Checking the insulin flow (Priming)
2
Keep the pen upright and tap the
cartridge holder gently to raise any
air bubbles. Press the dose button all
the way in until it clicks and the dial
returns to 0.
If insulin does not appear at the tip of
the needle, repeat steps 4 to 6.
If no drop appears, replace the needle
with a new needle and repeat the
process. You can also conduct a
Function Check (instructions overleaf).
6
Pull out the dose button. Turn the
dose button to select:
• 4unitswithanewPenll®cartridge.
• 1unitwithacartridgealreadyinuse.
5
Insert the Penll®cartridge into the
cartridge holder, colour-coded cap
endrst.
3
Put the pen back together and
twist it until you feel or hear a click.
If the cartridge contains an insulin
suspension (cloudy insulin) you
must always mix (resuspend) it
evenly before use (instructions
overleaf).
Before injection, always attach a
new NovoFine®needle.
Remove both the outer and inner
needle caps, and dispose of the
inner needle cap
.
Keep the outer needle cap to remove
the needle from the pen after injection.
4
NovoPen®4
1
Preparing NovoPen® 4
Pull off the pen cap and twist the pen
apart.
If the rod is sticking out, push it all the
way back in.
Pull out the dose button and turn the dial until the white indicator line is level with the required number of units
you need to inject. The odd units are shown as lines only (maximum 60 units).
Adjust the dose by dialling up or down, if necessary. No insulin will be released until the dose button is pressed.
If you need more than 60 units, you can divide your dose into two injections. Refer to the user manual for instructions.
Push the pen into your skin, then
press the dose button until you
feel or hear a click and the dose
indicator lines up with 0.
Leave the NovoFine®needle under
the skin for 6 seconds (keeping the
button pushed down), and then
remove it.
After you have withdrawn the needle
carefully put the outer needle cap
back on, then remove and dispose of
the needle safely.
Replace the pen cap after use.
78109
Setting your dose Injecting your dose
How to use NovoPen® 4
Pen cap
Piston rod head Display Dose button
Piston rod Dose indicator
Mechanical part
Penll®cartridge
Colour-coded cap
Penll®cartridge holder
5 units
selected
24 units
selected
Six
seconds
Use the injection technique shown to
you by your Healthcare Professional.
Always
use a new
NovoFine®
needle every time you inject
Remember
Outer
needle cap
Inner
needle cap
Needle Paper tab
Colour-coding
(example)
4mm 32G
NEW
NovoPen®4 is designed to be used with NovoFine® needles
NovoFine® needles 6mm & 8mm
Outer
needle cap
Inner
needle cap
Needle Paper tab