ASSEMBLING METAL TIP UNIT:
1. Loosen needle nut until reasonably loose.
2. Slide needle tip base into the lower slot of the needle
nut.
3. Screw the needle nut tightly back onto the main
threaded mount.
ASSEMBLING SILICONE TIP UNIT:
1. The silicone tip MUST always be sterilised prior to
use.
2. Remove the metal tip and needle nut.
3. Push the silicone tip onto the main threaded mount.
Turn clockwise as you push to ensure it is firmly in
place.
Packs of 25 silicone tips are available from your
distributor – Order Code: SP1370B.
CLEANING OR STERILISATION:
The following items may be autoclaved:
Ear tips needle; needle nut; the entire lever action hand
piece – however it is important to apply a small drop of
NJ Phillips Lubricant to any moving part post autoclave
as good maintenance. To apply NJ Phillips Lubricant to
moving parts within the handle grip, carefully unscrew
counter clock wise.
It is unnecessary, nor desirable, to dismantle any part of
the lever action unit for any purpose other than to lubricate
moving parts.
LIQUID MEASURE DISPERSMENT SETTINGS:
1. To set liquid dispersment
adjust the circular nut
located at the bottom of the
handle. Turn the circular
nut clockwise or counter
clockwise to adjust the
(ml) calibrations printed at
the top of the body post. Adults generally require 3-5 (ml)
setting. The dose is set from the bottom of the lever arm.
2. Prime instrument to ensure all air is removed prior to
use on the patient.
EAR SYRINGE LEVER ACTION
For use in the removal of cerumen along the outer ear canal.
CIRCULAR NUT
DOSE SETTING
BODY POST
EARDRUM
EAR CANAL
CAUTION:
1. DO NOT insert the ear tip end into the
outer ear canal.
2. Take due care - rapid jets of water
directed at the tympanic membrane could
be harmful.
3. This device should only be used by a
health care professional familiar with ear
syringing techniques.
ARTG No: 145575
The rubber tip is designed to be autoclaved
at temperatures of up to 250 degrees
celsius for up to 3.5 hours.
NOTE
Make sure the thread is properly engaged
when reconnecting the handle grip. To
reconnecthandlegrip,gentlypulllevertrigger
until the handle grip cover slides up and over
the handle body and screw clock wise until
tight – DO NOT OVER TIGHTEN!
IMPORTANT
!
DO NOT autoclave the hose piece, rather
wash in a good quality Medical Equipment &
Instrument detergent, then hang to drip dry.
NOTE
AUSTRALIA
NSW / ACT / VIC / WA / TAS
ALL STATES MEDICAL SUPPLIES
Contact: John Davies
35 Ellen Street
Wollongong
NSW 2500
Tel: 1300 520 510 Fax: 1300 509 089
Email: [email protected]
QLD
MEDICAL AND SURGICAL REQUISITES PTY. LTD.
Contact: MEDI + Plus - Service Advisor
33 Fulcrum Street
Richlands, Brisbane
QLD 4077
Tel: (07) 3859 2900 Fax: (07) 3859 2995
Email: [email protected]
SA / NT
Atlas McNeil Healthcare
Unit 5
32 – 50 Weaver Street
Edwardstown
SA 5039
Tel: 1300 741 085 Fax: 1300 365 408
Web: atlasmcneil.com.au
NEW ZEALAND
ESTEEM HEALTHCARE LTD
Contact: Mr. Brent Mitchell
PO Box 18-040, Glen Innes
or
24B Morrin Road
Tamaki
Auckland NZ
Tel: 0800 427 968 Fax: 0800 337 833
AS OUR POLICY IS ONE OF CONTINUOUS IMPROVEMENT THE MANUFACTURER RESERVES THE RIGHT TO ALTER THESE SPECIFICATIONS AT ANY TIME.
ALL PRODUCTS PRODUCED BY NJ PHILLIPS PTY LIMITED, ARE IDENTIFIED BY A UNIQUE BATCH NUMBER. THIS IDENTIFICATION NUMBER IS AFFIXED
TO THE PRODUCT TO ALLOW TRACEABILITY BY THE MANUFACTURER AND MUST NOT BE REMOVED IF PRODUCT INTEGRITY IS TO BE MAINTAINED.
NJ PHILLIPS PTY LTD ABN 36 000 082 002
ADDRESS: LOCKED BAG 8, GOSFORD, NSW 2250 AUSTRALIA
TELEPHONE: +61 2 4340 2044 FAX: +61 2 4340 1991
EMAIL: [email protected]
MADE IN AUSTRALIA
PAS1043 | QL561-R24
NJ PHILLIPS EAR SYRINGE DISTRIBUTION NETWORK
CARE AND MAINTENANCE
1. Clean ear tip and needle nut by brushing with a non
wire brush in a good quality Medical Equipment &
Instrument detergent prior to autoclave process.
2. Autoclave the unit as indicated in ‘CLEANING OR
STERILISATION’ when required.
3. Draw a small quantity of NJ Phillips Lubricant into the
lever action unit through the inlet fitting (where the hose
attaches to the back of the unit) to maintain a well-
lubricated action.
4. Place a few drops of NJ Phillips Lubricant in all moving
parts to ensure a smooth action.
KIT-PAS1043
BLUE
VALVE YELLOW
VALVE
SILVER
SPRING BROWN
SPRING
When using KIT-PAS1043 to service
your Ear Syringe, ensure the valves
are placed in the correct sequence,
as shown in this diagram.
WARNINGS
1. THIS APPLICATOR IS DESIGNED TO
INJECTANIMALS.THEOPERATORMUST
TAKE CARE TO AVOID CONTAMINATING
THEMSELVES WITH THE PRODUCT
BEINGUSEDORCAUSINGTHEMSELVES
