Minebea Intec PR 6130/68S User manual

Cable Junction Box PR 6130/68S
Installation Manual
Translation of the Original Installation Manual 9499 053 06801 Edition 3.1.1 08/24/2017
Minebea Intec GmbH, Meiendorfer Str. 205 A, 22145 Hamburg, Germany
Phone: +49.40.67960.303 Fax: +49.40.67960.383

Foreword
Must be followed!
Any information in this document is subject to change without notice and does not represent a commitment on
the part of Minebea Intec unless legally prescribed. This product should be operated only by trained and
qualied personnel. In correspondence concerning this product, the type, designation and version number as
well as all license numbers relating to the product have to be cited.
Note
The complete product is protected by copyright. It may not be changed or copied, and it may not be used
without purchasing or written permission from the copyright owner (Minebea Intec). Its use constitutes
acceptance by you of the above-mentioned provisions.

Table of contents
1 Introduction......................................................................................................................................... 3
1.1 Read the manual.........................................................................................................................................................3
1.2 This is what instructions look like.............................................................................................................................3
1.3 This is what lists look like...........................................................................................................................................3
1.4 This is what menus and soft keys look like.............................................................................................................3
1.5 This is what safety instructions look like.................................................................................................................3
1.6 Hotline..........................................................................................................................................................................4
2 Safety instructions...............................................................................................................................5
2.1 General notes..............................................................................................................................................................5
2.2 Intended use................................................................................................................................................................5
2.3 Incoming goods inspection.......................................................................................................................................6
2.4 Before operational startup........................................................................................................................................6
2.5 Repairs and maintenance..........................................................................................................................................6
2.5.1 General information ........................................................................................................................................6
3 Specications ...................................................................................................................................... 7
3.1 Equipment supplied................................................................................................................................................... 7
3.2 Technical Data.............................................................................................................................................................8
3.3 Electromagnetic Compatibility (EMC) .....................................................................................................................9
3.4 Marking for the Ex Zone............................................................................................................................................9
3.5 Dimensions................................................................................................................................................................. 11
4 Installation and connection information .............................................................................................12
4.1 General information..................................................................................................................................................12
4.2 Connecting intrinsically safe circuits......................................................................................................................12
4.3 Terminals and jumpers .............................................................................................................................................13
4.4 Cable gland................................................................................................................................................................ 14
5 Cable connections...............................................................................................................................15
5.1 General information..................................................................................................................................................15
5.2 Connecting cables .....................................................................................................................................................15
5.3 Cable connections .....................................................................................................................................................15
5.4 Equipotential bonding conductor...........................................................................................................................17
5.5 Electronic corner correction.................................................................................................................................... 18
5.5.1 General information ...................................................................................................................................... 18
5.5.2 Procedure........................................................................................................................................................ 18
6 Maintenance/repairs/soldering work/cleaning .................................................................................20
6.1 Maintenance..............................................................................................................................................................20
6.2 Repairs........................................................................................................................................................................20
6.3 Soldering work..........................................................................................................................................................20
Cable Junction Box PR 6130/68S Table of contents
EN-1 Minebea Intec

6.4 Cleaning .....................................................................................................................................................................20
7 Disposal ..............................................................................................................................................21
8 Appendix ........................................................................................................................................... 22
Cable Junction Box PR 6130/68S Table of contents
Minebea Intec EN-2

1 Introduction
1.1 Read the manual
-Please read this manual carefully and completely before using the product.
-This manual is part of the product. Keep it in a safe and easily accessible location.
1.2 This is what instructions look like
1. - n. are placed before steps that must be done in sequence.
1.3 This is what lists look like
-indicates an item in a list.
1.4 This is what menus and soft keys look like
[ ] frame menu items and soft keys.
Example:
[Start]- [Applications]- [Excel]
1.5 This is what safety instructions look like
Signal words indicate the severity of the danger involved when measures for preventing
hazards are not followed.
DANGER
Warning of personal injury
DANGER indicates that death or severe, irreversible personal injury will occur if
appropriate safety measures are not observed.
Take appropriate safety measures.
WARNING
Warning of potential health risk or risk of personal injury!
WARNING indicates that death or severe, irreversible injury may occur if appropriate
safety measures are not observed.
Take appropriate safety measures.
CAUTION
Warning of personal injury and/or property damage.
CAUTION indicates that minor, reversible injury or damage to property may occur if
appropriate safety measures are not observed.
Take appropriate safety measures.
is placed before a step.
describes the result of a step.
1 Introduction Cable Junction Box PR 6130/68S
EN-3 Minebea Intec

