natus Olympic Warmette 48 User manual

602027A
Instruction Manual
Olympic Warmette®
Models 48/49 and 55/56
CAUTION
Read and be familiar with this manual before installing,
operating, or servicing this device. To ensure operator,
technician, and patient safety, use only as specified in
this manual.


Product Manufacturer:
Natus Medical Incorporated
5900 First Avenue South
Seattle, WA 98108 USA
Global sales & support: 1-800-303-0306
Phone: +1-206-767-3500
Fax: +1-206-762-4200
Web: www.natus.com
The information in this manual is subject to change without notice.
No part of this manual may be photocopied, reproduced, translated, or reduced to
any electronic medium without the express written permission of Natus Medical
Incorporated.
All tradenames and trademarks mentioned herein are property of their respective
owners.
Copyright ©2009 Natus Medical Incorporated. All rights reserved.


Olympic Warmette Models 48/49 and 55/56 Instruction Manual i
Contents
Section 1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
About this Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Description. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Section 2 Installation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Assembling the Warmette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
Inspecting the Heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Testing the Warmette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Electrical Requirements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Section 3 Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Operating the Warmette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Controls and Signals. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Section 4 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Removing the Trays. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Quarterly Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Cleaning the Lint Filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Annual Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Cleaning the Inside of the Upper Cabinet . . . . . . . . . . . . . . . . . . . . . 4-3
Cleaning the Blower Housing and Blower Wheel . . . . . . . . . . . . . . . 4-4
Cleaning the Heater Plenum and Box . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Testing the Thermal Cut-out Function . . . . . . . . . . . . . . . . . . . . . . . . 4-4
Section 5 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Technical Support . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Section 6 Service. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Servicing the Warmette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Repair Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Repair Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Temperature Measurement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Replacing the Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Removing the Access Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Replacing the Mains Power Switch . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Replacing the Heater Indicator Lamp . . . . . . . . . . . . . . . . . . . . . . . . 5-6
Replacing the Thermostat. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-6

ii Olympic Warmette Models 48/49 and 55/56 Instruction Manual
Replacing the Thermostat Label . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Replacing the Blower . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Replacing the Thermal Cut-out. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Replacing the Heater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-12
Section 7 Replacement Parts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Section 8 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-1
Appendix A Drawings and Schematics . . . . . . . . . . . . . . . . . . . . . . . . A-1

Olympic Warmette Models 48/49 and 55/56 Instruction Manual iii
List of Figures
Model 48/49 and Model 55/56 dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Model 48/49 anchor holes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Model 48/49 anchor hole placement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Model 55/56 anchor holes. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Model 55/56 anchor hole placement. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Screw locations inside Warmette door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Removing the front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Blower/Filter assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Control panel. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Removing the tray for cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Location of the lint filters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Blower assembly. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Access panel assembly—exploded view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Access panel assembly —side view . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Screw locations inside Warmette door . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Removing the front panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Blower/Filter assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Model 48/49/55/56, Olympic Warmette Wiring Diagram, Drawing No. 700192 Rev. C . .
A-1
Model 48/49/55/56, Olympic Warmette Schematic, Drawing No. 700193 . . . . . . . . . . A-2
Model 48/49/55/56, Olympic Warmette Drawing No. 700194 . . . . . . . . . . . . . . . . . . . A-3
Model 48/49/55/56 Olympic Warmette Drawing No. 700195 . . . . . . . . . . . . . . . . . . . . A-4

iv Olympic Warmette Models 48/49 and 55/56 Instruction Manual

Olympic Warmette Models 48/49 and 55/56 Instruction Manual v
List of Tables
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Product symbology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Troubleshooting chart . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
Replacement parts list. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1