NEEDLE-STICK INJURIES OR SELF-
INJECTION.
2.ITISESSENTIALTHATTHEOPERATOR
READSANDFOLLOWSTHEDIRECTIONS
FOR USE AND SAFETY INSTRUCTIONS
ON THE LABEL OF THE PRODUCT
BEING APPLIED. THE OPERATOR IS
AT RISK OF INJURING THEMSELVES,
OTHERPEOPLE,ORTHEANIMALS BEING
TREATED IF THE PRODUCT THEY ARE
APPLYINGISNOTHANDLEDSAFELY AND
USED CORRECTLY.
3. THE OWNER OF THIS APPLICATOR
MUST ENSURE THAT THE OPERATOR
IS TRAINED IN ITS CORRECT USE. THE
OPERATOR IS AT RISK OF INJURING
THEMSELVES, OTHER PEOPLE, OR
THE ANIMALS BEING TREATED IF THE
APPLICATORISNOTUSEDCORRECTLY.
4. WHEN FITTING OR REMOVING
NEEDLES AND DURING USE, DUE CARE
AND ATTENTION MUST BE GIVEN TO
ENSURE THAT ACCIDENTAL NEEDLE
STICK INJURIES ARE AVOIDED.
5. WHEN SETTING UP, CLEANING,
MAINTAINING OR USING THIS
APPLICATOR, TAKE ALL APPROPRIATE
PRECAUTIONS TO AVOID SPILLING,
DIRECTCONTACTWITH,ORINHALATION
OF, THE PRODUCT BEING USED WITH
THIS APPLICATOR. REFER TO THE USE
AND SAFETY INSTRUCTIONS ON THE
LABEL OF THE PRODUCT FOR SPECIFIC
ADVICE AND WEAR APPROPRIATE
SAFETY CLOTHING AS RECOMMENDED
BY THE MANUFACTURER OF THE
PRODUCT BEING APPLIED.
6. KEEP PACKAGING MATERIAL AWAY
FROMCHILDRENANDANIMALS. PLASTIC
BAGS AND PACKAGING CAN CAUSE
SUFFOCATION.PLEASEDISPOSEOFALL
PACKAGING CORRECTLY.
7. BEFORE USE THE EQUIPMENT MUST
BE STERILISED.
8. SEEK ADVICE ON THE STANDARD
PRACTICES TO BE USED TO MINIMISE
THE RISKS OF CROSS-INFECTION
BETWEEN TREATED ANIMALS. SUCH
ADVICE SHOULD BE AVAILABLE FROM
SOURCES SUCH AS PRODUCER
AND INDUSTRY ORGANISATIONS,
GOVERNMENT EXTENSION SERVICES
OR THE MANUFACTURER OF THE
PRODUCT BEING APPLIED.
9. BEFORE DOSING, ENSURE THAT THE
COMPONENTS OF THE APPLICATOR
ARE SECURE AND THAT IT IS IN GOOD
WORKING ORDER.
10. DO NOT USE THIS APPLICATOR
IF IT IS LEAKING. STOP IMMEDIATELY
AND CORRECT THE FAULT BEFORE
CONTINUING.
11. THE DOSE ADJUSTMENT
MECHANISM OF THIS APPLICATOR MAY
BE DISLODGED DURING HANDLING.
REGULARLY CHECK THAT THE DOSE
SETTING IS CORRECT.
12. DO NOT LEAVE THE EQUIPMENT
EXPOSED TO DIRECT SUNLIGHT
OR EXTREMES OF HIGH OR LOW
TEMPERATURES AS THE APPLICATOR
OR THE PRODUCT BEING APPLIED MAY
BE AFFECTED.
13. DURING THE WORKING DAY, WHEN
THE APPLICATOR IS NOT IN USE, LEAVE
IT IN A SAFE PLACE OUT OF REACH OF
CHILDREN, ANIMALS AND PEOPLE NOT
TRAINED IN ITS CORRECT USE.
14. DURING THE WORKING DAY, WHEN
THE APPLICATOR IS NOT IN USE, AND IF
THE APPLICATOR IS OF A TYPE WHICH
ISCONNECTEDBYAFEEDTUBETOTHE
CONTAINER, ENSURE THE APPLICATOR
IS ABOVE THE CONTAINER. IF IT HANGS
BELOW THE CONTAINER FOR LONG
PERIODS IT MAY DRIP.
15. FOLLOW THE SEPARATE CLEANING
AND STORAGE INSTRUCTIONS SHOWN
BELOW.
16. STORE THE APPLICATOR IN
A SAFE PLACE OUT OF THE REACH
OF CHILDREN. STORE NEEDLES
SEPARATELY FROM THE APPLICATOR.
17.ALWAYSWASHHANDSTHOROUGHLY
AFTER USING OR SERVICING THIS
EQUIPMENT.
STERILISING EQUIPMENT BEFORE USE
The recommendations for sterilisation
vary according to the material from which
the injector is made. Specific sterilisation
instructionsshouldbefoundonthe packaging
or leaflet accompanying this injector. If in
doubt, contact us at [email protected]
CLEANING & STORAGE
DO NOT store your applicator or feed tube
full of product. To ensure long service from
this applicator it should be properly cleaned
and stored after use:
1.Afteruse,emptytheremainingproductand
dispose of it appropriately.
2. Flush through with warm, soapy water,
drawing through the applicator feed tube or
bottle attachment.
3.Thenflush through withclean water.Expel
any excess water.
4. Before storing, disconnect the feed tube
and inject a small quantity of NJ Phillips
Lubricant into the inlet. Operate the gun
severaltimesuntilthelubricantappearsinthe
barrel and the seals are lubricated.
5. Dry the outside surfaces of the applicator
and store in a plastic bag in a safe place out
of the reach of children.
This informationisonlyrelevant to equipment
produced by NJ Phillips Pty Ltd. The
manufacturer will accept no responsibility if
this applicator is used in any way contrary to
the information contained in both this leaflet
and the other information and instructions
enclosed with the applicator.
NJ PHILLIPS PTY LTD
ABN 36 000 082 002
MADE IN AUSTRALIA
ADDRESS: LOCKED BAG 8,
GOSFORD, NSW 2250 AUSTRALIA
TELEPHONE: +61 2 4340 2044
FAX: +61 2 4340 1991
EMAIL: [email protected]
YOU CAN CONTACT NJ PHILLIPS,
ANDOBTAINMOREINFORMATION
ABOUT THIS EQUIPMENT AND
SPARE PARTS BY GOING TO
WWW.NJPHILLIPS.COM
QL620A-R5
ENG, GER, FR, SP.
PLEASE READ THIS LEAFLET BEFORE USE. KEEP THIS LEAFLET AND
OTHER INSTRUCTIONS IN A SAFE PLACE FOR FUTURE REFERENCE.
GENERAL ADVICE FOR USE OF NJ PHILLIPS
INJECTORS & INJECTOR DRENCHERS

