Met One Instruments BAM 1022 Installation manual

Met One Instruments, Inc.
1600 NW Washington Blvd.
Grants Pass, OR 97526
Telephone: (541) 471-7111
Facsimile: (541) 471-7116
www.metone.com
BAM 1022 Particulate Monitor Operation Manual - © Copyright 2012 Met One Instruments, Inc. All Rights Reserved
worldwide. No part of this publication may be reproduced, transmitted, transcribed, stored in a retrieval system, or translated into any
other language in any form without the written permission of Met One Instruments, Inc.
BAM 1022
PARTICULATE MONITOR
OPERATION MANUAL
BAM 1022-9800 REV E

Page 2 BAM 1022-9800 REV E

BAM 1022-9800 REV E Page 3
Table of Contents
1INTRODUCTION 5
1.1 About This Manual .....................................................................................................5
1.2 Technical Service.......................................................................................................6
1.3 BAM: Beta Attenuation Monitor..................................................................................7
1.4 Safety Statements...................................................................................................... 8
1.5 BAM 1022 Specifications ...........................................................................................9
1.6 U.S. EPA Federal Equivalency Method Notifications ............................................... 10
2INSTALLATION & COMMISSIONING 11
2.1 Unpacking................................................................................................................ 11
2.2 Accessories..............................................................................................................12
2.3 BAM 1022 Installation for FEM Monitoring Applications........................................... 12
2.4 BAM 1022 Power and Electrical Connections.......................................................... 17
3USER INTERFACE 18
3.1 Main Operating Screen ............................................................................................ 18
3.2 Menu Hierarchy and Navigation............................................................................... 19
3.3 Operate Menu .......................................................................................................... 21
3.4 Test Menu................................................................................................................ 22
3.5 Setup Menu..............................................................................................................26
3.6 Alarms Menu............................................................................................................ 30
4OPERATION 31
4.1 Initial Procedures...................................................................................................... 31
4.2 Loading Filter Tape .................................................................................................. 31
4.3 Warm-up Period....................................................................................................... 33
4.4 Commissioning......................................................................................................... 33
5The BAM 1022 Measurement Cycle 34
5.1 Hourly Measurement Cycle...................................................................................... 34
5.2 Short Term Averaging.............................................................................................. 34
6MAINTENANCE, ALARMS & TROUBLESHOOTING 36
6.1 Periodic Maintenance............................................................................................... 36
6.2 Basic Leak Check..................................................................................................... 37
6.3 Advanced Leak Checks............................................................................................ 38
6.4 Flow Audit and Calibration ....................................................................................... 40
6.5 Background Determination (Mass Offset) ................................................................ 43
6.6 Nozzle, Vane & Pinch Roller Cleaning..................................................................... 45
6.7 Internal Nozzle Cleaning.......................................................................................... 46
6.8 Span Mass Audit...................................................................................................... 47
6.9 Filter Sensor Testing................................................................................................ 49
6.10 Basic Problems & Solutions ..................................................................................... 51
6.11 Alarms...................................................................................................................... 53
6.12 Hardware Failure Screen ......................................................................................... 54
6.13 Sensor Out of Range Event ..................................................................................... 54
6.14 Alarm Relay..............................................................................................................55
7DATA COLLECTION AND COMMUNICATIONS 56
7.1 Analog & Digital I/O..................................................................................................56
7.2 Clock Sync ............................................................................................................... 57
7.3 Transfer Data to USB Flash Drive............................................................................ 57
7.4 Serial Communications ............................................................................................ 58

Page 4 BAM 1022-9800 REV E
7.5 Data Format & Example Reports .............................................................................61
8THEORY OF OPERATION 64
9ACCESSORIES and PARTS 66
9.1 Consumables, Replacement Parts, and Accessories............................................... 66
APPENDIX A: ACRONYMS AND ABBREVIATIONS 69
APPENDIX B: CONFIGURING NETWORK IP ADDRESS 70

BAM 1022-9800 REV E Page 5
1 INTRODUCTION
1.1 About This Manual
This document is organized with the most important information toward the front of the manual,
such as site selection, installation, setups, and field calibrations, which all BAM 1022 owners
and operators should read and understand. Toward the back are sections that provide in-depth
information on subjects such as theory, diagnostics, accessories, and alternate settings. These
sections provide valuable information that should be consulted as needed. Electronic versions of
this manual are also available. Operators of the monitoring system are encouraged to study this
manual to ensure correct performance, safe operation, and to prevent equipment damage.

