Metso MCA Use and care manual

- Shows error log
- ENTER = scroll error log
1
2
abc
MCA FT 200 QT-1234
Cs = 3.12 %
T = 56.7 C
Status = OK
% Range = 53 %
mA = 12.48 mA
Cabin T = 52.3 C
Meas Att = 45.6 dB
1 = Calibration
2 = Conguration
3 = Diagnostics
4 = Special functions
1 = Sample taking
2 = Enter lab
3 = Oset correction
4 = Filler
5 = Calib. history
6 = Sample history
1 = Current output
2 = User settings
3 = Device info
4 = Set clock
5 = Address
1 = Error log
2 = Diag. values
3 = Diag. alarms
1 = Chemical comp
2 = Temperature comp
3 = Sensitivity corr
4 = Recipes
Press sample key to
start sampling
CS = 0.00 %
Status: WAITING
Esc
1
2
abc
3def 4ghi
2
abc
4ghi
5jkl
Lo Cs (4mA) = 0.00 %
Hi Cs (20mA) = 10.00 %
Filtering = 1 s
Fault sig = 3.7 mA
1
Metso MCA
– Menu structure
& key functions
Measurement
Display
- 2 displays,
toggle with Enter
Main Menu
Enter
1
3def
3def
K10953 V1.0 EN © Metso Automation, December 2011. Subject to change without prior notice
Metso MCA
Calibration Guide
3def
20.12.11 08:15:00
MCA = 3.00 %Cs
Att = 45.2 dB T = 53.6 C
...ENTER
6mno
20.12.11 08:15:00
Lab = 3.08 % MCA = 3.00
Cum.oset = 0.00 %Cs
T = 53.6 C Att = 45.2 dB
Filler = 0.00 %
Kaolin = 0.0 %
CaCO3 = 0.0 %
Talc = 0.0 %
TiO2 = 0.0 %
Filler sens. = 1.00
Oset = 0.0 %
Cum. Oset = 0.0 %
20.12.11 14:35
MCA = 0.00 %Cs
Lab = 0.00 %Cs
Filler = 0.00 %
Kaolin = 0.0 %
CaCO3 = 0.0 %
Talc = 0.0 %
TiO2 = 0.0 %
Chem = 0.0 %
MCA con = 0.0 %
Lab con = 0.0 %
Filler sens. = 0.0 %
2
abc
Trend Int: 10 MIN
Position: POS-1234
Language: ENGLISH
Temp unit: DEG C
Password: N
Act siglev: ON
1 = Sensor Electronics
2 = TCU
Address = 0
5jkl
4ghi
Date 20.12.11
Time 19:17:24
Enter to accept...
Cab temp = 0.0 C
Proc temp = 0.0 C
Mlev = 0.0 mV
Rlev = 0.0 mV
Mstab = 0.0
El drift = 0.0
VCO drift = 0.0
Ctmp max = 90.0 / OFF
Ptemp max = 100.0 / ON
Ptmp min = 0.0 / ON
M.Lev min = 10.0 / ON
R.Lev min = 100.0 / ON
Mstab max = 20.0 / ON
Rstab max = 2.0 / ON
Eldr abs = 80.0 / ON
VCOdr abs = 50.0 / ON
1
2
abc
3def
Chem.comp. = OFF
Comp.value = 0.00 %
MCA con = 0.00 mS/cm
Chem = NaOH
Lab.con = 0.00 mS/cm
Temp MCA
1. 0.0 C 0.00 %
2. 0.0 C 0.00 %
3. 0.0 C 0.00 %
4. 0.0 C 0.00 %
5. 0.0 C 0.00 %
6. 0.0 C 0.00 %
Input new sens.
correction = 1.00
4ghi
1 = Choose recipe
2 = Clear recipe
1 = Calibration
2 = Conguration
3 = Diagnostics
4 = Special functions
1 = Current output
2 = User settings
3 = Device info
4 = Set clock
5 = Address
1
2
abc
Basic Settings: A - Collect lab. sample
Calibrating:
1 = Calibration
2 = Conguration
3 = Diagnostics
4 = Special functions
Main Menu
Calibration Menu
1
Calibration Menu
- Sample taking
- Waiting for sampling
- Sample = start sampling
- Collect samples for lab.
evaluation
- Sample = stop sampling
- Stores average values
during sampling
- Exit sampling result:
Esc
Calibration Menu
- Enter lab
- Enter lab. value
- Edit
Save to change value
- To save: Edit
Save
If pulp contains llers:
- Enter amount of ller as a
percentage of total cons.
- Also enter the proportional
amount of each ller type
For chemical compensation:
- Enter conductivity of lab.
sample to calibrate MCA
conductivity measurement
Sample
B - Enter lab. values to sensor
2
abc
Documentation:
- Always document all basic settings, calibration data and
compensations/adjustments that have been entered to the
transmitter.
MCA FT 200 QT-1234
Cs = 3.12 %
T = 56.7 C
Status = OK
1 = Calibration
2 = Conguration
3 = Diagnostics
4 = Special functions
Esc
Measurement display
Main Menu
Main Menu
Conguration Menu
- Set 4mA & 20mA output
signal limits
- Set ltering time
- Edit
Save to change value, move
to next value with
Enter
- To save: Edit
Save ,
Esc
Set position (tag no.)
Set date and time:
- Edit
Save to change value, move
to next value with
Enter
- To save: Edit
Save ,
Esc
Change HART address
if other than 0
2
abc
Lo Cs (4mA) = 0.00 %
Hi Cs (20mA) = 10.00 %
Filtering = 1 s
Fault sig = 3.7 mA
Trend Int: 10 MIN
Position: POS-1234
Language: ENGLISH
Temp unit: DEG C
Password: N
Act siglev: ON
Address = 0
Date 20.12.11
Time 19:17:24
Enter to accept...
5jkl
4ghi
1 = Sample taking
2 = Enter lab
3 = Oset correction
4 = Filler
1
1 = Sample taking
2 = Enter lab
3 = Oset correction
4 = Filler
Press sample key to
start sampling
CS = 0.00 %
Status: WAITING
Press sample key to
stop sampling
CS = 3.37 %
Status: SAMPLING
Sample Sampling ready
Average Cs = 3.34 %
Min: 3.30 Max: 3.38
T = 55.3 C
1 = Sample taking
2 = Enter lab
3 = Oset correction
4 = Filler
20.12.11 14:35
MCA = 0.00 %Cs
Lab = 0.00 %Cs
Filler = 0.00 %
Kaolin = 0.0 %
CaCO3 = 0.0 %
Talc = 0.0 %
TiO2 = 0.0 %
Chem = 0.0 %
MCA con = 0.0 %
Lab con = 0.0 %
Filler sens. = 0.0 %
Oset adjustment:
Calibration Menu
- Oset correction
- Edit
Save to change value
- To save: Edit
Save ,
Esc
- Oset = Lab. – MCA
1 = Sample taking
2 = Enter lab
3 = Oset correction
4 = Filler
Oset = 0.00 %
Cum.oset = 0.00 %
3def
Compensations / adjustments:
Special functions menu
1 = Chemical comp
2 = Temperature comp
3 = Sensitivity corr
4 = Recipes
- Chemical compensation
Compensates for conductivity variations in the pulp suspension
- Temperature compensation
Included from factory, but can be ne tuned if necessary
- Sensitivity correction
Set for wood bers, can be tuned for other materials
- Recipes
If one calibration cannot cover the entire consistency span, up to
4 recipes can be used
For more detailed information refer to Metso MCA manuals!
General:
Edit
Save
Enter
Esc
Sample
0
...
9wxyz
Start editing, change values, save changes
Move to next value, scroll function
Move up in menu structure
Scroll in display, change values
Store values for calibration
Select menus, edit values

