E. Adjusting the limit contacts
Notice: The hysteresis can only adjusted by programming software Mpro (see H and I, page 6)
1. Push and hold the button <ENTER> for 5 seconds.
The display will show up the program menu.
2. Change to menu SP_1 for limit contact #1 or SP_2 for limit contact #2,
by the button <+> or <->, then press <ENTER>.
The display shows the actual value of the limit contact.
3. Change the value for the limit contact by the button <+> or <->, then press <ENTER>.
4. Change to menu HST_1 for the hysteresis of the limit contact #1 or HST_2 for the hysteresis of the
limit contact #2, by the button <+> or <->, then press <ENTER>.
The display shows the actual value of the hysteresis.
5. Change the value for the hysteresis by the button <+> or <->, then press <ENTER>.
6. Change to menu LEd_1 for limit contact #1 or LEd_2 for limit contact #2,
by the button <+> or <->, then press <ENTER>.
7. Change the lamp status by the button <+> or <-> , then press <ENTER>.
"nor":LED lamp ON, when input > setpoint
"Inr": LED lamp ON, when input < setpoint
8. Change to menu rEL_1 for limit contact #1 or rEL_2 for limit contact #2,
by the button <+> or <->, then press <ENTER>.
9. Change relay operating mode by the button <+> or <-> , then press <ENTER>.
"nor":Relay ON, when input > setpoint (normally open contact closed)
"Inr": Relay ON, when input < setpoint (normally open contact open)
10.Press button <ESC>
The display shows the actual value.
F. Change current output
1. Push and hold the button <ENTER> for 5 seconds.
The display will show up the program menu.
2. Change to menu OutI, by the button <+> or <->, then press <ENTER>.
The display shows the actual value.
3. Change the value to 0_20 or 4_20 (for 4…20mA) by the button <+> or <->, then press <ENTER>.
4. Change to menu OutP, by the button <+> or <->, then press <ENTER>.
The display shows the actual value.
5. Change the value to 2_12 or 0_14 (for 0…14pH) by the button <+> or <->, then press <ENTER>.
6. Press button <ESC>
The display shows the actual value.
G. Keyboard lock
The values of the user menu can be protected from unintended access using an access code. After activating
the access code, menu values can be observed but no longer modified. To modify setpoints and other
parameters, the access code has to be set to the value 0000. If the code has a value different from 0000,
writing to the user menu is suppressed. Did you forget your code? Call us we have a special unlock code
available.
Set the access code:
1. Press and hold the button <ENTER> for 5 seconds.
The display show up the program menu.
2. Now navigate to the Code menu item with the button <->, then press <ENTER>.
a) A code has already been entered: The display flashes between Code und 0. Enter now the valid
access code with the buttons <+> and <-> and confirm with the <ENTER> button. If the code is wrong,
the instrument quits the user menu (in case you do not know your code any more, call MOSTEC we have
a unlock code). If the code is correct, it can be modified with the button <+> and <->, or in order to make
changes to the user menu, set to 0000.
b) No code has been entered: The display shows 0000. A code can be entered with the button <+>
and <-> then confirm with the <ENTER> button. If you do not need the user menu protection, leave the
value 0000 unchanged and quit the code menu with the <ESC> button.