Page 6 BAM 1022-9800 REV E
1.2 Technical Service
This manual is structured by customer feedback to provide the required information for setup,
operation, testing, maintaining, and troubleshooting your BAM 1022 mass monitor. Should you
still require support after consulting your printed documentation, we encourage you to contact
one of our expert Technical Service representatives during normal business hours of 7:00 a.m.
to 4:00 p.m. Pacific Standard Time, Monday through Friday. In addition, technical information
and service bulletins are often posted on our website. Please contact us and obtain a Return
Authorization (RA) number before sending any equipment back to the factory. This allows us to
track and schedule service work and to expedite customer service. Please have your instrument
serial number available when contacting the manufacturer.
Contact
Information:
Address:
Met One Instruments, Inc.
1600 Washington Blvd
Grants Pass, Oregon
97526 U.S.A.

BAM 1022-9800 REV E Page 7
1.3 BAM: Beta Attenuation Monitor
The Met One Instruments Model BAM 1022 Continuous PM Monitoring System utilizes the
principal of beta ray attenuation to accurately measure and report the concentration of airborne
particulate matter (PM) in ambient air. The centerpiece of the measurement system consists of
a beta source that emits a consistent supply of electrons, and a sensitive detector that counts
the incident electrons.
A vacuum pump draws air through a size selective inlet, down the inlet tube, and deposits the
airborne particulate on a filter tape that is located between the beta source and detector. The
accumulation of mass onto the filter tape increasingly attenuates beta ray transmission through
the media.
Beta attenuation through the filter tape is continuously monitored throughout the measurement
cycle. The degree of beta ray attenuation is used to determine the mass of particulate matter
deposited on the filter tape. During sampling, the flow rate is precisely controlled.
Having determined both mass and sample volume, the BAM 1022 calculates and reports the
ambient PM concentration, expressed as µg/m3or mg/m3.
A detailed description of the measurement is provided in Section 8 of this Operating Manual.
BAM 1022 Measurement System
NOTE: For proper performance, the BAM 1022 must be operated outdoors and should not
be installed inside of a building, trailer, or other shelter. It must be allowed to run at
ambient conditions. This is required for the BAM 1022 to operate as a U.S. EPA PM2.5 Class
III Equivalent Method monitor.
Sample Entry
Nozzle
Filter Tape
Detector
Vane

Page 8 BAM 1022-9800 REV E
1.4 Safety Statements
1.4.1 Beta Radiation
The Met One Instruments BAM 1022 contains a small 14C (Carbon 14) beta radiation-emitting
source. The nominal activity of the source is 60 μCi ±15μCi (microcurries), which is below the
“Exempt Concentration Limit” as defined in USC 10 CFR Section 30.71 –Schedule B. The
owner of the BAM 1022 is not required to obtain any license in the United States to own or
operate the mass monitor. The owner of a BAM 1022 may elect to return the entire mass
monitor to Met One Instruments for recycling of the 14C source when the mass monitor has
reached the end of its service life, although the owner is under no obligation to do so. Under no
circumstances should anyone but factory technicians attempt to remove or access the beta
source. The beta source has a half-life of about 5,730 years, and should never need to be
replaced unless it becomes damaged or corroded. Neither the 14C source nor the beta ray
detector are serviceable in the field. Should these components require repair or replacement,
the BAM 1022 must be returned to the factory for service and recalibration. The BAM 1022 is
manufactured in compliance with the U.S. NRC safety criteria in 10 CFR 32.27.
1.4.2 Other Safety Items
If the BAM 1022 is operated in a manner not specified by the manufacturer, the performance
and protection provided by the equipment may be impaired.
Any repairs or adjustments made to the BAM 1022 that are not outlined within this
document must be carried out only by factory-trained personnel. Under no circumstances
should repairs be attempted by personnel not qualified to service electronic
instrumentation.
Disconnect power to the BAM 1022 power before working on electrical components.
Failure to do so may result in an electrical hazard and could damage the equipment.
Grounding and antistatic procedures must be followed when handling any internal
components. Internal circuitry of the BAM 1022 may be damaged by static electricity.
Do not operate the BAM 1022 in potentially explosive environments.
If safe operation of the BAM 1022 has been impaired, the instrument must be
disconnected from power and secured against unintended operation.
The BAM 1022 should be secured against accidental tipping. Anchoring holes are
provided in the feet at the base of pump box assembly for this purpose.
For procedures not outlined within this document, consult your local Met One Instruments
technical service representative or contact the manufacturer.

