LR DM 80-UMS User manual

OPERATING INSTRUCTIONS
Digital
Pressure Gauge
DM 80-UMS

Described product
Manufacturer
LEITENBERGER GmbH, Bahnhofstrasse 32/33
72138 Kirchentellinsfurt, Germany
Legal notice
This work is protected by copyright. All associated rights remain with LEITENBERGER GmbH. Reproduc-
tion of the work or parts thereof is permitted only within the limits of the statutory provisions of copyright
law. Any modication, abridgement or translation of the work without the express written consent of
LEITENBERGER GmbH is prohibited. We reserve the right to make product changes with respect to
appearance, dimensions, weights, functionality, properties and performance in the course of improvement
and adaptation to the state of the art. No right to correction or subsequent delivery of already delivered
products is thereby implied. Deletions may be made at any time without any right to a legal claim arising.
All information on use and safety are non-binding. They do not replace any legal provisions or regulations
of the employers' liability insurance association. The brand names mentioned in this document are the
property of their respective owners.
©LEITENBERGER GmbH. All rights reserved. Translation of the original document. This document is a
translation of the original LEITENBERGER GmbH document.
- 2 - DM80UMS_MDM08_V2.0_0720_en

- 3 -
DM80UMS_MDM08_V2.0_0720_en
Table of Contents
1. Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.1 Safety notes and regulations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
1.2 Explanation of symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.3 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.4 Disclaimer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
1.5 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
2. Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.1 Device versions / delivery scope . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.2 Device view, buttons and functions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
2.3 Menu functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3. Prior to commissioning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.1 Check the scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.2 Prepare the digital pressure gauge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
3.3 Connect the digital pressure gauge. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
4. Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.1 Commissioning (switching on) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.2 Switching on/o . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.3 Switching the lighting on/o . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.4 Setting/resetting the measurement unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
4.5 Setting the lter function (attenuation) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.6 Logging (setting/changing the sampling rate) . . . . . . . . . . . . . . . . . . . . . . . . . . 10
4.7 FS (function menu reserved by the manufacturer) . . . . . . . . . . . . . . . . . . . . . . . . .11
4.8 Setting the date and time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
4.9 Displaying the software version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.10 Displaying the MIN/MAX pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.11 Recording measurement values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.12 Charging the battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
4.13 Transferring the measurement data to the PC . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5. Handling instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.1 Care and storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
5.2 Environmentally friendly disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
6. Technicaldata,specications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
7. Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16

- 4 - DM80UMS_MDM08_V2.0_0720_en
1. Safety
1.1 Safety notes and regulations
To ensure safe and reliable work with the digital pressure gauge and to protect the user from
injuries, the following points are to be observed:
- Read the operating instructions carefully.
- The digital pressure gauge may only be operated and installed in systems by
trained specialists who possess the necessary expertise and qualications
to perform these tasks, and who are familiar with the applicable standards,
regulations and requirements of the eld of application.
- The operating instructions are an integral part of the digital pressure gauge.
Store the operating instructions in a safe location so that you can quickly nd the
desired information if necessary.
- The legal provisions and regulations as well as those of the system/vehicle
manufacturer are generally to be observed.
- Prior to commissioning, perform a check to ensure that the digital pressure gauge
is in faultless condition. Do NOT operate the digital pressure gauge if damaged.
- No technical modications to the digital pressure gauge may be made.
- Protection class according to DIN 60529: The ambient conditions at the place
of use must not exceed the requirements for the specied protection class (see
Technical data).
- Do not subject the digital pressure gauge to any extreme temperatures, direct
sunlight, extreme humidity or wet.
- Observe the measurement range limits and temperature ranges.
- Do not exceed the maxim measurement range limits and temperature ranges of
the digital pressure gauge at any time.
- The maximum pressure/load range of the system or component with the lowest
value is the decisive maximum value for the overall pressure/load range.
- Do not exceed the maximum pressure/load value of the smallest component of
the system at any time.
- Always observe the nominal pressure specications of the built-in measuring
connections and lines as well as the specied safety factors.
- Repairs to the digital pressure gauge may only be performed by the manufac-
turer.
- Use only original spare parts and accessories.
- Non-specied use of the product or failure to observe the safety and operating
notes may lead to severe malfunctions as well as personal injury and property
damage.
- The discharge of media or hot liquids that are under pressure may cause injuries.
- Pressurised systems may burst and cause parts to be ejected.
- Wear personal protective equipment, such as protective goggles, gloves, safety
shoes, etc. when working on pressurised systems and components.

