Maxtec BlenderBuddy 2 User manual

BlenderBuddy 2
Operating Manual & Instructions For Use
R233M01 REV. B
ENGLISH

866.4.MAXTEC WWW.MAXTEC.COMii
Maxtec
2305 South 1070 West
Salt Lake City, Utah 84119
USA
TEL (800) 748.5355
FAX (801) 973.6090
www.maxtec.com
E-mail: [email protected]
AUTHORIZED REPRESENTATIVE:
For the most current manual revision please visit our website: www.maxtec.com
WARRANTY
Under normal operating conditions, Maxtec warrants the BlenderBuddy 2 to be free from defects of
workmanship or materials for two (2) years from receipt. Warranty does not cover breakage/abuse.
These warranties are from the date of receipt, provided that the product is properly operated and
maintained in accordance with Maxtec's operating instructions. Based on Maxtec product evaluation,
Maxtec's sole obligation under foregoing warranty is limited to making replacements, repairs, or issu-
ing credit for equipment found to be defective. This warranty extends only to the buyer purchasing
the equipment directly from Maxtec or through Maxtec's designated distributors and agents as new
equipment. Routine maintenance items, such as O-rings, are excluded from warranty. Maxtec and any
other subsidiaries shall not be liable to the purchaser or other persons for incidental or consequential
damages or equipment that has been subject to abuse, misuse, mis-application, alteration, negligence
or accident. These warranties are exclusive and in lieu of all other warranties, expressed or implied,
including warranty of merchantability and fitness for a particular purpose.
For product warranty returns, please contact Maxtec Customer Service for a Return Material
Authorization (RMA).
EC REP QNET BV
Kantstraat 19
NL-5076 NP Haaren
The Netherlands

866.4.MAXTEC WWW.MAXTEC.COMiii
TABLE OF CONTENTS
WARRANTY ...................................................................................II
DESCRIPTION .................................................................................1
INTENDED USE...............................................................................1
WARNINGS
!
..............................................................................1
1.0 INSTALLATION/OPERATION INSTRUCTIONS................... 3
2.0 FLOWMETER OPERATION...................................................4
3.0 CLEANING & MAINTENANCE ..............................................4
3.1 Cleaning ............................................................................................4
3.2 Maintenance...................................................................................4
4.0 SPECIFICATIONS ..................................................................4
4.1 Symbol Guide .................................................................................5

866.4.MAXTEC WWW.MAXTEC.COM1
DESCRIPTION
The BlenderBuddy 2 is an accessory designed for use with air/oxygen blenders. It utilizes a Maxtec
Designed for Blenders (DFB) flowmeter which is designed and calibrated specifically for air/oxygen
blenders to provide increased accuracy compared to standard flowmeters. By utilizing dual-scale gradu-
ations, the flowmeter provides two flowmeters in one for increased accuracy at lower flows. The flow-
meter features a high-quality acrylic block body and precision valve. The BlenderBuddy 2 also provides
a sensor analysis port which may be used to measure the gas concentration from the blender. A small
continuous sensor bleed is provided which enables gas analysis regardless of which outlet port is used
on the blender.
INTENDED USE
The BlenderBuddy 2 is a flowmeter intended for use with an air/oxygen blender and is to be used
by physicians, respiratory therapists and other authorized hospital personnel to administer selected
doses of medical gases to a patient The BlenderBuddy 2 also contains a gas analysis port which may
be used with a suitable oxygen analyzer to measure the gas concentration.
WARNINGS
!
Read this User Manual before installing or operating the BlenderBuddy 2.
This manual instructs a professional to install and operate the BlenderBuddy 2. This is provided
for your safety and to prevent damage. If you do not understand this manual, DO NOT USE the
BlenderBuddy 2 and contact your provider.
Use BlenderBuddy 2 only for its “Intended Use” as described in this manual.
This product should only be used under proper supervision of a healthcare professional.
The BlenderBuddy 2 is for use with air/oxygen blenders only. The accuracy of the flowmeter will be
affected if the BlenderBuddy 2 is used in any other way.
Follow all manufacturer instructions for proper air/oxygen blender and bleed operation.
ALWAYS activate the blender bleed when necessary according to the blender manufacturer instruc-
tions. Failure to do so may result in inaccurate concentrations from the blender.
The flowmeter is capable of delivering flows greater than the indicated value (flush flow). Adjusting
the flow beyond the indicated range will result in an undetermined flow.
The BlenderBuddy 2 is capable of delivering gas mixtures at pressures equal to the blender outlet.
Always confirm proper setup before patient use.
Check for leaks and proper operation before placing the BlenderBuddy 2 into service.
ALWAYS confirm prescribed flow and oxygen concentration before administering to patient and
monitor on a frequent basis.
The BlenderBuddy 2 may contain magnetic, ferrous material and is NOT for use in MRI Environments.
DO NOT connect to source pressure greater than 100 psi.

866.4.MAXTEC WWW.MAXTEC.COM2
DO NOT disassemble the BlenderBuddy 2 while pressurized.
DO NOT use substitute parts.
DO NOT use lubricants on O-rings.
DO NOT use the BlenderBuddy 2 if you suspect components are damaged, altered, or missing.
Inspect unit before each use.
To Reduce the Risk of Fire or Explosion:
ALWAYS follow ANSI and CGA standards for Medical Gas Products and Oxygen Handling.
DO NOT use or store oils, greases, organic lubricants or any combustible materials on or near the
BlenderBuddy 2.
DO NOT use near any type of flame or flammable/explosive substances, vapors or atmosphere.
DO NOT smoke in an area where oxygen is being administered.
CAUTION
The BlenderBuddy 2 must be operated with the Flowmeter in a vertical, upright position.
Gas temperatures other than 70° F (21°C) may affect the accuracy of the indicated flow.
DO NOT drop the BlenderBuddy 2.
DO NOT attempt to sterilize the BlenderBuddy 2
DO NOT clean with aromatic hydrocarbons.
DO NOT immerse the BlenderBuddy 2 in any kind of liquid.
DO NOT over-tighten knob when turning off. This may damage the flow meter.
DO NOT occlude or plug the gas sensor port.