vi Olympic Warmette Models 48/49 and 55/56 Instruction Manual

Olympic Warmette Models 48/49 and 55/56 Instruction Manual 1-1
Section
1
Overview
This section provides an overview of the Olympic Warmette®, including the
intended use and a description of the controls and symbols.
About this Manual
This manual provides the necessary information to install, maintain, and service the
Olympic Warmette®. The operating instructions in this manual are intended for use
under the direct supervision of a licensed medical practitioner. The installation and
service instructions in this manual are intended for use by qualified service
technicians.
Conventions
Table 1.1 describes the conventions used in this manual.
Table 1.1
Conventions
Convention Description
NOTE
#
Notes provide additional information to clarify a point in the text.
CAUTION
Cautions indicate situations that, if not avoided, could result in minor
to moderate injury to the patient or operator, or damage to the equipment.
WARNING
Warnings indicate situations that, if not avoided, could result in
serious injury or death to the patient or operator.
BUTTON
This character style represents buttons and controls that the user
can touch or press.

1-2 Olympic Warmette Models 48/49 and 55/56 Instruction Manual
Intended Use
Symbols
Table 1.2 describes the symbols used on this device, packaging, and its accessories.
Symbol Description Symbol Description
Ground Line (hot conductor)
Read instructions Neutral (neutral conductor)
Mains power ON Alternating current (AC)
voltage
Mains power OFF Hot surface
Intended Use
The Warmettes are intended for warming fluids, blankets or other linens in a
hospital setting.
Table 1.2
Product symbology

Olympic Warmette Models 48/49 and 55/56 Instruction Manual 1-3
Description
Description
The Olympic Warmette is a large-capacity, free-standing warming cabinet. It is
available in two sizes with the option of convenient see-through doors on the front
and back for either size.
The Warmette has a fully-enclosed, recirculating air system that warms blankets
and bottles quickly and uniformly. Forced-air distributes heat evenly throughout
the entire cabinet, so supplies are heated rapidly no matter where they are placed.
Mesh shelves allow air to move under and around supplies.
Figure 1.1
Model 48/49 and Model 55/56 dimensions
21 in.
86 in.
86 in.
26 in. 51.8 in.
Leveling feet
Allow approx. 1/2 in.
vertical adjustment
Temperature
controls
12-ft. Power cord
located on back needs
2-in. clearance (one for
each cabinet)
21 in.
Model 48/49 Model 55/56

1-4 Olympic Warmette Models 48/49 and 55/56 Instruction Manual
Description

Olympic Warmette Models 48/49 and 55/56 Instruction Manual 2-1
Section
2
Installation
This section describes how to install the Olympic Warmette Models 48/49
and 55/56.
Assembling the Warmette
When you first receive your Warmette from the factory, minimal assembly
is required.
Required Items:
Level
Soap-and-water or liquid disinfectant
Towel
To assemble the Warmette:
1After uncrating, install the leveling feet and stand the Warmette upright.
2Unpack the bottom tray(s) and put in place in the warming cabinet (see
Figure 1.1 on page 1-3).
3Clean all metal surfaces with a towel moistened with soap-and-water or liquid
disinfectant. Plexiglas®doors must be cleaned with soap and water.
CAUTION
Read and be familiar with this manual before installing, operating, or servicing this
device. To ensure operator, technician, and patient safety, use only as specified in
this manual.
Only qualified technicians should assemble and install this product.
NOTE
#
The Warmette Models 48 and 55 have clear doors on the front (see Figure 1.1 on page 1-3).
The Models 49 and 56 have an additional set of doors on the back side. The Models 48/49
and 55/56 are assembled in the same manner.
CAUTION
The bottom tray(s) must be in place for the Warmette to operate properly.
NOTE
#
Never use alcohol to clean the doors as alcohol will damage the Plexiglas®.

2-2 Olympic Warmette Models 48/49 and 55/56 Instruction Manual
Assembling the Warmette
4After the Warmette is in place, adjust the leveling feet.
5Use the level to ensure the Warmette is evenly balanced.
To install floor anchors:
1Access the anchor holes by removing the bottom tray (for Models 48/49, see
Figure 2.1 and Figure 2.2; for Models 55/56, see Figure 2.3 and Figure 2.4).
2Pivot the bottom tray up and out to remove.
3Ensure that the Warmette is level and that the jam nuts for the leveling feet are
locked in place before using the Warmette.
Figure 2.1
Model 48/49 anchor holes
Figure 2.2
Model 48/49 anchor hole placement
CAUTION
The Warmette must be bolted to the floor.
Leveling feet
Anchor holes
Center line
9.92 in.
6 in.
Anchor holes
1.45 in.
17.5 in.