BAM 1022-9800 REV E Page 9
1.5 BAM 1022 Specifications
PARAMETER
SPECIFICATION
Measurement Principle
Beta Attenuation.
US EPA Designations
EPA Class III Federal Equivalency Method (EQPM-1013-209)
Measurement Range
-15 g/m3–10,000 g/m3
Accuracy
Meets US-EPA Requirements for Class III PM2.5 FEM
Data Resolution
1.0 µg/m3
Lower Detection Limit
< 4.8 µg/m3(Hourly, 2σ) / < 1.0 µg/m3(24 Hour, 2σ)
Sampling Time
Continuous Air Sampling with Hourly Tape Advance
Measurement Cycles
Primary: Automatic Hourly PM Measurement (Required for PM2.5 FEM Operation)
Secondary: User Selectable Short Term Averages (15 to 60 minute)
Sample Flow Rate
16.7 liters/minute
Flow Accuracy
±2.0 %
Filter Tape
Continuous glass fiber filter, 30mm x 21m roll. Up to 2 months operation per roll.
Maintenance Interval
8 Weeks (minimum)
Span Check
Manual Audits Performed with Zero and Span Foils (Included)
Beta Source
14C (carbon-14), 60 µCi ±15 µCi (< 2.22 X 106Beq), Half-Life 5730 years.
Detector Type:
Scintillation Photomultiplier Tube.
Operating Temperature Range
-30° to +60° C
Operating Humidity Range
0 to 90% RH, non-condensing
Enclosure Requirements
Monitor housing meets all requirements for FEM sampling.
User Interface
4.3” graphical touch screen
Analog Output
Two channels; optically isolated; switch settable voltage or current
0-1, 0-2.5, 0-5 VDC, 4-20 mA
Alarm Contact Closure
1 channel; dry NO contact; 1 A at 125 VAC or 60 VDC maximum.
Serial Interfaces
1 channel; full duplex RS-232, USB, and Ethernet serial (Shared common serial
output). Baud rates 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200
Compatible Software
Comet™ (included), Air Plus™, terminal programs such as HyperTerminal®
Data Logger Memory
11,827 records (1.4 years @ 1 record/hr. 8 days @ 1 record/min).
Unit Dimensions
55.9cm wide x 45.7cm deep x 139.7cm high (22”w x 18”d x 55”h)
Weight
38.6 kg (85 lbs) fully assembled
Power Supply
100 - 230 VAC; 50/60 Hz universal AC power
Power Consumption
Approximately 300 W (including BAM 1022, Inlet Heater, Pump, and Vent Fan)
Approvals
NRC, ISO-9001, ROHS, US-EPA
(Specifications may be subject to change without notice.)

Page 10 BAM 1022-9800 REV E
1.6 U.S. EPA Federal Equivalency Method Notifications
The BAM 1022 Beta Attenuation Mass Monitor is US-EPA designated for PM2.5 under the
following designation number:
•Designation Number: EQPM-1013-209 (PM2.5 with BGI/Mesa Labs VSCC™ or Tisch
Cyclone)
US-EPA designated methods using the BAM 1022 are modified from time to time in order to
reflect hardware or software improvements. These modifications do not impact previously
designated configurations of the BAM 1022 but may provide the end user with a product
upgrade path that will allow the monitor to continue to be operated as a US-EPA designated
method. For further details, please contact our service department. Details concerning USEPA
designated configurations of the BAM 1022 may be found on the US-EPA website:
https://www3.epa.gov/ttn/amtic/files/ambient/criteria/AMTIC%20List%20Dec%202016-2.pdf

