Lode Valiant Ultra User manual

Manual - Valiant Ultra
Sports Medicine
Lode BV
Zernikepark 16
NL 9747 AN Groningen
The Netherlands
t: +31 50 571 28 11
f: +31 50 571 67 46
@: ask@lode.nl
www.lode.nl
M938945 / VERSION: 1.1 - 2015-01-09

33
33
77
77
88
99
9
99
1010
1010
1010
1010
1111
1111
1212
1313
1515
Table of contentsTable of contents
1 - Intended Use1 - Intended Use
2 - Precautions2 - Precautions
3 - Contra Indication3 - Contra Indication
4 - Validity4 - Validity
5 - Definition of parts5 - Definition of parts
6 - Installation6 - Installation
6.1 - Emergency stop installation
7 - Control with an External Device7 - Control with an External Device
8 - Control with Control Unit8 - Control with Control Unit
9 - Control with the Lode Ergometry Manager (LEM)9 - Control with the Lode Ergometry Manager (LEM)
10 - Maintenance10 - Maintenance
11 - Troubleshooting11 - Troubleshooting
12 - Cleaning12 - Cleaning
13 - List of symbols used13 - List of symbols used
14 - Specification remarks14 - Specification remarks
15 - Accessories15 - Accessories
16 - Specifications16 - Specifications

1 - Intended Use1 - Intended Use
This device is intended to be used as a stress test device in a medical environment. The main goal of the
use of the devices is to create reproducible stress tests. With cycling ergometers typically workload
(watt) is imposed. With treadmills typically speed (km/h) and inclination (% grade) are imposed. This
product is designed both for manual operation and for control by external ECG-, pulmonary equipment.
Most likely this device will be used in conjunction with another medical device to obtain other important
physiological data, allowing a physician to evaluate a test subjects physical status.
The product may also be used for rehabilitation or active aging therapy.
As such the intended use of the device solely cannot be defined specifically.
The device has to be operated under the supervision of well-trained medical specialists in the field of
use.
The ultimate judgment whether a test subject should undertake a stress test with the device must be
made by the responsible medical specialist, based on the limitations of each individual, the medical
history and all other applicable circumstances. Neither the manufacturer nor its distributors assume any
responsibility for the final use of its equipment.
2 - Precautions2 - Precautions
Read this manual before using the device and follow it
carefully.
The operator should instruct the test subject prior to
performing an exercise protocol. If, at any time during
exercise, the test subject feels faint, dizzy, or experiences
pain, stop the test and he or she should be consulted by the
physician.
Service of this device is restricted to factory trained personnel
only.
3/16M938945 / VERSION: 1.1 - 2015-01-09

Do not jump onto the rotating belt. Do not jump off the
rotating belt (not even to the front). Do not stop moving on
the running belt. Do not turn around on the running belt. Do
not move sideways or backwards. Do not make movement
that could get you out of balance.
The operator should not touch accessible parts and the
patient simultaneously.
The test subject should always wear the belt of the lanyard
safety belt.
Replacement of the power supply cord should be installed by
authorised service personnel, instructions for correct
connection and anchoring must be ensured, securing that the
requirements of clause 8.11.3 of IEC 60601 1:2005 are met.
Set up and operate the device on a solid level surface.
The test subject should not wear loose or dangling clothing
while using the device.
The equipment has a safety earth (ground) connection and
must be connected to a (grounded) wall socket with
protective earth to avoid the risk of electric shock. The
functional earth connection is for potential equalisation only.
The operator must keep away from the footrail while
lowering.
4/16M938945 / VERSION: 1.1 - 2015-01-09

The test subject shall wear clothes when using the harness or
the lanyard belt to prevent skin irritation.
Replacement of parts can only be done by the manufacturer
or designated service personnel.
Read all warnings posted on the device.
Inspect the device for worn or loose components prior to
use. Tighten / replace any loose or worn components prior to
use.
Do not place any loose parts or objects on the device before
and during use.
The test subject should wear proper sports or running shoes
(no spikes).
If this equipment is modified, appropriate inspection and
testing must be conducted to ensure continued safe use of
the equipment.
Before each use, inspect the vest for damage. Do not use if
the integrity is questionable.
Cleaning and user maintenance shall not be made by children
without supervision.
5/16M938945 / VERSION: 1.1 - 2015-01-09

Children being supervised are not to play with the appliance.
Children shall not play with the appliance.
Heart rate monitoring systems may be inaccurate. Over
exercise may result in serious injury or death. If the test
person feels faint stop exercising immediately.
Keep away from the rollers.
Move the entrance plate away from the treadmill before
switching on
When using a bike on the treadmill, the subject needs to be
instructed never to use the brake. It is advised to disable
brakes on the bike.
No modification of this equipment is allowed.
Not suitable for use in the presence of flammable
anaesthetics.
When using skates or skeelers, subjects must be instructed
not to assume the braking position by place skates skeelers
angled to the rolling direction of the belt
6/16M938945 / VERSION: 1.1 - 2015-01-09

