MDM300 & MDM300 I.S. Quick Start Guide
97244 Issue 2.2, September 2013
This guide provides an overview of the correct measurement procedure for the MDM300 at dew points above -70°C.
For any further information regarding these procedures, measurement below -70°C, or any other aspect of operation,
please refer to the User’s Manual.
Quick Start
1. Follow procedure A and note the result.
2. Turn the instrument off.
3. If you wish to measure at sample pressure then use the Michell Humidity Calculator to convert this result back to
the equivalent at your sample pressure.
4. Using this converted result, and the table below, nd the appropriate procedure to measure your sample.
Dew Point Measure at System Pressure >2.5 barg (36 psig) Measure at Atmospheric Pressure
> -40°C Procedure B Procedure A
< -40°C Procedure C Procedure A
Procedure A: Standard Purge (Atmospheric Pressure)
1. Connect the MDM300 to the sample gas stream.
2. Adjust the ow rate, using the inlet ow control valve, to between 0.2 & 1.2 Nl/min (0.5 & 2.5 scfh) {MDM300
Version} OR 0.2 & 0.5 Nl/min (0.5 & 1 scfh) {MDM300 I.S. Version} and ensure that the outlet ow control
valve is fully open.
3. Turn the instrument on - the initialization period begins automatically. The screen will show a countdown of the
remaining initialization time. The status display will read ‘Initializing internal sensor’.
4. After the Initialization period is complete, the instrument will begin measurement. The ow rate should now be
adjusted, using the inlet ow control valve, to 1.0 Nl/min (2.1 scfh).
5. During the rst few minutes of measurement, the displayed moisture value will drop rapidly below the actual
moisture value, before returning to a stable reading. The instrument display will show an initial measured value,
and the status display will show ‘Measurement in progress’.
6. Once ‘Measurement in progress’ has disappeared, the instrument is close to the nal dew point. Depending on the
conditions, it may continue to respond for a number of minutes before tracking the measured dew point.
Procedure B: Standard Purge (System Pressure)
1. Connect the MDM300 to the sample gas stream.
2. Adjust the ow rate, using the outlet ow control valve, to between 0.2 & 1.2 Nl/min (0.5 & 2.5 scfh) {MDM300
Version} OR 0.2 & 0.5 Nl/min (0.5 & 1 scfh) {MDM300 I.S. Version} and ensure that the inlet ow control
valve is fully open.
3. Turn the instrument on - the initialization period begins automatically. The screen will show a countdown of the
remaining initialization time. The status display will read ‘Initialising internal sensor’.
4. After the Initialization period is complete, the instrument will begin measurement. The ow rate should now be
adjusted, using the outlet ow control valve, to 1.0 Nl/min (2.1 scfh).
5. During the rst few minutes of measurement, the displayed moisture value will drop rapidly below the actual
moisture value, before returning to a stable reading. The instrument display will show an initial measured value,
and the status display will show ‘Measurement in progress’.
6. Once ‘Measurement in progress’ has disappeared, the instrument is close to the nal dew point. Depending on the
conditions, it may continue to respond for a number of minutes before tracking the measured dew point.
Procedure C: Enhanced Purge
1. Connect the MDM300 to the sample gas stream.
2. Adjust the ow rate, using the inlet ow control valve, to between 0.2 & 1.2 Nl/min (0.5 & 2.5 scfh) {MDM300
Version} OR 0.2 & 0.5 Nl/min (0.5 & 1 scfh) {MDM300 I.S. Version} and ensure that the outlet ow control
valve is fully open.
3. Turn the instrument on - the initialization period begins automatically. The screen will show a countdown of the