BAM 1022-9800 REV E Page 11
2 INSTALLATION & COMMISSIONING
2.1 Unpacking
Any damages incurred to the equipment during shipping are the responsibility of the carrier. If
any damage to the shipment is noticed before unpacking, a claim must be filed with the
commercial carrier immediately. You should follow any special unpacking instructions
provided by the carrier as you then carefully remove all items from the containers and inspect
each component. It is recommended to document and photograph all damaged packages and
items before, during, and after unpacking them. Contact Met One Instruments (see section 1.2
of this manual) to arrange for any replacement items needed.
Unpack the mass monitor and accessories and compare them to the packing list to make sure
all items are present.
The BAM 1022 is shipped with a white plastic shim pinched under the nozzle of the mass
monitor which prevents the moving parts of the nozzle assembly from being damaged in transit.
The shim must be replaced anytime the mass monitor is being transported in order to avoid
damaging the nozzle control mechanism. Do not ship or transport the BAM 1022 with filter tape
installed.
Please keep all of the special shipping items (box, foam packing material, etc.) used to ship your
BAM 1022. They should be re-used if you must transport your monitor (changing site locations,
returning to the factory, etc.). Met One is not responsible for damage to the mass monitor if
shipped in non-original packaging, or without the shim in place. Contact Met One Instruments
(see section 1.2 of this manual) for replacement packing materials if necessary.

Page 12 BAM 1022-9800 REV E
2.2 Accessories
Every BAM 1022 requires the use of ancillary components. The items supplied with a standard
BAM 1022 PM2.5 FEM monitoring system are as follows:
Standard Components
Part No.
PM10 Size Selective Inlet
BX-802
PM2.5 Very Sharp Cut Cyclone
BX-808
Inlet Tube Heater
80559
Inlet Tube Heater Insulation Sleeve
9123-2
Inlet Tube (27 inches)
8112-2.25
Ambient Combination Sensor (AT, BP, RH)
597
Filter Tape
460180
Accessory Kit*
82055
*See section 9.1 for a complete listing of all items contained in the Accessory Kit
Application and installation dependent accessories for the BAM 1022 are available from Met
One Instruments.
Common Optional Component
Part No.
Volumetric Flow Calibration Kit
BX-307
Zero Filter Kit
BX-302
2.3 BAM 1022 Installation for FEM Monitoring Applications
The BAM 1022 is a weatherproof monitor and must be operated outside of any additional
enclosures in order to maintain its US-EPA Class III designation. Its housing is designed to
maintain a near constant temperature difference between ambient and measurement conditions.
The required constant temperature difference will not be achieved if the BAM 1022 is installed
inside of any structure where the temperature is not allowed to vary with ambient temperature.
There are no special precautions or handling concerns except for the normal level of care
required for handling scientific equipment.
1. Remove the cover from the top of the pump box
assembly and set it aside. Remove all packing
material from inside the box and gather the
loose ends of the cables and tubing.
2. Invert the cover so that the white nipple is
protruding upward, feed the cables and tubing
through it from the bottom, and then replace it
atop the pump box assembly. It should look like
the picture shown here.

BAM 1022-9800 REV E Page 13
3. Open the front door of the main enclosure and remove the protective packing material from
around the touch screen display. On the back of the enclosure, undo the four wing nuts
holding the cable access cover in place and then remove the cover. Set it aside for now.
4. Behind the cable access panel, locate the ingress hole on the right hand side of the bottom
of the enclosure. Lift the monitor and carefully feed the cabling through this hole and
continue to lower it until the white nipple extends in to the housing and the monitor is resting
flat on top of the pump box assembly. Raise the side latches and secure them in place by
tightening the large knob. See pictures below.
CAUTION: Do not lift the BAM 1022 monitor by the angled solar radiation shield mounted to the
top of the enclosure. It is not designed to hold the weight of the full assembly and can easily be
damaged if used to lift the monitor. Instead, firmly grasp the sides of the housing and lift. It
would be easiest to have at least two people for this step.
5. Route the pump cable, the pump box fan cable, and the pump tubing to the appropriate
locations and plug them in to the monitor. Be sure to connect the ground lead tied to the
yellow pump cable to the grounding connection on the BAM 1022 case. Note that the pump
cable connection is a twist lock type. Once inserted, rotate it slightly in a clockwise direction
to lock it in place. To insert the tubing, simply press it in to the fitting. See photos below.

