NeoMedLight BiliCocoon User manual

Instructions for use: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 1
EN/FR
BiliCocoon™ Fixation
System

Instructions for use: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 2
EN
Table of contents
1. Product description and application......................................................................................................................5
A. Description........................................................................................................................................................5
B. Intended use of the Device...............................................................................................................................5
C. Additional information .......................................................................................................................................5
2. Safety information.................................................................................................................................................5
3. Technical specifications........................................................................................................................................7
4. Installation and method for use.............................................................................................................................9
A. Preliminary precautions........................................................................................................................................9
B. Installing the fixation system...........................................................................................................................10
5. Maintenance and after-care................................................................................................................................13
5.A List of authorized cleaning agents and disinfectants..................................................................................13
6. End of service life................................................................................................................................................14
7. Product reference ...............................................................................................................................................14
8. Applicable standards and directives...................................................................................................................14
9. Warranty .............................................................................................................................................................14

Instructions for use: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 3
FR
Table des matières
1. Description du produit et application 16
A. Description 16
B. Usage prévu du Dispositif 16
C. Informations supplémentaires 16
2. Informations de sécurité 17
3. Caractéristiques techniques 18
4. Installation et mode d’emploi 20
A. Précautions préalables 20
B. Mise en place du système de fixation 21
5. Entretien et Maintenance 24
5.A Liste des agents nettoyants et désinfectants autorisés 24
6. Fin de vie 25
7. Référence Produits 25
8. Normes et directives appliquées 25
9. Garantie 25

User manual:
BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 4
EN
NeoMedLight
88-90 rue Frédéric Faÿs 69100 VILLEURBANNE, FRANCE
April 2017
User liability
This user manual describes how the single-use BiliCocoon™ Fixation System – referred to simply as the “fixation
system” (mounting system) – should be installed and used. The fixation system should be used with the BiliCocoon™
1
phototherapy system: the system comprises the fixation system and the phototherapy system, and is also referred
to as the “Device” in this document.
The fixation system should only be used appropriately by a trained user and should not be used if it is defective,
damaged or if there are any missing parts. In this case, NeoMedLight recommends you dispose of the device and
replace it.
The user will be considered solely responsible for the risks incurred by the patient, caregivers, third parties or goods;
or if the results differ from those expected due to improper or unsuitable use, or for any improper modifications made
to the fixation system.
Any serious incident which occurs in relation to the device should be reported to the manufacturer and the local
competent authority of the state in which the user and/or patient is established. You can contact NeoMedLight at the
following address: quality@neomedlight.com.
According to U.S. Federal Law, the Product may only be sold to authorised professionals and upon their request.
NeoMedLight declares that this product complies with Directive 93/42/EEC.
Year the CE marking was added: 2016.
1
Also referred to as the BiliCocoon™ System in this document.

User manual:
BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 5
EN
1. PRODUCT DESCRIPTION AND APPLICATION
A. Description
The fixation system –or BiliCocoon™ Fixation System – is a fixation system designed for exclusive use with the
BiliCocoon™ System, a medical device offering a treatment solution for unconjugated hyperbilirubinemia in new-born
babies and which uses phototherapy.
The BiliCocoon™ System consists of a light pad connected to a Light Box emitting blue light in the absorption spectrum
of bilirubin, thereby enabling its concentration to be reduced. For more information, see the BiliCocoon™ System user
manual.
The BiliCocoonTM Fixation System should only be operated by caregivers or a person who has been trained by a
qualified trainer.
For more information, see “3. Technical specifications”.
B. Intended use of the Device
The BiliCocoonTM Phototherapy System is intended for the treatment of unconjugated hyperbilirubinemia in the
population of neonates and infants under 3 months old and weighing less than 10kg. It can be used in the clinical
setting or in the home.
C. Additional information
-Only for the treatment of unconjugated hyperbilirubinemia, which is responsible for the majority of cases of
neonatal jaundice.
-Only with medical prescription
-Only for the duration indicated in the treatment protocol
-Target patient population: neonates and infants
2
.
-Over the patient's entire body except for genitals and eyes –eye protection must be worn.
-To be used with a single patient
-To be used in hospital or at home: adapted to different environments for use (in an incubator, on a table, in a
cradle, in the arms of a parent or caregiver).
-Not to be used on damaged skin
2. SAFETY INFORMATION
2
Under 3 months old and weighing less than 10 kg –definition according to EN 60601-2-50. The term “infant” used in this
document is used in compliance with this definition according to EN 60601-2-50.