The ergometer should not be used by persons (including
children) with reduced physical, sensory or mental
capabilities, or lack of experience and knowledge.
The test subject should wear proper sports clothing. The
operator should check possible entrapment of clothing,
fingers or feet, before the start of an exercise test.
This appliance can be used by children aged from 8 years and
above and persons with reduced physical, sensory or mental
capabilities or lack of experience and knowledge if they have
been given supervision or instruction concerning use of the
appliance in a safe way and understand the hazards involved.
This device should only be sold by, or under the supervision
of authorized persons.
Make sure you always fix the handrails by tightening the
spanners on the side handrails before anybody starts using
the treadmill.
3 - Contra Indication3 - Contra Indication
The device is to be operated by classified personnel only. As stated in the intended use, the device is
intended to be used in a medical environment. During the intended use the test subject will deliver
energy. Application of the wrong dosis of energy could lead to permanent damage of the test subject
health. Only use this device under supervision of a medical physician.
4 - Validity4 - Validity
7/16M938945 / VERSION: 1.1 - 2015-01-09

This manual covers all Valiant 2 Ultra versions starting with the following serial numbers:
938925 Valiant Ultra: S/N 20150001 - ..
938930 Valiant Ultra: S/N 20150001 - ..
938945 Valiant Ultra: S/N 20150001 - ..
Date of issue: 2015-01-09
5 - Definition of parts5 - Definition of parts
1. Service Connector
2. Communication Module
3. On/Off switch
4. Equipotential bonding plug
5. Safety symbols tag
6. Serial number tag
7. Motor compartment for
height adjustment and drive
8. Motor compartment for
height adjustment
9. Walking surface [APPLIED
PART]
10. Front bar
11. Connector for external
emergency stop
12. Emergency Stop
13. Lanyard [APPLIED PART]
14. Magnetic contact for lanyard
15. Fallstop with harness
(optional for speed < 25 km/h)
[APPLIED PART]
16. Entrance plate (optional)
[APPLIED PART]
17. Extra external emergency stop (optional)
18. Control Unit (optional)
19. Colour Display 3.5" (optional)
20. Height adjustment positive +25% (optional)
21. Height adjustment negative -25% (optional)
22. Wheelchair guidance (optional)
23. Side bar (optional)
24. Network module (optional)
25. Floor protection frame (optional)
8/16M938945 / VERSION: 1.1 - 2015-01-09

6 - Installation6 - Installation
Unpack the device while leaving the transport straps in place.
Place the device in the location where it will be used.
Remove transport straps and remaining packaging materials.
Connect the power cord to the line output at the required voltage (see type plate).
Remove all objects from the device.
Switch on the unit with the on/off button and wait until the self-test is carried out. The product is ready
for use after you have heard a series of “beeps”.
- Use the on/off button in order to switch the ergometer off.
6.1 - Emergency stop installation6.1 - Emergency stop installation
When your treadmill is delivered with an emergency stop, please note the
following:
The plug of the emergency stop button needs to be placed in the back
connector of the console on the front handrail.
Some emergency stop buttons have a connector for another emergency
stop. You can place the spare dummy there in case you have one button.
Also make sure that the magnet of the emergency lanyard is connected to
the console.Otherwise the treadmill will not operate.
7 - Control with an External Device7 - Control with an External Device
If your device is fitted with a communication card you can connect to many external devices.
1. Connect the device with the supplied cable to the external device. You can control the device with e.g.
an exercise ECG device or Metabolic Card.
2. Download the interfacing instructions from the internet at www.lode.nl after you have registered your
device.
3. If your device has a control unit, select “Terminal”- mode to allow external commands. When the
device does not have a control unit the terminal mode is automatically selected. (See for detailed
instructions the user manual of your external device).
9/16M938945 / VERSION: 1.1 - 2015-01-09