remaining initialization time. The status display will read ‘Initialising internal sensor’.
4. After 2-3 minutes of Initialization have elapsed, close the outlet ow control valve and fully open the inlet ow
control valve. The ow rate should now be adjusted, using the outlet ow control valve, to 1.0 Nl/min (2.1
scfh).
5. After the Initialization period is complete, the instrument will now begin measurement. During the rst few minutes
of measurement, the displayed moisture value will drop rapidly below the actual moisture value, before returning
to a stable reading. The instrument display will show an initial measured value, and the status display will show
‘Measurement in progress’.
6. Once ‘Measurement in progress’ has disappeared, the instrument is close to the nal dew point. Depending on the
conditions, it may continue to respond for a number of minutes before tracking the measured dew point.
MDM300 & MDM300 I.S. Quick Start Guide
97244 Issue 2.2, September 2013
Atmospheric
Atmospheric
System
System
Yes
No
Connect the MDM to
the sample gas stream
Determine the approximate
dew point by using the ‘initial check’
procedure on the reverse
During the first few minutes of
measurement, the displayed moisture
value will drop rapidly below the actual
moisture value, before returning
to a stable reading.
The instrument’s display will show an initial
measured value, and the status display
will show ‘Measurement in progress’.
Do you
want to measure
at Atmospheric pressure
or System
pressure?
Do you
want to measure
at Atmospheric pressure
or System
pressure?
Is the
equivalent
dew point
> - 40Cdp?
Turn the instrument
on. The status display
will read ‘Initializing
internal sensor’. After
2-3 mins of initialization
have elapsed,
close the outlet flow
control valve and
fully open the inlet
flow control valve.
Then adjust the flow
rate using the outlet
flow control valve
to 1.0 Nl/min.
After initialization period
is complete, the
instrument will begin
measurement.
Turn the instrument on. The status display
will read ‘Initializing internal sensor’.
After initialization period is complete, the
instrument will begin measurement.Then
adjust the flow rate using the inlet
flow control valve to 1.0 Nl/min.
Adjust the flow rate
using the
inlet flow control valve
to between
0.2 & 1.2 Nl/min
{MDM300 Version}
0.2 & 0.5 Nl/min
{MDM300 I.S. Version}
Ensure outlet flow control
valve is fully open.
Adjust the flow rate
using the
outlet flow control valve
to between
0.2 & 1.2 Nl/min
{MDM300 Version}
0.2 & 0.5 Nl/min
{MDM300 I.S. Version}
Ensure inlet flow
control valve is
fully open
Adjust the flow rate using the
inlet flow control valve
to between
0.2 & 1.2 Nl/min
{MDM300 Version}
0.2 & 0.5 Nl/min
{MDM300 I.S. Version}
Ensure outlet flow control
valve is fully open.
Typical MDM300 Sampling Arrangement
VENT
1
1
2
2
3
3
4
4
SAMPLE FLOW
Coalescing Filter
Particulate Filter
Inlet Flow Control Valve
Outlet Flow Control Valve Once ‘Measurement in progress’
has disappeared, the instrument is close
to the final dew point.
Depending on the conditions, it may
continue to respond for a number of
minutes before tracking the measured dew point.
Turn the instrument on. The
status display will read
‘Initializing internal sensor’.
After initialization period is
complete, the instrument will
begin measurement. Then
adjust the flow rate using
the outlet flow control valve
to 1.0 Nl/min.
!
HIGH PRESSURE! High pressure gases are potentially hazardous. Energy stored in these
gases can be released suddenly and with extreme force. High pressure systems should be
assembled and operated only by people who have been trained in proper safety practices.

