NEULOG VOLTAGE LOGGER SENSOR GUIDE
NeuLog voltage logger sensor
NUL-201
The NeuLog voltage sensor can be used for any science experiment which utilizes
voltage readings. It is used in the fields of Physics, Electronics, Chemistry, Biology,
etc.
The sensor comes pre-calibrated so you can start experimentation right out of the box
using this guide.
Among hundreds of possible experimental subjects that can be studied with the NUL-
201 sensor are: fuel cells, battery chemistry, electrical parallel circuits, electrical series
circuits and electronics.
The voltage sensor's measurement units are:
Volts (V): The SI derived unit for electrical potential difference across two determined
points.
Quick start procedure:
PC or Mac Computer
Materials needed:
NUL-201 Voltage Sensor
USB-200 USB Module
USB to mini USB cable (included with the USB-200 module)
Your voltage sensor needs to be connected to a USB-200 module. The USB-200
module then connects to a computer via a USB to mini-USB cable. Please note that
you cannot plug the voltage sensor directly into the computer.
The browser based application can be downloaded for free at www.NeuLog.com as
well as a full application user guide.
Choose "Downloads" on the main menu and then choose "Software and Application".
Procedure:
1. Install the NeuLog application.
2. Connect the USB-200 module to the PC or Mac.
3. Connect the voltage sensor to the USB-200 module (they directly plug together).
Please note that no calibration is required for this sensor.
4. Open the NeuLog application by clicking on the shortcut on the screen.
5. Once a voltage sensor module box appears on the left side of the screen, the
probe has been automatically identified and you can begin experimentation.
6. If the voltage sensor is not automatically identified, click the “Search for sensors”
icon to find the sensor.
7. Click on the “On-line experiment” icon; this will open a graph below.
8. Click on the “Module setup” button located in the voltage sensor module box to
change the sensor's settings if need be.
9. Click on the "Experiment setup" icon to change the experiment settings if need
be (experiment duration for example).
10. The voltage sensor will give a live reading in its module box while plugged in.
11. To run an experiment and collect data, click the “Run experiment” icon.
12. To end data collection early, click the “Stop experiment” icon.
Tablet, smart phone device
Materials needed:
NUL-201 Voltage Sensor
WIFI-201 WiFi Module
BAT-200 Battery Module
Your voltage sensor needs to be connected to a WIFI-201 module. The WIFI-201
module will create a closed NeuLog WiFi network which will stream the NeuLog data
to the device of your choosing. Once your device is wirelessly connected to the
NeuLog network, you can run experiments and collect data through a browser of your
choosing.
Procedure:
1. Connect the voltage sensor directly to the left side of a WIFI-201 module (no
wires required).
2. Connect a BAT-200 module to the right side of the WIFI-201 module.
3. Although not required, we recommend plugging the BAT-200 to an outlet using a
USB to mini USB charger (such as a typical cell phone charger). The WIFI-201
module will run for 60-90 minutes (depending on the sensor) without being
plugged in.
4. The WIFI-201 module can be powered directly using a mini to USB cord and
plugging it into your computer or a wall charger. Please note this function is only
available on WIFI-201 units with the USB icon on the cover.
NEULOG VOLTAGE LOGGER SENSOR GUIDE
5. For further WIFI-201 instructions or the WIFI-201 quick start guide please visit:
www.NeuLog.com; Choose "Products" on the main menu,
then choose "WiFi Communication module".
6. The WIFI-201 indicator lights will flash; take no action until the LED to the far left
turns blue; this can take up to one minute.
7. Take your tablet or smart phone, go to the WiFi settings and select the NeuLog
network (NeuLog0184 for example) which matches the ID found on the back of
the WIFI-201 module (0184 for example).
8. Give your device about 20 seconds to connect to the WIFI-201 module.
9. Once the device is connected, go to your browser and type the website
wifi201.com into the URL bar, then wait for 30-60 seconds.
10. You will see a “Control mode” icon in the application, click on it.
11. The application will then load a new screen and begin to auto detect the
sensor(s); this can take a minute. (You can stop the search when the sensor is
found.)
12. If the application does not auto detect the sensor(s), select “Search for sensors”.
13. Once the sensor is found, you will see a module box on the left side of the
screen for the voltage sensor. In this box, real time data will be displayed.
14. Click on the “Module setup” button located in the voltage sensor module box to
change the voltage probe settings if need be.
15. Click on the “Experiment setup” icon to change the experiment settings if need
be (Experiment duration for example).
16. To run an experiment and collect data, click on the “On-Line experiment” icon,
then click the "Run experiment" icon.
17. If you have a newer generation WIFI-201 module (with the USB icon on the
cover) this can be used as a USB-200, meaning you can tether the probes
