
....
I) Set ANALOG OUTPUT
Analog outputs are designed for interfacing with recorders, pumps and computers.
!!WARNING!!
controller could
be
damaged
and
void
warranty!
Analog outputs are
self
powered.
Do
not try to
externally
loop
power.
Externally
powered
outputs
will
damage
your controller!
The programmable option is an Isolated Proportional4-20 mA output,
designedto interfacewith an analog recorder,meteringpump or computer.
The outputis 4-20mA over a programmable range. For example,
if
the
analog input was pH, the output couldbe programmed to be 4 mA at 6.7
pH, the "LO RANGE"set point, and 20 mA at 8.2pH, the "HI RANGE"
set point".
Analog outputs require calibration upon initial start up. Calibration is re-
commended after changing scales or after factory reinitialization. For
accurate calibrationof the analog output,a milliamp meter is required.
Calibration
Note: LO
=
4 mA
HI =20 mA
TIP:
For help with
menu
locations,
please
refer to the
"Menu Map"
supplied
with your
controller.
1. Remove signal wires if installed and connect a milliamp meter
to output connections on daughter board. Observe proper polarity
(+/-).
Refer to Diagram 8, pg. 35, for connections. Note: The Diagram
reflectsa dual or 2 channelboard. Depending on your controller model,
there may be connectionsfor onlyone output.
2. SCROLL to CONFIGURE-D/A OUT SELECT under the
MAIN MENU-CONFIGURE menu. Press ENTER.
3. D/A OUT #1 PARAM will be displayed on the first line. The sec-
ond line will display system conductivity, pH, or ORP (depending
on model of your controller) to be monitored or controlled. If more
than one analog output is present, use the ARROW keys to display
the analog input to be tagged to analog output. Present analog
input is designated by an asterisk (*). Press ENTER.
4. Press ENTER and SCROLL DOWN to LO ADJUST. As indi-
cated on the display, use the ARROW keys to calibrate the
"lo"
mA
reading (4
mA
) on the external milliamp meter. Press ENTER
to enter the adjustment.
5. SCROLL DOWN to HI ADJUST.As indicated on the display,
use the ARROW keys to calibrate the "hi"
mA
reading (20
mA
) on
the external milliamp meter. Press ENTER to enter the adjustment.
6.
If
controller is equipped with the programmable option,
SCROLL DOWN to LO RANGE.As indicated on the display, use
the ARROW keys to set the "lo range" set point. Press ENTER to
enter the adjustment. SCROLL DOWN to HI RANGE.As indicat-
ed on the display, use the ARROW keys to set the "hi range" set
point. Press ENTER to enter the adjustment.
7. Press SCROLL DOWN for additional analog outputs to be set,
or press HOME repeatedly to return to MAIN MENU-CONFIGURE.
8. Remove milliamp meter and connect analog recorder, metering
pump or computer. Be sure to maintain proper polarity when con-
necting external devices!
Configure
D / A Out Select
D / A out #1 param
•
sys cond
D / A #1 lo adjust
ARROWS to change
D / A #1 hi
adjust
ARROWS to change
D / A #1 lo range
0
us/cm
D / A #1 hi range
5000
us/cm
16