User manual:
BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 6
EN
Types of safety advice
Warning
Signals danger or risk of injury for the child or the operator
Important
States the necessary actions for obtaining good clinical results
Warning
Prerequisites
The Device must only be used by qualified medical staff who are fully aware of the risks and
advantages of this type of treatment and this type of apparatus, or by individuals who have been
trained by medical staff.
Use at home: Before using the Device for the first time, it is vital that home users be trained in
how to install and use the Device, and be made aware of the risks associated with using it. It is of
the utmost importance that they follow the recommendations and instructions provided in the
instructions (system and single use) and by healthcare professionals.
The device must not be used if contraindicated, or for medical indications or parts of the body for
which it is not recommended or authorised. This could either render the system completely
ineffective (no treatment benefits) or could put the patient's health at risk (adverse effects).
Warning
Keep out of reach of children.
Warning
Variation in ambient parameters
Variation in ambient parameters (room temperature and humidity, exposure to sunlight,
surrounding equipment) may have a negative effect on the child’s health, especially their
temperature or fluid balance.
If the Device is used other than in accordance with instructions, it may malfunction.
Warning
Inflammability
Keep away from flames or sources of ignition.
For more information on the risks associated with phototherapy, see the BiliCocoon™ System user manual.
Warning
Label integrity and readability
If labels are damaged, you can find information concerning the safety of patients, users, third-
parties or property in this document. Contact the supplier if information
concerning the identification of the product or of associated products is not visible anymore.

User manual:
BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 7
EN
Symbol
Meaning
Symbol
Meaning
Manufacturer
CE marking declaration
according to Directive
93/42/EEC
See instructions for use
Date of manufacture
Catalogue reference
Serial number
Keep dry
Storage and handling
temperature range
Storage and handling humidity
range
Fragile, to be handled with care
Abbreviations and units of measure
Abbreviations
Meanings and units of measure
°C
Degrees Celsius
(Unit of temperature)
kg
Kilogram
(Unit of mass)
cm
Centimetre
(Unit of length)
%RH
Humidity
(Percentage of humidity in the air)
3. TECHNICAL SPECIFICATIONS
The fixation system –or BiliCocoon™ Fixation System – is a fixation system designed for exclusive use with the
BiliCocoon™ System, a medical device offering a treatment solution for unconjugated hyperbilirubinemia neonates
and infants and which uses phototherapy.
The BiliCocoon™ System consists of a light pad connected to a Light Box emitting blue light in the absorption spectrum
of bilirubin, thereby enabling its concentration to be reduced. For more information, see the BiliCocoon™ System user
manual.
Parameter
Value
Unit
Dimensions
18 x 15 x 11
cm
Weight
0.6
kg
Colour
Grey
Min. and max. temperature for use
From 0°C to 40°C
°C
Min. and max. humidity for use
From 30% to 75%
%RH
Min. and max. temperature for
storage
From -20°C to 60°C
°C
Min. and max. humidity for storage
From 0% to 90%
%RH
0%
90%
-20 °C
+60 °C