Olympic Warmette Models 48/49 and 55/56 Instruction Manual 2-3
Inspecting the Heater
Figure 2.3
Model 55/56 anchor holes
Figure 2.4
Model 55/56 anchor hole placement
Inspecting the Heater
If a Warmette fails to heat, a heater might have loosened during shipping. The
Warmette has three 500W heaters per Plenum Assembly.
Required Items:
Heater (part no. 401572)
1/4-in. nut driver
To check the heater:
1Press the power switch to the
OFF
(O) position, and unplug the power cord.
2Remove the uper white door seal, and then loosen—but do not remove—the
screws inside the door, behind the door stop (see Figure 2.5).
Leveling feet
Anchor holes
9.92 in.
17.5 in.
1.45 in.
6 in.
6 in.
Center
line
25.97 in.

2-4 Olympic Warmette Models 48/49 and 55/56 Instruction Manual
Inspecting the Heater
Figure 2.5
Screw locations inside Warmette door
3Pull out and remove the front panel (see Figure 2.6).
Figure 2.6
Removing the front panel
4Unhook and remove the plenum hold-down spring and bridge (see Figure 2.7).
Figure 2.7
Blower/Filter assembly
5Separate the plenum from the casing.
6Lift the plenum, and then unscrew the heater from its socket. Ensure that the
socket cap is tight, and then screw the heater firmly back into place.
7Reinstall the plenum. Ensure that the plenum and case are seated firmly. Install
the plenum hold-down spring and bridge.
8Reinstall the front panel and tighten the screws.
9Test the Warmette for proper operation (see Testing the Warmette).
Door stop
Screws (4)
Access panel
NOTE
#
Push together the hold-down bridge and the spring, and then separate them.
Plenum
Plenum hold-down
spring
FRONT
Blower
Plenum hold-down
bridge
BACK
Terminal strip

Olympic Warmette Models 48/49 and 55/56 Instruction Manual 2-5
Testing the Warmette
Testing the Warmette
Test the Warmette after installing it or performing any maintenance or service
procedure to ensure that all systems are functioning properly.
To prepare the Warmette for testing:
1Plug in the power cord, then press the power switch to the
ON
(|) position.
2Set the temperature to 120°F.
3Allow the Warmette to warm up for 5 minutes.
To test the Warmette:
Ensure that the
HEATER ON
lamp is illuminated.
Ensure that the air blowing from the top of the Warmette is noticeably warm.
Listen for the blower to make sure it is running.
Check that the
HEATER ON
lamp turns off within 60 minutes. It will then cycle
on and off to maintain the set temperature.
Electrical Requirements
The Warmette product line has several voltage/frequency options:
120V~, 60 Hz, 14A
220V~, 50 Hz, 8A
220V~, 60 Hz, 8A
CAUTION
This product must be connected directly to a hospital-grade electrical outlet(s) rated for
minimum 20-ampere (A) service. Each heating cabinet is electrically separate. Two
separate electrical outlets are required for the Model 55/56.

2-6 Olympic Warmette Models 48/49 and 55/56 Instruction Manual
Electrical Requirements
This manual suits for next models
3
Table of contents
Other natus Laboratory Equipment manuals
Popular Laboratory Equipment manuals by other brands

Belden
Belden HIRSCHMANN RPI-P1-4PoE installation manual

Koehler
Koehler K1223 Series Operation and instruction manual

Globe Scientific
Globe Scientific GCM-12 quick start guide

Getinge
Getinge 86 SERIES Technical manual

CORNING
CORNING Everon 6000 user manual

Biocomp
Biocomp GRADIENT MASTER 108 operating manual