Lutron Electronics PM-9102 User manual

200 mbar, differential input
MANOMETE
R
Model : PM-9102
Y
our purchase of this
MANOMETER marks a step
forward for you into the
field of precision
measurement. Although
this MANOMETER is a
complex and delicate
instrument, its durable
structure will allow many
years of use if proper
operating techniques are
developed. Please read
the following instructions
carefully and always keep
this manual within easy
reach.
OPERATION MANUAL

TABLE OF CONTENTS
1.FEATURES..........................................................1
2.SPECIFICATIONS................................................1
2-1 General Specifications................................... 1
2-2 Electrical Specifications.................................
.
3
3.FRONT PANEL DESCRIPTION..............................
.
4
3-1 Display.......................................................4
3-2 Power Off/On Button..................................
.
4
3-3 Hold Button................................................4
3-4 Zero Button................................................4
3-5 " MAX./MIN. " Button.................................. 4
3-6 Unit Button............... 4
3-7 P1 input socket...........................................4
3-8 P2 input socket...........................................4
3-9 RS-232 Output Terminal..............................4
3-10 Battery Compartment/Cover........................
.
4
3-11 PLug/quick coupler......................................4
4.MEASURING PROCEDURE...................................
.
5
5.AUTO POWER OFF DISABLE................................6
6.RS232 PC SERIAL INTERFACE.............................
.
7
7.BATTERY REPLACEMENT.....................................8

1. FEATURES
* Dual & differential input, 200 mbar max. range.
* Application : Industrial, laboratory, heating,
ventilation, medical hospital, used for air or not
corrosive and not ionized gas & liquid.
* Sensor is built inside the housing.
* Single lugs for pipe connection.
*
8 kind display units ( mbar, psi, , mm Hg,
inch Hg, meter , inch , Atmosphere ) select
by push button on the front panel
* Auto shut off saves battery life.
* Zero button on the front panel, easy to offset the
zero value.
* Microprocessor circuit assures maximum possible
accuracy, provides special functions and features,
* Super large LCD display with contrast adjustment
for best viewing angle.
* Records maximum & minimum readings with recall.
* Data Hold function for stored the desired value on
display.
* Built-in low battery indicator.
* RS232 PC serial interface, can match the personal
computer used as the Data Logger, Recorder.... &
other modern pressure measuring system.
2. SPECIFICATIONS
2-1 General Specifications
Circuit Microprocessor LSI circuit.
Display 61 mm x 34 mm supper large LCD
display. 15 mm ( 0.6" ) digit size.
1

Display units
mbar, psi, , mm Hg, inch Hg,
meter , inch , Atmosphere.
Function Dual & differential input, data hold,
zero/relative, memory.
Zero adjust Push button on the front panel.
Sensor * Sensor is built inside the housing.
* Piezoelectric sensor.
* Used for dry, non-corrosive and
non-ionic air and gas only.
Liquid is prohibited.
Data hold By push button.
Data record Record maximum & minimum readings.
Data output RS 232 PC serial interface.
Sampling time Approx. 0.8 second.
Power off Auto shut off, saves battery life or manual off
by push button.
Operating 0 to 50 ( 32 to 122 ).℃℉
temperature
Operating Less than 80% R.H.
humidity
Power supply 006P DC 9V battery ( heavy duty ).
Power current Approx. DC 6.0 mA.
Weight 345 g/0.76 LB .
Dimension 180 x 72 x 32 mm ( 7.1 x 2.8 x 1.3 inch ).
Accessories * Instruction manual....................... 1 PC.
included * Hard carryin
g
case....................... 1 PC.
* Plu
g
for quick coupler................... 2 PCs.
Optional * Data acquisition software
accessories ( Windows version ).........SW-U801-WIN
* USB cable........................USB-01
* RS232 cable.................... UPCB-01
2

2-2 Electrical Specifications
Unit Max. range Resolution
mbar ± 200 mbar 0.1 mbar
psi ± 2.9 psi 0.001 psi
± 0.204 0.001
mm Hg ± 150 mm Hg 0.1 mm Hg
inch Hg ± 5.906 inch Hg 0.002 inch Hg
meter
± 2.040
meter
0.001 meter
inch ± 80.2 inch 0.05 inch
Atmosphere ± 0.197 Atmosphere 0.001 Atmosphere
Unit Max. range Accuracy
mbar ± 200 mbar ± 2 % F. S.
psi ± 2.9 psi
± 0.204
Note :
mm Hg ± 150 mm Hg
* 23± 5 .℃
inch Hg ± 5.905 inch Hg
* F.S. : full scale
meter
± 2.040
meter
* Included linearity,
inch ± 80.2 inch
hysteresis and
Atmosphere ± 0.197 Atmosphere
repeatability
Remark :
Measuring Display unit
unit
mbar m Bar
psi Psi
mm Hg mm /Hg
inch Hg in/Hg
meter
m
inch
inch
Atmosphere ATP
3