8 - Control with Control Unit8 - Control with Control Unit
You can operate the device as a stand-alone device with the (optional) control unit.
1. After switching on, follow the instructions on the screen for the settings
2. Consult the manual of the control unit for more information
9 - Control with the Lode Ergometry Manager (LEM)9 - Control with the Lode Ergometry Manager (LEM)
1. If your treadmill ergometer has an RS232 or USB port connect it directly to the PC LEM is installed on.
If your device has a RJ45 connector plug it with an UTP cable to the special interface connector that was
delivered with your treadmill ergometer and to a USB port in your PC.
2. Start Lode Ergometry Manager on your PC
3. If your treadmill has a control unit, select “Terminal”- mode to allow external command. If your
treadmill does not have a control unit the terminal mode is automatically selected. (See for detailed
instructions the user manual of LEM).
10 - Maintenance10 - Maintenance
Maintenance should be carried out on a regular and planned basis. We recommend to check the unit
annually. This may be done by your local dealer. It is also recommended that a record of the service
history is kept for all activities relating to service and maintenance.
Maintenance and all repairs should only be carried out by an authorized agency. The manufacturer will
not be held responsible for the results of maintenance or repairs by unauthorized persons.
The check up and/or technical maintenance must be carried out conform the procedure described in the
service manual of the unit.
Opening of the equipment by unauthorized agencies is not allowed and will terminate any claim to
warranty.
Lode will make available on request circuit diagrams, component part lists, descriptions, calibration
instructions, or other information that will assist authorized service personnel to repair those parts of
the ergometer that are designated by Lode as repairable by authorized service personnel.
11 - Troubleshooting11 - Troubleshooting
For a complete overview of possible error message and the solutions belonging to these messages,
please consult the service manual of the ergometer.
10/16M938945 / VERSION: 1.1 - 2015-01-09

Read manufacturer’s guide, advises and instructions and manual
Emergency Stop
USB
RS232
Network (Proprietary Lode)
12 - Cleaning12 - Cleaning
Before cleaning and other maintenance of any part of the device, first switch off the device and
disconnect it from the mains. The surfaces can be cleaned with a damp cloth and a liquid (non-abrasive)
cleaning product.
General Cleaning
Wipe the device surface down with a cloth moistened with soap water or a disinfectant. The cloth should
not be dripping wet; do not allow liquids to enter the device
Cleaning the Saddle
Clean the saddle with a soft and dry or moist cloth. Disinfectants used should not contain any alcohol.
Cleaning the Upholstery (e.g. couch ergometer)
Wipe the upholstery down with a soft cloth moistened with soap water. The cloth should only be moist
and not dripping wet. If the cleaning agents and disinfectants used are caustic or contain alcohol, they
may damage and/or discolor the upholstery.
If you are not certain about the discolouring effect of a cleaning agent, you may try a little on a part of
the product that is not visible during normal use.
Lode cannot advise a specific cleaning agent, since local recipes may differ.
13 - List of symbols used13 - List of symbols used
11/16M938945 / VERSION: 1.1 - 2015-01-09

Potential Equalization Conductor
Symbol for collection, treatment, recycling and disposal of waste electrical and
electronic equipment (WEEE) as set out in Directive 2002/96/EC of 27 January 2003 of
the European Parliament and of the Council on waste electrical an electronic equipment
are necessary to reduce the waste management problems linked to the heavy metals
concerned and the flame retardants concerned
Class I MEDICAL EQUIPMENT Type B Electrical Safety IEC 60601-1
Trapping zone. Warning: do not jump onto the rotating belt. Do not jump off the
rotating belt (not even to the front). Do not stop moving on the running belt. Do not
turn around on the running belt. Do not move sideways or backwards. Do not make
movement that could get you out of balance. Keep away from rollers. The operator
must keep away from the footrail while lowering.
External Control Connector
"OFF" position of on/off switch: IEC 60417-5008 “OFF” (power)
"ON" position of the on/off switch: IEC 60417-5007 “ON” (power)
14 - Specification remarks14 - Specification remarks
12/16M938945 / VERSION: 1.1 - 2015-01-09

The specification in this manual is of the basic version of the product at the time of the publication of
this manual. The specification of all actual versions can be found on the Lode website. www.lode.nl
Changes to specifications can be made without prior notice.
Mentioned speed at the specifications in this manual is the standard speed supplied on the treadmill.
When optional speed upgrade or speed downgrade is installed on the treadmill, the actual speed of your
treadmill may differ from the speed mentioned at the specifications.
15 - Accessories15 - Accessories
Remote Control for
treadmill
Partnumber: 945831Partnumber: 945831
Ease of use
Remote Control for treadmill
Colour Display 3.5" -
2nd screen
Partnumber: 945819Partnumber: 945819
Multifunctionality
Colour Display 3.5"
SpO2 for control unit
with touch panel
(treadmill)
Partnumber: 945822Partnumber: 945822
Oxygen saturation
SpO2 for control unit with
touch panel
Blood Pressure
Measurement with ECG
trigger for ...
Partnumber: 945824Partnumber: 945824
with ECG trigger
Blood Pressure Measurement
with ECG trigger
Heart rate for
treadmills
Partnumber: 945820Partnumber: 945820
Heart rate in beats per
minute
Heart rate for treadmills
Emergency Stop
Button
Partnumber: 945804Partnumber: 945804
Ultimate safety
Emergency Stop Button
Autospeed
Partnumber: 945840Partnumber: 945840
Flexibility in exercising
Autospeed
Adjustable Acceleration
& Deceleration for
Katana & ...
Partnumber: 945847Partnumber: 945847
Flexibility in exercising
Acceleration & Deceleration
Reverse walking for
Valiant Ultra
Partnumber: 938843Partnumber: 938843
walking backwards
Reverse walking for Valiant
Ultra
Wheelchair support for
Valiant Ultra 450
Partnumber: 938872Partnumber: 938872
Versatile wheelchair
testing
Wheelchair support for
Valiant Ultra 450
13/16M938945 / VERSION: 1.1 - 2015-01-09