NOTICE
Warning of property and/or environmental damage.
ATTENTION indicates that damage to property and/or the environment may occur if
appropriate safety measures are not observed.
Take appropriate safety measures.
Note:
User tips, useful information and notes.
1.6 Hotline
Phone: +49.40.67960.444
Fax: +49.40.67960.474
E-mail: [email protected]
Cable Junction Box PR 6130/68S 1 Introduction
Minebea Intec EN-4

2 Safety instructions
2.1 General notes
CAUTION
Warning of personal injury.
The product was in perfect condition with regard to safety features when it left the
factory.
To maintain this condition and to ensure safe operation, the user must follow the
instructions and observe the warnings in this manual.
2.2 Intended use
The cable junction box is designed for tank and hopper weighing applications.
It can also be operated in potentially explosive atmospheres.
Product operation, commissioning and maintenance must be performed by trained and
qualied personnel who are aware of and able to deal with the related hazards and take
suitable measures for self-protection.
The device reects the state of the art.
The manufacturer does not accept any liability for damage caused by third-party system
components or due to incorrect use of the product. The use of this product signies
recognition of the stipulations listed above.
The following table shows the load cells that should and should not be used for dierent
applications.
Load cells to be used Load cells not to be used
PR 6201/..L, /..LE,/..N, /..NE, /..D1, /..D1E,
/..C3, /..C3E, /..C4, /..C4E, /..C5, /..C5E,
/..C6, /..C6E
PR 6201/..LA
PR 6201/..NDB, /..NDBE, /..LDB, /..LDBE
PR 6202/..C1, /..C1E, /..C3, /..C3E, /..C4,
/..C4E
PR 6204 Pendeo® Process
PR 6207/..D1, /..C3
PR 6211/..D1, /..D1E, /..LT
PR 6212/..C1, /..C1E, /..LT
PR 6221/.., all types
PR 6224 Pendeo® Truck
PR 6241/..D1, /..D1E, /..C3, /..C3E, /..C6,
/..C6E
Contego®/..D1, /..D1Ex, /..C3, /..C3Ex
PR 6246/..D1, /..D1E, /..C3, /..C3E, /..C6,
/..C6E
2 Safety instructions Cable Junction Box PR 6130/68S
EN-5 Minebea Intec

Load cells to be used Load cells not to be used
PR 6251/..L, /..LE PR 6251/..LA, /..LAC
Novego®/..D1, /..D1E, /..C3, /..C3E
MP 55/..C3MR+, /..C3MR+E
MP 58(T)/..C3MR, /..C3MRE
MP 70/..C3MR
MP 71/..C3MR
MP 72/..C3MR, /..C3MRE
MP 76/..N, /..C3
MP 77/..C3MR, /..C3MRE
MP 79(T)/..C3MR, /..C3MRE
2.3 Incoming goods inspection
The shipment must be checked for completeness. A visual inspection must be performed
to determine if the shipment has been damaged. If there are grounds for a complaint, this
must be brought to the attention of the delivery company immediately. A Minebea Intec
sales or service point must be informed.
2.4 Before operational startup
NOTICE
Perform visual inspection.
Before operational startup as well as after storage or transport, inspect the device
visually for signs of mechanical damage.
2.5 Repairs and maintenance
2.5.1 General information
Repairs are subject to inspection and must be carried out at Minebea Intec.
In case of defect or malfunction, please contact your local Minebea Intec dealer or service
center for repair.
When returning the device for repair, please include a precise and complete description of
the problem.
Maintenance work may only be carried out by a trained technician with expert knowledge
of the hazards involved and the required precautions.
Cable Junction Box PR 6130/68S 2 Safety instructions
Minebea Intec EN-6

