multicomp pro MP730001 User manual

1
Pen Type Multimeter
Model: MP730001
IMPORTANT SAFETY INFORMATION
SYMBOL GUIDE
WHAT’S INCLUDED
FUNCTIONS
Please read these instructions carefully before use and
retain for future reference.
• This instrument is designed and manufactured in compliance
with: lEC61010-1, CAT III 300V Pollution Degree 2 and
Double Insulation standards.
• Check the test leads, probes and case insulation before
using. If you nd any breakage or abnormality, or you
consider the device is broken, stop using the device
immediately.
• When using the test probes, keep your ngers behind the
nger protection ring.
• Replace damaged test leads with identical type and
electrical specications before using the Meter.
• Do not use or store the Meter in an environment of high
temperature, humidity, explosives, ammable materials and
strong magnetic elds.
• Do not use the meter with the back cover open.
• Select appropriate test range for measurements.
• Ensure all inputs are less than the range selected otherwise
it may cause electrical shock or meter damage.
• Do not change the range selector position during voltage
measurements.
• Disconnect circuit power and discharge all high voltage
capacitors before testing resistance, diodes or continuity.
• Do not apply a voltage over 300V between input terminal
and ground.
• Take caution when working voltages are above 60V DC or
30V AC rms.
• Do not connect the meter to voltage signals when the range
selector is on resistance, diode or continuity range.
• When a measurement has been completed, disconnect the
testing probes from the circuit under test.
• Replace the batteries as soon as the low battery indicator
appears on the display.
• Remove dead batteries from the meter or if it is not going to
be used for a long time.
• Never mix old and new batteries together, or different types
of batteries.
• Never dispose of batteries in a re, or attempt to recharge
ordinary batteries.
• Before replacing the battery, turn off the meter and
disconnect all the test probes.
• To prolong battery life turn off the meter after use.
Low Battery Ground
AC Current DC Current
Double Insulated Continuity
AC or DC Diode
• Digital pen meter with battery
• 1 pair of test leads
• Operating instructions
9. Minimum reading
10. Maximum reading
11. Capacitance
12. Diode test
13. Continuity buzzer
14. mv/V: Volts
15. Ohm.(resistance) k:kilohm; M:Megaohm.

