MVP ICON 78088-00 User manual

pH/TemperatureProbe 
_________________________________________________________________________________________________
Part No. 78088-00
55226.R004.042016
1 of 3
GENERALDESCRIPTION
TheMVPICON®pH/TemperatureProbeissemi‐conductor
designedforgeneralpurposepHmeasurementofliquidsor
semi‐solidsrangingfrom0–14pH.Temperature
measurements,rangingfrom0–80°C,maybetaken
simultaneously.Theprobefeaturesasilver/silverchloride/
potassiumchloridereferencesystemandathermistorwith
automatictemperaturecompensation.Solid‐state
componentrybuiltintotheMVPICONinstrumentcontrols
thissensingelement.TheMVPICONpH/Temperature
ProbeisdesignedforusewiththeMVPICONinstrument
only.Anyothercombinationmightcauselossof
performanceorirreversibledamagetobothprobeand
instrument.
PartNo:78088‐00
KITCOMPONENTS
EachkitcontainsoneMVPICONpH/TemperatureProbe
andasoftbristlebrushforcleaningthesensoronthetip.
TESTPROCEDURE
A. UsingtheMVPICONpH/TemperatureProbe
Note:TheMVPICONpH/TemperatureProbemustbe
calibratedforpHpriortofirstuse.SeeB.Calibrationfor
pHfordetails.
(a) Removeandsavetheprotectioncaportubefromthe
probetippriortouse.
Note:Referencegelmaybeobservedasaviscous
materialonthetipoftheprobe.Somegelseepagefrom
anewprobeisnormalandwillnotaffectthelongevity
orperformanceoftheprobe.
(b) Priortoeachdayofuse,scrubtheprobetipwithasoft
bristledbrush(included)anddeionizedwatertoremove
possibleresidues.Useamilddetergentifrequired.
(c) Iftheprobehasnotbeenusedfor2ormoredays,clean
asabove,soaktipinsaturatedKCI(potassiumchloride)
solution(seeOPERATINGTIPS)for10‐15min,then
soakinpH7.00orpH4.00bufferforatleast10min.
(d) EnsuretheprobeissecurelyconnectedtotheMVPICON
instrumentbyinsertingtheprobe’sconnectorwiththe
arrowsfacingtowardthebackoftheinstrumentandthe
whitecircularstickerfacingthefront.Insertprobeinto
thesample.FollowinstructionsontheMVPICONto
completereading.BothpHandtemperaturereadings
willbedisplayedontheinstrumentscreen.
Note:Todisconnecttheprobe,squeezebothsidesof
theconnectorwhileunpluggingfromtheMVPICON.
(e) WhenusingtheMVPICONpH/Temperatureprobewith
semi‐solids,insertprobesothatbothsensorandporous
whiteplugonoppositesideofprobetiparecontacting
orimmersedinthesample.Rotatelefttorightseveral
timesandtilttoensureconstantcontactwiththe
sample.
(f) Toensurecorrectmeasurementvalues,samplesor
buffersshouldbemixedwell.Thismaybedoneusinga
magneticstirrerorbygentlystirringtheprobeinthe
sampleforatleast5sec.Youmaycontinuetostirwhile
thereadingistakingplace.
(g) Theprobeshouldberinsedthoroughlywithdeionized
waterandblotteddrybetweensamples.
B. CalibrationforpH
TheMPVICONpH/TemperatureProbefeaturesa3‐point
calibrationusingpH7.00,thenpH10.00,andfinallypH4.00
buffersolutions.
Note:Useaborate‐basedpH10.00buffer(notacarbonate
basedbuffer)toensurearobustcalibrationslopeforpH
valuesabove7.00.
Note:InordertoperformaccuratepHmeasurements,the
MVPICONpH/TemperatureProbemustbecalibrateddaily
priortouse,orwhenusingadifferentMVPICONpH/
TemperatureProbe.
(a) RinsetheMVPICONpH/TemperatureProbewith
deionizedwater.SoakprobeinfreshpH7.00bufferfor
atleast2minpriortocalibration.
(b) FromtheMVPICON,enterpHmode,pressDevice,then
pressCalibration.
(c) InserttheprobeintopH7.00buffertobegin.Repeat
withfreshpH10.00andpH4.00bufferswhenprompted
ontheMVPICONscreen.Ensurebuffersarewellmixed.
(d) Rinseprobewithdeionizedwaterandblotdrybetween
buffers.
C. CalibrationforTemperature
TheMVPICONpH/TemperatureProbeusesafactory‐
calibratedthermistor.Itonlyneedstobere‐calibratedifit
hasbeenusedtomeasuretemperatures20°Caboveor
belowroomtemperature.Theprobefeaturesasingle‐point
calibrationthatcanbedoneatanyknowntemperature
withintheoperatingrange.
Note:Thecalibrationismostaccuratewhenusingaknown
temperatureclosetothesampletemperaturerange.
(a) FromMVPICON,enterTEMPERATUREmode,press
Devicemenu,thenpressCalibration.
Part No. 78088-00
55226.R004.042016
2 of 3