- 5 -
DM80UMS_MDM08_V2.0_0720_en
1.2 Explanation of symbols
Warning notices and important information are marked in this document with symbols for better
identication.
The notices begin with signal words that indicate the extent of the danger. You must absolutely comply
with these notices and information and handle the device with care to prevent accidents and avoid perso-
nal injury and property damage.
DANGER
Notice indicating a dangerous situation which, if not avoided, will result in death or severe
injury.
WARNING
Notice indicating a dangerous situation which, if not avoided, could result in death or severe
injury.
CAUTION
Notice indicating a dangerous situation which, if not avoided, could result in minor to
moderate injury.
Attention
Notice indicating a dangerous situation which, if not avoided, could result in property
damage.
Note
Notice including useful tips, recommendations and information for ecient and trouble-free
operation of the digital pressure gauge.
1.3 Intended use
The DM 80-UMS digital pressure gauge is used for the mobile and stationary measurement of pressures,
within the approved ranges (see technical data).
WARNING
Danger from improper use.
Any improper use and failure to observe safety and operating notices can result in serious
malfunctions as well as personal injury and property damage.
–Only use the DM 80-UMS digital pressure gauge as intended.
–All information in the operating instructions and safety data sheets must be strictly complied with.
1.4 Disclaimer
Failure to comply with the intended use will immediately void any warranty and guarantee claims against
the manufacturer. No liability is accepted for damage or malfunctions caused by improper use, assembly
errors or failure to observe the operating instructions.
1.5 Ambient conditions
The DM 80-UMS digital pressure gauge is only intended for use in commercial workshops. It is not
approved for ATEX areas and therefore must not be used in potentially explosive areas.
The climatic requirements correspond to the conditions typically prevailing in Central Europe. Optimum
function is ensured between -10 °C and +50 °C. The device must be stored within a temperature range of
-20 °C to + 50 °C.
The DM 80-UMS digital pressure gauge must be secured against falling down.

- 6 - DM80UMS_MDM08_V2.0_0720_en
2. Product description
The DM 80-UMS is a digital pressure gauge with actual value display as well as min/max display and
accuracy class 0.5 FSO. The MIN/MAX memory is permanently updated and overwritten. The sampling
rate is 10 milliseconds (100 measurement values/second) for measuring dynamic pressure peaks. The
integrated data logger saves the measured data. The data storage is 8 GB. The data is accessed via the
Micro-USB interface in the form of a .csv le.
2.1 Device versions / delivery scope
– DM 80-UMS Digital pressure gauge with rubber protection cap (green) and Micro USB cable
2.2 Device view, buttons and functions
Display:
4.5-digit display with background lighting. Display of measurement values and menu functions.
Display mode:
The actual value display shows the current pressure. The measured value is displayed in the set unit. The
MIN, MAX or FS value and battery status are shown at the bottom of the display.

- 7 -
DM80UMS_MDM08_V2.0_0720_en
Item Designation Button Function
①Bargraph
The bar graph graphically depicts the maximum measuring span.
The stored MAX value is displayed as a static, individual bar
(= drag pointer function). The display is updated at an interval of
10 milliseconds. (100 measurements / second)
②
Actual value display,
measured value and
pressure unit
Displays the current pressure.
This display is updated at an interval of 300 milliseconds. (3
displays / second) Displays the unit set.
③Display MIN/MAX/FS Displays the display value selected: MIN, MAX or full scale (FS).
④REC/RESET/OK
button
Activation of data recording or data logger.
Deleting the MIN and MAX measured data storage.
Conrmation of the MENU functions
⑤MENU/ZERO button Short press of button: menu
Long press of button (3 seconds): nullication
⑥MIN/MAX/FS button
Selection of the display value: MIN, MAX or full scale (FS)
minimum value / maximum value:
Pressure peak / Display of the maximum measuring span
(e.g. 60,00 bar)
⑦ON/OFF button
Short press of button: switch device on/o
Long press of button (3 seconds)
Background lighting on/o (for 20 seconds)
⑧Battery symbol Displays the battery status (fully charged = 5 segments)
⑨Charge control
indicator
Illuminates red when the battery is charging and green when the
battery is fully charged.
⑩Micro USB socket
(female) For battery charging and data transmission
⑪Rubber protective
cap Protection against damage to the device

