LogTag UTRID-16 User manual

This quickstart guide covers preparation,
use and maintenance for the following model:
UTRID-16
1. SOFTWARE SETUP
2. LOGTAG®CONFIGURATION
Software requirements vary depending on the type of file you want it to
generate when plugged into a USB port.
Loggers can be ordered pre-configured, ready to start. Different
profiles are available; if none of these suit or the logger needs different
configuration settings to those already installed, you will need to
download and install our freely available companion LogTag®Analyzer
(version 2.8 or higher) software from our website: www.logtagrecorders.
com/software. Simply browse to our software page, fill in your details and
begin your download.
Once it has finished downloading, install LogTag®Analyzer to your
computer and start the program once it is installed. (For detailed
software download and install instructions, please refer to the LogTag®
Analyzer User Guide).
It is recommended that you connect each USB PDF Temperature logger
via a permanently plugged in USB extension cable to protect your
computer’s USB socket. Please refer to
www.logtagrecorders.com/support if you are having issues.
With the Software running, the easiest way to configure your LogTag®
for use is with the ‘LogTag Wizard’ which can be accessed by
pressing the ‘F2’ key on your keyboard or you can browse via the top
navigation menu: ‘LogTag’ >‘Wizard’
Note: You can configure many devices at the same time, however,
it is practical to limit the number of devices to about 10, using a
powered USB hub. It is not possible to configure different models at
the same time.
Advanced Configuration Options
Click Advanced Settings for additional configuration settings.
These settings decide which files -if any- are generated at the end
of the trip and also determine the appearance and contents of the
generated files.
Alarm Configuration
You need to observe some basic rules when entering alarm
conditions into the configuration screen in LogTag®Analyzer:
• A primary alarm condition must be entered before the
corresponding secondary alarm condition.
• Equally, primary and secondary alarm conditions must be entered
before the corresponding tertiary alarm condition.
• You can enter a different number of upper and lower conditions,
or only upper, or only lower conditions, or none at all.
• A secondary upper alarm cannot be entered with a lower
threshold value than the primary upper alarm.
• A tertiary upper alarm cannot be entered with a lower threshold
value than the secondary alarm. Threshold values can, however,
be equal and combined with different time-to-activation values.
For example, you can enter a primary upper alarm with 8°C
threshold and 10 accumulative readings, plus an alarm with 8°C
threshold and 5 consecutive readings. In this instance the alam
will be triggered, if either 10 readings in total are above 8°C, or 5
consecutive readings above 8°C have occurred.
Any alarm condition that is activated will be shown on the PDF, in the
software and on the display.
“Standard Configuration”
UTRID-16 Quickstart Guide
Page 1 of 3
1. Provide a ‘Description’ to identify the logger being configured.
2. Select Push button or Time/Date start.
3. Enter the recording interval you wish to use. The duration
of the shipment and/or number of readings to record changes
according to the recording interval entered.
4. (Optional) Enter a start delay before the logger begins recording.
5. LogTag®Analyzer allows up to 6 different alarm trigger conditions
when configuring a UTRID-16 for recording, depending on the
model. Each alarm trigger condition consists of a threshold
temperature value, an activation type (which can be instant,
1
2
3
4
consecutive or accumulative) and a delay time, if it is not an
instant alarm. The UTRID-16 can have up to three upper alarms,
and up to three lower alarms.
5
UTRID-16 Quickstart Guide
Page 2 of 3
After configuration with LogTag®Analyzer for push button start the word READY is shown,
together with the current time in 24 hour format.
Press the START/Mark button until READY disappears and REC starts flashing.
Release START when REC stops flashing.
REC will be displayed to confirm the logger is recording.
The UTRID-16 now records temperature data. This start option works the same as in
LogTag®LED products. If the logger was configured for push button start with a start delay
(regardless of quickstart settings), the word DELAY is shown instead of the symbol.
During normal operation the display shows the most recently recorded temperature. The
current time is also displayed in 24 hour format. A tick symbol is shown as long as no
alarm event has occured.
When you press the START/Mark button while the UTRID-16 is recording temperature,
an inspection mark is recorded against the next temperature reading. The MARK symbol
appears and remains shown until the next reading is taken.
What happens when you plug your LogTag®into a USB port depends on the operating
system of the computer, the settings made during configuration and whether or not
LogTag®Analyzer is running.
If the LogTag®was configured to generated files, a new drive or mounted device will
appear. The device name will be created from the serial number of the logger. You can
access the files by browsing to the newly created drive and double-clicking the PDF, CSV
or LTD files. For PDF files you need Adobe Acrobat Reader or a similar PDF viewer.
To open the LTD file you need to install the free LogTag®Analyzer software. CSV files can
be opened with a text editor, or imported into a spreadsheet program such as Microsoft
ExcelTM.
You can unplug the logger at any time; there is no need to stop the device.
3. START / INSPECT
4. DOWNLOAD / RESULTS
DATA EVALUATION - REPORT
HIBERNATION
To prolong the life of your logger you can hibernate it between trips via
LogTag®Analyzer when it is not being used for extended periods. Please
refer to the User Guide in the section about Hibernation.
CALIBRATION
All LogTag®products can be calibrated with special utility software,
which will be made available to calibration labs upon request.
Re-calibrated loggers can be identified and traced back to the
calibrating lab thus reducing the risk of fraudulent data tampering.
WATER INGRESS
Please carefully observe the IP rating for your unit and remember that
none of the LogTag®products can be immersed into liquids.
ALERT/OK INDICATOR
The symbol is shown as soon as the UTRID-16 has registered an
alarm event. While there are no alarms, the symbol is shown.
RECORDING INDICATORS
The recording indicators show what data the UTRID-16 is currently
collecting.
• If the word READY is shown, the UTRID-16 is ready to be started.
Depending on the configuration it may record pre-start readings.
• If the symbol is shown, the UTRID-16 is recording temperatures
at the sample interval defined during configuration with LogTag®
Analyzer.
• If is shown together with the word PAUSED, the product is also
recording, but the recorded values are not taken into account when
calculating alarm events and durations.
If the word STOPPED is shown, the UTRID-16 has finished recording
temperature data.
MARK INDICATOR
The word MARK is shown when the next temperature reading is
identified with an inspection mark in the recorded data.
CLOCK
The clock display is used to show either:
• current time,
• time remaining to start of a delayed start or
• duration, for example of an alarm.
The display identifies which of those is displayed as follows:
• If the word DELAY is shown, the time value represents a start delay,
or the time remaining until a date/time start will occur.
• If the word DURATION is shown, the time value represents the
duration, for example the time above the upper alarm limit.
• If the word DAYS is shown, the clock represents the number of days
in a delay or duration.
DISPLAY OVERVIEW
Page 3 of 3
• If the word HOURS is shown, the clock represents the number of
hours and minutes in a delay or duration. If none of the above 4 words
are displayed, the clock shows the current time.
MIN/MAX INDICATOR
The word MAX is shown when the temperature on the display represents
the maximum recorded temperature for the current trip.
The word MIN is shown when the temperature on the display represents
the minumum recorded temperature for the current trip.
ALARM THRESHOLD MARKERS
The upper alarm threshold marker is shown if an alarm event occurred
at one of the upper temperature limits. It is also shown when one of the
upper thrershold values is shown on the review screen.
The lower alarm threshold marker is shown if an alarm event occurred
at one of the lower temperature limits. It is also shown when one of the
lower threshold values is shown on the review screen.
TEMPERATURE UNITS
Depending on the selected PDF temperature units, this shows either °F
or °C.
TEMPERATURE VALUE
This 4 character, 7-segment display shows the last recorded temperature
while the UTRID-16 is recording readings. Once the logger has stopped,
nothing will be displayed.
ALARM ARROWS
The up-arrow is shown when the temperature displayed is above the
specified upper temperature limit. The down-arrow is shown when the
temperature displayed is below the specified lower temperature limit.
The four smaller up/down-arrows are secondary and tertiary alarm
symbols. They are used when more than one upper or more than one
lower alarm is defined and are shown then the current temperature is
above or below the corresponding alarm levels.

