Lascar EL-SIE-2+ User manual

EL-SIE-2+
High Accuracy Temperature and Humidity
USB Data Logger, EasyLog Cloud Compable
Monitor the environment you live and work in with the EasyLog EL-SIE-2+. Conguraon is simple, with no soware to install
on your PC or Mac – just connect the logger with a USB cable, and use your standard web browser to congure the device for
logging. You don’t even need internet access to set up and use the EL-SIE-2+, it really couldn’t be easier.
The logging interval can be set between 10 seconds and 24 hours, with immediate, delayed, triggered or push-to-start logging.
Alarms are fully user congurable, with funconality including cumulave alarms, pre-alarms, a delay before alarm triggering,
and an alarm hold opon, which connues showing the alarm condion even if the reading returns to an acceptable level.
The display shows current, maximum and minimum readings, and three coloured LEDs indicate device status at a glance.
Once logging is complete, re-connect to your computer and use your browser to view, analyse and save your data. You can also
choose to upload your data to an Easy Log Cloud account, making the data accessible online for powerful graphing, analysis and
report generaon.
Typical baery life is over 1 year using standard AAA alkaline baeries, and a wall mounng bracket is supplied with the device.
• -18 to +55°C (-0.4 to +131°F) temperature measurement range
• 0 to 100%RH humidity measurement range
• Stores over 1,000,000 readings (As per spec table)
• No soware to install - congure using your normal web browser
• Use with a PC or Mac
• Display shows current, maximum and minimum readings
• Status and alarm indicators
• On-board alarm sounder
• Data can be uploaded to the EasyLog Cloud
ACCESSORIES
BAT 1V5 AAA 2 x AAA 1.5V alkaline
baeries
CABLE USB 3.1 C 1M USB A to USB C Cable
INCLUDED IN THE BOX
BAT 1V5 AAA 2 x AAA 1.5V alkaline
baeries
EL-SIE WALL BRACKET Mounng bracket
CABLE USB 3.1 C 1M USB A to USB C Cable
Temperature Measurement range -18 to +55°C (-0.4 to +131°F)
Resoluon 0.01°
Accuracy ±0.2°C (±0.36°F) typical
Long term stability <0.03°C (<0.054°F) / year
Measurement units °C, °F or K
Relave Humidity Measurement range 0 to 100%RH
Resoluon 0.1%
Accuracy ±1.5% typical (0 to 80%RH)
Long term stability <0.25%RH / year typical
Logging rate 10 seconds to 24 hours, user selectable
Start modes Immediate, push to start, triggered start,
parameter triggered
Memory capacity Over 1,000,000 total readings >500,000 readings per
channel
Sounder Integrated alarm sounder
Power source 2 x AAA 1.5V baery
Baery life >1 year (at 25°C with 10 minute logging rate)
Dimensions 93 x 42 x 17 mm (excluding bracket)
Operang temperature range -18 to +55°C (-0.4 to +131°F)
Environmental rang IP4X
SPECIFICATIONS
CALIBRATION CERTIFICATES AVAILABLE
Lascar oers a Traceable Calibraon Cercate Service on Temperature Data Loggers. Using reference equipment
which has been calibrated by a UKAS/NIST/CNAS accredited laboratory and using apparatus traceable to naonal or
internaonal standards. For more informaon please see www.lascarelectronics.com.