User manual:
BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 8
EN
Warning
The disregard of the operating and storage conditions may lead to risks for the patient, the
users, third parties or goods.
Warning
Conditions for use and storage are different for the BiliCocoon™ System and the
BiliCocoon™ Fixation System. Please ensure that the product is stored and used in
accordance with the recommended conditions.
Warning
No not alter the BiliCocoon™ Fixation System. Do not use any accessories, spare parts and
equipment not described in the conditions for use. If the fixation system is modified or used
improperly, or used with other parts, patients, users, third parties and goods may be at risk.
BiliCocoon™ Fixation System
1: Holding claw
2: Fixation plate
3: Fixation hooks
4: Groove
5: Fixation screw
6: Tightening hinge
1
3
4
5
6
2

User manual:
BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 9
EN
4. INSTALLATION AND METHOD FOR USE
A. Preliminary precautions
Warning
Before positioning the BiliCocoon on the fixation system, check the Pad's irradiance (see the
BiliCocoon™ System user manual).
If the check is not carried out before each use, there is risk of under-exposure, and therefore
ineffectiveness of the treatment, or over-exposure.
Warning
Ensure that the fixation system is not chipped or damaged, and that there are no sharp edges
that could cause harm.
Warning
Ensure that the fixation system is clean and unmarked. If necessary, clean it with a suitable
cleaning product, such as water or the products specified in this manual.
Warning
Ensure that the holding claw is properly attached to the fixation bracket and that there are no
missing screws. Check that the mobile hinge can be tightened without any obstruction.
Warning
Check that the holding claw is not damaged nor chipped.
Warning
Also check that the two fixation hooks are in good condition and are not damaged nor
chipped.
Warning
The fixation system is designed to be attached to a stand (mounting pole or bed). Check that
this stand is solid, stable and sturdy, and that it is between 1 and 5 cm wide. Check the stand
(fixation pole or bed) is stable before and after mounting the system. Check the full
BiliCocoon™ System is stable before using the system to avoid the fall of the BiliCocoon™
System.
Warning
Check that the holding claw is properly screwed on and that it is flush with the stand.
Warning
Check that the fixation system is mounted at a suitable height and that it is positioned so it
can be easily reached. Check the Light Box™ ventilation holes are not obstructed.
Warning
Check that the Light Box™ is correctly positioned on the fixation system.

User manual:
BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page
10
EN
Warning
Check the Light Box™ power lead is not caught on anything.
Warning
Failure to follow the conditions of use, maintenance and storage may put patients, users,
third parties or goods at risk.
Warning
The fixation system –or BiliCocoon™ Fixation System – is designed for exclusive use with
the BiliCocoon™ System, a medical device offering a treatment solution for unconjugated
hyperbilirubinemia in new-born babies using phototherapy.
Warning
While BiliCocoon™ Fixation System is used to support BiliCocoon™ system, the support
where the fixation system is fixed should not be moved. Improper use may actually lead to
an ineffective (not beneficial) treatment or to risks for the patient's health (unwanted adverse
events) and the users.
B. Installing the fixation system

Instructions for use: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 11
EN
Warning
The BiliCocoon™ Light Box can only be installed on the BiliCocoon™ Fixation System once the pads have been removed from the Light Box. The
pads should only be connected or disconnected when the BiliCocoon™ Light Box is mounted on the BiliCocoon™ Fixation System. Users must
only disconnect or connect the pads while securely holding the Light Box tightly with their hand.
Presentation of the fixation system
Hanging the fixation system
The fixation system consists of a fixation plate (1) with two fixation hooks, as
well as a holding claw (2). Completely unscrew the screw so that the tightening
hinge can be fully opened. If the hinge cannot be fully opened, replace the
fixation system.
Place the tightening hinge on a stable stand (3), so that the hinge can be tightened on
to it. The stand must be stable, solid, vertical and must be between 1 and 5 cm thick.
Then, tighten the hinge by fully screwing the tightening screw in as far as it will go. The
fixation system must be flush with the stand and must not slide or move.
1
2
3