- 8 - DM80UMS_MDM08_V2.0_0720_en
2.3 Menu functions
The following settings are made using the MENU function
–unit: Set the measurement unit (bar, mbar, PSI, kPa, MPa)
–Filt: Set the lter function (attenuation)
–LOG: Set the sampling rate
–FS: Function menu reserved by the manufacturer (locked)
–date: Set the date and time
–Sw: Shows the software version
–Press the "MENU" button to adjust the parameters and the
"REC/RESET/OK" button to save the respective parameter.
Will then jump to the next parameter.
–Press the "REC/RESET/OK" button continuously
(approx. 3 seconds) to exit the function menu.
If the user is in the function menu and does not activate any action
for 20 seconds, he also exits from the function menu and all set
parameters are saved.
Overow
If the device displays "OFL" (overow), this means that the existing
pressure is too high.
The device switches o automatically after 5 minutes if there is a
maximum pressure change of 2 % of the respective end value.
3. Prior to commissioning
3.1 Check the scope of delivery.
3.2 Prepare and connect the digital pressure gauge.
NOTE
Charge the battery fully before using the digital pressure gauge for the rst time.
- Charge the battery fully, see point 4.13
CAUTION
Risk of injury due to leaking liquids or gases.
The discharge of media or hot liquids that are under pressure may cause injuries.
Risk of injury due to ejected parts.
Pressurised systems may burst and cause parts to be ejected.
Do not exceed the maximum pressure/load value of the smallest component of the system
at any time.
Never open pressurised ttings.
Wear personal protective equipment, such as protective goggles, gloves and safety shoes when working
on pressurised systems.

- 9 -
DM80UMS_MDM08_V2.0_0720_en
Attention
Follow the instructions and specications of the system manufacturer regarding the
tightening torques for ttings, adapters, etc.
Install the pressure gauge at the measuring point of the system. To do so, loosen or open the tting and
install the pressure gauge. Tighten ttings to the specied torque. Do not exceed the maximum torque of
the pressure gauge connection.
4. Operation
4.1 Commissioning (switching on)
Switching on the pressure gauge:
Briey press the ON/OFF button. The pressure gauge is ready.
4.2 Switchingo
Briey press the ON/OFF button.
4.3 Switchingthelightingon/o
Press and hold the ON/OFF button for approx. 3 seconds.
4.4 Setting/resetting the measurement unit
Measurement units: bar, PSI, kPa, MPa, mbar
Press the MENU/ZERO button to start the menu function.
Press the REC/RESET/OK button and open the menu item.
Briey press the MENU/ZERO button to call up the desired measurement unit.
(The next unit is called up by briey pressing the MENU/ZERO button again)
Briey press the REC/RESET/OK button, save the selection and call up the lter function.
Option: Press and hold the REC/RESET/OK button for 3 seconds to exit the menu.

- 10 - DM80UMS_MDM08_V2.0_0720_en
4.5 Settingthelterfunction(attenuation)
The lter function serves to attenuate the measurement value display.
The following lter functions can be set:
0 = no attenuation
1 = weak attenuation
2 = medium attenuation,
3 = strong attenuation
Press the MENU/ZERO button to start the menu function.
(This does not apply if the menu item is already open)
Keep pressing the REC/RESET/OK button until the menu item is displayed.
(This does not apply if the menu item is already open)
Briey press the MENU/ZERO button to call up the desired attenuation value.
(The next value is called up by briey pressing the MENU/ZERO button again)
Briey press the REC/RESET/OK button, save the selection and call up the logging function.
Option: Press and hold the REC/RESET/OK button for 3 seconds to exit the menu.
4.6 Logging (setting/changing the sampling rate)
For capturing and logging the pressure peaks.
Selection: 10 ms, 100 ms, 1 s, 10 s.
Press the MENU/ZERO button to start the menu function.
Keep pressing the REC/RESET/OK button until the menu item is displayed.
Briey press the MENU/ZERO button to call up the desired sampling rate.
(The next rate is called up by briey pressing the MENU/ZERO button again)
Briey press the REC/RESET/OK button, save the selection and call up full scale.
Option: Press and hold the REC/RESET/OK button for 3 seconds to exit the menu.