866.4.MAXTEC WWW.MAXTEC.COM3
1.0 INSTALLATION/OPERATION INSTRUCTIONS
NOTE: Low flow BlenderBuddy 2 models (3, 15 and 30 LPM) should only be used with Low Flow
blenders. High Flow BlenderBuddy 2 models (70 LPM) should only be used with High Flow blenders.
Use outside these applications may result in flowmeter inaccuracies.
ATTACH THE
BLENDERBUDDY
2 TO THE LEFT
SIDE DISS PORT OF
THE AIR/OXYGEN
BLENDER USING AN
11/16" WRENCH.
ALIGN THE FLOWMETER IN AN UPRIGHT, VERTICAL POSITION WITH
THE BLENDER AND FIRMLY TIGHTEN THE FITTING.
NOTE: Ensure the fitting is completely tightened to prevent rotation during use.
INSERT OXYGEN ANALYZER
WITH SENSOR DIVERTER
INTO THE SENSOR PORT TO
MEASURE THE DELIVERED
GAS FROM THE BLENDER.
NOTE: Use the sensor
diverter provided with your
BlenderBuddy 2 or genuine
Maxtec replacement.

866.4.MAXTEC WWW.MAXTEC.COM4
WARNING
!
Follow all manufacturer instructions for proper blender bleed operation. A blender bleed is typically
required to maintain mixing accuracy for flows less than 3LPM on low flow blenders and below 15LPM
for high flow blenders. Failure to activate the blender bleed may result in inaccurate concentrations
from the blender.
NOTE: A small continuous sensor bleed is provided to the analysis port which enables gas analysis
regardless of which outlet port is used on the blender.
2.0 FLOWMETER OPERATION
Adjust the flowmeter to the desired set-point as read by the center of the float ball .
To increase flow—turn knob counter-clockwise
To decrease flow—turn knob clockwise
!
WARNING: The flowmeter is capable of delivering flows greater than the indicated (flush flow).
Adjusting the flow beyond the indicated range will result in an undertermined flow.
3.0 CLEANING & MAINTENANCE
3.1 Cleaning
Exterior surfaces may be cleaned using a cloth and mild detergent, isopropyl alcohol solution, or
germicidal wipe.
3.2 Maintenance
The BlenderBuddy 2 does not require any periodic maintenance or contain any user serviceable
components.
4.0 SPECIFICATIONS
Flowmeter Accuracy:
MODEL FLOWMETER
GRADUATIONS ACCURACY FLUSH FLOW
3 LPM 0.1 (0.1-1) LPM
0.5 (1-3) LPM ± 0.5 LPM 20–30 LPM
15 LPM 0.25 (0.5-3) LPM
1 (5-15) LPM
0.5-3: ± 0.5 LPM
5-15: ± 10% of indicated value 20–30 LPM
30 LPM 0.25 (0.5-3) LPM
2.5 (5-30) LPM
0.5-3: ± 0.5 LPM
5-30: ± 10% of indicated value 35–45 LPM

866.4.MAXTEC WWW.MAXTEC.COM5
70 LPM 1 (2-15) LPM
5 (15-70) LPM
2-4: ± 0.5 LPM
5-70: ± 10% of indicated value 70–80 LPM
WEIGHT .88 lbs. (.4 kg)
TOTAL SYSTEM RESPONSE TIME
≤
30 seconds
STORAGE TEMPERATURE -40°F (-40°C) to 140°F (60°C)
The BlenderBuddy 2 is calibrated for air/oxygen gas mixtures at 70° F (21° C) and is compensated
for the pressure drop of a typical blender with inlet pressures at 50 psig. Specifications are subject to
change without prior notice.
4.1 Symbol Guide
The following symbols and safety labels are found on the BlenderBuddy 2:
FEDERAL LAW
(USA) RESTRICTS
THIS DEVICE
TO SALE BY OR
ON ORDER OF
NOT
MANUFACTURED
WITH NATURAL
RUBBER LATEX
!
WARNING
60°C
(140°F)
-40°C
(-40°F)
STORAGE
TEMPERATURE
CONSULT
INSTRUCTIONS
FOR USE
NON-STERILE
MANUFACTURER AUTHORIZED
REPRESENTATIVE
MADE IN USA
CONTAIS NO
POLYVINYL
CHLORIDE
DO NOT DESIGNED FOR
BLENDERS

866.4.MAXTEC WWW.MAXTEC.COM6
REF CATALOG NUMBER LOT LOT NUMBER

2305 South 1070 West
Salt Lake City, Utah 84119
800.748.5355
Table of contents
Other Maxtec Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Thermo Scientific
Thermo Scientific 88882013 quick start guide

Philips
Philips Avent SCF293 manual

Molecular Devices
Molecular Devices SpectraMax MiniMax 300 installation guide

Agilent Technologies
Agilent Technologies 7890A Instruction guide

CAMAG
CAMAG AMD 2 instruction manual

Thermo Scientific
Thermo Scientific Fisherbrand GT4 instruction manual

SICCO
SICCO V 1982-08 operating instructions

MICROLIT
MICROLIT E-BURETTE Operation manual

Lucky Lifestyle
Lucky Lifestyle CW1827S manual

WIKA
WIKA CTD9100-165 operating instructions

Fritsch
Fritsch PULVERISETTE 19 operating instructions

Thermo Scientific
Thermo Scientific PRO-LINE Series Operating and service manual