3. FRONT PANEL DESCRIPTION
Fig. 1
3-1 Display 3-7 P1 input socket
3-2 Power Off/On Button 3-8 P2 input socket
3-3 Hold Button 3-9 RS-232 Output
3-4 Zero Button Terminal
3-5 " MAX./MIN. " Button 3-10 Battery Compartment
3-6 Unit Button /Cover
3-11 Plug/ quick coupler
4

4. MEASURING PROCEDURE
1)Power on the meter by pressing the " Power Off/On
Button " ( 3-2, Fig. 1 ).
2)Select the desired temperature units ( mbar, psi,
mm Hg, inch Hg, meter , inch , Atmosphere )
by pushing the " Unit Button " ( 3-6, Fig. 1 ).
3)Zero adjusting
Adjust the display reading to zero value by pushing
the " Zero Button " ( 3-4, Fig. 1 )
4)Install the measuring pipe to " Plug/quick coupler "
( 3-11, Fig. 1 ).
5)The meter is build the two input socket ( P1 input socket,
P2 input socket) for accepting the differential pressure
input.
Connecting the pipe along the " Plug " ( 3-11, Fig. 1) to
a.
" P1 input socket " ( 3-7, Fig. 1 ) only
b.
" P2 input socket " ( 3-8, Fig. 1 ) only
c.
Both P1 & P2 input socket
The LCD will show the measuring pressure value.
Note :
*
If the P1 pressure > P2 pressure, the display will
get positive reading.
*
If the P1 pressure < P2 pressure, the display will
get negative reading.
6)Data Hold
* During the measurement, pressing the " Hold Button "
( 3-3, Fig. 1 ) will freeze the measured value & the LCD
will show " HOLD " symbol.
* Press the " Hold Button " again to cancel the data hold
function.
5

7)Data Record ( Maximum, Minimum reading )
* The DATA RECORD function displays the maximum
and minimum readings. To start the DATA RECORD
function, press the " MAX./MIN. Button " ( 3-5, Fig. 1 )
once. " REC " symbol will appear on the LCD display.
* With the " REC " symbol on the display :
(a) Press the " MAX./MIN. Button " ( 3-5, Fig. 1 )
once, the " Max " symbol along with the maximum
value will appear on the display.
(b) Press the " MAX./MIN. Button " again, the " Min "
symbol along with the minimum value will appear
on the display.
(c) To exit the memory record function, press the
" MAX./MIN. Button " continuously for at least 2
seconds. The display will revert to the current reading.
8)For quick measurement, follow the procedures
shown below :
Main procedures :
POWE
R
ZERO ADJUS
T
ON
DETERMINE UNI
T
Optional measurin
g
procedures :
DATA HOLD MEMORY RECORD RS232 OUTPUT
Max., Min.
Power mana
g
ement
AUTO POWER OFF or MANUAL POWER OFF
(Not activated durin
g
Memory Record Selection)
6

5. AUTO POWER OFF DISABLE
The instrument has built-in " Auto Power Shut-off " in
order to prolong battery life. The meter will switch off
automatically if none of the buttons are pressed within 10
min.
To de-activate this feature, Select the memory record
function during measurement, by pressing the " MAX./MIN.
Button " ( 3-5, Fig. 1 ).
6. RS232 PC SERIAL INTERFACE
The instrument features an RS232 output via 3.5 mm
Terminal ( 3-9, Fig. 1 ).
The connector output is a 16 digit data stream which
can be utilized to the user's specific application.
An RS232 lead with the following connection will
be required to link the instrument with the PC
serial input.
Meter PC
(3.5 mm jack plug) (9W 'D" Connector)
Center Pin..................................Pin 2
Ground/shield...............................
.
Pin 5
7

The 16 digit data stream will be displayed in the
following format :
D15 D14 D13 D12 D11 D10 D9 D8 D7 D6 D5 D4 D3 D2 D1 D0
Each digit indicate the following status :
D15 Start Word
D14 4
D13 1
D12 & D11 Annunciator for Upper Display
mbar = 86 psi = 23
mm Hg = 78 inch Hg = 80
inch H20 = 25 Kg/cm^2 = 77
Atmosphere = 26 meter H20= 79
D10 Polarity 0 = Positive 1 = Negative
D9 Decimal Point ( DP ) for Upper display.
0 = No DP, 1= 1 DP, 2 = 2 DP, 3 = 3 DP
D1 to D8 Display reading, D1=LSD, D8=MSD
For example : If the display reading is 1234, then
D8 to D1 is 00001234
D0 End Word
RS232 FORMAT : 9600, N, 8, 1
Baud rate 9600
Parity No parity
Data bit no. 8 Data bits
Stop bit 1 Stop bit
7. BATTERY REPLACEMENT
1)When the left corner of LCD display show " LBT ",
it is necessary to replace the battery. However,
in-spec measurement may still be made for several
hours after low battery indicator appears before the
instrument become inaccurate.
2)Slide the Battery Cover ( 3-10, Fig. 1) away from the
instrument and remove the battery.
3)Install a 9 V battery (PP3 type) and replace the cover.
80906-PM9102
Table of contents