3 Specications
3.1 Equipment supplied
No. Description
1 Cover
2 Box incl. electronics
3 Cable gland M12 (8×)
4 Cable gland M20 (9–13 mm)
5 Ex-zone locking pin 6 mm (3×)
6 Cable gland M20 (6–12 mm)
7 Pressure compensation element
3 Specications Cable Junction Box PR 6130/68S
EN-7 Minebea Intec

No. Description
8 Jumpers for corner correction impedances
The following items are not shown:
9 Drilling template
10 Installation manual 9499 053 06801
11 Certicates
3.2 Technical Data
Protection classes per DIN EN 60529
IP68, IP69:
Dust-proof and leak-proof against water, with harmful
effects when immersed, (1.5 m water depth, 10,000 h)
and water jets (high pressure and temperature).
Installation position Cable entry from below
Quantity of load cells max. 8
Clamping area screw connection M12 36 mm
Clamping area screw connection M20 612 mm/913 mm
Overvoltage protection 2× Surge arrestor Type A81C90X
Pressure equalization Stainless steel pressure equalization element
Material of the junction box High-grade stainless steel 1.4404 (X2CrNiMo17-12-2)
according to EN 10088-3, AISI 316L, JIS SUS316
Surface of the junction box 2R according to DIN EN 10088-2 (cold-rolled, bright
annealed, at, blank, reective)
Net weight 1.4 kg
Shipping weight 1.5 kg
Service temperature area -30°C+80°C
Service temperature area, in explosion-prone area -20°C+60°C
Only with Authorization RU С-DE.МЮ62.В.05021:
-52+60 °C
Storage temperature range -30°C+80°C
Cable screw connections Metal Ex EMC cable screw connections
Brass CuZn39Pb3, galv. nickel-plated
Insulation impedance (in service temperature range at
95% air humidity and 500 V DC) >1000 MΩ
Ex-connection values See Chapter 8.
Cable Junction Box PR 6130/68S 3 Specications
Minebea Intec EN-8

3.3 Electromagnetic Compatibility (EMC)
All data corresponds to EN 61326 industrial section
High frequency electromagnetic elds
(803000 MHz) EN 61000-4-3 10 V/mHousing
Electrostatic discharge (ESD) EN 61000-4-2 6/8 kV
Fast transients (burst) EN 61000-4-4 1 kV
Peak voltages (surge) 1.2/50 µs EN 61000-4-5 1 kV
Signal and control
lines
Conducted disturbances by high fre-
quency coupling (0.1580 MHz) EN 61000-4-6 10 V
3.4 Marking for the Ex Zone
Zone Marking Certicate No.
0 and 1 II 1 G Ex ia IIC T6/T4 Ga
Ex ia IIC T6/T4 Ga
0Ex ia IIC T6/T4 X
DEKRA 13ATEX0133 X
IECEx DEK 13.0042X
RU С-DE.МЮ62.В.05021*
20 II 1 D Ex ia IIIC T90 °C Da
Ex ia IIIC T90 °C Da
Ex ia IIIC T90 °C X
DEKRA 13ATEX0133 X
IECEx DEK 13.0042X
RU С-DE.МЮ62.В.05021*
21 and 22 II 2 D tb IIIC T50 °C/T70 °C Db
Ex tb IIIC T50 °C/T70 °C Db
Ex tb IIIC T50 °C/T70 °C X
DEKRA 13ATEX0133 X
IECEx DEK 13.0042X
RU С-DE.МЮ62.В.05021*
2 II 3G Ex nA IIC T6 Gc
Ex nA IIC T6 Gc
2Ex nA IIC T6 X
DEKRA 13ATEX0133 X
IECEx DEK 13.0042X
RU С-DE.МЮ62.В.05021*
* Certifying body: Prommash Test LLC
(Accreditation Code МЮ62)
NOTICE
Installation in Ex Area
The Ex safety instructions in the appendices must be observed when installing in the Ex area.
3 Specications Cable Junction Box PR 6130/68S
EN-9 Minebea Intec