Page 14 BAM 1022-9800 REV E
6. Wrap the lower threads of the pipe seal with Teflon
tape and install it in the top of the enclosure. See
picture to the right.
7. Insert the black sealing ring and loosely attach the
white clamping ring (also shown in the picture at
right) to the pipe seal.
8. Locate the red cap covering the inlet tube receiver
inside the BAM 1022 enclosure. It is found just to the left of the touchscreen display.
Remove the large red cap and a smaller red insert will be revealed beneath it. Remove this
cap as well. See image below left.
9. Slide the inlet tube through the pipe seal on top of the enclosure just until you can see it
inside the enclosure. Position the inlet heater beneath the inlet tube and slide the tube
through the heating element until it just comes in contact with the o-rings. Raise the heater
so it is near (but not touching) the top of the enclosure and fasten it in position with the
supplied hex wrench. See image above center.
10.Rotate the tube so that the heater power cable is facing the back left corner of the enclosure
and route the cable down the large empty hole located near that corner. Push the inlet tube
down all the way and wrap the heater element in the supplied insulating material as shown in
the above right image.
11.Route the inlet tube heater to the power connection labeled
HEATER in the back of the BAM 1022 and plug it in. Note
that this connection is a twist lock type. Once inserted,
rotate it slightly in a clockwise direction to lock it in place.

BAM 1022-9800 REV E Page 15
12.Install the 597 probe on the inlet tube above the roof of the BAM 1022 enclosure using the
U-bolt with the nuts and washers provided. Connect the silver connector of the cable to the
silver connection on the bottom of the sensor. See pictures below.
13.Route the sensor cable to the back of the enclosure and attach it to the connection labeled
as EXTERNAL AMBIENT SENSOR. See photo below left.
14.Install the AC power supply cable to the remaining connection on the bottom of the internal
power strip. It is located next to the ON/OFF switch. See photo above right. Route the
power cable and the sensor cable out of the bottom of the cable access opening.
15.Connect the 9528 grounding cable (included in the accessory kit) to the Ground connection
shown in the above right photo. Route the ground cable out of the bottom of the cable
access opening and connect it to the isolated site grounding rod.

Page 16 BAM 1022-9800 REV E
16.Remove the inlet protector plugs from the BX-802 PM10 inlet head and BX-808 PM2.5 Very
Sharp Cut Cyclone. Install the BX-808 on top of the inlet tube and then the BX-802 directly
above it. See photos below.
17.Plug the AC power cable in to your AC source and turn the power switch to ON. Verify the
fans start up in both the pump box and the main enclosure. The touchscreen display should
also turn on as described in section 4.1.
18.Replace the cable access cover panel on the back of the enclosure and affix it in place with
the four wing nut fasteners. Be sure the flange is on the bottom so that the power cable and
the sensor cable will be pinned between the foam strips and held in place.

BAM 1022-9800 REV E Page 17
2.4 BAM 1022 Power and Electrical Connections
The BAM 1022 should be powered by either a 110 VAC 50/60Hz or 220 VAC 50/60 Hz source.
Before proceeding with electrical connections, review the following notifications:
Ensure that the monitoring site mains power (voltage and frequency) is compatible with
the version of BAM 1022 monitor and accessories that has been purchased. Consult the
electrical labels on the devices.
All electrical connections must be in accordance with national codes. Always use
grounded receptacles and power cords.
A dedicated 15 Amp electrical circuit is adequate to run a single BAM 1022 system.
A summary of instrument power consumption is provided below:
Description
Amps
Watts
BAM 1022 on, pump and inlet heater off (115 VAC).
0.4 A
46 W
BAM 1022, pump and inlet heater on (115 VAC).
2.6 A
299 W
BAM 1022 on, pump and inlet heater off (230 VAC).
0.3 A
69 W
BAM 1022, pump and inlet heater on (230 VAC).
3.7 A
851 W