2
OPERATION
AC/DC VOLTAGE MEASUREMENT
DC VOLTAGE MEASUREMENT
AC VOLTAGE MEASUREMENT
EF MEASUREMENT
AUTO FUNCTION MEASUREMENT
• Press SELECT to switch between resistance, AC/DC
voltage, continuity buzzer and diode measurement modes.
Press and hold to exit “sleep” mode.
• Press HOLD to enter and exit hold mode (except under auto
scan mode).
• Press and hold the HOLD button, the meter displays
Hautomatically holds the value currently displayed for
approximately 6 secs. If the Meter enters “sleep” while in
hold mode, the meter will still be in the hold mode when it is
powered back on.
• The MAX/MIN mode stores minimum (MIN) and maximum
(MAX) input values (except under auto scan mode). Manual
ranging is default when you select this function.
• Press MAX/MIN button to switch between MAX, MIN, and
MAX/MIN options.
• If under both HOLD mode and MAX/MIN mode, exit HOLD
mode rst then press and hold MAX/MIN for more than 1
second to exit MAX/MIN mode.
• Press BACKLIGHT to turn the display backlight and test
lead LED on. Press again to turn the display backlight and
test lead light off. It will automatically turn off after 1 minute.
• Auto power off is set to approximately 10 minutes to
preserve battery life if no button is pressed. The meter can
be activated again by pressing any button, following which it
returns to the previously active function.
• Before making any measurement the meter test input
terminal has to be exposed by pressing down and twisting
the red cover counter clockwise.
• Connect the supplied COM test lead to the connector on the
end of the meter.
• When measurement is complete restore the cover by
twisting it clockwise to protect the terminal.
• Set the mode to V~. Auto measurement mode is the default.
• This mode can measure AC voltage and DC voltage.
• Connect the test leads with the object being measured. The
measured value shows on the display.
• When voltage measurement has been completed,
disconnect the connection from the test leads and the circuit
under test, and remove the testing leads from the input
terminal of the meter.
Note: The threshold voltage of AC voltage is around 400mV.
• Set the mode to V~
• Press SELECT to set DC voltage measurement mode.
• Connect the test leads with the object being measured.
• The measured value shows on the display.
• When voltage measurement has been completed,
disconnect the connection between the testing leads and
the circuit under test, and remove the testing leads from the
input terminal of the meter.
Note: Do not input greater than 300V voltage although it is
possible to measure higher voltage, accuracy and safety are
reduced.
• Set the mode to V~
• Press SELECT to set AC voltage measurement mode.
• Connect the test leads with the object being measured.
• The measured value shows on the display.
• When voltage measurement has been completed,
disconnect the connection between the testing leads and
the circuit under test, and remove the testing leads from the
input terminal of the meter.
Note: Do not input greater than 300V voltage to avoid damage
to the meter. The threshold voltage of AC voltage is around
400mV.
• Set the mode to V~EF and remove the test lead from the
input terminal.
• Press SELECT to set EF measurement mode.
• Place the red tip probe toward the object being measured.
• The measured value shows on the display.
• Set the mode to Ω
• Auto measurement mode is the default, in which the meter
can measure Resistance, Diode, Continuity and Capacitance
automatically.
• For best accuracy, remove the component being measured
from the circuit before measurement.
• Connect the test leads with the object being measured.
• The measured value shows on the display.
• When measurement has been completed, disconnect the
connection between the testing leads and the item under
test, and remove the testing leads from the input terminal of
the meter.
Note: To avoid damages to the Meter or to the devices under
test, disconnect circuit power and discharge all the high-voltage
capacitors before measurement.
Note: Under auto measurement mode, when measuring
Resistance of <15Ω or >10MΩ or Capacitance of <400pF or
>1mF the Meter may display an inaccurate value. Use the
dedicated mode to avoid this for these values.