3.5 Dimensions
all dimensions in mm
3 Specications Cable Junction Box PR 6130/68S
EN-11 Minebea Intec

4 Installation and connection information
4.1 General information
Note:
When used in protection class "Ex nA" (non-sparking), a transient protective device must
be set to a level that does not exceed 140% of the peak voltage of 85 V.
When used in environments with ammable dust, electrostatic discharge from the
plastic label must be avoided.
-The safety instructions in Chapter 8must be followed!
-Only use the metal EX EMC cable glands provided by the manufacturer.
-Install the junction box so that the cable glands are on the bottom.
-Do not open the junction box when connected to the voltage supply.
Note:
At ambient temperatures >45°C, cables that are suitable for at least 85°C must be used.
4.2 Connecting intrinsically safe circuits
-The housing grounding or equalization line must be connected to the outside of the
housing under the ground screw.
-The junction box is suitable for use in potentially explosive atmospheres for
connection to intrinsically safe circuits. The circuits consist of
-the connection load cells (passive)
-the connection cable to an interface with an (active) intrinsically safe circuit, e.g.,
PR 1626/60 in connection with a downstream device.
The intrinsically safe circuit consists of the circuits for the supply, sense, and
measuring voltage.
-It is not permissible to connect multiple active intrinsically safe circuits in the junction
box.
-When used in zone 20, the accumulated dust must not exceed a thickness of 5 mm.
-When used in zone 2, ensure that no voltage peaks more than 40% above the
nominal voltage can occur.
Cable Junction Box PR 6130/68S 4 Installation and connection information
Minebea Intec EN-12

4.3 Terminals and jumpers
Terminal contacts
18+ measuring voltage (load cells)Signal + green
Terminal contact
IND + measuring voltage (connection cable)
Terminal contacts
18- measuring voltage (load cells)Signal - gray
Terminal contact
IND - measuring voltage (connection cable)
Terminal contacts
18+ supply voltage (load cells)
Terminal contact
IND + supply voltage (connection cable)
Supply + red
Terminal contact S + sense (connection cable – white)
Terminal contacts
18- supply voltage (load cells)
Terminal contact
IND - supply voltage (connection cable)
Supply - blue
Terminal contact S - sense (connection cable – black)
Terminal contacts
12Screen (connection cable)Screen yellow
Terminal contacts
19Screen (load cells)
4 Installation and connection information Cable Junction Box PR 6130/68S
EN-13 Minebea Intec

4.4 Cable gland
The cables have to be fed into the device via glands to ensure leak-tightness. The
following cable diameters are suitable: 612 mm for gland M20 and 36 mm for cable
gland M12.
The enclosed M20 cable gland is suitable for cable diameters 913 mm.
The cable wires are connected to the terminals inside the device.
NOTICE
For protection against dust and moisture during transport and installation, the
cable glands are tted with a polyethylene cover.
For full IP protection, operation with the dust protection cover tted is not permitted.
Remove the dust protection cover.
If a cable gland is not used, it must be sealed with a supplied locking pin.
NOTICE
Property damage is possible.
Regularly check the tted cable gland for tightness and re-tighten it, if necessary.
Cable Junction Box PR 6130/68S 4 Installation and connection information
Minebea Intec EN-14