Instructions for use: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 12
EN
Hanging the BiliCocoon™ Light Box
Using the BiliCocoon™ Fixation System groove
Once the fixation system is properly attached to the stand, hang the
BiliCocoon™ Light Box (4) by resting the Light Box unit handle on the two
fixation hooks. Everything must be stable and well-centred.
The BiliCocoon™ Nest Pad or the BiliCocoon™ Bag Pad (5) can be placed in the groove
of the fixation system.
4
5

Instructions for use : BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 13
EN
*
5. MAINTENANCE AND AFTER-CARE
Warning
Please clean the BiliCocoon™ Fixation System after each use.
Never immerse in a bath and never spray a solution directly onto the BiliCocoon™ Fixation
System’s surface.
Do not autoclave or gas sterilise.
If you do not sufficiently clean the device and/or do not follow the cleaning instructions, you
risk contaminating or damaging it.
Maintenance should not be carried out on the BiliCocoonTM Fixation System when it is
positioned on the fixation system.
Remove the BiliCocoon ™ Fixation System from its stand before cleaning it. Use a soft cloth
and cleaning and disinfectant agent (see the list of cleaning and disinfectant agents used in
paragraph 5.A), then carefully rinse with a soft, wet cloth. Perform a visual inspection of the
BiliCocoon ™ Fixation System to assure that it is clean. Otherwise repeat the previous
instructions until there is no more soil observed. Only use the fixation system once it is
completely dry.
Maintenance must not be carried out when the system is connected to the power supply.
5.A List of authorized cleaning agents and disinfectants
Warning
The system can be cleaned without being damaged using the following products:
-An EPA-registered cleaning and disinfectant product, such as CaviWipes™, METREX
-A product containing didecyldimethylammonium chloride, such as Surfa’Safe,
Laboratoires ANIOS
Do not use products containing:
-Caustic or abrasive cleaning products
-Powerful cleaning solutions (acidic and alkaline)
-Iodine-based solutions
-Phenolic compound-based products, which can lead to increased bilirubin levels in new-
born babies

Instructions for use : BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 14
EN
*
6. END OF SERVICE LIFE
If the fixation system becomes damaged or worn, or no longer works, you should stop using it and replace it; it should
be disposed of in accordance with the law in force. The fixation system is made primarily of aluminium.
Warning
Do not use the BiliCocoon™ Fixation System if its integrity and/or its functions are
compromised by a defect or worn parts.
7. PRODUCT REFERENCE
BCC010 BiliCocoon™ Light Box
BCC101 BiliCocoon™ Nest System
BCC111 BiliCocoon™ Nest Pad
BCC12150 BiliCocoon™ Disposable Nest x50
BCC201 BiliCocoon™ Bag System
BCC211 BiliCocoon™ Bag Pad
BCC22150 BiliCocoon™ Disposable Bag x50
BCC301 BiliCocoon™ Fixation System
8. APPLICABLE STANDARDS AND DIRECTIVES
EN 60601-1
Directive 93/42/EEC
REACH 1907/2006
9. WARRANTY
The BiliCocoon™ Fixation System comes with a one-year contractual warranty.