directly to the computer using the WiFi-201. To engage operation in this fashion,
press 3 times on the panel button. Repeating this operation will return the unit to
WiFi mode.
Operation with NeuLog Graphic Display Module (viewer)
Materials needed:
NUL-201 Voltage Sensor
VIEW-101 Graphic Display Module
BAT-200 Battery Module
Procedure:
1. Connect the voltage sensor to the left side of the viewer.
2. Connect the BAT-200 module to the right side of the viewer.
3. The VIEW-101 will auto detect the sensor and you will see it appear on the left
side of the screen.
4. Once the sensor appears, it will be monitoring data in real time.
5. To run an experiment and collect data, press the run icon (little green person).
Off-line experiments
(Off-line experiments are when you do not have a sensor connected directly to a
computer, tablet, smartphone, or NeuLog viewer).
Materials needed:
NUL-201 Voltage Sensor
BAT-200 Battery Module
Materials needed to configure your offline experiment:
USB-200 module, WIFI-201 module or VIEW-101 module
USB to mini USB cable (included with the USB-200 module)
Procedure:
1. To determine the experiment settings, first connect to your chosen device (PC,
Mac, tablet, smart device) and click on the “Off-line experiment” icon after the
sensor search has ended. (For more information on how to change the
experiment settings, view the quick start procedure section for your chosen
device in this document.)
2. Connect the voltage sensor directly to the left side of a charged BAT-200
module.
3. When ready to collect data, press the “Start/Stop” button on the voltage sensor;
a red light will turn on for the duration of the experiment.
4. After the experiment has concluded (the “Start/Stop” button is pressed again or
the data collection period ends) connect the sensor to your chosen device (you
can find full instructions in each device’s quick start procedure section).
5. Open the NeuLog application.
6. Click the “Off-line experiment” icon.
7. Click the “Load data from sensors” icon.
8. Select which experimental data to upload (5 experimental runs can be stored at
one time on the NUL-201 voltage sensor).
NEULOG VOLTAGE LOGGER SENSOR GUIDE
Restoring sensor’s factory default settings:
Procedure:
1. Connect the USB-200 module to a PC, Mac, tablet, or smart device.
2. Connect the voltage sensor to a USB-200 module (they directly plug together).
3. Open the NeuLog application.
4. Click the “Tools” icon.
5. Click “Restore sensor’s factory defaults” on the menu.
Included with the sensor:
NeuLog NUL-201 Voltage Sensor Guide (this document)
Red (positive probe) and black (negative probe) wires with 4 mm plugs connected
directly to the sensor's body
Sensor's specifications
Range and operation
modes
±20 V
ADC resolution
15 bit
Accuracy
1%
Resolution
0.01 V
Max sample rate (S/sec)
3000
Experiment Duration: 50 milliseconds to 31 days.
Sensor's features:
Fully digital data
Rugged plastic ergonomic case
4 mm plugs for easy connectivity
Push button switch for Start/Stop experiments in off line mode
LED indicator of experiment status (blinks while collecting data)
Pre-calibrated sensing equipment
Note: NeuLog products are intended for educational use.
Videos and experiment examples:
Videos, literature and other probes can be found at www.NeuLog.com.
In order to access the voltage sensor's page, choose "Products" on the main
menu and then "Voltage logger sensor".
In order to find the science experiments, choose "Experiments" on the main
menu of the NeuLog website, then choose "Physics", "Chemistry", "Biology" or
"Environmental Science".
Physics Experiment:
oOhm’s Law (P-10)
Chemistry Experiment:
oProducing Electricity (C-1)
Technical background:
The philosophy behind NeuLog’s plug and play technology is based on each sensor’s
ability to store its own data due to an internal flash memory chip and micro-controller
in each plastic NeuLog body. This technology allows the sensor to collect and then
store the digital data in the correct scientific units (oC, oF, Lux, %, ppm, for example).
The sensor is pre-calibrated at the factory. The built-in software in the logger can be
upgraded for free at any time using the provided firmware update.
The sensor is based on a very fast, high input impedance differential operational
amplifier with protected inputs.
Maintenance and storage:
Never submerge the NeuLog plastic body in any liquid.
Do not allow liquid into the voltage sensor’s body.
After use, gently wipe away any foreign material from the sensor.
Store in a box at room temperature out of direct sunlight.
NEULOG VOLTAGE LOGGER SENSOR GUIDE
Warranty:
We promise to deliver our sensor free of defects in materials and workmanship. The
warranty is for a period of 3 years from the date of purchase and does not cover
damage of the product caused by improper use, abuse, or incorrect storage. Sensors
with a shelf life such as ion selective probes have a warranty of 1 year. Should you
need to act upon the warranty, please contact your distributor. Your sensor will be
repaired or replaced.
Thank you for using NeuLog!
Flexible, simple, fast, forward thinking.
W: www.neulog.com
E: info@neulog.com
A: 850 St Paul Street, Suite 15, Rochester, NY 14605
P: 1.866.553.8536
V11042014