3
RESISTANCE MEASUREMENT
CONTINUITY MEASUREMENT
DIODE TEST
CAPACITANCE MEASUREMENT
SPECIFICATION
• Press SELECT to select Ω measurement mode.
• Connect the test leads with the object being measured.
• The measured value shows on the display.
• When resistance measurement has been completed,
disconnect the testing leads from the circuit under test, and
remove the testing leads from the input terminals of the
Meter.
• Press SELECT to select
measurement mode.
• Connect the test leads with the object being measured.
• The measured value shows on the display.
• The buzzer sounds continuously if the resistance of a
circuit under test is <30Ω, indicating the circuit has a good
connection.
• When continuity measurement has been completed,
disconnect the testing leads from the circuit under test, and
remove the testing leads from the input terminals of the
Meter.
• Press SELECT to select measurement mode.
• Connect the test leads with the object being measured.
• The measured value shows on the display.
• For best accuracy, remove the component being measured
from the circuit before measurement.
• When diode measurement has been completed, disconnect
the testing leads from the item under test, and remove the
testing leads from the input terminals of the Meter.
Note: Open Circuit Voltage is around 3V, the display shows
approximate forward voltage drop.
A good diode should produce a forward voltage drop of 0.5V
to 0.8V.
• Press SELECT to select measurement mode.
• Connect the test leads with the object being measured.
• The measured value shows on the display.
• For best accuracy, remove the component being measured
from the circuit before measurement.
• When capacitance measurement has been completed,
disconnect the testing leads from the item under test, and
remove the testing leads from the input terminals of the
Meter.
Warning: To avoid damage to the Meter or to the equipment
under test, disconnect circuit power and discharge all high
voltage capacitors before measuring capacitance. Use the DC
Voltage function to conrm that the capacitor is discharged.
Function Range/description
Operating Temperature 0o~40oC (32oF~104oF)
Storage Temperature -10oC~50oC (14oF~122oF)
Battery Type 3V Li-MnO2 Button cell
Altitude Operating: 2000m
Display/Update Max 3000/4 times per second
AC/DC Voltage 0 to 300V rms
Dimensions (H x W x L) 20.2 x 26.5 x 181.5mm
Weight 90g incl battery
Range Resolution Accuracy O/L Protection
3V AC 0.001V
± (1%+4) 300Vrms
30V AC 0.01V
300V AC 0.1V
Remarks: Input Impedance: 10MΩ Frequency Response: 40Hz~400Hz
3V DC 0.001V
± (1%+3) 300Vrms30V DC 0.01V
300V DC 0.1V
Remarks: Input Impedance: ≥10MΩ
300Ω 0.1Ω
± (1%+3) 300Vrms
3kΩ 1Ω
30kΩ 10Ω
300kΩ 100Ω
3MΩ 1kΩ
30MΩ 10kΩ ± (1.5%+5)
Remarks: Under auto scan mode the maximum range is 3MΩ
3nF 0.001nF
±(3%+5)
300Vrms
30nF 0.01nF
300nF 0 .1nF
3µF 1nF
30µF 10nF
300µF 100nF ±(5%+5)
3mF Reference
Remarks: Under auto scan mode the maximum range is 300µF.
There is a residual reading when the circuit is open. To measure a small
value of capacitance, subtract it from the reading to ensure accuracy.
Range Resolution Remark
0.1Ω
Open Circuit Voltage around -1.2V.
Buzzer sounds at resistance ≤10Ω. Buzzer
does not sound when resistance >70Ω.
Range Resolution Overload Protection
1mV 300V rms
Accuracy Specications: ±(a% reading+b digits) guaranteed for 1 year.
Operating temperature: 18oC~ 28oC. Relative humidity: <75%.
Range Remark
EF Test
220V/50Hz
<10 up to 50mm: buzzer sounds
>50mm: buzzer silent

4
CLEANING
BATTERY REPLACEMENT
• Clean the meter with a clean, soft cloth.
• Do not use any chemicals, abrasives or solvents that could
damage the meter.
• Clean the terminals using a cotton swab with mild detergent
or contact cleaner, as dirt or moisture in the terminals can
affect readings.
• Turn the Meter off, disconnect the connection between the
testing leads and the circuit under test and remove the test
leads from the input terminals of the Meter.
• Remove the screw in the rear casing and remove the battery
cover from the casing.
• Remove the battery from the battery compartment.
• Replace the battery with a new 3V battery.
• Ret the battery cover and reinstall the screw.
INFORMATION ON WASTE DISPOSAL FOR CONSUMERS OF
ELECTRICAL & ELECTRONIC EQUIPMENT
These symbols indicate that separate collection of Waste Electrical and Electronic
Equipment (WEEE) or waste batteries is required. Do not dispose of these
items with general household waste. Separate for the treatment, recovery and
recycling of the materials used. Waste batteries can be returned to any waste
battery recycling point which are provided by most battery retailers. Contact your
local authority for details of the battery and WEEE recycling schemes available
in your area. Made in China. LS12 2QQ
Man Rev 1.0
Table of contents
Other multicomp pro Multimeter manuals

multicomp pro
multicomp pro MP730657 User manual

multicomp pro
multicomp pro MP730026 User manual

multicomp pro
multicomp pro MP730425 User manual

multicomp pro
multicomp pro MP730651 User manual

multicomp pro
multicomp pro MP730424 User manual

multicomp pro
multicomp pro MP730119 User manual

multicomp pro
multicomp pro MP730835 User manual

multicomp pro
multicomp pro MP730889 User manual

multicomp pro
multicomp pro MP731065 User manual

multicomp pro
multicomp pro MP730988 User manual