Ambient sensor pack
Partnumber: 945826Partnumber: 945826
Check environmental
conditions during test
Ambient sensor pack for
treadmills
Bottle holder
Partnumber: 401200Partnumber: 401200
Convenient beverage
storage
Bottle holder
Extension for
emergency lanyard
Partnumber: 945931Partnumber: 945931
Original Accessory
Extension for emergency
lanyard
Table foot for Control
Unit with Touchscreen
Partnumber: 945818Partnumber: 945818
Control from any
desired place
Support for Control Unit with
Touchscreen
Entrance plate Valiant
Ultra
Partnumber: 938852Partnumber: 938852
Easy entrance
Entrance plate Valiant Ultra
14/16M938945 / VERSION: 1.1 - 2015-01-09

16 - Specifications16 - Specifications
Workload
Maximum speed 40 km/h 24.9 mph
Minimum operational speed 01 km/h 0.6 mph
Speed adjustment steps 0 km/h 0 mph
Positive elevation 15 %
Negative elevation -15 %
Comfort
Allowed user weight 225 kg 496 lbs
User Interface
English user interface
Norwegian user interface
Czech user interface
Danish user interface
Dutch user interface
Finnish user interface
French user interface
German user interface
Italian user interface
Japanese user interface
Korean user interface
Polish user interface
Portugese user interface
Russian user interface
Spanish user interface
Turkish user interface
Ukrainian user interface
Readout Angle
Readout Target Angle
Readout Distance
Readout Speed
Readout Energy
Readout Time
Set Display
Customized Screenviews
Manual operation mode
Quick start operation mode
Touchscreen
15/16M938945 / VERSION: 1.1 - 2015-01-09

Connectivity
Lode 38K4 interface protocol
Lode interface protocol
Lode WLP interface protocol
"HP Cosmos" interface protocol
"Woodway" interface protocol
"Trackmaster" interface protocol
"GE" interface protocol
USB connector
Analog connector
SpO2 connector
RS232 in connector
RS232 out connector
Dimensions
Walking surface length 450 cm 177.2 inch
Walking surface width 300 cm 118.1 inch
Step up height 35 cm 13.8 inch
Product length (cm) 455 cm 179.1 inch
Product width (cm) 395 cm 155.5 inch
Product height 130 cm 51.2 inch
Power requirements
Power cord 3L+N+PE
Power cord NEMA
Maximum rated power input 8000 VA
3 x 400 V
Maximum motor power 8 kW
Standards & Safety
ISO 9001:2008 compliant
Integrated emergency stop
Standard emergency lanyard
Environmental conditions
Maximum operational temperature 40 °C
Minimum operational temperature 14 °C
Maximum operational air pressure 106 kPa
Minimum operational air pressure 70 kPa
Maximum operational non-condensing
humidity 90 %
Minimum operational non-condensing
humidity 30 %
Maximum storage & transport temperature 70 °C
Minimum storage & transport temperature -25 °C
Maximum storage & transport air pressure 106 kPa
Minimum air pressure storage & transport 50 kPa
Max. humidity storage & transport 95 %
Min. humidity storage & transport 10 %
Order info
Partnumber 938945
16/16M938945 / VERSION: 1.1 - 2015-01-09
Other manuals for Valiant Ultra
1
Table of contents
Other Lode Treadmill manuals

Lode
Lode Katana Sport User manual

Lode
Lode Valiant 2 cpet User manual

Lode
Lode Edan ST-1212 Stress User manual

Lode
Lode CORIVAL CPET Operator's manual

Lode
Lode OEM User manual

Lode
Lode Schiller AT-10 Operator's manual

Lode
Lode ERGOMETERS User manual

Lode
Lode Edan EC-12 EXPRESS Operator's manual

Lode
Lode Valiant Ultra User manual

Lode
Lode Valiant 2 Rehab User manual