(201) 343-5202, Ext. 1
If external cleaning of the lter is needed, use 1% bleach. AVOID contact
between the lter and the chemical alkyl dimethyl benzyl ammonium
chloride (ADBAC).(1)
WALL BOX
TURN OFF
BI-CARB PUMP
BI-CARB
INLET
INSERT
INSERT
INSERT
INSERT
CONNECT
TEFLON TAPE
CONNECT
TEFLON TAPE
SSUmini
1 2 3
WASH hands before starting installation.
WEAR gloves before handling lter.
Hand sanitizers used in healthcare can affect
performance of product material.(1)
TURN OFF the central bi-carb pump.
Behind the wall box cut the bicarb tube between
the loop outlet port and the wall box bi-carb
receptacle inlet.
If necessary, CONNECT the John-Guest to
threaded connectors onto the inlet and outlet
side of the SSUmini.
INSERT the tube tting into the opening of the
John-Guest adapter. Be sure to insert the tube
tting so that it is secure into the connector and
check by tugging at the tube tting.
FLOW
INSERT
INSERT
4
(1) The chemical ADBAC can cause structural damage to lter.
The FLOW of the lter should be in the direction
towards the wall box.
INSERT the barb tting from the inlet of the lter
into the bi-carb pump line.
INSERT the barb tting of the outlet of the
SSUmini into the tube from the wall box bi-carb
receptacle.
PRIME the bi-carb line by connecting a tting to
the bi-carb receptacle on the wall box. PRIME
the line till bi-carb is seen leaving the tube.
WALL
BOX
TURN OFF
BI-CARB PUMP
CONNECT
CONNECT
SSUmini
INSERT
INSERT
INSERT
INSERT
FLOW
INSERT
INSERT
WALL
BOX
For SSUmini