Chapter 2. Performing Measurements
Fig. 2-2
c. In the control panel of Resonance tab make the following changes (Fig. 2-3):
-In Response signal drop-down list select Mag.
-Set the range (From … To)to include several cantilever resonance frequencies.
The second and third resonance frequencies are related to the first one as
f1/f0=6.22 and f2/f0=17.5. So, with the exemplary frequency f0=28 of the contact
probe the operating range should be taken as From = 0 kHz, To = 600 kHz.
-Clear the Auto peak find check box.
-Set Amplitude parameter equal to 1V × 1.
-Set the lock-in amplifier gain (Gain parameter) equal to 10.
Fig. 2-3
d. Click
Run button to plot the frequency response of cantilever oscillation amplitude
(Mag signal), see Fig. 2-4, Fig. 2-5. On the resulting response apart from the peaks
corresponding to the cantilever resonance frequencies, spurious peaks may appear
owing to resonance oscillations of the piezodriver exciting the cantilever.
e. By narrowing From … To range adjust it to cover the frequency range
corresponding to a chosen AFAM resonance mode (n=1, n=2, …). In the given
example the range was chosen between the first, f0, and the second, f1 , resonance
peaks related to the first mode. The cantilever contact resonance frequencies
( , ) shall lie between the resonance frequencies of the cantilever
oscillations in the free state: f0< < f1; f1< < f2.
AFAM
n
f1
AFAM
n
f2
AFAM
n
f1
AFAM
n
f2
9