5 Cable connections
5.1 General information
Cable entry must be from below.
Fit the wires and screen of the connection cable with wire end ferrules. This is not
necessary for the load cell cables.
Connect the wires to the terminals according to the color coding .
Place the cable screens of the connection cable and load cell cables on the yellow
terminals. The terminals are connected with the housing via overvoltage conductor.
Connect the screens on the other side of the connection cable with the equipotential
bonding terminal of the downstream device (see instrument manual) as described in
Chapters 5.3 and 5.4.
WARNING
When installing in potentially explosive atmospheres:
It is imperative that you follow the application-dependent installation instructions!
Always check whether it is permissible to bilaterally connect the screens to the
equipotential bonding.
Note:
The calibration certicate, installation manual and data sheet of the load cell contain
information about how to connect the load cell cable screen with the load cell housing.
5.2 Connecting cables
To connect the junction box to the device, connection cable PR 6135/..or PR 6136/.. (for Ex
zone) must be used; max. length: 300 m.
5.3 Cable connections
Note:
All components are only shown schematically.
5 Cable connections Cable Junction Box PR 6130/68S
EN-15 Minebea Intec

Color code
bk = black
bu = blue
gn = green
gy = gray
rd = red
wh = white
Cable Junction Box PR 6130/68S 5 Cable connections
Minebea Intec EN-16

5.4 Equipotential bonding conductor
NOTICE
It is especially important that the ground is connected correctly to the components
and the cable junction box.
You must also ground the device separately and ensure the power supply is properly
shielded against the eects of lightning. Simply connecting the protective grounding
conductor is not enough!
If the installation is not carried out according to our instructions, this voids the
warranty. In particular, the entire installation, including the power supply, must be
suciently protected against lightning.
In Ex areas, equipotential bonding must be established to avoid any compensatory
currents between the individual conducting system components.
5 Cable connections Cable Junction Box PR 6130/68S
EN-17 Minebea Intec

5.5 Electronic corner correction
5.5.1 General information
Load cells from Minebea Intec are manufactured according to high quality standards and
have very precisely balanced output values.
Nevertheless, mechanical imbalances may cause impermissibly large corner load errors,
which must be compensated for by soldering in resistors.
Note:
The soldering studs for the resistors are short-circuited at the factory.
The load cells must be correctly installed and precisely aligned in order to achieve good
measuring results and good results with corner loads. For this reason, if you notice a
corner load error, always check the installation and alignment of the load cell rst.
If necessary, mechanically adjust the height (see load cell installation manual).
NOTICE
Loss of verication accuracy
Contact problems cause unstable signals. The consequences are wandering zero points
that lead to loss of verication accuracy.
Do not use a potentiometer.
Only use resistance values of 05.62 Ω (1%, P70 = 0.6 W) of size MBB0207
(approx. 2.5 × 6.5 mm) or CECC B.
5.5.2 Procedure
The corner load error can be minimized by selecting reference cell to be load cell with the
lowest displayed load.
The other load cells are adjusted to value displayed on the reference cell by additional
resistors at the input of the load cells.
The electronic corner correction makes it possible to reduce the sensitivity of each
separate load cell individually.
The following formula is used:
R Resistivity added to the input circuit of the load cell
(the resistivity is normally <7 Ω)
Weight nom Weight value in the display
Weight target Weight of applied load
RLC Input impedance of the load cell
Example:
Loaded piece 12,000 kg
Cable Junction Box PR 6130/68S 5 Cable connections
Minebea Intec EN-18
Other manuals for PR 6130/68S
1
Table of contents
Languages:
Other Minebea Intec Cable Box manuals

Minebea Intec
Minebea Intec PR 6130/64Sa User manual

Minebea Intec
Minebea Intec IE11357 User manual

Minebea Intec
Minebea Intec PR 6021/08 User manual

Minebea Intec
Minebea Intec PR 6130/68S User manual

Minebea Intec
Minebea Intec PR 6130/04N User manual

Minebea Intec
Minebea Intec PR 6024/68S User manual

Minebea Intec
Minebea Intec PR 6130/65S User manual

Minebea Intec
Minebea Intec Weighbridge PR 6021/68S User manual

Minebea Intec
Minebea Intec Weighbridge PR 6021/18 User manual

Minebea Intec
Minebea Intec PR 6130/64Sa User manual