(b) Insertprobeintoaliquidofknowntemperature,enter
valueintotheReferenceValuefield,thenpress
Calibrate.
(c) ThedateoflastcalibrationwillappearontheMVPICON
Homescreen.
STORAGE
Cleanprobethoroughlywithwater,asoftbristledbrush
(included),andamilddetergentsolution.Rinsewith
deionizedwater.Leave1‐2dropsofdeionizedwaterinthe
blackprotectivecaporclearprotectivetube,thencover
probetiptopreventdehydrationofthereferenceelectrode
andallowforfasterstart‐uptimes.
WARNING:DONOTIMMERSEPROBEINANYSOLUTIONS
WHENSTORING!
Forolderprobesthatusetheprotectivetube,slidethescrew
capontotheprobebodyfollowedbythesealingO‐ring.
Insertprobetipintoprotectivetube,thenslidescrewcap
andO‐ringdownontotherimoftheprotectivetube.Slightly
pressthecapdownandonlythentightenthescrewcap.
Storeprobeinoriginalpackagingwhennotinuse.
PRECAUTIONS
TheMVPICONpH/TemperatureProbeisruggedand
durableandrequireslittlemaintenance.Toensure
lastingperformance,readandfollowalloperating
guidelines.
AvoidprolongedexposureinsamplescontainingTrisor
proteins.Trisbuffersandsamplescontainingproteins
shouldbereadquicklyandtheprobeshouldberinsed
thoroughlywithdeionizedwaterbetweensamples.
Whentestingiscomplete,cleanwithwateranda
laboratorydetergentandrinsewithdeionizedwater.
Avoidprolongedexposureinsamplesateitherendof
thepHrange.Whenunavoidable,rinsewithample
waterbetweensamples.Rinsewithneutralizingagents
anddeionizedwateraftercompletionandpriorto
storage.
Avoidprolongedexposuretoextremetemperatures.
Above50°C,limitprobeexposuretotheminimumtime
neededtoobtainastablereading.Donotusethe
probeoutsidethespecifiedtemperaturerange(see
SPECIFICATIONS)asthismightresultinprobe
performancefailureorirreversibledamage.
Samplesmustbeaqueousliquidsorsemi‐solidsand
compatiblewiththeprobe’swettedmaterials.
TheusefullifeoftheMVPICONpH/Temperature
Probeisdeterminedbythefrequencyandtypeof
samplestestedandthelevelofcleaningbetweenuses.
Ifinformationisrequiredregardingthechemical
resistanceoftheprobe,contactBioControl.
OPERATINGTIPS
TopreparesaturatedKCI(potassiumchloride)solution,add
KCIgranulestodistilledwateruntilnomorewilldissolve.
Adding38gofKCIto100mLwaterissufficient.Letsolution
standforatleast2hrsanddecanttheclearsolution.Useas
describedabove.
Toensurecorrectmeasurementvalues,mixsamplesor
bufferswell.
Proteins,fats,andoilsmayberemovedbygentlyscrubbing
inasolutionofTerg‐A‐Zyme(AlconoxCompany),apepsin
solution,orsimilarproduct.Afterwards,rinsethoroughly
withdeionizedwater.Cleaningagentsareavailablefrom
yourlaboratorysupplyvendor.
Whentestingindirectsunlightoronabrightreflecting
surface,usebrown,opaque,orshielded‐samplecontainers.
Verybrightlightmayaffectsensorperformance.
Forbestresults,usebuffersthathavealreadybeen
reconstituted(notpowderedtabletsorpackets).Alsouse
bufferswithspecifiedvaluesof4.00,7.00,and10.00(each
±0.02)at25°C.Useaborate‐basedpH10.00buffer.
Buffer‐handling:pH7.00buffers(phosphate‐based)andpH
4.00buffers(biphthalate‐based)arelesssusceptibleto
carbondioxidecontamination(fromexposuretoair)thanpH
10.00buffers(borateorcarbonate‐based).Keepbuffers
sealedwhennotinuse.Inabilitytocalibrateprobeusually
indicatesafailingprobeoracontaminatedbuffer.Try
calibratingwithfreshbuffersfirst.
Bestresultsareobtainedbystirringtheprobeinthesolution
for1minbeforecalibratingortakingareading,then
continuingtostirwhilereadingisbeingtaken.
SPECIFICATIONS
Sensor:
pH: Semi‐conductorIonSensitiveFieldEffect
Transistor(ISFET)sensorwithpatentedESD
protectioncircuit
Temperature:Thermistor
Operating
range:
pH:
Temperature:
0.00
–
14.00
0.0–80.0°C
ResolutionpH:
Temperature:
0.01
0.1°C
AccuracypH:
Temperature:
±0.02
±0.5°C
Reference
compartmentSaturatedKCIgel,non‐refillable
Wetted
materials:
Barrelandtip:
Packagingmaterial:
Referenceliquid:
Diaphragm:
PEI
Epoxy‐resin
GelledKCI
PorousPTFE
Part No. 78088-00
55226.R004.042016
3 of 3

