METHOD FOR COLLECTING WATER SAMPLES
1. To set the trigger mechanisms, hold the sampler by the brass handle
with triggers in the “up” position. Pull the cable attached to Plunger
A outward and hook the loop over Trigger A. This will remove
Plunger A from sampling chamber.
2. After the loop is hooked over Trigger A, attach the other loop to
Trigger B. This will remove Plunger B from sampling chamber.
3. The water sampler is now cocked and ready to be lowered into the
water.
4. While holding the brass messenger in one hand, steadily lower the
water sampler by the calibrated line until the desired depth is
reached. The line is marked at 0.5 meter and 1 meter increments.
5. Hold the line in a vertical position over the sampler and release the
messenger to travel down the graduated line. After the messenger
trips the closing mechanism, the sample within the collection
chamber is sealed from mixing with unsampled water.
NOTE: A slight tug on the line after the messenger had been
released signals the sampler has closed.
6. Gradually pull the water sampler to the surface with the line and
carefully rest sampler on level surface.
7. Aliquots or portions of the sample are carefully taken by standing the
water sampler upright on Plunger A and unclamping the outlet
tubing. Partially remove Plunger B to prevent formation of a vacuum.
NOTE: Using the JT-1 sampler at a depth of less than 5 meters may
damage the trigger.
LaMOTTE COMPANY
Helping People Solve Analytical ChallengesSM
PO Box 329 • Chestertown • Maryland • 21620 • USA
800-344-3100 • 410-778-3100 (Outside U.S.A.) • Fax 410-778-6394
Visit us on the web at www.lamotte.com
01.10