Manuel d’utilisation
: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 15
FR
*
NeoMedLight
88-90 rue Frédéric Faÿs 69100 VILLEURBANNE, FRANCE
Avril 2017
Responsabilité de l’utilisateur
Ce manuel d‘utilisation décrit la manière dont doit être installé et utilisé l’usage unique BiliCocoon™ Fixation System
–nommé par simplicité « système de fixation ». Le système de fixation est à utiliser avec le système de photothérapie
BiliCocoon™
3
: l’ensemble formé du système de fixation et du système de photothérapie est aussi nommé
« Dispositif» dans ce document.
Le système de fixation ne doit être utilisé que par un utilisateur formé de façon opportune et ne doit pas être utilisé
s’il est défectueux, abîmé ou si des parties sont manquantes. Dans ce cas, NeoMedLight recommande de le mettre
au rebut et de le remplacer.
L’utilisateur sera considéré comme seul responsable des risques encourus par le patient, le personnel soignant, les
tiers ou les biens, ou si les performances étaient amenées à être différentes de celles attendues provoqués par un
emploi abusif, inadapté ou des modifications inadéquates du système de fixation.
Tout incident grave survenant en lien avec le dispositif doit faire l’objet d’une notification au fabricant ainsi qu’à
l’autorité compétente locale de l’Etat dans lequel l’utilisateur et/ou le patient est établi. Vous pouvez contacter
NeoMedLight à l’adresse suivante : quality@neomedlight.com.
Selon la loi fédérale des Etats-Unis la vente du Produit est permise exclusivement aux professionnels autorisés ou
suite à leur demande.
NeoMedLight déclare que ce produit satisfait aux exigences de la Directive 93/42/CE.
Année d’apposition du marquage CE : 2016.
3
Aussi nommé BiliCocoon™ System dans ce document

Manuel d’utilisation
: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 16
FR
*
1. DESCRIPTION DU PRODUIT ET APPLICATION
A. Description
Le système de fixation - ou BiliCocoon™ Fixation system - est un dispositif de fixation destiné à être utilisé comme
support exclusivement pour le BiliCocoon™ System, dispositif médical apportant une solution de traitement par
photothérapie à l’hyperbilirubinémie non-conjuguée du nouveau-né.
Le BiliCocoon™ System est composé d’un pad lumineux connecté à une Light Box émettant de la lumière bleue dans
le spectre d’absorption de la bilirubine, permettant ainsi de réduire sa concentration. Pour plus d’informations, se
référer au manuel d’utilisation du BiliCocoon™ System.
Le BiliCocoon™ Fixation system ne doit être manipulé que par le personnel soignant ou une personne qui a reçu une
formation par un formateur qualifié.
Pour plus d’informations, se référer au « 3. Caractéristiques techniques ».
B. Usage prévu du Dispositif
Le BiliCocoon™ Phototherapy System est indiqué pour le traitement de l’hyperbilirubinémie non-conjuguée
4
chez la
population des nouveau-nés à terme et prématurés –Agés de moins de trois mois et pesant moins de 10 kg –
définition selon la EN 60601-2-50. Il peut être utilisé en environnement hospitalier ou à domicile.
C. Informations supplémentaires
-Exclusivement dans le cadre du traitement de l’hyperbilirubinémie non-conjuguée, responsable de la majorité
des jaunisses néonatales
-Uniquement sur prescription médicale
-Uniquement pendant la durée prescrite dans le protocole thérapeutique
-Population Patients Cible : Nouveau-nés à terme et prématurés
5
.
-Sur tout le corps du patient à l’exception des parties génitales et des yeux – une protection oculaire doit être
utilisée.
-Utilisation sur un seul patient
-Usage hospitalier ou usage à domicile : adapté aux différents environnements d’utilisation (en incubateur, sur
table, en berceau, dans les bras d’un parent ou du personnel soignant).
-Utilisation sur une peau non lésée
4
Le Dispositif n’est pas destiné à traiter l’ictère cholestatique
5
Agés de moins de trois mois et pesant moins de 10 kg –définition selon la EN 60601-2-50. Les termes « nouveaux-nés » et
« nourissons » utilisés dans ce document sont utilisés selon cette définition en accord avec EN 60601-2-50.