Other NeuLog Data Logger manuals

NeuLog NUL-233 User manual

NeuLog

NeuLog NUL-233 User manual

NeuLog NUL-226 User manual

NeuLog

NeuLog NUL-226 User manual

NeuLog NUL-217 User manual

NeuLog

NeuLog NUL-217 User manual

NeuLog NUL-204 User manual

NeuLog

NeuLog NUL-204 User manual

NeuLog NUL-226 User manual

NeuLog

NeuLog NUL-226 User manual

NeuLog NUL-208 User manual

NeuLog

NeuLog NUL-208 User manual

Popular Data Logger manuals by other brands

ST HT22IoT user manual

ST

ST HT22IoT user manual

ELPRO LIBERO CE Operation manual

ELPRO

ELPRO LIBERO CE Operation manual

T&D TR41 user manual

T&D

T&D TR41 user manual

Elitech RC-5+ user manual

Elitech

Elitech RC-5+ user manual

globalwater GL500-7-2 manual

globalwater

globalwater GL500-7-2 manual

GFZ DiGOS Omnirecs DATA-CUBE3 Series user manual

GFZ

GFZ DiGOS Omnirecs DATA-CUBE3 Series user manual

Rotronic RMS-LOG-915 Short instruction manual

Rotronic

Rotronic RMS-LOG-915 Short instruction manual

Dahua EEC300D8-N1 quick start guide

Dahua

Dahua EEC300D8-N1 quick start guide

Clarke CIXTLI1 User instructions

Clarke

Clarke CIXTLI1 User instructions

Eagle Tree Systems eLogger V4 instruction manual

Eagle Tree Systems

Eagle Tree Systems eLogger V4 instruction manual

Sungrow Logger3000 user manual

Sungrow

Sungrow Logger3000 user manual

MadgeTech pHTemp2000 Product user guide

MadgeTech

MadgeTech pHTemp2000 Product user guide

Lascar EL-PDF-1 Series quick start guide

Lascar

Lascar EL-PDF-1 Series quick start guide

Gallagher HR5 instructions

Gallagher

Gallagher HR5 instructions

Hioki LR8432-20 instruction manual

Hioki

Hioki LR8432-20 instruction manual

Dynojet POWER VISION PV3-29-01 installation guide

Dynojet

Dynojet POWER VISION PV3-29-01 installation guide

Dostmann Electronic LOG40 operating instructions

Dostmann Electronic

Dostmann Electronic LOG40 operating instructions

AEMC Simple Logger II user manual

AEMC

AEMC Simple Logger II user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.