TROUBLESHOOTING
Ifanyofthefollowingareobserved—driftand/orinstability
ofthereading,slowcalibration,probewillnotcalibrate,or
pHvaluedoesn’tchangeasexpectedwhenchanging
sample—cleanprobe(seeSTORAGE),soaktipinhot(40°C)
tapwaterfor5–10min,blotdry,thensoaktipina
saturatedKCl(potassiumchloride)solution(seeOPERATING
TIPS)atroomtemperaturefor10‐15min.
ContactBioControlTechnicalSupportat800.245.0113for
moreinformation.

WARRANTY
BioControlSystems,Inc.(BCS)warrantsthisproducttobefreefromdefectsin
materialsandworkmanship,whenstoredunderlabeledconditionsandused
asintended,for6monthsfromdateofpurchase.BCSagreesduringthe
applicablewarrantyperiodtoreplacealldefectiveproductsafterreturnto
BCS.BCSshallnothaveobligationunderthisLimitedWarrantytomake
replacementswhichresult,inwholeorinpart,fromnegligenceoftheBuyer,
orfromimproperuseoftheproducts,oruseoftheproductinamannerfor
whichitwasnotindicated.BuyershallnotifyBCSofanyproductswhichit
believestobedefectiveduringthewarrantyperiod.AtBCSoption,such
productsshallbereturnedtoBCS,transportationandinsuranceprepaid.BCS
shallreplaceanysuchproductfoundtobedefectiveatnocharge.ShouldBCS
examinationnotdiscloseanydefectcoveredbytheforegoingwarranty,BCS
shallsoadviseBuyersanddisposeoftheproductinaccordancewithBuyer’s
instructions.

©BioControlSystems,Inc.2016.PrintedinUSA.
LIGHTNINGMVPandMVPICONaretrademarksofBioControlSystems,Inc.,
USpatentsNo.7,399,984;7,544,961;7,030,403;6924498.Australianpatents
2002357107;2003231770.NewZealandpatents533051,535819.Japanese
patents4,467,304;4,431,034.Mexicopatents259897;246269;272356.All
rightsreserved.

Popular Measuring Instrument manuals by other brands

Simex STI-638 user manual

Simex

Simex STI-638 user manual

Emerson Rosemount 9295 quick start guide

Emerson

Emerson Rosemount 9295 quick start guide

Milwaukee MW802 user manual

Milwaukee

Milwaukee MW802 user manual

Test Equipment Depot AMPROBE WT-80 user manual

Test Equipment Depot

Test Equipment Depot AMPROBE WT-80 user manual

Chauvin Arnoux DECADES BOXES user manual

Chauvin Arnoux

Chauvin Arnoux DECADES BOXES user manual

ELMACRON 1005RE instructions

ELMACRON

ELMACRON 1005RE instructions

Doyon WM35 manual

Doyon

Doyon WM35 manual

ABB NETA-21 Installation and startup guide

ABB

ABB NETA-21 Installation and startup guide

Prexiso PR500HV Safety instructions and the User Manual

Prexiso

Prexiso PR500HV Safety instructions and the User Manual

FRER C15WPM100 operating manual

FRER

FRER C15WPM100 operating manual

Greenlee AM-600T instruction manual

Greenlee

Greenlee AM-600T instruction manual

Mitsubishi Electric EcoMonitorPro EMU2-RD3-F instruction manual

Mitsubishi Electric

Mitsubishi Electric EcoMonitorPro EMU2-RD3-F instruction manual

Exotek Instruments BDD mini user manual

Exotek Instruments

Exotek Instruments BDD mini user manual

T3 Motion MotionTrak 330 4CH user guide

T3 Motion

T3 Motion MotionTrak 330 4CH user guide

ANDERSON-NEGELE IZMSA operating manual

ANDERSON-NEGELE

ANDERSON-NEGELE IZMSA operating manual

YOKOGAWA WT1800 user manual

YOKOGAWA

YOKOGAWA WT1800 user manual

Chelonia F-POD user guide

Chelonia

Chelonia F-POD user guide

wilks InfraCal 2 ATR-SP quick start

wilks

wilks InfraCal 2 ATR-SP quick start

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.