LogBox-RHT-LCD
NOVUS AUTOMATION 5/7
Fields are:
Title: In this field, the user identifies the logger by assigning it
a name.
General Information: Area with information about the logger,
such as Model, Serial Number, Logger current Date/Time, PC
Date/Time, firmware version, memory capacity and number of
acquisitions stored in memory.
In this field, time is constantly updated while the communication
between logger and computer is taking place.
Acquisitions:Presents a series of parameters that define
how measurements will take place.
Interval: Defines the interval between readings: Minimum
interval is 1 (one) second.
Note: When the type of value logged is mean, maximum and
minimum values, the minimum interval is 10 seconds.
In Daily repetitions the user defines the time that daily logs will
take place.
Estimated time: In this parameter, the logger informs the
user how long it will take to occupy the full memory, in the
conditions set up during configuration.
Start of Readings: Readings can be started in one of five
different modes:
•Immediately: start as soon as programming is
considered to be ready, and is then sent (OK) to the
logger.
•Date: readings start at predefined date and time.
•Through Start Button: starts and interrupts readings by
pressuring the Start button, in the frontal part of the
logger, for two seconds. See Fig. 1.
•Setpoint: measurements start when a temperature
setpoint is reached. In this option, the setpoint value is
defined in the Channels field, where the Alarm parameter
is replaced by setpoint.
Daily Repetition: In Daily repetitions the user defines the
time that daily logs will take place.
Stop loggings: Options for the end of readings are:
•Full Memory: readings can be stored up to the full
memory capacity is reached.
•Wrap around: readings are continuous, replacing old
records with new ones as the number of readings
overpass the memory capacity.
•After: the logger stops readings after a certain number of
readings.
•Date: Readings stop at user-predefined date and time. In
case the logger memory capacity is reached before the
date defined, readings are stopped.
Channels: Shows parameters referring to each channel
separately. Channel 1 is the Temperature channel, and Channel 2
the Relative Humidity channel.
Fig. 12 – Logger Set Up (Channels)
Tag: Defines a name for the temperature registers.
Offset:Makes possible to correct the value logged.
Value: It defines how the value measured will be
registered. Options are:
•Instantaneous: The value logged will be exactly the
value measured at each interval defined. The minimum
interval between measurements is 1 seconds.
•Average: The value logged will be the mean of ten
measures taken within a predefined interval. The
minimum interval between measurements is 10 seconds.
•Maximum: The value logged will be the maximum value
found within ten measurements taken within a predefined
interval. The minimum interval between measurements is
10 seconds.
•Minimum: The value logged will be the minimum value
found within ten measurements taken within a predefined
interval. The minimum interval between measurements is
10 seconds.
Unit: Defines the unit of the value measured: ºC or ºF for
channel 1 (temperature) and only %for channel 2 (relative
humidity or dew point).
Alarm: Defines limit values that, when exceeded,
characterize an alarm condition. Alarm events are informed to the
user through the Alarm Indicator. The alarm limit value is shown
with a dotted line with the same color of the channel in the chart
from LogChart II software.
Input: Parameter available in channel 1 (temperature)
and channel 2 (humidity or dew point). It is also possible to
deactivate channel 1 and kepp the entire memory for humidity.
It is possible to select the Dew Point logging.
After filling all the fields select OK and settings are sent to the logger.
7OFFLOAD AND DATA VISUALIZATION
The transference of data to a PC is accomplished with the LogChart
II software. Measurements are displayed in tabular and graphical
form. Data can be saved in files for future analysis and
comparisons.
7.1 Offloading data
To offload data, click the icon Offloading Data, and turn the
interface towards the logger front. During data transference, a status
bar indicates remaining data to be transferred. Data offloading time
is proportional to the number of readings logged.
7.2 Visualizing data
At the end of values transfer, data can be displayed in a graphical
form.