www.lascarelectronics.com
Issue 1 10/2019 Page 2 of 6
EL-SIE-2+
High Accuracy Temperature and Humidity
USB Data Logger, EasyLog Cloud Compable
NO SOFTWARE TO INSTALL
All the soware needed to congure your EL-SIE-2+, and view and analyse the data it logs, is contained within the product itself. Just
connect the logger to your PC or Mac with a USB cable, open any web browser and in the address bar type “hp://EasyLog.local”. No
internet connecon is needed, and you can save this address in your bookmarks or favourites as normal.
The interface is easy to use and allows the following parameters to be congured:
• Logger and channel names
• Measurement units
• Logging rate and start mode
• Up to 16 separate alarms (high/low/pre-alarm/cumulave)
with thresholds, delay and hold
• Display and LED indicator modes
Once the logger is running, you can plug it back
into your computer and see the latest data, device
status and the event log. You can also chose to
stop the logger and change the conguraon, or
just let it connue logging.
EASYLOG CLOUD DATA STORAGE
Store your data securely, and make it available from
any internet-connected PC or mobile device, with
EasyLog Cloud. The EL-SIE-2+ can upload logged data
to the Cloud from your PC or Mac, making sharing
and analysis easier than ever. For more informaon
and to set up an account on EasyLog Cloud, visit
www.easylogcloud.com.
Display Logger Status Explanaon
USB Connected The logger is connected via the USB cable
Push to Start The logger is set up for Push to Start logging, a long
press of either buon will start logging
Delayed Start The logger is set up for Delayed Start logging and will
automacally start logging at the specied me
Triggered Start The logger is set up for Triggered Start logging and
will start logging as soon as the specied limit (either
temperature, humidity or pressure) is reached
Low Baery The baeries are low and should be changed when
possible
Memory Full The memory is full and logging has stopped
Memory 90% The memory is 90% full, and data should be down-
loaded when possible
Calibraon Pend-
ing
The calibraon will expire in < 30 days
Calibraon
Expired
The calibraon has expired
Logger Running The logger is sll logging but can be stopped by a long
press of the boom buon
Alarm Triggered An alarm is currently acve on the logger
Cumulave Alarm A cumulave alarm is acve on the channel currently
being displayed
High Alarm A high alarm is acve on the channel currently being
displayed
Low Alarm A low alarm is acve on the channel currently being
displayed
Held High Alarm A high alarm is being held on the channel currently
being displayed
Held Low Alarm A low alarm is being held on the channel currently be-
ing displayed
High Pre-alarm A high pre-alarm is acve on the channel currently
being displayed
Low Pre-alarm A low pre-alarm is acve on the channel currently be-
ing displayed

www.lascarelectronics.com
Issue 1 10/2019 Page 3 of 6
EL-SIE-2+
High Accuracy Temperature and Humidity
USB Data Logger, EasyLog Cloud Compable
On power up, the LCD runs through a test sequence in which all elements are acvated, the LEDs light up and the sounder beeps.
Calibraon
reminder
DISPLAY STATUS INDICATION
The high-contrast LCD shows current,
maximum and minimum readings, as
well as alarm and logger status:
Min Reading Max Reading
Baery level
Reading Alarm status
Channel No Warning Units
Display Logger Status Explanaon
USB Connected The logger is connected
via the USB cable
Push to Start The logger is set up for
Push to Start logging, a
long press of either
buon will start logging
Delayed Start The logger is set up for
Delayed Start logging and
will automacally start
logging at the specied
me
Triggered Start The logger is set up for
Triggered Start logging
and will start logging as
soon as the specied
limit (either temperature,
humidity or pressure) is
reached
Low Baery The baeries are low and
should be changed when
possible
Memory Full The memory is full and
logging has stopped
Memory 90% The memory is 90%
full, and data should
be downloaded when
possible
Calibraon Pending The calibraon will expire
in < 30 days on the date
shown (can be DD/MM/
YYYY or MM/DD/YYYY)
Calibraon Expired The calibraon has
expired on the date
shown (can be DD/MM/
YYYY or MM/DD/YYYY)
Display Logger Status Explanaon
Logger Running The logger is sll logging
but can be stopped by a
long press of the boom
buon
Alarm Triggered An alarm is currently
acve on the logger
Cumulave Alarm A cumulave alarm is
acve on the channel
currently being displayed
High Alarm A high alarm is acve on
the channel currently
being displayed
Low Alarm A low alarm is acve on
the channel currently
being displayed
Held High Alarm A high alarm is being held
on the channel currently
being displayed
Held Low Alarm A low alarm is being held
on the channel currently
being displayed
High Pre-alarm A high pre-alarm is acve
on the channel currently
being displayed
Low Pre-alarm A low pre-alarm is acve
on the channel currently
being displayed