- 11 -
DM80UMS_MDM08_V2.0_0720_en
4.7 FS (function menu reserved by the manufacturer)
Press the MENU/ZERO button to start the menu function.
Keep pressing the REC/RESET/OK button until the menu item is displayed.
Briey press the REC/RESET/OK button and call up the date/time.
Option: Press and hold the REC/RESET/OK button for 3 seconds to exit the menu.
4.8 Setting the date and time
Press the MENU/ZERO button to start the menu function.
Keep pressing the REC/RESET/OK button until the menu item is displayed.
Press the MIN/MAX/FS button & MENÜ/ZERO MENU/ZERO button simultaneously
and call up the year (YY).
Press the MIN/MAX/FS button to lower the value. ▼
Press the MENU/ZERO button to increase the value. ▲
Press the REC/RESET/OK button and month (MM); day (TT); time [HH] and minutes (MM).
Set the values as described in how to set the year.
Press the REC/RESET/OK button and save the settings. The menu item is displayed.
Briey press the REC/RESET/OK button, save the selection and call up the software version.
Option: Press and hold the REC/RESET/OK button for 3 seconds to exit the menu.

- 12 - DM80UMS_MDM08_V2.0_0720_en
4.9 Displaying the software version
Press the MENU/ZERO button to start the menu function.
Keep pressing the REC/RESET/OK button until the menu item is displayed.
The software version is displayed.
Briey press the REC/RESET/OK button, conrm the measurement unit and call up
the measurement unit.
Option: Press and hold the REC/RESET/OK button for 3 seconds to exit the menu.
4.10 Displaying the MIN/MAX pressure
Press the MIN/MAX/FS button to call up the following values:
MIN (lowest measured pressure)
MAX (highest measured pressure)
FS (full scale = maximum measuring span)
4.11 Recording measurement values
Press the MIN/MAX/FS button to start logging.
REC is shown on the right of the display.
Note: During logging it is not possible to call up the menu with the MENU/ZERO button.
4.12 Charging the battery
Use the USB cable and a commercially available USB charger* (230 VAC). Plug the USB charger into a
220 V socket. Insert the A plug of the USB cable into the USB charger and the micro-plug into the USB
socket on the digital pressure gauge. The battery charge control indicator lights up red when charging and
green when the battery is fully charged, at which point the charger should be removed.
(*USB charger not included in scope of standard delivery)
4.13 Transferring the measurement data to the PC
Use the USB cable. Insert the A plug of the USB cable into a USB socket on the PC and the micro-plug
into the USB socket on the digital pressure gauge. Switch on the digital pressure gauge. The digital pres-
sure gauge is recognised as a USB mass storage device. The measurement data is saved in a CSV le.

- 13 -
DM80UMS_MDM08_V2.0_0720_en
5. Handling instructions
5.1 Care and storage
To enable eective working, we designed the digital pressure gauge to be low maintenance. Never-
theless, you should still observe a few notes. This helps to ensure trouble-free operation and to preserve
the value of the technology.
Note:
Aggressive cleaning agents or solvents may damage the digital pressure gauge or individual
components.
Do not use aggressive or abrasive cleaning agents, solvents or similar chemicals for cleaning.
–Keep dust and dirt away from the digital pressure gauge.
–Clean the digital pressure gauge with a soft, slightly moistened cloth.
–Pay attention to the self-discharging of the battery. (We recommend charging the battery every 8 weeks).
–Battery warranty 6 months / device warranty 12 months.
–Store your product in a dry, dust-protected environment. Avoid places with higher temperatures and
moisture or places which can become wet, also for maintenance.
–Keep the original packaging to avoid damage during transportation.
5.2 Environmentally friendly disposal
Recycling according to WEEE (EU Directive 2002/96 EC)
You can optionally return the digital pressure gauge to us for disposal.
This digital pressure gauge or its components must not be disposed of as normal waste.
If you prefer not to return the digital pressure gauge to us for disposal, you are required to bring the
device to a specialised centre for the separate collection and disposal of hazardous and special waste.
The digital pressure gauge contains an electronic circuit board and a battery. These parts must be
removed before the device is scrapped.