This manual suits for next models

1

Popular Medical Equipment manuals by other brands

LINET Eleganza 1 LE Service manual

LINET

LINET Eleganza 1 LE Service manual

Drive PreserveTech Harmony True 14200N user manual

Drive

Drive PreserveTech Harmony True 14200N user manual

Weinmann ACCUVAC Basic Description and operating instructions

Weinmann

Weinmann ACCUVAC Basic Description and operating instructions

bort medical Valco SOFT quick guide

bort medical

bort medical Valco SOFT quick guide

Dr. Mach LED 115 Instructions for use

Dr. Mach

Dr. Mach LED 115 Instructions for use

Link Endo-Model 15-8007/01 manual

Link

Link Endo-Model 15-8007/01 manual

Boscarol OB 2012 FA operating instructions

Boscarol

Boscarol OB 2012 FA operating instructions

Future Mobility Healthcare AQUA REHAB COMMODE TL operating instructions

Future Mobility Healthcare

Future Mobility Healthcare AQUA REHAB COMMODE TL operating instructions

Thomashilfen SWIFTY user manual

Thomashilfen

Thomashilfen SWIFTY user manual

Relign R-10002 Instructions for use

Relign

Relign R-10002 Instructions for use

AGFA WS-Motorized-T-001 user manual

AGFA

AGFA WS-Motorized-T-001 user manual

Ubiotex TALLA S user manual

Ubiotex

Ubiotex TALLA S user manual

Valon Multispot Laser user manual

Valon

Valon Multispot Laser user manual

GEIGER TCU 150 Operation manual

GEIGER

GEIGER TCU 150 Operation manual

Arjo Prioma Instructions for use

Arjo

Arjo Prioma Instructions for use

Chiltern Invadex WISPA 100 Plus user guide

Chiltern Invadex

Chiltern Invadex WISPA 100 Plus user guide

UMF Medical 5005 Service manual

UMF Medical

UMF Medical 5005 Service manual

mychway WL-SW5 user manual

mychway

mychway WL-SW5 user manual

manuals.online logo
manuals.online logoBrands
  • About & Mission
  • Contact us
  • Privacy Policy
  • Terms and Conditions

Copyright 2025 Manuals.Online. All Rights Reserved.