Be sure to use a GCQ-100 mercury burner.
Be sure to mount positive pole (the wide head) before
the other, or the damage to the burner may occur.
Never subject the burner to excessive force when
mounting the Mercury Burner.
Be careful and avoid leaving fingerprints or dirt on the
mercury burner. Attached stain may cause distortion in
glass which could result in a ruptured burner. If stained,
wipe it a way gently with clean gauze.
To prevent any hazard, always turn the main switch on
the power supply unit to “O” (OFF), unplug the power
cord plug from the mains outlet, and wait for at least 10
minutes before replacing the burner.
2.2.8. Assembly of the Fluorescent Attachment, Cable and
Cord Connections (Figure 12, Figure 13)
1. Mount the lamp housing into the other end of the
attachment and fix it with two screws ④.
2. Plug the connector ⑤from the burner socket securely
into the connector on the power supply unit and make
sure the cord is correctly connected. (Make sure that the
main switch ④of the power supply is set to “O” (OFF)
before connecting cables)
3. Connect the power cord connector ⑥into connector on
the power supply unit and make sure the cord is
correctly connected.
Verify that the voltage and the frequency of the AC
mains outlet match the setting of the voltage switch and
the frequency switch on the rear of the power supply
units and improper setting may degrade burner
performance, or in the worst case(although very rare ),
cause the burner to explode.
It is better to use the power cord provided by Labomed
and the same type power cord should be used if you lose
or damage the old one.
2.2.9. Fuse Replacement (Figure 12, Figure 13)
1. Set the main switch to “O” (OFF) and unplug the
power cord before replacing fuses.
2. Using a flat-blade screwdriver, remove each of the
fuses holder ⑦by tuning it counter- clockwise and
pulling out.
3. Replace both fuses with new ones.
Always use the designated fuses (8A). And make sure
the voltage of the fuse match the voltage of the AC
mains outlet.