- 14 - DM80UMS_MDM08_V2.0_0720_en
6. Technicaldata,specications
Feature Unit Measurement range
Measurement range relative bar -1…+16 0…25 0…40 0…60 0…100 0…250 0…400 0…600 0…1000 0…2500
Measurement range absolute bar - - - -------
Measurement range relative mbar - - - -------
Overpressure Pmax bar 32 50 80 120 200 500 800 1200 2000 3750
Bursting pressure bar 80 125 200 300 500 1000 2000 3000 5000 5000
Accuracy % FS ±0.5
Ambient temperature °C -10…+50
Storage temperature °C -20…+60
Media temperature °C -40…+125
Maximum relative humidity % RH 85
Materials
Pressure
connection 1.4301
Membrane 1.4548
Seal welded
Test connection - 1.078 - - - 1.078 - 1.078 1.078 1.078 -
Seal for test
connection -NBR- - - NBR - NBR NBR NBR -
Compensated range °C 0…+50
Error in compensated range % FS / °C ≥ ±0.035
Pressure cycles minimum - 100 x 10⁶
Battery - Li-Ion
Max. battery life 500 load cycles @ 23 ±5°C ≥ 800 mAh
IP protection class - IP40
Units - bar, PSI, kPa, MPa, mbar bar, PSI, kPa, MPa bar, PSI, MPa
Standard pressure connection - G1/2’’ EN837
Test connection - M16x2 ---M16x2 -M16x2 M16x2 M16x2 -
Approx. weight g 450
Approx. dimensions L x W x H mm 140 x 90 x 45
Approx. dimensions L x W x H
(with test connection) mm 180 x 90
x 45 ---180 x 90
x 45 -180 x 90
x 45
180 x 90
x 45
180 x 90
x 45 -

- 15 -
DM80UMS_MDM08_V2.0_0720_en
6. Technicaldata,specications
Feature Unit Measurement range
Measurement range relative bar -1…+16 0…25 0…40 0…60 0…100 0…250 0…400 0…600 0…1000 0…2500
Measurement range absolute bar - - - -------
Measurement range relative mbar - - - -------
Overpressure Pmax bar 32 50 80 120 200 500 800 1200 2000 3750
Bursting pressure bar 80 125 200 300 500 1000 2000 3000 5000 5000
Accuracy % FS ±0.5
Ambient temperature °C -10…+50
Storage temperature °C -20…+60
Media temperature °C -40…+125
Maximum relative humidity % RH 85
Materials
Pressure
connection 1.4301
Membrane 1.4548
Seal welded
Test connection - 1.078 - - - 1.078 - 1.078 1.078 1.078 -
Seal for test
connection -NBR- - - NBR - NBR NBR NBR -
Compensated range °C 0…+50
Error in compensated range % FS / °C ≥ ±0.035
Pressure cycles minimum - 100 x 10⁶
Battery - Li-Ion
Max. battery life 500 load cycles @ 23 ±5°C ≥ 800 mAh
IP protection class - IP40
Units - bar, PSI, kPa, MPa, mbar bar, PSI, kPa, MPa bar, PSI, MPa
Standard pressure connection - G1/2’’ EN837
Test connection - M16x2 - - - M16x2 -M16x2 M16x2 M16x2 -
Approx. weight g 450
Approx. dimensions L x W x H mm 140 x 90 x 45
Approx. dimensions L x W x H
(with test connection) mm 180 x 90
x 45 - - - 180 x 90
x 45 -180 x 90
x 45
180 x 90
x 45
180 x 90
x 45 -

Subject to technical alteration. 07/2020
07/2018/DD02XXLR
7. Accessories
– Rubber protective cap (not shown)
blue, green, black, red, orange
– Rubber protective cap
with hook and wire rope 340 mm
blue, green, black, red, orange
– Hydaulic adapter
①M 16x2 I x G1/4 I
②M 16x2 A x G1/4 I ①
②
A green rubber cap is included but
must have been ordered seperately.
https://www.druck-temperatur.de
Table of contents
Other LR Measuring Instrument manuals
Popular Measuring Instrument manuals by other brands

Mid-Continent Instrument
Mid-Continent Instrument MD215 Series Installation manual and operating instructions

Ankom
Ankom TDF28 Service Procedure

PPM Technology
PPM Technology PPM-400ST Operation manual

TESTO
TESTO 310 instruction manual

Testec
Testec TT-SX 200 user manual

Seametrics
Seametrics iMAG 4700r quick start guide

Electro Industries/GaugeTech
Electro Industries/GaugeTech Shark 100T user manual

Mastech
Mastech MS6503 instruction manual

Ecolab
Ecolab D16 operating instructions

promethean
promethean ActivExpression quick start guide

Pioneer
Pioneer SGY-PM90 quick guide

PCB Piezotronics
PCB Piezotronics IMI SENSORS 607A01 Installation and operating manual