Popular Transmitter manuals by other brands

Distech Controls HS-OT 2X Series Commissioning and hardware installation manual

Distech Controls

Distech Controls HS-OT 2X Series Commissioning and hardware installation manual

PEHA Easyclick FU-BLS 20.450 Series Installation and operating instructions

PEHA

PEHA Easyclick FU-BLS 20.450 Series Installation and operating instructions

Acromag DT233-0600 user manual

Acromag

Acromag DT233-0600 user manual

Leviton 57A00-1 installation instructions

Leviton

Leviton 57A00-1 installation instructions

HK Instruments DPT Series installation instructions

HK Instruments

HK Instruments DPT Series installation instructions

Retekess TA001 user manual

Retekess

Retekess TA001 user manual

Broadata LBC-H-T/R-4P-LITE user manual

Broadata

Broadata LBC-H-T/R-4P-LITE user manual

PR electronics 3532 manual

PR electronics

PR electronics 3532 manual

Bison Goal Tender IPGTS Operation manual

Bison

Bison Goal Tender IPGTS Operation manual

Emerson Rosemount 5400 Series quick start guide

Emerson

Emerson Rosemount 5400 Series quick start guide

Siemens US-100-1 operating instructions

Siemens

Siemens US-100-1 operating instructions

SenseAir eSENSE Ind installation manual

SenseAir

SenseAir eSENSE Ind installation manual

Directed Video WFM100 Owner's installation guide

Directed Video

Directed Video WFM100 Owner's installation guide

Absolute Process Instruments API 4380 G HV3 quick start guide

Absolute Process Instruments

Absolute Process Instruments API 4380 G HV3 quick start guide

Handsender express BHS140 quick start guide

Handsender express

Handsender express BHS140 quick start guide

Transmitter Solutions Firelly 3 SU7L318ALD21K3 manual

Transmitter Solutions

Transmitter Solutions Firelly 3 SU7L318ALD21K3 manual

Lectrosonics LMb/E01 quick start guide

Lectrosonics

Lectrosonics LMb/E01 quick start guide

Inovonics Frequency Agile FA200W installation instructions

Inovonics

Inovonics Frequency Agile FA200W installation instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.