www.lascarelectronics.com
Issue 1 10/2019 Page 4 of 6
EL-SIE-2+
High Accuracy Temperature and Humidity
USB Data Logger, EasyLog Cloud Compable
LED AND SOUNDER STATUS INDICATION
The EL-SIE-2+ has three LEDs and a sounder to clearly indicate the status:
BUTTON FUNCTIONS
The two buons are used to navigate between display screens and control other funcons, some of which also create a record in an Event
Log, which can be viewed using the web browser.
If the display mode is set to Buon Press, pressing any buon wakes the display up, aer which it operates as described above.
LEDs Sounder Status
Acve Alarm / Memory Full /
Calibraon Expired
O Baery Low
LEDs Sounder Status
O Logger primed but not
yet logging / Pre-alarm
/ Memory 90% Full /
Calibraon Pending
(check display for
specic warning)
O Logger in operaon, no
alarms or warnings
Slow Flash
Flashing
Flashing
Flashing
Screen Buon Press Funcon Event Recorded
USB n/a n/a n/a n/a
Push to Start Any Long Start logging n/a
Triggered Start n/a n/a n/a n/a
Delayed Start n/a n/a n/a n/a
Channel –
Current reading
Top Short Move to next channel or STOP LOG n/a
Long Clear alarm hold for all channels Clear Held Alarms
Boom Short Show Min reading for this channel n/a
Long Mute alarm sounder Mute Alarm
Channel –
Min reading
Top Short Show Min reading for next channel or STOP
LOG
n/a
Long Reset Max/Min reading for all channels Clear Max/Min
Boom Short Show Max reading for this channel n/a
Long Mute alarm sounder Mute Alarm
Channel –
Max reading
Top Short Show Max reading for next channel or STOP
LOG
n/a
Long Reset Max/Min reading for all channels Clear Max/Min
Boom Short Show current reading for this channel n/a
Long Mute alarm sounder Mute Alarm
STOP LOG? Top Short Move to Channel 1 (or Warning Screen), also
generates an Audit Mark when cycling back
through to Channel 1
Audit Mark
Boom Long Stop logging, returns to Push to Start n/a
Warning Screen Any Short Move to next warning or Channel 1 n/a

www.lascarelectronics.com
Issue 1 10/2019 Page 5 of 6
EL-SIE-2+
High Accuracy Temperature and Humidity
USB Data Logger, EasyLog Cloud Compable
SENSOR ACCURACY AND INFORMATION
Typical and maximal tolerance for the temperature sensor in °C:
Operang Condions
The performance of the humidity sensor can be aected by long-term exposure to operang condions at the extents of the logger’s
range. The sensor shows best performance when operated within the recommended normal temperature and humidity range of 5 to 55°C
and 20 to 80%RH, respecvely. Long-term exposure to condions outside the normal range, especially at high humidity, may temporarily
oset the RH signal (e.g. +3%RH aer 60h kept at >80%RH). Aer returning into the normal temperature and humidity range the sensor
will slowly come back within the calibraon state by itself. Prolonged exposure to extreme condions may also accelerate ageing.
Typical tolerance of RH over Temperature:
Typical and maximal tolerance for the relave humidity sensor at 25°C:

www.lascarelectronics.com
Issue 1 10/2019 Page 6 of 6
EL-SIE-2+
High Accuracy Temperature and Humidity
USB Data Logger, EasyLog Cloud Compable
When tracking changes in ambient condions, the response me of the humidity sensor in your data logger is approximately 20 minutes
to reach 90% of the reading. However, if you are measuring step changes in humidity (for example if calibrang the product) it is advised
that you leave the unit for up to four hours to ensure that it has enough me to sele at the new level.
It is worth remembering that the value of relave humidity is of course sensive to temperature variaon. As an example, at a relave
humidity of ~90%RH at ambient temperature, a variaon in temperature of 1°C will result in a change of up to -5%RH. Therefore when
comparing mulple devices or calibrang them, any temperature variaons must be considered.
The humidity measuring element in the humidity data loggers can be contaminated through exposure to a variety of compounds. These
products should not be kept in proximity to volale chemicals such as solvents and other organic compounds. Generally speaking, if a
material or compound emits a strong odour you should not keep your humidity data logger in close proximity to it. If you would like more
informaon, please contact your local Lascar Electronics oce.
High levels of pollutants may cause permanent damage to the internal sensor.
BATTERY INFORMATION
We recommend that you replace the baeries annually, or prior to logging crical data. Use only AAA 1.5V alkaline baeries. Before
replacing the baeries, disconnect the logger from your computer.
The logger does not lose its stored data readings when the baeries are discharged or replaced. However, the logging process will stop
and will not resume unl the baeries are replaced and the logger is connected to your computer and the logger is started.
Note that while the logger is plugged into a computer it draws power from the USB port instead of the baeries, which can raise the
temperature of the logger slightly. It will return to normal shortly aer disconnecon.
Table of contents
Other Lascar Data Logger manuals