Page 18 BAM 1022-9800 REV E
3 USER INTERFACE
The BAM 1022 user interface is a touchscreen display used to control almost all of the features
and functionality of the BAM 1022. It is mounted on a hinged bracket located inside the
enclosure near the inlet tube in the upper right hand corner of the monitor. The center screw of
the mounting bracket hinge adjusts the amount of force necessary to reposition the bracket.
BAM 1022 User Interface
3.1 Main Operating Screen
In addition to the last hourly concentration reading, this screen shows the current real time
values being measured and the operational state of the BAM 1022. This image below on the
left hand side is the screen that will normally be displayed.
Note that the display has a limited amount of space and cannot show all of the real time data on
one screen. Tap the up or down arrow keys in the lower left corner of the display to navigate
between the two screens shown above.
Tension Adjustment

BAM 1022-9800 REV E Page 19
3.2 Menu Hierarchy and Navigation
The BAM 1022 menu structure is outlined in the following table.
Main Menu
Sub Menu Options
Overview
Operate
See section 3.3
Start Sample
Load Filter Tape
Transfer Data
About
Begin or resume monitoring (will indicate Stop Sample, if appropriate)
Load and properly tension the filter tape
Download stored data to a USB memory stick
Details the current BAM 1022 firmware type and version number
Test
See section 3.4
Leak Test
Ambient Temperature
Ambient Pressure
Flow Calibration
Run Self-Test
Parameters
Filter Sensors
Span Mass Audit
Tape Test
Nozzle/Count Test
Analog Output Cal
Analog Output Test
Inlet Heater
Relay Test
Input Test
Digital Link
Perform the leak test
Calibrate ambient temperature or restore default settings
Calibrate ambient pressure or restore default settings
Calibrate flow rate or restore default settings
Run the BAM 1022 Self-Test
Review current settings
Calibrate filter temp, filter RH, and pressure or restore default settings
Run the zero and span foil tests
Verify tape travels expected distance
Test nozzle operation and verify beta counting
Calibrate the analog output
Test the analog output
Manually turn the inlet heater on and off
Manually open and close the alarm relay
Test the digital clock sync input
Test digital communications with the 597 sensor
Setup
See section 3.5
Set Clock
Clear memory
Background
Serial Port
Location
Data Average
Real-Time Period
Hourly Report
Analog Outputs
Conc. Units
Inlet Heater
Tape Advance
Change Password
Sound Volume
Touch Calibrate
Set the date and time
Clear all stored data
Change the concentration offset
Set the baud rate and connection type for serial communications
Set the location number used to identify the BAM 1022
Set the averaging interval for collecting other than hourly data
Set the averaging interval for collecting real time data
Set the type of time stamp to use for the hourly report
Set the parameters for both analog outputs
Set the concentration units
Set the RH threshold for the inlet heater to activate
Set the differential pressure threshold for advancing the filter tape
Change the master password
Adjust the volume of the touchscreen sounds
Calibrate the touch screen
Alarms
See section 3.6
No sub menu
View alarms
Menu selections and instructions are detailed in the following sections of this operating manual
as detailed in the Main Menu column of the table above.

Page 20 BAM 1022-9800 REV E
To access the various main menus, press
the three horizontal lines in the top left
corner. A drop down menu will appear (see
right) to allow selection of any of the four
main menus. This option is available on all
main menu screens (such as the Setup
Menu shown below) and on the main
operating screen.
To return to the main operating screen (see
section 3.1), press the Home icon located in
the upper right corner of all main menu
screens. This icon can clearly be seen in the
Setup Menu image shown here on the left.
To cancel an action and return to the
previous menu screen, press the X icon
located in the upper right corner of all sub
menu screens. This icon can clearly be
seen in the Set Clock Menu image to the
right.
Some parameters, such as the Date and
Time settings (shown above) or a Location
value, require numeric entry. When you
press the button to edit such a field, a visual
keypad (shown left) will open up and allow
you to input the value. Press the OK key to
accept the changes or the Cancel key to
return to the previous screen. The X key on
the far right performs a backspace operation.
Table of contents
Other Met One Instruments Monitor manuals

Met One Instruments
Met One Instruments E-Sampler User manual

Met One Instruments
Met One Instruments AEROCET-380 User manual

Met One Instruments
Met One Instruments C-12 User manual

Met One Instruments
Met One Instruments BAM-1020 Installation and operating instructions

Met One Instruments
Met One Instruments BAM-1020 User manual

Met One Instruments
Met One Instruments BAM-1020 User manual