Manuel d’utilisation
: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 17
FR
*
2. INFORMATIONS DE SÉCURITÉ
Types de consignes de sécurité
Avertissement
Signale un danger ou un risque de blessure pour l’enfant ou l’opérateur
Important
Mentionne les actions nécessaires à l’obtention de bons résultats cliniques
Avertissement
Prérequis
Le Dispositif ne doit être utilisé que par du personnel médical qualifié qui connaît bien les risques
et avantages de ce type de traitement et ce type d’appareils ou par des individus formés par ce
personnel médical.
Utilisation à domicile : Avant la première utilisation du Dispositif, l’utilisateur à domicile doit
impérativement être formé à son installation, son utilisation ainsi que sensibilisé aux risques liés à
l’utilisation du Dispositif. Il est impératif de suivre les recommandations et instructions fournies dans
les notices (système et usage unique) et par le professionnel de santé.
L'utilisation du Dispositif sur des contrindications, ou pour des indications médicales ou des parties
du corps différentes de celles préconisées n'est pas autorisée. En effet, cela pourrait conduire soit
à un système totalement inefficace (aucun bénéfice thérapeutique), soit à une situation
potentiellement à risque pour la santé (effets indésirables).
Avertissement
Tenir hors de la portée des enfants.
Avertissement
Variation des paramètres ambiants
La variation des paramètres ambiants (température et humidité de la pièce, exposition au soleil,
matériel environnant) peut potentiellement affecter négativement la santé de l’enfant, notamment
sa température et son bilan hydrique.
Une utilisation du Dispositif dans des conditions hors de celles spécifiées peut par ailleurs conduire
à un mauvais fonctionnement
Avertissement
Inflammabilité
Conserver à l’écart de toute flamme ou source d’étincelles.
Pour consulter l’ensemble des risques liés à la photothérapie, se référer au manuel d’utilisation du
BiliCocoon™ System.
Avertissement
Intégrité et lisibilité des étiquettes
Dans le cas où les étiquettes sont dégradées, les informations concernant la sécurité du patient,
des utilisateurs, de tierces parties ou de propriétés peuvent être retrouvées dans ce document.
Contacter le distributeur si les informations relatives à l’identification du produit
ou des produits associés ne sont pas visibles.

Manuel d’utilisation
: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 18
FR
*
Symbole
Signification
Symbole
Signification
Fabricant
Déclaration de marquage CE
selon la Directive 93/42/CEE
Consulter les précautions
d’emploi
Date de fabrication
Référence catalogue
Numéro de série
Craint l’humidité
Limites de températures de
stockage des produits
Limites d’humidité de stockage
des produits
Fragile, manipuler avec soin
Abréviations et unités
Abréviations
Significations et unités
°C
Degré Celsius
(Unité de température)
kg
Kilogramme
(Unité de masse)
cm
Centimètre
(Unité de longueur)
%RH
Humidité
(Pourcentage d’humidité de l’air)
3. CARACTÉRISTIQUES TECHNIQUES
Le système de fixation - ou BiliCocoon™ Fixation system - est un dispositif de fixation destiné à être utilisé
exclusivement avec le BiliCocoon™ System, dispositif médical apportant une solution de traitement par photothérapie
à l’hyperbilirubinémie non-conjuguée du nouveau-né.
Le BiliCocoon™ System est composé d’un Pad lumineux connecté à une Light Box émettant de la lumière bleue dans
le spectre d’absorption de la bilirubine, permettant ainsi de réduire sa concentration. Pour plus d’informations, se
référer au manuel d’utilisation du BiliCocoon™ System.
Paramètre
Valeur
Unité
Dimension
18x15x11
cm
Poids
0,6
kg
Couleur
Gris
Température min et max d’utilisation
De 0°C à 40°C
°C
Humidité min et max d’utilisation
De 30% à 75%
%RH
Température min et max de stock.
De -20°C à 60°C
°C
Humidité min et max de stock.
De 0% à 90%
%RH
Le non-respect des conditions de stockage peut entrainer des risques pour les patients, les
utilisateurs, les tiers ou les biens.
0%
90%
-20 °C
+60 °C

