
FPD-IG-25 User Manual 3
II. Operation of your Ophir Fast Photodiode Detector
A. Caution: Eye safety precautions must be followed with any equipment used in the vicinity of laser
beams. Laser beams may reflect from the surface of the detector or the optical mount and caution
must be exercised.
B. Mount the detector to an optical stand by the mounting holes on the bottom of the detector housing.
Each detector is provided with an 8-32 threaded nylon standoff. If the detector is mounted to a metal
post, it is recommended to insert the standoff between the detector and the post in order to minimize
the potential for electrical noise pickup.
C. Adjust the voltage scale of the oscilloscope to 20mV/division before connecting the detector.
D. Connect the detector to the oscilloscope using a coaxial cable designed for at least 15 GHz operation.
E. Use the 50Ω termination input of the oscilloscope.
F. After being certain that the damage threshold of the detector is not exceeded, place the detector in
the laser beam and turn the detector power switch to “On”.
G. There is an internal 50Ω resistor in the output stage of the detector circuit. This will reduce the
output current from the detector to 50% of the photodiode output current. For example, the output to
your equipment will be 450µA for a 1mW optical input at a wavelength at which the responsivity is
0.9A/W.
Note: This 50Ω resistor presents a constant load to the photodiode so anytime the detector
power switch is turned on, current will be drawn from the batteries to flow through the
photodiode. To prevent premature battery drainage, turn off the power switch whenever the
detector is not in use.
III.Batteries
Batteries will typically operate for several years, but operation with CW or high rep rate lasers can
drain the batteries much faster. Leaving the detector output connected to a load for long periods of
time can also drain the batteries. It is recommended to disconnect the output whenever the detector
will not be in use for long periods of time.
As the batteries become depleted, the bias voltage on the photodiode will decrease. This will
increase the rise and falls times of the detector and also reduce its saturation level. The ability of the
battery to supply current quickly will also be reduced which is manifested as saturation effects in the
output signal.
The batteries can be accessed for checking their voltage and replacement by removing the top plate
of the housing. If necessary, replace with type two (2) CR2430 lithium cells. Install batteries with
positive side down.