Other manuals for MDM300 I.S.

5

This manual suits for next models

1

Other Michell Instruments Measuring Instrument manuals

Michell Instruments QMA401 User manual

Michell Instruments

Michell Instruments QMA401 User manual

Michell Instruments Easidew Portable Standard User manual

Michell Instruments

Michell Instruments Easidew Portable Standard User manual

Michell Instruments MDM50 User manual

Michell Instruments

Michell Instruments MDM50 User manual

Michell Instruments Pura User manual

Michell Instruments

Michell Instruments Pura User manual

Michell Instruments MDM300 I.S. User manual

Michell Instruments

Michell Instruments MDM300 I.S. User manual

Michell Instruments XTP601 User manual

Michell Instruments

Michell Instruments XTP601 User manual

Michell Instruments XTP501 User manual

Michell Instruments

Michell Instruments XTP501 User manual

Michell Instruments Cermet II User manual

Michell Instruments

Michell Instruments Cermet II User manual

Michell Instruments Liquidew EExd User manual

Michell Instruments

Michell Instruments Liquidew EExd User manual

Michell Instruments QMA601 User manual

Michell Instruments

Michell Instruments QMA601 User manual

Michell Instruments Liquidew I.S. User manual

Michell Instruments

Michell Instruments Liquidew I.S. User manual

Michell Instruments Easidew Online User manual

Michell Instruments

Michell Instruments Easidew Online User manual

Michell Instruments Condumax II User manual

Michell Instruments

Michell Instruments Condumax II User manual

Michell Instruments XZR250 User manual

Michell Instruments

Michell Instruments XZR250 User manual

Michell Instruments PC33 User manual

Michell Instruments

Michell Instruments PC33 User manual

Michell Instruments Promet I.S. User manual

Michell Instruments

Michell Instruments Promet I.S. User manual

Michell Instruments MDM300 I.S User manual

Michell Instruments

Michell Instruments MDM300 I.S User manual

Michell Instruments Optidew User manual

Michell Instruments

Michell Instruments Optidew User manual

Michell Instruments XZR400 Series User manual

Michell Instruments

Michell Instruments XZR400 Series User manual

Michell Instruments QMA Series Parts list manual

Michell Instruments

Michell Instruments QMA Series Parts list manual

Michell Instruments MDM25 User manual

Michell Instruments

Michell Instruments MDM25 User manual

Michell Instruments S8000 User manual

Michell Instruments

Michell Instruments S8000 User manual

Michell Instruments XTC601 User manual

Michell Instruments

Michell Instruments XTC601 User manual

Michell Instruments Optidew User manual

Michell Instruments

Michell Instruments Optidew User manual

Popular Measuring Instrument manuals by other brands

CYP CUSB-V605H Operation manual

CYP

CYP CUSB-V605H Operation manual

Ebtron GTA116e-P Startup guide

Ebtron

Ebtron GTA116e-P Startup guide

Intron Plus INTROMET ITM-525 operating manual

Intron Plus

Intron Plus INTROMET ITM-525 operating manual

Agilent Technologies ESA-L1500A Installation notes

Agilent Technologies

Agilent Technologies ESA-L1500A Installation notes

Air Monitor Corporation VOLU-flo / OAM INSTALLATION PROCEDURE

Air Monitor Corporation

Air Monitor Corporation VOLU-flo / OAM INSTALLATION PROCEDURE

Masimo PRONTO-7 Operator's manual

Masimo

Masimo PRONTO-7 Operator's manual

Hanna Instruments HI 931100 instruction manual

Hanna Instruments

Hanna Instruments HI 931100 instruction manual

Leica RANGEMASTER CRF 1000-R instructions

Leica

Leica RANGEMASTER CRF 1000-R instructions

MONARCH INSTRUMENT Pocket Tach 99 instruction manual

MONARCH INSTRUMENT

MONARCH INSTRUMENT Pocket Tach 99 instruction manual

Sekonic DUAL MASTER L-558 Opearting manual

Sekonic

Sekonic DUAL MASTER L-558 Opearting manual

MFJ MFJ-836H instruction manual

MFJ

MFJ MFJ-836H instruction manual

Earthworks M30BX owner's manual

Earthworks

Earthworks M30BX owner's manual

Hesai PandarXT-16 user manual

Hesai

Hesai PandarXT-16 user manual

Knick Stratos PROFIBUS 2221X Cond user manual

Knick

Knick Stratos PROFIBUS 2221X Cond user manual

Kamstrup MULTICAL 402 Installation and user guide

Kamstrup

Kamstrup MULTICAL 402 Installation and user guide

MRC EMF-831SD operating manual

MRC

MRC EMF-831SD operating manual

R.M. Young 52202 instruction sheet

R.M. Young

R.M. Young 52202 instruction sheet

Danaher Hengstler tico 734 operating instructions

Danaher

Danaher Hengstler tico 734 operating instructions

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.