2www.mastercool.com
File Management Browse images from the micro SD card
Note: If you want to manage images, please use the USB
cable to connect the device and the computer)
Settings
Emissivity
The default emissivity is 0.95, or select other default emissivity:
Blackbody: 1.0, Matt: 0.95, Semi-matt: 0.8, Semi-glossy:
0.6, Glossy: 0.3
Select Set Value to adjust the emissivity (0.1 to 1, in 0.01 steps)
HAL When the reading is outside the High Alarm (HAL) or Lo Alarm
(LAL) limit, the reading (HAL: red texts, LAL: blue texts) shown
at the top left corner of the screen will flash with two short
beep sounds constantly.
Select Set HA temp or Set LA temp to adjust the limits or
select ON/OFF to turn on/off the alert.
LAL
Save file as Save the file as photograph or record
Auto off
The default setting is “Auto off 1 min”, you can select other de-
fault values: 5, 10, 20 minutes or “No auto off”. Manually power
off the device by pressing Power []key (2) for 3 seconds.
Target
distance
The maximum distance between the thermal imaging camera
and the target: 1m, 2m or 3m
Languages Select: English/French/Spanish/Portuguese/German/
Dutch/Italian/Traditional Chinese/Simplified Chinese
Set time
and date
Set the date and time in MM/DD/YYYY (month/day/year) and
HH/MM/SS (hour/minute/second) formats.
STORAGE & CLEANING
The thermal imaging camera should be stored at room temperature. The sensor lens is
the most delicate part of the thermometer. The lens should be kept clean at all times, care
should be taken when cleaning the lens using only a soft cloth or cotton swab with water
or medical alcohol. Allowing the lens to fully dry before using the thermometer. Do not
submerge any part of the thermometer.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
“Battery OK”:
Measurements are
possible
“Battery Low”:
Battery needs to be replaced,
measurements are still possible
“Battery Exhausted:
Measurements are not
possible
• When the ‘Low Battery’ icon indicates the battery is low, the battery should be recharged
by a USB port.
• Dispose of used batteries properly and keep away from children.
• If the device is not to be used for a long time, turn the power off, remove and store the
batteries in a cool, dry place.
• Please Note: It is important to turn the instrument off before replacing the battery
otherwise the thermometer may malfunction
SPECIFICATIONS:
Thermal Image Resolution: 32 x 32
Visual Image Resolution: 640 x 480
Measurement Range: -20 to 350°C (-4 to 662°F)
Accuracy(Tamb=20~26°C): ±2% of reading, or 2°C (4°F) whichever is greater
Thermal Sensitivity: 0.1°C (0.1˚F)
Frame Rate: 9 Hz
Field of View: 33 x 33 deg
Emissivity Range: 0.95 default - adjustable 0.1 to 1 (in 0.01 steps)
Focus: Focus free
LCD (inch): 2.8
LCD type: Color
Memory Type: Support micro SD card up to 32 gb (16 gb micro SD card
provided)
Spectral Range: 8 - 14 um
Interface: USB
Battery Type: 3.7 V Li-ion rechargeable battery
Battery Life (with backlight): 3 hrs continuous use
Dimensions: 96.29 x 74.03 x 233.68 mm (3.79 x 2.91 x 9.20 inch)
Weight: 345 grams (12.7 oz) Including batteries
EMC/CRFI: Reading may be affected if the unit is operated within radio frequency elec-
tromagnetic field strength of approximately 3 volts per meter, but the performance of the
instrument will not be permanently affected.