Manuel d’utilisation
: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 19
FR
*
Avertissement
Avertissement
Les conditions d’utilisation et de stockage sont différentes entre le BiliCocoon™ System et
le BiliCocoon™ Fixation system. Veuillez ainsi vérifier que les conditions de stockage et
d’utilisation soient compatibles avec les conditions conseillées.
Avertissement
Aucune modification du Bilicocoon™ Fixation system n’est permise. Ne pas utiliser aucun
accessoire, pièce détachée et matériel non décrit dans les conditions d’utilisations. L’usage
du système de fixation ayant subi une modification en terme d’utilisation ou ayant été
indûment modifié avec d’autres pièces peut entraîner des risques pour le patient, les
utilisateurs, les tiers et les biens.
BiliCocoon ™ Fixation system
1 : Pince de fixation à vis dite holding clow
2 : Plaque de fixation
3 : Crochets de fixation
4 : Gorge
5 : Vis de montage
6 : Mors de serrage
1
3
4
5
6
2

Manuel d’utilisation
: BiliCocoon™ Fixation System
IFU-BCC301-EnFr-2017-04-v03 Page 20
FR
*
4. INSTALLATION ET MODE D’EMPLOI
A. Précautions préalables
Avertissement
Avant de positionner le BiliCocoon sur le système de fixation, vérifier l’irradiance du Pad (se
référer au manuel d’utilisation du BiliCocoon™ System).
Si la vérification n’est pas réalisée avant chaque utilisation, il existe un risque de sous-
exposition et donc de non-efficacité du traitement, ou de surexposition.
Avertissement
S’assurer que le système de fixation n’est ni ébréché ni abîmé, et qu’il ne présente pas d’arêtes
vives pouvant blesser.
Avertissement
S’assurer que le système de fixation soit propre et sans tâche. Le nettoyer si nécessaire avec
un nettoyant adapté, comme de l’eau ou les nettoyants précisés dans ce manuel.
Avertissement
S’assurer que la pince de fixation à vis est bien fixée au support de fixation et qu’aucune vis
ne manque. Vérifier que le mors mobile puisse assurer sans obstacle sa fonction de serrage
Avertissement
Vérifier que la pince de fixation à vis ne soit ni abimée ni ébréchée.
Avertissement
Vérifier également que les deux crochets de fixations soient en bon état, et ne soient ni abimés
ni ébréchés.
Avertissement
Le système de fixation est appelé à être attaché à un support (mât de fixation ou lit). Veillez à
que ce support soit solide, stable et résistant, et que la largeur soit entre 1 et 5 cm. Veuillez
vérifier la stabilité du support (mât de fixation ou lit) avant la fixation et après la fixation. Veuillez
vérifier la stabilité de l’ensemble avant toute utilisation pour éviter la chute du BiliCocoon™
System.
Avertissement
Vérifier que le serrage de la pince de fixation à vis est suffisant et qu’elle est solidaire du
support.
Avertissement
Vérifier que le système de fixation est fixé à une hauteur utilisable, et qu’il soit orienté pour être
facilement accessible. Vérifier que les orifices de ventilation de la BiliCocoon™ Light Box ne
sont pas obstrués.
Avertissement
Vérifier que la Light Box ™ est correctement positionnée sur le système de fixation.
Avertissement
Vérifier de ne pas coincer le cordon d’alimentation de la Light Box ™.
Avertissement
Le non-respect des conditions d’utilisations, de maintenance et de stockage peut entrainer des
risques pour le patient, les utilisateurs, les tiers ou les biens
Table of contents
Languages:
Other NeoMedLight Medical Equipment manuals
Popular Medical Equipment manuals by other brands

Maquet
Maquet Datascope CS300 Operator's manual

Chattanooga
Chattanooga LIGHTFORCE XLi Instructions for use

Kolibri
Kolibri StarDevice user manual

Allegion
Allegion Normbau NY.347 Series Fixing instructions

Mizuho
Mizuho MES-CK08-117-16EN-0 Instructions for use

Medic Therapeutics
Medic Therapeutics YM201 operating instructions