MAINTENANCE
CHANGING PRE-FILTERS
Regular checks should be carried out to ensure the airflow performance is
maintained. Should the module fail to achieve adequate performance, this normally
indicates that the Pre-Filter/s and / or HEPA filter require changing.
(For replacement filter part numbers, see page 5)
1. Turn the system off and disconnect from mains.
2. Remove the screws securing the pre-filter retaining frame.
3. Remove the pre-filter frame and filter.
4. Replace pre-filter with the correct type (see page 5).
5. Replace the screws securing the pre-filter retaining frame.
6. Reconnect the mains supply and switch on.
7. Check the airflow and adjust if necessary (as detailed in the setup section)
CHANGING HEPA FILTER
1. Turn the system off and disconnect from mains.
2. Remove the screws securing the filter retaining frame.
3. Remove the filter frame, cover filter (if fitted) and HEPA filter.
4. Replace HEPA filter with the correct type (see page 5).
5. Replace cover filter and filter frame using the screws.
6. Reconnect the mains supply and switch on.
7. Check the airflow and adjust if necessary (as detailed in the setup section)
8. Filter should be DOP tested to verify filter integrity and results recorded.
REPLACEMENT FILTER PART NUMBERS
Replacement filters are available from Monmouth Scientific:
CAM1000:
Pre-Filter Part No- PF-0081
HEPA –filter Part No- HF-0127
SERVICING
An annual service is recommended to maintain optimum operating conditions and will
include the following points:
•Check / replace pre-filter
•DOP test the main HEPA filter if required
•Particle count check if required
•Check airflow monitor and re-calibrate if necessary
•Check and record air velocity readings
•Check general condition of system.
•Issue test report and airflow certificate.
For parts or service information please contact Monmouth Scientific on:
+44 (0) 1278 458090