Low airflow alarm calibration
For factory calibration purposes, the unit is set to alarm at 0.5m/sec with a
boundary point 10mm from the filter face. It is recommended the user sets
their own parameter to suit their specific application and limits.
•Turn off the cabinet and turn on again whilst holding down the ‘ALARM
MUTE’ button. Release the button and the airflow indicator lights will flash
Red/Green alternately.
•Using a 100mm rotating vane anemometer positioned at the required
minimum capture velocity boundary point from the pre-filter face and by
rotating the adjustment knob, adjust the fan speed to the desired minimum
allowable velocity. Check in at least 3 positions per pre-filter to obtain an
average and adjust as necessary.
•Wait approximately 20 seconds for the airflow to stabilise and press MUTE
again to record the set point.
•The low airflow alarm point is now set and the system calibrated.
•Re-Adjust the fan speed back up to the normal operating velocity as required.
Section 5
Servicing
An annual service is recommended to maintain optimum operating conditions and
will include the following points:
•Check / replace pre-filter
•DOP test the main HEPA filter
•Check airflow monitor and re-calibrate if necessary
•Check and record air velocity readings
•Check general condition of cabinet.
•Inspect electrical components, lighting, cables etc.
•Issue test report and airflow certificate.