Other LogTag Other manuals

LogTag TRED30-7 User manual

LogTag

LogTag TRED30-7 User manual

LogTag TREX-8 User manual

LogTag

LogTag TREX-8 User manual

Popular Other manuals by other brands

Reer 46760 manual

Reer

Reer 46760 manual

Formech IMD508 Installation, operating and service manual

Formech

Formech IMD508 Installation, operating and service manual

Honeywell BFW112 Operating and maintenance manual

Honeywell

Honeywell BFW112 Operating and maintenance manual

Deer Fence Access Gate installation instructions

Deer Fence

Deer Fence Access Gate installation instructions

Velleman VMM500 user manual

Velleman

Velleman VMM500 user manual

Sicor MR17 Operation and maintenance manual

Sicor

Sicor MR17 Operation and maintenance manual

GEZE SecuLogic GCER 100 manual

GEZE

GEZE SecuLogic GCER 100 manual

DMP Electronics 1116 installation guide

DMP Electronics

DMP Electronics 1116 installation guide

Ironman4x4 BBT034 installation guide

Ironman4x4

Ironman4x4 BBT034 installation guide

AeroScout T2 Tag user guide

AeroScout

AeroScout T2 Tag user guide

Snap-On Solus user manual

Snap-On

Snap-On Solus user manual

Promate Lumino user guide

Promate

Promate Lumino user guide

Edimax EU-4306 Declaration of conformity

Edimax

Edimax EU-4306 Declaration of conformity

ReelCraft 4000 Series operating instructions

ReelCraft

ReelCraft 4000 Series operating instructions

Red Sea REEFER DLX 425 Assembly manual

Red Sea

Red Sea REEFER DLX 425 Assembly manual

Arregui Supra Instruction manual and guarantee

Arregui

Arregui Supra Instruction manual and guarantee

Panasonic TY-42TM4Z installation manual

Panasonic

Panasonic TY-42TM4Z installation manual

GE HFA51A instructions

GE

GE HFA51A instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.