
5.2 Changeable mechanical
clockwork (Fig. 3)
Recording time 1 or 7 days (selectable by
changing the gear wheels 24)
Temperature range -35...+65 °C
5.3 Quartz clockwork (Fig. 4 or 5)
Recording time 1/ 7/ 31 days (can be changed
by gear wheels 15 resp. DIP switch 15)
Temperature range -10...+50 °C (Alkaline battery)
-35...+60 °C (rechargeables)
6 Thermo-Hygrograph 252 (Fig. 1)
The proven natural hair grid is the most frequently used device
to measure humidity.
Especially advantageous is the use of synthetic ber in rooms
and in temperatures between -10…+40 °C, since no regen-
eration is required.
6.1 Humidity measuring element N
The humidity measuring element N consists of human hair that
is subject to aging. To maintain its high measuring accuracy,
it should therefore be regenerated regularly (approx. every
14 days). To do so, moisten all hair of the measuring element
18 with distilled water. After about 2 minutes, the device will
indicate 95 % r.h.. If this value is not reached, a value of 95 % r.h.
must be adjusted using the humidity adjusting screw 16.
Due to storage/delivery times etc. we recommend to treat

Operating Instructions
Thermo-Hygrograph (252)
Enclosure: Operating Instructions “Drum Recorders“
Thermo-Hygrograph (252) OperatingInstructions
the humidity measuring element this way when the device is
rst put into operation. If the values at the measuring location
frequently reach 95 % r.h. the measuring element regenerates
automatically.
Measuring element standardized hair grid with
seven hair bundles
Measuring range 5...100 % r.h.
Accuray ± 2 % r.h. (with regular
regeneration)
Range of application -35...+70 °C
0...100 % r.h.
6.2 Humidity measuring element K
A soiled humidity measuring element K (synthetic ber) may
only be cleaned with a soft, dry brush. Do not regenerate the
humidity measuring element K!
Measuring element standardized harp with seven
synthetic ber bundles
Measuring range 25...70 % r.h.
Accuracy ± 3 % r.h.
Range of application -10...+40 °C
25...70 % r.h.
6.3 Temperature measuring element
From time to time a reference measurement should be made,
using a calibrated thermometer. For this purpose, place the
calibrated thermometer as close as possible to the tempera-
ture measuring element. If required, the felt-tipped pen can be
set to the correct temperature value by turning the temperature
adjustment nut 17.
Measuring element high quality, aged bimetal
Accuracy ± 0.3 °C
Range of application -40...+80 °C
7 Warranty
Please note the loss of warranty and non-liability by unau-
thorised manipulation of the system. You need a written
permission from LAMBRECHT meteo GmbH for changes
of system components. These activities must be operated
by a qualied technician.
The warranty does not cover:
1. Mechanical damages caused by external impacts (e. g.
icefall, rockfall, vandalism).
2. Impacts or damages caused by over-voltages or elec-
tromagnetic elds which are beyond the standards and
specications in the technical data.
3. Damages caused by improper handling, e. g. by wrong
tools, incorrect installation, incorrect electrical installation
(false polarity) etc.
4. Damages which are caused by using the device beyond
the specied operation conditions.
8 Disposal
LAMBRECHT meteo GmbH is listed and registered at the
Stiftung Elektro-Altgeräte Register ear under:
WEEE-Reg.-No. DE 45445814
In the category of monitoring and control instruments, device
type: “Monitoring and control instruments for exclusively
commercial use”.
Within the EU
The device has to be disposed according
to the European Directives 2002/96/EC and
2003/108/EC (Waste Electrical and Elec-
tronic Equipment). Do not dispose the old
device in the household waste! For an
environmentally friendly recycling and dis-
posal of your old device, contact a certied
disposal company for electronic waste.
Outside the EU
Please follow the regulations in your country regarding the
appropriate disposal of waste electronic equipment.

Operating Instructions
Thermo-Hygrograph (252)
Enclosure: Operating Instructions “Drum Recorders“
Thermo-Hygrograph (252) OperatingInstructions

LAMBRECHT meteo GmbH Tel +49-551-4958-0
Friedländer Weg 65-67 Fax +49-551-4958-312
Germany Internet www.lambrecht.net
Abb. / Fig. 5
02520_b-de.indd 03.21
Operating Instructions
Thermo-Hygrograph (252)
Enclosure: Operating Instructions “Drum Recorders“
Thermo-Hygrograph (252) OperatingInstructions
Subject to change without notice.
Clockwork with DIP switch
Clockwork with adjustable gear wheels
31 days 14 days 7 days 1 day
OFF
Table of contents
Other Lambrecht Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

AUDIOropa
AUDIOropa PROLOOP FSMplus Operation instructions

Kamstrup
Kamstrup OMNIPOWER installation manual

Flomotion Systems
Flomotion Systems LT260 Operation and maintenance manual

Autonics
Autonics MX4W Series product manual

K-Patents
K-Patents PR-33-AC instruction manual

Moeller
Moeller ps4-201-mm1 installation instructions